TELEVAC MC 300 2A User manual

INSTRUCTION MANUAL
MC 300 2A & 4A
DUAL CHANNEL THERMOCOUPLE / CONVECTION
VACUUM MEASUREMENT INSTRUMENT
TELEVAC
A DIVISION OF THE FREDERICKS COMPANY
2400 PHILMONT AVENUE,
HUNTINGDONVALLEY, PA 19006
Phone:(215) 947-2500 fax:(215) 947-7464 e-mail: [email protected] web site:www.televac.com
6-8000-002 REV A
1

PARTS LIST
1
32
6
MOUNTING
KIT
5
4
# QTY ITEM DESCRIPTION
1 1 MC300 Thermocouple, Convection, Measurement Instrument
2 1 Instruction Manual
3 3 Sensor Cables (sold separately)
4 4 Sensors (sold separately)
5 1 Mounting kit
6 1 AC Power Cable
Phone:(215) 947-2500 fax:(215) 947-7464 e-mail: [email protected] web site:www.televac.com
6-8000-002 REV A
2

TABLE OF CONTENTS
SECTION & TITLE
01. DESCRIPTION
02. SAFETY INSTRUCTIONS
03. PRINCIPLES OF OPERATION
04. INSTALLATION & OPERATION
05. ANALOG RECORDER OUTPUTS
06. MAINTENANCE
07. CALIBRATION
08. CALIBRATION VERIFICATION
09. TROUBLESHOOTING
10. DIMENSION
11. SPECIFICATIONS
12. UNPACKING & INSPECTION
13. WARRANTY INFORMATION
14. ORDERING INFORMATION
PLEASE READ THIS MANUAL THOROUGHLY BEFORE USING THIS
INSTRUMENT. REPORT ANY PROBLEMS IMMEDIATELY.
Phone:(215) 947-2500 fax:(215) 947-7464 e-mail: [email protected] web site:www.televac.com
6-8000-002 REV A
3

CHAPTER 1 TELEVAC
Description
The TELEVAC Series MC300 vacuum-measuring instrument is a combination sensor,
which incorporates various range options, as shown in the table below:
Range of Options
Model Range Sensors
MC300 2A 0 - 20,000 millitorr 2AThermocouple
MC4A 0 – Atmosphere 4A Convection
The 2A thermocouple gauge tube will monitor pressures between 1 millitorr
(1 micron) and 20,000 microns ).
The 4A convection sensor tube will monitor pressures between 1 millitorr (1
micron) and 999 Torr. (1 atmosphere = 760 Torr @ sea level ).
The vacuum measuring system consists of various gauge tubes, which sense the vacuum,
an electronics unit and interconnecting signal/power cables. The customer must supply
power at 115VAC 50-60 Hz. (220 Volt operation is a factory selectable option). The
measured vacuum is displayed on the front panel LED displays. (See figure 1)
The power cable, gauge tubes, relay outputs and output voltages of 0 to 10 VDC
connections are on the back panel. (See figure 2)
Phone:(215) 947-2500 fax:(215) 947-7464 e-mail: [email protected] web site:www.televac.com
6-8000-002 REV A
4

CHAPTER 1(cont) TELEVAC
Figure 1
Figure 2
The MC300 FEATURES
Large bright LEDs
Built-in self-diagnostics
Selectable Units (Torr, microns, mBar or Pascal)
Gas compensation (Argon/Air-Nitrogen)
Four process relays with individual assignable set points
Analog outputs for each sensor
Operator lockout switch
Phone:(215) 947-2500 fax:(215) 947-7464 e-mail: [email protected] web site:www.televac.com
6-8000-002 REV A
5

