Systec Controls deltawaveC-F User manual

deltawaveC
deltawaveCdeltawaveC
deltawaveC-
--
-F
FF
F 2CH
2CH 2CH
2CH
Quickstart
QuickstartQuickstart
Quickstart
V1.01
A
User Interface
User InterfaceUser Interface
User Interface
deltawaveC-F will be switched on automatically after supply voltage has been plugged on.
Activates / deactivates the background lighting
To navigate through the different menus please use the keys located on the left and right beside the
display.
Move cursor in specific direction
Confirm your settings
Back to previous menu
XYZ
Activates a certain function (depending on choosen menu)
B
Set
SetSet
Set-
--
-up y ur fl w measu
up y ur fl w measuup y ur fl w measu
up y ur fl w measur
rr
rement in 5 steps
ement in 5 stepsement in 5 steps
ement in 5 steps
1. Choose suited mounting position for your transducers
2. Parameterize your flow transmitter
3. Mounting of ultrasonic transducers on your pipe
4. Set zero point (if possible)
5. Start your flow measurement
Pipe need to be filled completely for clamp-on measurement
C
Parameterizati n and m unting
Parameterizati n and m unting Parameterizati n and m unting
Parameterizati n and m unting
1. Switch on flow transmitter (automatically when plugging on supply voltage, see below)
2. Choose menu „quick setup“
3. Select measurment channel 1 or 2
4. Select diameter or circumference (depending on what figure you have) of your pipe
5. Put in value for diameter or circumference
6. Put in value for wall thickness of your pipe
7. Select pipe material
8. Select lining (if there is a lining)
9. Select type of ultrasonic transducer
Nr. 1 = XUC-FW-F21 DN10…..DN100
Nr. 2 = XUC-FW-F20 DN10…..DN100 (replaced by ultrasonic transducer model XUC-FW-F20 )
Nr. 3 = XUC-FW-F10 DN32…..DN400
Nr. 4 = XUC-FW-F05 DN200…DN6000
10. Select method of mounting. Standard is V-mounting.
Confirm your settings and moves to next menu
Increase of values
Decrease of values
No function

11. Mount ultrasonic transducers on a pipe
After parameterization the display is showing the required distance between the two transducers in
mm and also as Index-No. Index-No. means the number of holes to be used for mounting with the
spacer bar. The first hole after the first transducers is hole No. 1, the last hole is the hole where
second transducer is put into the spacer bar (threaded bar). Picture 3 shows an example for an index-
No. 4.
Together with deltawaveC-F unit itself you will get one pair US- transducers an a spacer bar
with matches together with the delivered US- transducers .
If you have ordered more deltawaveC-F units and US- transducers for different pipe diameters
it could be that you got spacer bars with different length (short or long). Pls. see in the chart below
which spacer bar could be operated together with the corresponding transducer model.
Picture 1 spacer bar short (length 250mm, grid hole distance 7,5mm)
Picture 2 spacer bar„long“ (length 400mm, grid hole distance 15mm)
Picture 3 Example for spacer bar with bar Index 4 means Picture 4 example for distance in mm without
3 holes spare between us- transducers (V or W mode)
Some acoustic coupling gel (app. size of a peanut, picture 5) or acoustic coupling foil has to be put at the part which
touches the pipe wall after installation (picture 6).
Picture 5 Using acoustic coupling gel Picture 6 Using acoustic coupling foil
(Magnalube)
Spacer bar F10 F20 F21
Short (25 cm) X
Long (40 cm) X X
0
1
2
3
4

12. Chose best location for US- transducer installation.
If you install one pair of transducers on each pipe straight run should be 10x diameter (inlet) and 3x
diameter (outlet) in order to achieve best accuracy. If you install 2 pairs of transducers on the same
pipe it could be used to; increase accuracy, for redundant operation or to reduce effects of flow
inclination (see pictures 8 and 9). It is mandatory that both transducers are installed exactly shifted
about 180°.
When using horizontal pipe we recommend mounting the transducers at 10 o’clock or 2 o’clock
position to avoid any influence of gas bubbles / sedimentation might accumulated on top / on bottom
of pipe.
Picture 7 Example Mounted Transducers with spacer
bar and mounted at app. 9 o`clock
Picture 8 Example two transducer Picture 9 Example two transducer one
pairs on the same pipe to increase line. Second transducer pair lies exactly
accuracy or to reduce effects of 180° shifted to the first transducer pair
flow inclination or for redundant
operation.
13. The transducers are fixed to the pipe by using metal belts. The metal belts are tightened by using
tightener (picture in the middle). The length of the belts are designed for the biggest pipe size your
transducer can cover. Please feed the belt into the tightener as shown in picture on
the left. Proceed in the same way for PT100. PT 100 Nr. 2 must be installed on the hotter pipe.
PT Nr. 1 on the colder pipe. Put between PT100 and surface of the pipe also acoustic
coupling gel in between. PT100 should after installation insulated tom minimize the
influence from ambience air temperature.
Pictures 10 Mounting US- Transducers, PT100 with stainless steel belts on a pipe
Gas bubbles
Pipe
Sedimentation

