TESTO 184 User manual

testo 184 · Data logger
Instruction manual

2

1 Safety and the environment
3
1 Safety and the environment
1.1. About this document
Use
> This instruction manual is an essential component of the
product.
> Please read this documentation through carefully and
familiarize yourself with the product before using it. Pay
particular attention to the safety instructions and warning
notices in order to prevent injuries and damage to the product.
> Keep this document to hand so that you can refer to it when
necessary.
> Make sure that the instruction manual is read by subsequent
users of the product.
1.2. Ensure safety
> Only operate the product properly, for its intended purpose and
within the parameters specified in the technical data. Do not
use any force.
> Do not use the product if there are signs of damage to the
housing.
> Carry out only the maintenance and repair work on this
instrument that is described in the documentation. Follow the
prescribed steps exactly. Use only original spare parts
from Testo.
1.3. Protecting the environment
> Dispose of faulty rechargeable batteries/spent batteries in
accordance with the valid legal specifications.
> At the end of its useful life, send the product to the separate
collection for electric and electronic devices (observe local
regulations) or return the product to Testo for disposal.
WEEE Reg. Nr. DE 75334352

2 Specifications
4
2 Specifications
2.1. Use
The testo 184 data loggers are used to save and read out individual
readings and measurement series. They have been specifically
designed for monitoring the transportation of products subject to
cold chain requirements.
Temperature and humidity readings are saved throughout the
entire duration of the measurement program.
Acceleration readings are monitored throughout the duration of the
measurement program and saved when the set limit value is
exceeded.
Data logger programming and measurement report output are
implemented via PDF files, no software needs to be installed.
Product versions T1 and T2 are single-use data loggers with a
time-limited service life.
2.2. Technical data
testo 184 T1, T2, T3, T4
Feature Values
Display T1, T4: no
T2, T3: yes
Probe type T1, T2, T3: internal NTC temperature sensor
T4: internal PT1000 temperature sensor
Measuremen
t channels
1 internal
Measuremen
t parameters
[unit]
Temperature [°C, °F]
Measuring
range
T1, T2, T3: -35 to 70 °C
T4: -80 to 70 °C
Accuracy T1, T2, T3: ±0.5 K
T4: ±0.8 K (-80 to -35.1 °C), ±0.5 K (-35.0 to 70 °C)
Resolution 0.1 °C

2 Specifications
5
Feature Values
Operating
temperature
T1, T2, T3: -35 to 70 °C
T4: -80 to 70 °C
Storage
temperature
T1, T2, T3: -35 to 70 °C
T4: -80 to 70 °C
Battery type T1: internal, non-replaceable
T2: internal, non-replaceable
T3: CR2450, replaceable
T4: TLH-2450, replaceable
Battery life
(reusable
data loggers)
T3: 500 days (15-minute measuring cycle, 25 °C)
T4: 100 days (15-minute measuring cycle, -80 °C)
Operating
time (single-
use data
loggers)
T1: 90 days from the first time the program is started
(5-minute measuring cycle, -35 °C)
T2: 150 days from the first time the program is
started (5-minute measuring cycle, -35 °C)
Protection
class
IP67
Measuring
interval
1 minute to 24 hours
Memory T1: 16,000 readings
T2, T3, T4: 40,000 readings
Dimensions T1: 33 x 9 x 74 mm
T2, T3, T4: 44 x 12 x 97 mm
Weight T1: 25 g
T2, T3, T4: 45 g
Directives,
standards,
certificates
2014/30/EC, EN 12830, HACCP-certified,
temperature calibration certificate traceable
according to ISO 17025
testo 184 H1, G1
Feature Values
Display yes
Probe type Internal NTC temperature sensor
H1: Internal digital humidity sensor
G1: Internal digital humidity sensor & internal 3-axis
accelerometer