CHAPTER 1 (cont) TELEVAC
Theory of Operation:
The MC300 is a microprocessor-controlled instrument utilizing modern data acquisition
techniques to achieve a high degree of accuracy at a reasonable cost.
The microprocessor in the instrument performs two basic tasks:
1. Controls the acquisition and conversion of all sensor data
2. Processes all operator interface inputs and requests
After power is applied to the instrument, the software initializes all of the circuitry
associated with acquiring and displaying sensor information then performs a self-
diagnostic. If there are no errors detected, the MC300’s data acquisition process is
started. Both input channels are read periodically and a reading for each sensor is stored.
After several readings the data is averaged, then converted using the appropriate sensor,
units and gas compensation settings. Set point relays and the microprocessor controlled
Analog outputs are also updated at the same time. This method of averaging samples
results in a smooth and accurate transition from reading to reading.
Each sensor has its own conversion algorithm. The data conversion process always
occurs regardless of the instrument mode (operation set-up). However, during the set-up
mode the converted data is not sent to the display but the set-point relay control and
analog output operations continue to operate.
Whenever the instrument software is not processing sensor data for the displays or
recorder outputs, the Set-Up, Test/Operate and Up & Down (if enabled) buttons are
checked for operator interaction. The Set-Up button is used to sequence through all of the
operator settable parameters. Each depression of the Set-Up button may change the
function of the displays as well as the Up & Down buttons. Each time the Set-Up button
is pressed, the data from the previous level is stored in non-volatile memory to ensure
that the instrument’s gas compensation, units and set point parameters are preserved. This
information is used to restore the instrument’s configuration each time power is applied.
For simplicity, the Up & Down buttons are enabled and disabled at the proper time to
prevent the operator from entering an invalid value. In addition, each time an On Point is
adjusted, the corresponding Off point is verified to ensure that it is higher then the new
On Point. If lower, the Off Point is adjusted to be one increment higher than the new On
Point. In this instance both the On and the Off Point values are updated in the non-
volatile memory.
When “GAS” is selected in SETUP, a correction is applied to the instrument readings to
compensate for the use of Argon as a gas instead of nitrogen/air. Both the displayed
readings and the set point calibrations are compensated as well as the linear thermocouple
and the linear by decade convection gauge recorder outputs at the back panel. Note,
however, that the “traditional” thermocouple, “linear” convection gauge outputs and ion
gauge outputs are NOT compensated.
Phone:(215) 947-2500 fax:(215) 947-7464 e-mail: [email protected] web site:www.televac.com
6-8000-002 REV A
6

CHAPTER 2 TELEVAC
Safety Instructions
START BY READING THESE IMPORTANT SAFETY INSTRUCTIONS AND
NOTES collected here for your convenience and repeated with additional information at
appropriate points in these instructions.
These safety alert symbols in this manual or on the Product rear panel mean
cautions - personal safety, property damage or danger from electrical shock.
Read these instructions carefully.
In these instructions the word “product” refers to the MC300 and all of its approved parts
and accessories.
NOTE: These instructions do not and cannot provide for every contingency that may
arise in connection with the installation, operation, or maintenance of this product.
Should you require further assistance, please contact Televac at the address on the title
page of this manual.
This product has been designed and tested to offer reasonably safe service provided
in it’s installed, operated and serviced in strict accordance with these safety
instructions.
Failure to comply with these instructions may result in serious personal injury, including
death, or property damage.
These safety precautions must be observed during all phases of operation,
installation, and service of this product.
Failure to comply with these precautions or with specific warnings elsewhere in
this manual violates safety standards of design, manufacture, and intended use of
the instrument. Televac disclaims all liability for the customer’s failure to comply
with these requirements.
Phone:(215) 947-2500 fax:(215) 947-7464 e-mail: [email protected] web site:www.televac.com
6-8000-002 REV A
7

CHAPTER 2(cont.) TELEVAC
The service and repair information in this manual is for the use of Qualified Service
Personnel. To avoid shock, do not perform any procedures in this manual or perform any
Servicing on this product unless you are qualified to do so.
READ Instructions:
Read all safety and operating instructions before operating the product.
RETAIN instructions:
Retain the Safety and Operating Instructions for future reference.
HEED warnings:
Adhere to all warnings on the product and in the operating instructions.
FOLLOW instructions:
Follow all operating and maintenance instructions.
ACCESSORIES:
Do not use accessories not recommended in this manual as they may require a technician
to restore the product to its normal operation.
To reduce risk of fire or electric shock, do not expose this product to rain or moisture.
`
Objects and Liquid Entry – Never push objects of any kind into this product through
openings as they may touch dangerous voltage points or short out parts that could result in
a fire or electric shock. Be careful not to spill liquid of any kind onto the products.
Do not substitute parts or modify instrument.
Because of the danger on introducing additional hazards, do not install substitute parts or
perform any unauthorized modifications to the product. Return the product to Televac fo
r
service and repair to ensure that safety features are maintained. Do not use this product
if it has unauthorized modifications.
Phone:(215) 947-2500 fax:(215) 947-7464 e-mail: [email protected] web site:www.televac.com
6-8000-002 REV A
8

