Mastech CSIMS8264 Assembly instructions

HANDHELD
DIGITAL MULTIMETER
OPERATOR’S
INSTRUCTION MANUAL

Table of Contents
TITLE PAGE
1. GENERAL INSTRUCTIONS 1
1.1 Precaution safety measures 1
1.1.1 Preliminary 1
1.1.2 During use 2
1.1.3 Symbols 4
1.1.4 Instructions 5
1.2 Protection mechanisms 6
2. DESCRIPTION 6
2.1 Instrument Familiarization 6
2.2 LCD Display 7
2.3 Key pad 8
2.4 Terminals
9
2.5 Accessories
9
3. FUNCTION DESCRIPTION 10
3.1 General Functions 10
3.1.1 DATA HOLD mode 10
3.1.2 Battery saver 10
3.2 Measurement Functions 10
3.2.1 AC and DC Voltage measurement 10
3.2.2 Resistance measurement 11
3.2.3 Diode Test 12
3.2.4 Continuity Check 13
Ⅰ
Table of Contents
TITLE PAGE
3.2.5 Capacitance measurement 14
3.2.6 Transistor measurement 15
3.2.7 Frequency measurement 15
3.2.8 Temperature measurement 16
3.2.9 Current measurement 16
4. TECHNICAL SPECIFICATIONS 18
4.1 General specifications.. 18
4.2 Measurement specifications 19
4.2.1 DC Voltage 19
4.2.2 AC Voltage 19
4.2.3 Frequency 19
4.2.4 Resistance 20
4.2.5 Diode Test 20
4.2.6 Continuity Check 20
4.2.7 Transistor 20
4.2.8 Temperature 20
4.2.9 Capacitance 21
4.2.10 DC Current 21
4.2.11 AC Current 21
5. MAINTENANCE 22
5.1 General maintenance 22
5.2 Battery replacement 22
Ⅱ

1. GENERAL INSTRUCTIONS
This instrument complies with IEC 1010-1 (61010-1@IEC:
2001), CAT. II 1000V and CAT. III 600V overvoltage
standards. See Specifications.
To get the best service from this instrument, read carefully
this user's manual and respect the detailed safety
precautions.
International symbols used on the Meter and in this manual
are explained in chapter 1.1.3
1.1 Precautions safety measures
1.1.1 Preliminary
* Measurement category III is for measurements performed
in the building installation.
NOTE: Examples are measurements on distribution boards,
circuit-breakers, wiring, including cables, bus-bars, junction
boxes, switches, socket-outlets in the fixed installation, and
equipment for industrial use and some other equipment, for
example, stationary motors with permanent connection to
the fixed installation.
* Measurement category II is for measurements performed
on circuits directly connected to the low voltage installation.
NOTE: Examples are measurements on household
appliances, portable tools and similar equipment.
* Measurement category I is for measurements performed on
circuits not directly connected to MAINS.
NOTE: Examples are measurements on circuits not
derived from MAINS, and specially protected (internal)
MAINS derived circuits. In the latter case, transient stresses
1
are variable; for that reason, requires that the transient
withstand capability of the equipment is made known to the
user.
*When using this Multimeter, the user must observe all
normal safety rules concerning:
―protection against the dangers of electric current.
―protection of the Multimeter against misuse.
*For your own safety, only use the test probes supplied with
the instrument. Before use, check that they are in good
condition.
1.1.2 During use
* If the meter is used near noise generating equipment, be
aware that display may become unstable or indicate large
errors.
* Do not use the meter or test leads if they look damaged.
* Use the meter only as specified in this manual; otherwise,
the protection provided by the meter may be impaired.
* Use extreme caution when working around bare conductors
or bus bars.
* Do not operate the meter around explosive gas, vapor, or dust.
* Verify a Meter's operation by measuring a known voltage.
Do not use the Meter if it operates abnormally. Protection
may be impaired. When in doubt, have the Meter
serviced.
* Uses the proper terminals, function, and range for your
measurements.
* When the range of the value to be measured is unknown,
check that the range initially set on the multimeter is the
highest possible or, wherever possible, choose the
autoranging mode. 2