2 Specifications
6
Feature Values
Measuremen
t channels
H1: 2 internal
G1: 5 internal
Measuremen
t parameters
[unit]
H1: temperature [°C, °F], relative humidity [%]
G1: temperature [°C, °F], relative humidity [%],
acceleration [g, m/s²]
Measuring
range
-20 to 70 °C
0 to 100 % (not for condensing atmospheres)1
G1: 0 to 27 g
Accuracy ±0.5 K (0.0 to 70 °C), ±0.8 K (-20 to -0.1 °C)
±1.8 % RH + 3 % of the reading (at 25 °C, 5 to
80 %), ±0.03 % RH / K (at 0 to 60 °C)
G1: ±1,1,1 m/s² + 5 % of the reading
Resolution 0.1 °C
0.1 % RH
G1: 0.1 g
Operating
temperature
-20 to 70 °C
Storage
conditions
-55 to 70 °C
30 to 60% RH
Battery type CR2450, replaceable
Battery life
(reusable
data loggers)
H1: 500 days (15-minute measuring cycle, 25 °C)
G1: 120 days (15-minute measuring cycle, 25 °C)
Protection
class
IP 30
Measuring
interval
1 minute to 24 hours (temperature and relative
humidity)
1 second (acceleration)
Scan
frequency
1600Hz (acceleration)
1For continuous application in high humidity (> 80 % RH at ≤30 °C for > 12 h,
> 60 % RH at > 30 °C for > 12 h), please contact us via www.testo.com/service-
contact
Regenerate the sensor for 12 hours at < 20% RH and 50 °C, or for 24 hours at
40 to 60% RH and 20 to 30 °C.

3 Product description
7
Feature Values
Memory 64,000 readings (temperature and relative humidity)
G1: 1,000 readings (acceleration)
Dimensions 44 x 12 x 97 mm
Weight 45 g
Directives,
standards,
certificates
2014/30/EC, HACCP-certified
Warranty
24 months, warranty conditions: see website
www.testo.com/warranty
EU conformity
The EU Declaration of Conformity can be found on the
testo homepage www.testo.com under the product specific
downloads.
EU countries:
Belgium (BE), Bulgaria (BG), Denmark (DK), Germany (DE),
Estonia (EE), Finland (FI), France (FR), Greece (GR), Ireland (IE),
Italy (IT), Latvia (LV), Lithuania (LT), Luxembourg (LU), Malta (MT),
Netherlands (NL), Austria (AT), Poland (PL), Portugal (PT),
Romania (RO), Sweden (SE), Slovakia (SK), Slovenia (SI), Spain
(ES), Czech Republic (CZ), Hungary (HU), United Kingdom (GB),
Republic of Cyprus (CY).
EFTA countries:
Iceland, Liechtenstein, Norway, Switzerland
3 Product description
3.1. Status LEDs
To increase the battery life, the status LEDs are not permanently
illuminated. They flash once every 5 seconds.
In hibernation mode, the status LEDs are disabled.

3 Product description
8
Alarm
Feature LED colour
No alarm green
Alarm red
Battery
Feature LED colour
Battery life > 10 days green
Battery life < 10 days red
Mode
Feature LED colour
WAIT mode (wait for program start) green and red
Rec mode (measurement program running) green
End mode (measurement program finished) red
3.2. Display (LCD)
Not available for all product versions.
1 Measurement program running
2 Measurement program finished

3 Product description
9
3 Wait for start of measurement program
4 Highest saved reading
5 Lowest saved reading
6 Reading
7 Status information: start criterion Date/time programmed/time
mark, xyz measurement axes for acceleration measurement,
Alarm set limit value(s) exceeded, set limit value(s) not
exceeded
8 Units
9 Battery capacity: sufficient, part empty, low,
(flashing) empty
10 Lower limit value exceeded
11 Upper limit value exceeded
For technical reasons, the display speed of liquid crystal
displays becomes slower at temperatures below 0 °C
(approx. 2 seconds at -10 °C, approx. 6 seconds at -20 °C).
This has no influence on the measuring accuracy.
For technical reasons, battery performance decreases at
lower temperatures. This has no influence on the
measuring accuracy. We recommend the use of fully
charged batteries in order to avoid a reset of the instrument
at low temperatures.
3.3. Key functions
Commissioning
The data loggers are delivered in hibernation mode to prolong the
battery life. In this mode, the status LEDs and the display are
disabled.
> Press the START key or the STOP key.
- Wait mode is activated.
START key
✓Wait mode and start criterion Button Start programmed.
> Press START key for approximately 3 seconds to start the
measurement program.
- The measurement program starts: Mode status LED flashes
green, Rec appears on the display.
✓Product versions with display:
> Press START key to toggle between the displays.
Display sequence (max. display range per version, certain data is
not displayed depending on the operating mode):

