Model 7712 3.5 GHz 50 Ohm RF Module Instructions for use with DAQ6510
077147700 / June 2018
Close and open channels
Careless multiple channel operation could create an electric shock hazard that could result in
severe injury or death. Improper operation can also cause damage to the switching modules
and external circuitry. Operating channels independently should be restricted to experienced
test engineers who recognize the dangers associated with multiple independent channel
closures. Do not attempt to perform this procedure unless qualified as described by the types
of product users in the Safety precautions (on page 14). Do not perform these procedures
unless qualified to do so. Failure to recognize and observe normal safety precautions could
result in personal injury or death.
Multiple channel operation provides direct control of the channels of one or two switching modules. OUT A
automatically switches as appropriate for channels 1 to 4 and OUT B automatically switches as appropriate for
channels 5 to 8 for each module. When you close two channels in a single module (one for OUT A and one for
OUT B), the appropriate channels will close; other channels or channels on other switching modules are not
affected by channel closures.
Performance verification
Do not attempt to perform this procedure unless qualified to do so, as described by the types
of product users in the Safety precautions (on page 14). Do not perform these procedures
unless qualified to do so. Failure to recognize and observe normal safety precautions could
result in personal injury or death.
The following tests verify the RF characteristics of the switching module. S-parameters are obtained by
measuring voltage standing wave ratio (VSWR) and insertion loss among the common output and channel.
The following equipment is recommended for the tests in this section:
▪Tektronix TTR506A Vector Network Analyzer (6 GHz)
▪Tektronix test port cables, Type-N(m) to Type-SMA(m), 1 m (012-1770-00)
▪Tektronix 3.5 mm SOLT mechanical calibration kit 0 to 32 GHz, Spinner BN 53 38 54 (CALMECH-35MM)
Prepare for verification
To prepare for verification:
1. Turn on the network analyzer and the DAQ6510 with the switching module installed.
2. Allow the analyzer to warmup for at least one hour.
3. Calibrate the network analyzer from 500 MHz to 3.5 GHz using the appropriate calibration kit. The
calibration routine should include a load, short, open, and pass-through procedure. Use SMA style
connectors in the calibration process.