* To avoid damages to the instrument, do not exceed the
maximum limits of the input values shown in the technical
specification tables.
* When the multimeter is linked to measurement circuits, do
not touch unused terminals.
* Caution when working with voltages above 60Vdc or 30Vac
rms. Such voltages pose a shock hazard.
* When using the probes, keep your fingers behind the finger
guards.
* When making connections, connect the common test lead
before connecting the live test lead; when disconnecting,
disconnect the live test lead before disconnecting the
common test lead.
* Before changing functions, disconnect the test leads from
the circuit under test.
* For all dc functions, including manual or auto-ranging, to
avoid the risk of shock due to possible improper reading,
verify the presence of any ac voltages by first using the ac
function. Then select a dc voltage range equal to or
greater than the ac range.
* Disconnect circuits power and discharge all high-voltage
capacitors before testing resistance, continuity, diodes, or
capacitance.
* Never perform resistance or continuity measurements on
live circuits.
* Before measuring current, check the meter's fuse and turn
off power to the circuit before connecting the meter to the
circuit.
3
*In TV repair work, or when carrying out measurements on
power switching circuits, remember that high amplitude
voltage pulses at the test points can damage the multimeter.
Use of a TV filter will attenuate any such pulses.
* Use the 9V NEDA battery, properly installed in the Meter's
battery case, to power the Meter.
* Replace the battery as soon as the battery indicator ( )
appears. With a low battery, the Meter might produce
false readings that can lead to electric shock and personal
injury.
* Do not measure voltages above 600V in Category III, or
1000V in Category II installations.
* Do not operate the Meter with the case (or part of the case)
removed.
1.1.3 Symbols:
Symbols used in this manual and on the instrument:
Caution:refer to the instruction manual.
Incorrect use may result in damage to the device
or its components.
~ AC (Alternating Current)
DC (Direct Current)
Earth ground
Double insulated
Fuse
Conforms to European Union directives
4

1.1.4 Instructions
* Remove test leads from the Meter before opening the
Meter case or battery cover.
* When servicing the Meter, use only specified replacement
parts.
* Before opening up the instrument, always disconnect from
all sources of electric current and make sure you are not
charged with static electricity, which may destroy internal
components.
*Any adjustment, maintenance or repair work carried out on
the meter while it is live should be carried out only by
appropriately qualified personnel, after having taken into
account the instructions in this present manual.
*A "qualified person" is someone who is familiar with the
installation, construction and operation of the equipment
and the hazards involved. He is trained and authorized to
energize and de-energize circuits and equipment in
accordance with established practices.
*When the instrument is opened up, remember that some
internal capacitors can retain a dangerous potential even
after the instrument is switched off.
*If any faults or abnormalities are observed, take the
instrument out of service and ensure that it cannot be used
until it has been checked out.
*If the meter is not going to be used for a long time, take out
the battery and do not store the meter in high temperature
or high humidity environment.
5
1.2 Protection mechanisms
*Fused by the resettable fuse (F200mA/250V) during
capacitance, temperature, mAand hFE measurements.
*A PTC resistor protects against permanent overvoltages of
up to 380V during resistance, Frequency, continuity and
diode test measurements.
2. DESCRIPTION
2.1 Instrument Familiarization
1
2
3
4
1. LCD display 2. Keypad 3. Rotary switch 4. Terminals
6