CHAPTER 3 TELEVAC
Principles of Operation
2A Thermocouple Sensor
The 2A thermocouple sensor tube consists of a
metal shell containing fine wire filaments of
dissimilar metals, which form a thermocouple.
The thermocouple sensor measures absolute
pressures by determining the temperature of a
thermocouple junction as the surrounding
pressure changes. The response of the sensor
depends on the gas pressure and type.
The range of these sensors is from 1 millitorr to
20,000 millitorr (10-3 Torr to 20 Torr). The
sensor housing is constructed of stainless steel or
alternately nickel plated brass. Both are capable
of withstanding 150 psi overpressure.
Figure 3
2A Thermocouple Sensor Tube
Figure 4
4A Convection Sensor Tube
Convection Sensor
The convection sensor measures absolute pressures
by determining the heat loss from a thermocouple
heated to a constant temperature by a current source.
A second thermocouple compensates for ambient
temperature changes and extends the vacuum
measurement range above 2 Torr by measuring the
convection effects. The range of these sensors is 1
millitorr to 1000 Torr
The convection sensor must be mounted vertically
for accurate measurements. Sensors are available in
Stainless Steel or Nickel plated Brass and can
withstand 150-psi overpressure.
Phone:(215) 947-2500 fax:(215) 947-7464 e-mail: [email protected] web site:www.televac.com
6-8000-002 REV A
9

CHAPTER 4 TELEVAC
Installation and Operation
Instrument Installation
The MC300 is pre configured for immediate installation and use. The unit should be first
mounted in the desired location and properly secured. In choosing a location, the display
should be clearly visible by an operator and positioned in such a way that the operator
can easily make adjustments to the instrument. All wiring should be neatly dressed and
bundled with special care not to have any wiring adjacent to AC power distribution. After
proper installation of sensors and wiring, the unit can be connected to an AC power
source and put into operation.
IT IS RECOMMENDED THAT POWER IS NOT APPLIED TO THE INSTRUMENT
DURING THE INSTALLATION OF ANY WIRING AND SENSORS.
Sensor Installation
It is desirable that the thermocouple and cold cathode sensor tubes be mounted on the
vacuum system in a VERTICAL POSITION open end extending downward. The 4A
convection sensor MUST be installed in the VERTICAL POSITION.
The electronic unit is factory calibrated at time of order. To ensure proper operating
conditions it is necessary to use factory-manufactured cables.
Contamination of thermocouples sensor can be caused by the process occurring in the
vacuum vessel, or by pump fluid back streaming and is a critical factor affecting the
measurement of high vacuum.
Depending upon the degree of contamination and the accuracy required, it is
recommended that sensor be periodically replaced and that filters (with line of sight
baffles (P/N 2-2100-50) be used to protect thermocouple sensors.
Phone:(215) 947-2500 fax:(215) 947-7464 e-mail: [email protected] web site:www.televac.com
6-8000-002 REV A
10

CHAPTER 4(cont.) TELEVAC
The thermocouple and convection sensor tubes incorporate a 1/8” National Pipe Thread
(1/8” NPT) fitting for connection to the vacuum system and other mounting
configurations are available.
The instrument may be turned on at any time, since atmospheric pressure will not harm
the thermocouple sensor tube elements or cause overdrive on the metering circuit. A brief
15-minute warm up is required before vacuum measurements can be made within stated
tolerances.
Wiring the instrument
Sensor wiring
The MC300 requires sensors & cables (sold separately) for each channel. Certain levels
of calibration require a specific cable and sensor be used with channel 1 or 2. In this case
the cable and sensor are marked for use as a set and must be used as such. Otherwise the
cables can be used with either rough vacuum channel. To ensure measurement integrity
and protection from ambient electrical noise, ground lugs are provided on each sensor
cable shield. These lugs should be connected to the ground screws provided on the rear
panel.
Phone:(215) 947-2500 fax:(215) 947-7464 e-mail: [email protected] web site:www.televac.com
6-8000-002 REV A
11

