
~
When connecting the probe
to
the PROBE connector,
the
two connectors
must
be correctly aligned.
Damage to the terminals can result from forcing the
connector and jack together. Refer to the probe
in-
struction manual.
6. Apply
the
voltage
to
be measured between
the
probe
tip and
the
common strap on the probe. Do
not
use the
COM
INPUT connector when making a measurement with
the probe.
7. The readout will display a
"+"
preceding the voltage
reading if
the
probe tip
is
positive with respect
to
the com-
mon strap, and
"-"
if
the
probe tip
is
negative with respect
to
the common strap.
Voltage measurements with
the
probe are essentially
the
same as with test leads. The probe
is
a straight-through
device; i.e., it provides no attenuation
to
extend the mea-
surement range.
The
probe
is
shielded
to
minimize pickup
of electrostatic interference. (For more information, see the
instruction manual for the probe.)
DC
Current
Measurements
The
7D13
will measure
DC
current
in
four ranges as
follows:
.000
to
2.000
mA;
0.00
to
20.00
mA; 00.0
to
200.0
mA; and
000
to
2000
mAo
To
measure
DC
current
with
the
7D13, use the following
procedure:
1. Install the
7D13
in
any available plug-in compart-
ment
of
a Tektronix 7000-series oscilloscope.
2. Turn the oscilloscope power on. Allow
about
twenty
minutes warmup.
3. Advance the oscilloscope Readout control
to
obtain a
usable readout display (if necessary, refer
to
the oscillo-
scope instruction manual).
4.
Set
the MODE/RANGE switch
to
the desired full-
scale
DC
CURRENT range.
5. Connect the
DC
current
to
be measured
to
the
IN-
PUT connectors.
The
polarity
of
the
current
is
sensed by the
7D13
cir-
cuitry and automatically displayed. The polarity symbol
@
Operating
Instructions-7D13
displayed
is
dependent
upon the relative voltage of the
two
INPUT connectors. If the
HIGH
input
connector
is
positive
with respect
to
the
COM
input connector, a
"+"
will be
displayed and vice versa. For current measurements, a cur-
rent (electron flow) into the
COM
connector and
out
of
the
HIGH
connector will result
in
a
"+"
being displayed.
Temperature Measurements
The 7D13,
in
conjunction with a temperature sensing
probe such
as
the P6058, can measure temperatures from
-55°C
to
+
150°C
in
one range.
To
measure temperature, use the following procedure:
1. Install the 7D13
in
any available plug-in compart-
ment
of
a Tektronix 7000-series oscilloscope.
2. Turn the oscilloscope power on. Allow
twenty
minutes warmup.
3.
Advance the oscilloscope Readout control
to
obtain a
usable readout display (if necessary, refer
to
the oscillo-
scope instruction manual).
4. Connect the temperature-sensing probe
to
the front-
panel PROBE connector.
~
When connecting the temperature probe to the
PROBE connector, the two connectors
must
be
cor-
rectly aligned.
Damage
to the terminals can result
from forcing the connector and jack together. Refer
to the probe instruction manual.
5.
Set
the
MODE/RANGE switch
to
TEMPERATURE.
6. Apply the probe sensor tip
to
the device being mea-
sured. For optimum temperature transfer, the surface of
the device being measured should be coated with silicon
grease and the probe tip should be applied squarely
to
the
surface.
7. Allow a sufficient
amount
of
time for the probe tip
to
"settle"
before taking a reading.
The
time required
depends upon several factors. Generally, when the tip
is
first applied
to
the device under test, the readings will
change rapidly. As the probe tip temperature approaches
the temperature
of
the device under test,
the
reading will
stabilize
or
"settle".
Using
a Transistor
as
a Temperature-Sensing Device
Certain
NPN
transistors such as a 2N2484 can be used as
a separate sensor
in
place of the probe tip with little or no
selection
of
the device.
2-5