2.2 LCD Display
See Table 1 indicated for information about the LCD display.
Figure 1.Display
Table 1. Display Symbols
Symbol Meaning
The battery is low.
Warning: To avoid false readings,
which could lead to possible electric shock
or personal injury, replace the battery as
soon as the battery indicator appears.
Indicates negative readings.
Indicator for ac voltage or current.
AC voltage and current are displayed as the
average of the absolute value of the input,
calibrated to indicate the equivalent rms
value of a sine wave.
Indicator for dc voltage or current.
The Meter is in the Diode Test mode
7
Table 1. Display Symbols (continued)
The Meter is in the Continuity Check mode.
The Meter is in the Data Hold mode
℃ or ℉ ℃: Celsius scale. The unit of temperature.
℉: Fahrenheit scale.
V, mV V:
mV: Volts. The unit of voltage.
Millivolt. 1x10-3 or 0.001 volts.
A, mA, μAA:
mA:
μA:
Amperes (amps). The unit of current.
Milliamp. 1x10-3 or 0.001 amperes.
Microamp. 1x10-6 or 0.000001 amperes
Ω, kΩ, MΩ
Ω:
kΩ:
MΩ:
Ohm. The unit of resistance.
Kilohm. 1x103or 1000 ohms.
Megohm. 1x106or 1,000,000 ohms.
kHz KHz: Kilohertz. 1x103or 1000 hertz.
μF, nF
F:
μF:
nF:
Farad. The unit of capacitance.
Microfarad.1x10-6 or 0.000001 farads.
Nanofarad. 1x10-9 or 0.000000001
farads.
2.3 Keypad
See Table 2 indicated for information about the keypad
operations. Table 2. Keypad
Key Function Operation performed
ON/OFF Any switch position turn the meter on or off
HOLD Any switch position Press HOLD to enter and
exit the Data Hold mode.
LIGHT Any switch position Press LIGHT to turn the
backlight on. After about
5 seconds, the backlight
is auto-off.
8

2.4 Terminals
See Table 4 indicated for information about the terminals.
Table 4. Terminals
Terminal Description
COM Return terminal for all measurements.
(Receiving the black test lead or the “com”
plug of the special multi-function socket)
VΩHz
Input for voltage, resistance, frequency, diode
and continuity measurements. (Receiving the
red test lead)
℃mA
hFE
Input for capacitance, Temperature, hFE and
0.001mA to 200mA current measurements.
(Receiving the red test lead or the “+” plug of
the special multi-function socket)
10A Input for 200mAto 10Acurrent measurements.
(Receiving the red test lead)
2.5 Accessories
Delivered with the multimeter:
¥User's manual One piece
¥Test leads One piece
¥Carry case One piece
¥"K" type bead Thermocouple One piece
¥Special Multi-function socket One piece
9

3. FUNCTION DESCRIPTION
3.1 General Functions
3.1.1 DATA HOLD mode
Data Hold mode makes the meter stop updating the
display. Data Hold function can be cancelled by changing
the measurement mode, or push HOLD key again.
To enter and exit the Data Hold mode:
1. Press HOLD key. Fixes the display on the current
value, is displayed.
2.A second short press returns the meter to normal mode.
3.1.2 Battery Saver
Turn on the meter. And then The Meter will be turned off
automatic after approx. 30 minutes.
3.2 Measurement Functions
3.2.1 AC and DC Voltage measurement
To avoid electrical shock and/or damage to the
instrument, do not attempt to take any voltage
measurement that might exceeds 1000Vdc or
750Vac rms.
To avoid electrical shock and/or damage to the
instrument, do not apply more than 1000Vdc or
750Vac rms between the common terminal and
the earth ground.
Voltage is the difference in electrical potential between two
points.
The polarity of ac (alternating current) voltage varies over
time; the polarity of dc (direct current) voltage is constant.
The Meter's DC voltage ranges are 200.0mV, 2.000V,
20.00V, 200.0V and 1000V;AC voltage ranges are 2.000V,
20.00V, 200.0V and 7500V. 10
To measure ac or dc voltage:
1. Set rotary switch to the proper range.
2. Connect the black and red test leads to the COM and
V terminals respectively.
3. Connect the test leads to the circuit being measured
4. Read the displayed value. The polarity of red test lead
connection will be indicated when making a DCV
measurement.
NOTE:
¥Unstable display may occur especially at DC200mV and
AC2V ranges, even though you do not put test leads into
input terminals, in this case, if an erroneous reading is
suspected, short the V terminal and the COM terminal,
and make sure the zero display.
3.2.2 Resistance measurement
To avoid electrical shock and/or damage to the
instrument, disconnect circuit power and
discharge all high-voltage capacitors before
measuring resistance.
Resistance is an opposition to current flow.
The unit of resistance is the ohm (Ω).
The Meter's resistance ranges are 200.0Ω, 2.000kΩ,
20.00kΩ, 200.0kΩ, 2.000MΩ, 20.00MΩand 200.0MΩ.
To measure resistance:
1. Set the rotary switch to proper range.
2. Connect the black and red test leads to the COM and
Ωterminals respectively.
3. Connect the test leads to the circuit being measured
and read the displayed value.
11
Some tips for measuring resistance:

¥The measured value of a resistor in a circuit is often
different from the resistor's rated value. This is because
the Meter's test current flows through all possible paths
between the probe tips.
¥In order to ensure the best accuracy in measurement of low
resistance, short the test leads before measurement and
memory the test probe resistance in mind. This necessary
to subtract for the resistance of the test leads.
¥The resistance function can produce enough voltage to
forward-bias silicon diode or transistor junctions, causing
them to conduct. To avoid this, do not use the 40MΩ
range for in-circuit resistance measurements.
¥On 20MΩand 200MΩranges, the meter may take a few
seconds to stabilize reading. This is normal for high
resistance measuring.
¥On 200MΩrange, the display is approx. 10 digits when test
leads are shorted. These 10 digits have to be subtracted
from measuring results. For example, when measuring 100
MΩresistance, the reading will be 101.0 and the correct
measuring result should be 101.0-1.0=100.0 MΩ.
¥When the input is not connected, i.e. at open circuit, the
figure "1" will be displayed for the overrange condition.
3.2.3 Diode Test
To avoid electrical shock and/or damage to the
instrument, disconnect circuit power and
discharge all high-voltage capacitors before
testing diodes.
Use the diode test to check diodes, and other semi-
conductor devices. The diode test sends a current through
the semiconductor junction, and then measures the voltage
drop across the junction; a good silicon junction drops
between 0.5V and 0.8V. 12
To test a diode out of a circuit:
1. Set the rotary switch to range.
2. Connect the black and red test leads to the COM and
terminals respectively.
3. For forward-bias readings on any semiconductor
component, place the red test lead on the
component's anode and place the black test lead on
the component's cathode.
4. The meter will show the approx. forward voltage of the
diode. If the test lead connection is reversed, only
figure "1" displayed.
In a circuit, a good diode should still produce a forward bias
reading of 0.5V to 0.8V; however, the reverse-bias reading
can vary depending on the resistance of other pathways
between the probe tips.
3.2.4 Continuity Check
To avoid electrical shock and/or damage to the
instrument, disconnect circuit power and
discharge all high-voltage capacitors before
testing for Continuity.
Continuity is a complete path for current flow.
The beeper sounds if a circuit is complete. These brief
contacts cause the Meter to emit a short beep.
To test for continuity:
1. Set the rotary switch to range.
2. Press the yellow key twice to activate Continuity Check.
3. Connect the black and red test leads to the COM and
Ωterminals respectively.
4. Connect the test leads to the resistance in the circuit
being measured. 13
5. When the test lead to the circuit is below approx. 30Ω,
a continuous beeping will indicate it.