CHAPTER 4(cont.) TELEVAC
Analog Output Wiring
Provisions are made to connect the instrument to a chart recorder or programmable logic
controller. Two different format signals are provided and are described in Section 5
Analog Outputs. In all cases follow the external device manufacturer’s instructions for
wiring and using these signals. Linear (LIN) output is computer generated so that the
voltage signal is plotted as a straight line as a function of pressure.
Traditional (TRDL) signal is one that represents the non-linear signal that is
representative of the actual raw sensor data. The 4A Convection version has
Linear/Decade and Linear outputs. By means of a factory settable internal jumper a full
range Traditional signal is available in lieu of the Linear/Decade recorder output. Please
refer to Section 5 Analog Output (4A) for a summary of the signals available
Relay Contact Wiring (Set point Relays)
There are four set point relays available for use to control ON/OFF or Open/close (binary
state) devices. The relays are assignable to any one of the vacuum signal input channels
through front panel software control.
There are three connections for each relay. They are NORMALLY CLOSED (NC),
COMMON (C) and NORMALLY OPEN (O). The operation mode of each is described
as follows.
NORMALLY CLOSED:
In this state there is an electrical continuity path between terminals Cand NC. This
continuity path is broken when the software set point values dictates an ON state for the
specific measurement channel. Continuity is returned when the OFF state is dictated by
software control. In the power off mode of the MC300 (due to failure of the main power
source or turning off the instrument), the relay will revert to this state.
NORMALLY OPEN:
In this state there is no electrical continuity path between terminals Cand NO. This Path
is established ONLY when the Instrument control software satisfies conditions for the
ON state. It will break continuity under the instrument control parameters for the OFF
condition or the power off mode of the MC300 (due to failure of the main power source
or turning off the instrument).
Phone:(215) 947-2500 fax:(215) 947-7464 e-mail: [email protected] web site:www.televac.com
6-8000-002 REV A
12

CHAPTER 4(cont.) TELEVAC
CAUTION:
In utilizing the process relays it is important not to exceed the relay’s power handling
capacity described in Section 10 Specifications and to fully understand the behavior of
the control relay state conditions. Failure to do so will result in premature relay or
equipment failure.
Operating the Instrument
This section describes the operation of a properly installed MC300 Instrument
Prior to using the instrument on a vacuum chamber, the unit can be set up in a test mode
using simulators to represent the vacuum gauges. This is done to gain familiarity with the
unit and to preset relay set points. The following describes operation of the instrument
with simulators or vacuum applied to the sensors.
POWER ON:
The instrument is turned on and off by means of a rear mounted power switch. The unit
will first display the Software version number and instrument type in a four-digit code.
The first two digits refer to the software version and the second two digits refer to the
Instrument Sensor type summarized below.
The instrument will then perform a self-diagnosis after which it will begin displaying
vacuum measurements. If the unit doesn’t pass the self-test, an error code will appear.
Pressing the “TEST OP ” button can clear the code, but proper operation of the
instrument may be curtailed. It is recommended not to use the instrument if the self
diagnosis fails. Contact the factory for instructions. Allow a minimum of 15 minutes
from an instrument “cold start” for the sensor readings to stabilize.
GAS CORRECTION FACTOR:
The instrument is calibrated for use in Nitrogen (dry air) but the MC300 has the ability to
adjust its pressure readings to compensate the thermal conductivity gauges for the use of
ARGON as a background gas. When Argon gas is selected, the GAS light will flash
during operation of the gauge.
Phone:(215) 947-2500 fax:(215) 947-7464 e-mail: [email protected] web site:www.televac.com
6-8000-002 REV A
13