3 Product description
10
Display T2 T3 H1 G1
Current temperature reading (°C / °F) X X X X
Current average value MKT (Mean Kinetic
Temperature)
X X X X
Current relative humidity reading (%) - - X X
Current acceleration reading, X axis (x, g) - - - X
Current acceleration reading, Y axis (y, g) - - - X
Current acceleration reading, Z axis (z, g) - - - X
Maximum temperature reading (Max, °C / °F) X X X X
Minimum temperature reading (Min, °C / °F) X X X X
Maximum relative humidity reading (Max, %) - - X X
Minimum relative humidity reading (Min, %) - - X X
Maximum acceleration reading , X axis (Max,
x, g)
- - - X
Maximum acceleration reading, Y axis (Max,
y, g)
- - - X
Maximum acceleration reading, Z axis (Max,
z, g)
- - - X
Time mark ( ) X X X X
Battery life in days ( ) X X X X
STOP key
✓Rec mode and stop criterion Button Stop programmed.
> Press STOP key for approximately 3 seconds to end the
measurement program.
- The measurement program ends: Mode status LED flashes red,
End appears on the display.
START + STOP key
The data loggers may be placed in hibernation mode to prolong the
battery life. In this mode, the status LEDs and the display are
disabled.
✓WAIT or End operating mode.
> Press the START key and the STOP key simultaneously for
approximately 3 seconds.

3 Product description
11
- Hibernation mode is activated.
3.4. Important information and glossary
of terms
• Single-use data logger (versions T1 and T2): the data logger
has a time-limited service life, which begins from the first time
the program is started.
• Start and stop setting: the criteria for starting and stopping the
program are defined in the configuration file.
One of the criteria must be selected for starting the program.
When selecting the criterion, a time delay can be entered
(program starts x minutes after pressing the key).
Both criteria can also be selected for stopping the program. The
criterion that occurs first stops the program.
• Measuring interval: the measurement interval defines the
intervals at which readings are saved.
• Time mark: time marks can be set for documentation by
pressing the START key for 3 seconds during the
measurement, e.g. when responsibility is switched to another
institution. A maximum of 10 time marks can be set. Setting a
time mark resets the statistical values Min, Max and MKT.
• Acceleration (shock): the (positive and negative) acceleration
is measured in 3 measurement axes. Only those readings that
exceed the set limit value (maximum value per 1 second) are
stored and displayed.
The acceleration readings of the 3 measurement axes are
displayed separately on the display of the data logger.
The maximum cumulative value (peak) of the 3 measurement
axes is displayed in the PDF report.
• Report time zone: defines the time zone to which all time
specifications in the measurement report refer. Any time zone
changes during the measurement are not taken into account.
If the logger was in rSt mode and could not be
reconfigured, configuration via a copy of an XML file
may result in the time & time zone not being correct.
• Reset mode (rSt): is triggered by an interruption to the power
supply, e.g. during battery replacement. A reconfiguration of the
logger is required to resume operation. Data that has already
been recorded will not be affected.
• MKT (mean kinetic temperature): the MKT is a single calculated
temperature. MKT may be considered an isothermal storage
temperature. It simulates the non-isothermal effects of storage
temperature variations.

4 Using the product
12
Calculation:
∆/
∆/
∆/
∆/
Tmkt = mean kinetic temperature in degrees Kelvin
∆E = activation energy (standard value: 83.144 kJ/mol)
R = universal gas constant (0.0083144 kJ/mol)
T1 = average temperature in degrees Kelvin during the first time
period
Tn = average temperature in degrees Kelvin during the nth time
period
• MKT activation energy: the default activation energy is set at
83.144 kJ/mol, as recommended in USP <1160>. If other
estimates are available as a result of studies that have been
carried out, the activation energy can be customised.
• Individual alarm: an alarm is triggered when the set limit value
is exceeded.
• Cumulative alarm (only for temperature and humidity
measurement): an alarm is not triggered when the set limit
value is first exceeded, but only once the overall duration,
during which limit values are exceeded, exceeds the set waiting
period (allowed time).
• Wall bracket (delivery includes version G1): for the
acceleration measurement, the data logger must be
permanently connected to the object to be monitored.
Mount the wall bracket using 2 screws or 2 cable ties and then
push the data logger into the wall bracket.
4 Using the product
4.1. Configuring the data logger
Showing/changing the configuration
Adobe Reader software (version X or later) is required.
The data logger must not be in Rec mode.
1. Connect the data logger to a PC via the USB port.
- The status LEDs are disabled, uSb is displayed (instruments
with display). The device drivers are installed automatically.
- The window Automatic playback is displayed.
2. Click on Open folder to view files.
- The file explorer opens.
3. Open the file testo 184 configuration.pdf.