Note:
¥Continuity test is available to check open/short of the
circuit.
3.2.5 Capacitance measurement
To avoid electrical shock and/or damage to the
instrument, disconnect circuit power and
discharge all high-voltage capacitors before
measuring capacitance. Use the dc voltage
function to confirm that the capacitor is
discharged.
Capacitance is the ability of a component to store an
electrical charge.
The unit of capacitance is the farad (F). Most capacitors are
in the nanofarad to microfarad range.
The Meter's capacitance ranges are 20.00nF, 200.0nF,
2.000μF, and 20.00μF.
To measure capacitance:
1. Set the rotary switch to proper range.
2. Connect the black and red test leads to the COM and
terminals respectively (or you can measure the
capacitance by using the special Multi-Function Socket).
3. Connect the test leads to the capacitor being measured
and read the displayed value.
Some tips for measuring capacitance:
¥The meter may take a few seconds to stabilize reading.
This is normal for high capacitance measuring.
¥To improve the accuracy of measurements less than 20nF,
subtract the residual capacitance of the Meter and leads.
14
3.2.6 Transistor measurement
To avoid electrical shock and/or damage to the
instrument, do not apply more than 250Vdc or
250Vac rms between the hFE
terminal and the
COM terminal.
1. Set the rotary switch to hFE range.
2. Connect the “com” plug and “+” plug of the special
multi-function socket to the COM and hFE terminals.
3. Determine whether the transistor to be tested is NPN or
PNP type and locate the Emitter, Base and Collector leads.
4. Insert leads of the transistor into proper holes of the
special multi-function socket.
5. The meter will show the approx. hFE value at test
condition of base current 10μA and Vce 2.8V.
3.2.7 Frequency measurement
Do not measure Frequency on high voltage
(>380V) to avoid electrical shock hazard and/or
damage to the instrument.
Frequency is the number of cycles a voltage or current
signal completes each second.
To measure frequency:
1. Set the rotary switch to 20kHz range.
2. Connect the black and red test leads to the COM and
Hz terminals respectively.
3. Connect the test leads to the circuit being measured
4. Read the displayed value.
15
3.2.8 Temperature measurement

To avoid electrical shock and/or damage to the
instrument, do not apply more than 250Vdc or
250Vac rms between the
℃
terminal and the
COM terminal.
To avoid electrical shock, do not use this instrument
when voltages at the measurement surface exceed
60v dc or 24v rms. Ac.
To avoid damage or burns. Do not make
temperature measurements in microwave ovens.
To measure temperature:
1. Set the rotary switch to ℃ range and the LCD will show
the current environment temperature.
2. Connect the “com” plug and “+” plug of the special
multi-function socket to the COM and hFE terminals.
3. Insert ‘K’ type thermocouples into the special multi-function
socket,Takings care to observe the correct polarity.
4. Touch the object with the thermocouple probe for
measurement.
5. Read the stable reading from LCD.
3.2.9 Current measurement
To avoid damage to the Meter or injury if the fuse
blows, never attempt an in-circuit current
measurement where the open-circuit potential to
earth is greater than 250V.
To avoid damage to the meter, check the meter's
fuse before proceeding. Use the proper terminals,
function, and range for your measurement. Never
place the test leads in parallel with a circuit or
component when the leads are plugged into the
current terminals. 16
Current is the flow of electrons through a conductor.
The Meter's DC current ranges are 20.00mA, 200.0mA
and 10.00A;AC current ranges are 2.000mA, 200.0mA
and 10.00A.
To measure current:
1. Turn off the power of the measured circuit. Discharge all
the high voltage capacitors.
2. Set the rotary switch to the proper range.
3. Connect the black test lead to the COM terminal and
the red test leads to the mA terminal for a maximum of
200mA. For a maximum of 10A, move the red test lead
to the 10A terminal.
4. Break the circuit path to be tested.
Connect the black test lead to the more negative side of
the break; connect the red test lead to the more positive
side of the break. (Reversing the leads will give a
negative reading, but will not damage the Meter.)
5. Turn on the power of the measured circuit, and then
read the display. Be sure to note the measurement
units at the right side of the display (mA or A). When
only the figure "1" displayed, it indicates overrange
situation and the higher range has to be selected.
6. Turn off the power of the measured circuit and discharge
all the high voltage capacitors. Remove the test leads and
recover the measured circuit.
17
4 TECHNICAL SPECIFICATIONS