CHAPTER 4(cont.) TELEVAC
To change the gas correction factor:
1) Assure the rear mounted operator lock out switch is in the “OFF” position
(the switch is in the down position)
2) Press the SETUP button once
3) The Nitrogen (N2) or Argon (Ar) display will flash
4) Press the UP or DOWN arrow buttons to select a gas correction, The flashing
display is the gas selected)
5) Press the TEST OP button to complete the selection
(Note: if the TEST OP switch is not pressed in 60 seconds the instrument will return to
normal operation and the change in gas correction is not accepted)
CHANGE MEASUREMENT UNITS:
The MC300 has the ability to display vacuum values in Pascals, MBAR or MICRONS
and TORR.
1) Assure the rear mounted operator lock out switch is in “OFF” position, (the
switch is in the down position)
2) Press the SETUP button once
3) Nitrogen (N2) and Argon (Ar) display will flash
4) Press the SETUP a second time
5) The Pascal, MBAR or TORR lamp will flash
6) Press the UP or DOWN arrow buttons to select the desired unit of measurement
7) Press the TEST OP button to complete the selection
(Note: if the TEST OP switch is not pressed in 60 seconds the instrument will return to
normal operation and the change in gas correction is not accepted)
ASSIGN SET POINT RELAYS AND VALUES:
The instrument has four relays that are assignable to any of the vacuum measurement
channels. Thermocouple (or Convection) Channel 1 is Channel 1 (CH1), thermocouple
(or Convection) Channel 2 is Channel 2 (CH2) and the Ion gauge is Channel 3 (CH3).
Setpoint SP 1 is relay #1, setpoint SP 2 is relay #2 etc.
Phone:(215) 947-2500 fax:(215) 947-7464 e-mail: [email protected] web site:www.televac.com
6-8000-002 REV A
14

CHAPTER 4(cont.) TELEVAC
1) Assure the rear mounted operator lock out switch is in the “OFF” position, (the
switch is in the down position)
2) Press the SETUP button once
3) The Nitrogen (N2) or Argon (Ar) display will flash
4) Press the SETUP a second time
The measurement units (PASCAL, MBAR or TORR)lamp will flash
5) Press the SETUP a third time
6) The set point SP 1 will appear in the left display, the channel number CH1, 2 or 3
will appear in the right display
7) Use the UP or DOWN arrows to select the channel number 1,2 or 3
8) Press the SETUP button again. “ON” will appear in the right display, the “ON”
value will appear in the lower left display
9) Press the UP/DOWN button to adjust the “ON” value. Press and hold the
UP/DOWN button to rapidly advance the “ON” value.
10)Press the SET UP button again. “OFF” will appear in the right display the relay
“OFF” value will appear in the lower left display
11)Press the UP/DOWN buttons to adjust the “OFF” value. Press and hold the
UP/DOWN button to rapidly advance the “OFF” value.
12)Press the SET UP button again and the next relay, SP 2 will appear. The sequence
listed in steps 6 and 11 are repeated until all four relays have been assigned.
The MC300 will return to its normal operating mode after the fourth relay has been set up
or 60 seconds have elapsed since the last button press. Changes are stored after the
complete set up of an individual relay channel. These set up parameters are stored in Non
volatile memory and are not lost during Power outages.
NOTE: Set point changes are also accepted by the MC300 by pressing the “TEST OP”
button at any point in the set up. The Instrument will not accept “OFF” values that are
lower than the “ON” value. “ON” values entered that are greater than the ”OFF”
values will automatically change the “OFF” value one unit higher than “ON”.
Phone:(215) 947-2500 fax:(215) 947-7464 e-mail: [email protected] web site:www.televac.com
6-8000-002 REV A
15