4 Using the product
13
4. Make changes to the configuration. Please note:
• The type of instrument used must be set correctly.
• Existing configuration data can be imported by clicking on
the Import button. The configuration data to be imported
must be available in XML data format.
• When using the Configuration Assistant, some functions are
predefined or filled in automatically. Expert Mode must be
enabled in order to use and manually set all instrument
functions.
5. Export changes to the configuration by clicking on the
button on the data logger.
- A window opens for exporting form data.
6. Select the data logger as the storage location
(Drive TESTO 184) and export the configuration data by
clicking on the Save button.
- The configuration is stored on the data logger as an XML file.
The XML file can be used as a template for other data loggers (via
the import function in the configuration PDF)
Configuring the data logger by copying/pasting the XML file is not
recommended, because the XML file does not contain any
information about the time and time zone and may therefore lead
to incorrect time settings.
7. Close file.
A message Do you want to save the changes to “testo 184
configuration.pdf” before closing? may appear. Answer this
question with No.
8. Disconnect the data logger from the PC.
- The logger switches to Wait mode, the Mode status LED
flashes green/red.
Configuring several data loggers with the same settings
> Configure your measurement protocol using the configuration
PDF or import an existing XML file.
> Connect the testo 184 data logger to a USB connection.
> Click to save the configuration on the connected
testo 184 data logger.
> Leave the configuration PDF open. Connect the next testo 184
data logger. Repeat the last step in order to export the identical
configuration

4 Using the product
14
Changing the logo for the measurement data report
A logo is inserted in the measurement data report. This can be
customised.
The logo must be available in JPEG data format, the file size
should not exceed 5 kB and the file name must be Logo.jpg.
> Create a logo that meets the above criteria and copy it to the
data logger.
Configuring the data logger using Testo PC software
Alternatively, the data logger can also be configured using testo
Comfort Software Professional (version 4.3 service pack 2 or later)
or testo Comfort Software CFR (version 4.3 service pack 2 or later).
Please consult the relevant software instruction manual.
In order to ensure conformity of the testo Comfort Software 21
CFR Part 11, configuration via the PDF file is no longer possible
once the testo 184 data logger has been configured using the
above-mentioned software.
4.2. Measuring
Starting measurement
Depending on the configuration of the data logger, the
measurement program is started via one of the following criteria:
• Button Start: hold down the START key for > 3 seconds.
• Time Start: the measurement starts automatically once the
configured time has been reached.
- The logger switches to Rec mode, the Mode status LED flashes
green.
Setting time marks
While the measurement program is running (Rec mode), up to
10 time marks can be set. These are used to document the transfer
of responsibility, for example.
> Hold down the START key for > 3 seconds.
- The number of set time marks is displayed for 3 seconds and
flashes three times (instruments with display), Mode status LED
flashes green/red three times.
Ending measurement
Depending on the configuration of the data logger, the
measurement program is ended via one of the following criteria:

4 Using the product
15
• Button Stop: hold down the STOP key for > 3 seconds.
• Time Stop: the measurement stops automatically once the
configured time has been reached.
- The logger switches to End mode, the Mode status LED
flashes red.
4.3. Reading out data
Displaying a measurement data report
Adobe Reader software (version 5 or later) or compatible software
is required to display PDF/A files.
1. Connect the data logger to a Windows PC via the USB port.
- The status LEDs are disabled, uSb is displayed (instruments
with display). The device drivers are installed automatically.
- The window Automatic playback is displayed.
2. Click on Open folder to view files.
- The file explorer opens.
3. Open the file testo 184 measurement report.pdf.
- The measurement data report is displayed.
> Save or print the report as required.
Detailed measurement data analysis
To carry out detailed analysis or to further process readings, the
software testo Comfort Software Professional (version 4.3 service
pack 2 or later) or testo Comfort Software CFR (version 4.3 service
pack 2 of later) is required (accessory). Please consult the relevant
software instruction manual.
Measurement data output via NFC
The data loggers are equipped with an NFC (Near Field
Communication) transmitter. This allows instrument data to be read
out via short-range radio using compatible devices (for example
a report printer with NFC).
• The NFC function on your data logger can be enabled/disabled
in the configuration file.
• You do not need any additional software to transfer data to
a compatible Testo report printer (e.g. mobile printer for data
loggers 0572 0576).
• To transfer data, the data logger must be placed on the NFC
transmitter ( ) of the target device.
• Please also consult the instruction manual for the target device.