4.1 GENERAL SPECIFICATIONS
¥Environment conditions:
1000V CAT. II and 600V CAT. III
Pollution degree: 2
Altitude < 2000m
Operating temperature:
0~40℃, 32℉~122℉(<80% RH, <10℃noncondensing)
Storage temperature:
-10~60 ℃, 14℉~140℉(<70% RH, battery removed)
¥Temperature Coefficient:
0.1×(specified accuracy) / ℃(<18℃or >28℃)
¥MAX. Voltage between terminals and earth ground:
750V AC rms or 1000V DC.
¥Fuse Protection: mA: Resettable fuse (F200mA/250V)
¥Sample Rate: 3 times/sec for digital data.
¥Display: 3 1/2 digits LCD display. Automatic indication of
functions and symbols.
¥Over Range indication: LCD will display "1".
¥Low battery indication:
The " " is displayed when the battery is under the
proper operation range.
¥Polarity indication: "−" displayed automatically.
¥Power source: 9V
¥Battery type: NEDA1604, 6F22, or 006P.
¥Dimensions: 195(L)×92(W)×55(H) mm.
¥Weight: 380g. Approx. (battery included).
18
4.2 Measurement specifications
Accuracy is specified for one year after calibration, at
operating temperatures of 18 ℃to 28 ℃, with relative
humidity at 0% to 75%.
Accuracy specifications take the form of: ±(% of Reading +
Number of Least Significant Digits)
4.2.1 DC Voltage
Range Resolution Accuracy
200mV 0.1mV
2V 1mV
20V 10mV
200V 100mV
±(0.5% of rdg +1 digit)
1000V 1V ±(0.8% of rdg +2 digits)
Input impedance: 10MΩ
Max. input voltage: 250Vdc or ac rms for 200mV range and
1000Vdc or 750V ac rms for other ranges,
4.2.2 AC Voltage
Range Resolution Accuracy
2V 1mV
20V 10mV
200V 100mV ±(0.8% of rdg +3 digits)
750V 1V ±(1.2% of rdg +3 digits)
Input impedance: 10MΩ
Max. input voltage: 250Vdc or ac rms for 200mV range and
1000Vdc or 750V ac rms for other ranges,
Frequency Range: 40Hz-200Hz for 750V range, 40Hz-400Hz
for other ranges.
Response:Average, calibrated in rms of sine wave
4.2.3 Frequency
Range Resolution Accuracy
20kHz 10 Hz ±(1.5% of rdg+5 digits)
Overload protection: 380V dc or 380V ac rms.
Input Voltage range: 200mV-10V ac rms
19
4.2.4 Resistance

Range Resolution Accuracy
200Ω0.1Ω±(0.8% of rdg+3 digits)
2kΩ1Ω
20kΩ10Ω
200kΩ100Ω
2MΩ1kΩ
±(0.8% of rdg+1 digit)
20MΩ10kΩ±(1.0% of rdg+2 digits)
200MΩ0.1MΩ±[5.0% of (rdg-10digits) +10 digits]
Overload protection: 380V dc or 380Vac rms.
Open Circuit Voltage: Less than 700mV.
4.2.5 Diode
Range Resolution Function
1mV
Display read approx. forward
voltage of diode
Forward DC Current: approx. 1mA
Reversed DC Voltage: approx. 2.8V
Overload protection: 380Vdc or 380Vac rms.
4.2.6 Audible continuity
Range Continuity beeper
≤30Ω
Open circuit voltage: Less than 700mV.
Overload protection: 380Vdc or 380Vac rms.
4.2.7 Transistor
Range Description Test Condition
hFE
Display read approx.
HFE value (0-1000) of
transistor under test (all
type).
Base Current approx.
10μA, Vce approx.
2.8V.
4.2.8 Temperature
Range Resolution Accuracy
-20℃~0℃±(5.0% of rdg+4 digits)
1℃~400℃±(1.0% of rdg+3 digits)
401℃~1000℃1℃
±2.0% of rdg
20
Overload protection: Resettable fuse (F200mA/250V).
* Temperature specifications do not include thermocouple
errors.
4.2.9 Capacitance
Range Resolution Accuracy
2nF 1pF
20nF 10pF
200nF 0.1nF
2μF 1nF
20μF 10nF
±(4.0% of rdg+3 digits)
Overload protection: Resettable fuse (F200mA/250V).
4.2.10 DC Current
Range Resolution Accuracy
20mA 10μA
200mA 0.1mA ±(1.5% of rdg+1 digit)
10A 10mA ±(2.0% of rdg+5 digits)
Overload protection: Resettable fuse (F200mA/250V). 10A
range unfused
Max. input current: 200mAdc or 200mA ac rms for mArange,
10A dc or 10A ac rms for 10A ranges.
For measurements>5A, 4 minutes maximum ON to measure
10 minutes OFF; Above 10Aunspecified.
4.2.11 AC Current
Range Resolution Accuracy
2mA 1μA ±(1.0% of rdg+3 digits)
200mA 0.1mA ±(1.8% of rdg+3 digits)
10A 10mA ±(3.0% of rdg+7 digits)
Overload protection: Resettable fuse (F200mA/250V). 10A
range unfused.
Max. input current: 200mAdc or 200mA ac rms for mArange,
10A dc or 10A ac rms for 10A ranges.
For measurements>5A, 4 minutes maximum ON to measure
10 minutes OFF; Above 10Aunspecified.
Frequency Range: 40Hz-400Hz
Response:Average, calibrated in rms of sine wave
21
5. MAINTENANCE