CHAPTER 5 TELEVAC
Analog Recorder Outputs
0-10VDC analog recorder output voltage/pressure conversions are provided for
each type of Sensor in the figures 5 & 6.
These voltage signals are available at the rear panel of the MC300. See figure 2
for locations of the connections.
Outputs #1 and #2 are for thermocouple or convection sensors. These outputs
are 0-10 volts and are available in a traditional or linear format. See figures 5 or 6.
MICRONS
ARGON
READS
TRAD’L
REC
OUT
VOLTS
LIN REC
OUT
OUT
(-5) 10.59 -0.05
(-1) 10.16 -0.01
0 10.00 0.00
1 9.84 0,01
5 9.41 0.05
10 8.92 0.10
20 8.03 0.20
30 7.23 0.30
40 6.63 0.40
50 6.13 0.50
60 5.69 0.60
70 5.31 0.70
80 4.95 0.80
90 4.63 0.90
100 4.37 1.00
150 3.37 1.50
200 2.89 2.00
300 2.23 3.00
400 1.88 4.00
500 1.67 5.00
600 1.58 6.00
700 1.50 7.00
800 1.42 8.00
900 1.34 9.00
1000 1.26 10.00
2000 1.02 10.20
5000 0.92 10.20
10000 0.90 10.20
20000 0.88 10.20
CABLE 0.00 10.20
Analog Output (2A) Linear
0
1
2
3
4
5
6
7
8
9
10
0 100 200 300 400 500 600 700 800 900 1000
PRESSURE (microns)
VOLTS
Analog Output (2A) Linear
0
1
2
3
4
5
6
7
8
9
10
0 100 200 300 400 500 600 700 800 900 1000
PRESSURE (microns)
VOLTS
Figure 5
Analog Output (2A)
Phone:(215) 947-2500 fax:(215) 947-7464 e-mail: [email protected] web site:www.televac.com
6-8000-002 REV A
16

Figure 6
Analog Output (4A)
CHAPTER 5(cont.) TELEVAC
TORR/
MICR
ARG
READS
TRDL
OUT
L/D REC
OUT
1 VOLT/
DEC
LIN OUT
VOLTS
(-5)-OL’D -0.065 3.500 <-.050
(-1) -0.013 3.900 -.010
0- 0.000 4.000 0.000
1 u 0.013 4.100 0.010
5 u 0.065 4.500 0.050
10 u 0.124 5.100 0.100
20 u 0.238 5.200 0.200
30 u 0.344 5.300 0.300
40 u 0.450 5.400 0.400
50 u 0.546 5.500 0.500
60 u 0.638 5.600 0.600
70 u 0.730 5.700 0.700
80 u 0.816 5.800 0.800
90 u 0.898 5.900 0.900
100 u 0.980 6.100 1.000
150 u 1.350 6.150 1.500
200 u 1.670 6.200 2.000
300 u 2.285 6.300 3.000
400 u 2.630 6.400 4.000
500 u 2.992 6.500 5.000
800 u 3.824 6.800 8.000
1 T 4.226 7.100 10.000
2 T 5.440 7.200 13.500
3 T 6.080 7.300 “
4 T 6.440 7.400 “
5 T 6.772 7.500 “
10 T 7.412 8.100 “
20 T 7.782 8.200 “
30 T 7.910 8.300 “
40 T 7.936 8.400 “
50 T 7.976 8.500 “
60 T 8.006 8.600 “
80 T 8.042 8.800 “
100 T 8.072 9.100 “
120 T 8.124 9.120 “
150 T 8.180 9.150 “
160 T 8.224 9.160 “
170 T 8.256 9.170 “
180 T 8.280 9.180 “
200 T 8.328 9.200 “
300 T 8.756 9.300 “
400 T 9.208 9.400 “
500 T 9.544 9.500 “
600 T 9.772 9.600 “
760 T 10.000 9.760 “
900 T 10.142 9.90 “
1000 T 10.232 10.000 “
CABLE >10.5 10.11 “
If L/DOUT = INT .DDD Volts
TORR = 0.DDD x 10 EXP (INT –6)
If .DDD < 0.1, use 0.1
Phone:(215) 947-2500 fax:(215) 947-7464 e-mail: [email protected] web site:www.televac.com
6-8000-002 REV A
17