5 Maintaining the product
16
5 Maintaining the product
5.1. Changing the batteries
It is not possible to change the batteries for instrument types T1
and T2 (single-use data loggers).
A battery replacement stops a measurement that is
currently running. However, stored measurement data are
preserved.
An interruption of the power supply leads to the testo 184
data logger's time settings being reset. In order to restore
the correct time setting, configuration via the PDF file or the
Comfort Software must be carried out.
1. Read out stored data.
2. Place the data logger on its front.
3. Open the battery compartment on the back of the data logger
by turning anti-clockwise. Use a coin for this purpose.
4. Take the spent battery out of the battery compartment.
5. Insert new batteries (see technical data for type required) into
the instrument so that the positive terminal is visible.
Only use new branded batteries. If a partially exhausted
battery is inserted, the battery capacity will not be
calculated correctly.
6. Place the battery compartment cover on the battery
compartment and close it by turning clockwise. Use a coin for
this purpose.
- The data logger is in reset mode, rSt is illuminated
(instruments with display), status LEDs are disabled.
7. Reconfigure the data logger, see the section Configuring the
data logger.
5.2. Cleaning the instrument
CAUTION
Damage to the sensor!
> Ensure that no liquid enters the inside of the housing.
> If the housing of the instrument is dirty, clean it with a damp
cloth.

6 Tips and assistance
17
Do not use any aggressive cleaning agents or solvents! Weak
household cleaning agents or soap suds can be used.
6 Tips and assistance
Questions and answers
Question Possible causes/solution
E0x is displayed
(instruments with display),
all status LEDs are flashing
red
An error has occurred.
• E01: Configuration unsuccessful /
PDF file faulty.
• E02, E03, E04 or E05: sensor
faulty.
• E06: maximum number of time
marks set, new time mark cannot
be set.
---- is displayed
(instruments with display)
• No reading available (after setting
a time mark)
• Reading invalid.
Hi is displayed
(instruments with display)
Reading is above the measuring
range.
Lo is displayed
(instruments with display)
Reading is below the measuring
range.
En is displayed
(instruments with display)
Set time marks function is disabled.
Err is displayed
(instruments with display)
Configuration not possible,
e.g. because Rec mode is active.
Configuration via the PDF
file is not possible
If you have used the Comfort Software
21 CFR Part 11 for the configuration,
configuration via the PDF file is
disabled.
The size of the PDF
configuration file is 0 kB or
the file is damaged.
Copy the PDF file of another testo 184
data logger or download the
configuration file from the Testo
website: http://www.testo.com/.

6 Tips and assistance
18
Question Possible causes/solution
Different time or time zone
in the report
> The testo 184 data logger has not
been configured after a battery
replacement. Repeat the
configuration to restore the correct
time settings.
> Check whether the PC which was
used for the configuration has the
correct time settings.
No measurement protocol
was created.
> Please check whether the data
logger is in record/end mode.
> Please reconnect the data logger
to the PC.
> Please check whether there is
enough free memory capacity
available on the data logger.
The PDF configuration is
not ready for use.
> Please check whether the data
logger is in record mode.
> Please check whether the correct
data logger mode is selected.
> Please check whether the
testo 184 data logger was
configured via the Comfort
Software 21 CFR Part 11. This
prevents configuration via the PDF
file.
There is no LCD display. Please check whether the LCD display
is disabled in the configuration.
There is no LED display. Please check whether the LED display
is disabled in the configuration.
There is no time mark
display.
Please check whether the time mark
display is disabled in the configuration.
There is no NFC. Please check whether the NFC
display is disabled in the configuration.

6 Tips and assistance
19
Question Possible causes/solution
The humidity value
measured is out of
tolerance.
a Was the t99 response time
achieved?
b Was the H1/G1 data logger stored
for longer than 60 hours at a
relative humidity of more than 80%
without an airtight bag?
c Was the H1/G1 data logger used
for longer than 60 hours at a
relative humidity of more than
80%?
Solution for points b and c:
> Store the device for 12 hours at
50°C and a relative humidity of
less than 20%.
or
> Store the device for 12 hours at
20°C to 30°C and a relative
humidity of around 75%.
If you have any questions please contact your local dealer or the
Testo Customer Service. Contact details can be found on the
internet: www.testo.com/service-contact

0970 1842 en 05
Other manuals for 184
2
This manual suits for next models
7
Table of contents
Other TESTO Other manuals
Popular Other manuals by other brands

IAG
IAG IAG-DRV-1000 Installation instruction

K&K Sound
K&K Sound BIG ISLAND SPOT product manual

Xytronix Research & Design
Xytronix Research & Design WebRelay user manual

EMG
EMG X Series Installation information

Edelbrock
Edelbrock 61905 installation instructions

Beko
Beko METPOINT BDL compact Instructions for installation and operation