11 Set Zero Point (if possible)
If there is the chance to create „zero flow“ please select „Zero Setup“ and set zero point.
12. Go back to main menu and select button „MEAS“ . You then entering the measuring
menu where you see the calculated flow, velocity,….
Change units:
->Go to main menu and select SETUP - > COMPL SETUP. Then select „UNITS SETUP“ to change units.
13. Activate Outputs
->Go to main menu and select SETUP - > COMPL SETUP. Then select “I/O-SETUP” and set digital and/or analog outputs.
Please check deltawaveC-F user manual for further information. The user manual can be found as pdf-
document on the CD which is included in the delivery.
Important Notice !
Heat transfer measurement: Works only in operation mode (CH1+CH2)/2
Two transducers on the sam pipe:
Increasing accuracy: Works only in operation mode (CH1+CH2)/2
Reducing flow inclination Works only in operation mode (CH1+CH2)/2
Wall mount of deltawaveC
To mount deltawaveC on a wall please create three boreholes corresponding to the red marked points on
the drawing (all length specifications are metric).

Wiring
To perform the wiring it is necessary to remove the cable cover from deltawaveC-F by removing the two
screws (marked with red arrows, picture 11)
Picture 11 Removing cover plate from cable terminal Picture 12 screw terminals
Before you install the ultrasonic sensor cables to the measurement transducer we highly recommend
pulling the sensor cable endings through the delivered ferrites.
Loop the cable once as shown in the photo below (The delivered ferrite could differ from the ferrite on the
photography below).
Picture 13 Ferrite mounted on cable with one loop
Power Supply (AC or DC
pls. see type plate)
I/O Terminal
For power supply please use cables with a cross
section of ca. 0.5…4mm². Solid cables should be
dismantled about 8mm to allow proper contact to
terminals.
For I/O terminals please use cables with a cross
section of ca. 0.5…2,5mm². Solid cables should be
dismantled about 6mm to allow proper contact to
terminals.
Loop
Ferrite

Terminalname Colour Description
UP1
Connection for upstream transducer
(red cable = + / black cable = -),
Channel 1
DWN1
ORANGE
Connection for downstream transducer
(red cable = + / black cable = -),
Channel 1
UP2
Connection for upstream transducer
(red cable = + / black cable = -),
Channel 2
DWN2
RED
Connection for downstream transducer
(red cable = + / black cable = -),
Channel 2
PT100 No. 1
Input for temperature sensor.
Left terminal is for the „colder“ temperature
(return pipe)
PT100 No. 2
YELLOW
Input for temperature sensor.
Right terminal is for the „hotter“ temperature
(feed pipe)
Relay
GREEN
Relay, potential-free
AO1 4-20mA
Analog output 1 (4…20mA), active, 24VDC
AO1 4-20mA
BLUE
Analog output 2 (4…20mA), active, 24VDC
Digital out DO1
Universal digital output 1 (transistor),
for configuration of the digital outputs pls.
read the user main manual.
Digital out DO2
WHITE
Universal Digital output 2 (transistor),
for configuration of the digital outputs pls.
read the user main manual.
Power Supply
Please provide either 90-240VAC or 18-
36VDC (depending on chosen model).
Please make sure that you use the correct
voltage.
Flow transmitter might be damaged when
using wrong supply.
Note: All in- and outputs (except relay) have defined potential on the
internal devices ground. For potential free operation of the in- and
outputs is additional hardware needed (with galvanic isolation). With the
normal in- and output it is not possible.
Other manuals for deltawaveC-F
2
This manual suits for next models
1
Table of contents
Other Systec Controls Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

Agilent Technologies
Agilent Technologies E4406A VSA Series Programmer's guide

Klein Tools
Klein Tools CL2200 instruction manual

Optima Home Scales
Optima Home Scales Figure user manual

AVCOM
AVCOM PSA-37XP owner's manual

Amprobe
Amprobe TMD Series operating instructions

Kessler-Ellis Products
Kessler-Ellis Products MRT Installation & operating instructions