CHAPTER 6 TELEVAC
Maintenance
The vacuum measuring sensors, if installed properly, require very little maintenance
under normal operating conditions.
If the unit should cease to operate properly, certain checks can be made:
1. If a sensor tube simulator is available, it can be used. The sensor tube simulators
are designed to be used as a check to see if the cable and the electronics are
working properly.
The simulators are marked for use with the various sensors. Although no
maintenance is required to keep the tube simulator in working condition, a
periodic check should be done against a calibrated standard unit.
To use the simulator, turn off and disconnect the cable from the sensor tube and
plug the sensor tube simulator in its place. Turn power on and check the output
voltage, it should correspond to the pressure that is written on the sensor tube
simulator’s label.
2. If you do not have a tube simulator, it is still possible to check the operation of the
instrument if you are able to reduce the system pressure to 10-5 Torr. At this
pressure, the thermocouple and the convection sensors should display zero and
give Analog output of zero volts. The “ZERO” adjust trimmers at the rear panel
can be used to set up the zero readings. If output still appears to respond
incorrectly, a new sensor tube should be replaced to restore the assembly to proper
operation.
3. The following table summarizes the messages on the MC300 display under several
different operating conditions as well as the expected Analog outputs.
Phone:(215) 947-2500 fax:(215) 947-7464 e-mail: [email protected] web site:www.televac.com
6-8000-002 REV A
18

CHAPTER 6(cont) TELEVAC
2A
Display
Conditions
Traditional Analog
Output (Volts)
Linear
Analog Output (Volts)
-OLd <-1 micron +10.59 -0.05
-1 Incorrect ZERO adjust +10.16 -0.01
0 microns Lowest Reading +10.00 0.00
20 Torr Highest Reading +0.88 +10.20
CabL > 20 Torr +0.00 +10.20
4A
Display
Conditions
Traditional
Recorder Output
Fact Option
(Volts)
Linear
Decade
Analog Output
(Volts)
Linear Analog
Output (Volts)
-OLd <-5 microns -0.07 +3.50 <-.05
-1 Incorrect ZERO
adjust
-0.01 +3.90 -0.01
0 microns Lowest Reading 0.00 +4.00 0.00
990 Torr Highest Reading +10.23 +9.99 +13.50
CAbL >990 Torr >+10.50 +10.11 +13.50
CHAPTER 7 TELEVAC
Calibration
All TELEVAC instrumentation is calibrated at the factory. No further calibration at the
customer’s facility should be required.
For a reliable re-calibration at a later date, it is necessary to have a reference standard
vacuum system whose pressure is known very accurately. For such a calibration, the
instrument, sensor tube and cable should be returned to TELEVAC for re-certification.
However, if the vacuum calibration certification is to be done at the customer’s facility,
equipment of certifiable accuracy should be used by a knowledgeable electronics
technician trained in vacuum calibration.
Phone:(215) 947-2500 fax:(215) 947-7464 e-mail: [email protected] web site:www.televac.com
6-8000-002 REV A
19

CHAPTER 8 TELEVAC
Calibration Verification
1. Connect sensor to the known vacuum source.
2. Plug unit into 115VAC. Allow a minimum of 15 minutes warm-up.
3. Pump the system down to high vacuum. At 10-5 TORR verify that the
thermocouple or convection gauge reads at “0” VDC. If not, adjust the “zero”
potentiometer.
4. Make a comparison table of indicated pressure on the instrument to that of the
reference vacuum standard. The comparisons should be made typically at “0”
(red line), 10, 100 and 1,000 millitorr and at 1, 10, 100, 500 and 760 Torr,
(depending on the 2A or 7A gauge type).
CHAPTER 9 TELEVAC
Troubleshooting
Troubleshooting of the circuit or components is similar to any industrial electrical
equipment, i.e., checking for circuit continuity, shorts, grounds, resistors, values, etc.
Generally, those familiar with electrical equipment should have no difficulty location
faults in either gauge or electronics unit by systematically proceeding through the circuit.
The enclosed diagram should facilitate maintaining the instrument. In order to
troubleshoot the unit, your gauge tube and cable must be in working condition and
plugged into the unit.
NOTE: If, after doing some or all of the above, the unit is still inoperative, return to
factory for repair.
Phone:(215) 947-2500 fax:(215) 947-7464 e-mail: [email protected] web site:www.televac.com
6-8000-002 REV A
20
This manual suits for next models
1
Table of contents
Other TELEVAC Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

HT
HT HT77N user manual

Honeywell
Honeywell C7355A1050 Mounting instructions

Brookfield
Brookfield TT100 Installation, operation and maintenance instructions

Extech Instruments
Extech Instruments 407703A user manual

PCE Instruments
PCE Instruments PCE-EM 890 user manual

Vaisala
Vaisala BAROCAP PTB330 user guide