Do not attempt to repair or service your Meter unless you are
qualified to do so and have the relevant calibration,
performance test, and service information.
5.1 General Maintenance
To avoid electrical shock or damage to the meter,
do not get water inside the case. Remove the test
leads and any input signals before opening the
case
Periodically wipe the case with a damp cloth and mild
detergent. Do not use abrasives or solvents.
Dirt or moisture in the terminals can affect readings.
To clean the terminals:
¥Turn the meter off and remove all test leads.
¥Shake out any dirt that may be in the terminals.
¥Soak a new swab with a cleaning and oiling agent (such as
WD-40).
¥Work the swab around in each terminal. The oiling agent
insulates the terminals from moisture-related contamination.
5.2 Battery replacement
To avoid false readings, which could lead to
possible electric shock or personal injury, replace
the battery as soon as the battery indicator ( )
appears.
Before replacing the battery, disconnect test leads
and/or any connectors from any circuit under test,
turn the meter off and remove test leads from the
input terminals.
To replace the battery (see Figure 2.):
¥Turn the meter off.
¥Disconnect test leads and/or any connectors from the
terminals. 22
¥Use a screwdriver to unscrew the two screws secured on
the battery cover.
¥Take out the battery cover from the meter.
¥Remove the used batteries.
¥Replace with the new 9V batteries (6F22).
¥Rejoin the battery cover and secure by the two screws.
9V Battery
Battery cover
Figure 2. Battery and Fuse Replacement
CAUTION
Using this appliance in an environment with a strong
radiated radio-frequency electromagnetic field (approx.
3V/m), may influence its measuring accuracy. The
measuring result can be strongly deviating from the
actual value. 23

HYS005517
Table of contents
Other Mastech Multimeter manuals

Mastech
Mastech MS8253 Series User manual

Mastech
Mastech MS8236 User manual

Mastech
Mastech MS8238C User manual

Mastech
Mastech ms8261 Assembly instructions

Mastech
Mastech MS2208 User manual

Mastech
Mastech MS8361D User manual

Mastech
Mastech MS8240A User manual

Mastech
Mastech MS8302F User manual

Mastech
Mastech MS8230B User manual

Mastech
Mastech ms8217 User manual

Mastech
Mastech m3900 User manual

Mastech
Mastech MS8251A User manual

Mastech
Mastech MY70 User manual

Mastech
Mastech MY-65 User manual

Mastech
Mastech MY-64 User manual

Mastech
Mastech my68 User manual

Mastech
Mastech MS8235 User manual

Mastech
Mastech MS2203 User manual

Mastech
Mastech MS8269 User manual

Mastech
Mastech MS8211 User manual