TESTO 316-1 User manual

Other manuals for 316-1

4

Other TESTO Other manuals

TESTO Gas detector User manual

TESTO

TESTO Gas detector User manual

TESTO 184 User manual

TESTO

TESTO 184 User manual

TESTO 184 User manual

TESTO

TESTO 184 User manual

TESTO 175-S1 User manual

TESTO

TESTO 175-S1 User manual

TESTO 175 User manual

TESTO

TESTO 175 User manual

TESTO 320 basic User manual

TESTO

TESTO 320 basic User manual

Popular Other manuals by other brands

Opticon EduMic Compatibility guide

Opticon

Opticon EduMic Compatibility guide

Kessil A360W owner's manual

Kessil

Kessil A360W owner's manual

TEFAL ULTRA COMPACT MIXER manual

TEFAL

TEFAL ULTRA COMPACT MIXER manual

Martin Audio WS18X - SCHEMATICS manual

Martin Audio

Martin Audio WS18X - SCHEMATICS manual

Hendrickson AR2 series Service instructions

Hendrickson

Hendrickson AR2 series Service instructions

GO Industries 38017 installation instructions

GO Industries

GO Industries 38017 installation instructions

Cissell 30LBCMAN2 manual

Cissell

Cissell 30LBCMAN2 manual

RIDGID K-60SP manual

RIDGID

RIDGID K-60SP manual

Gage Bilt GB751SH-5 Original instructions

Gage Bilt

Gage Bilt GB751SH-5 Original instructions

OVE SIERRA 60 installation manual

OVE

OVE SIERRA 60 installation manual

ReelCraft 7800 OLB21 operating instructions

ReelCraft

ReelCraft 7800 OLB21 operating instructions

Body Armor 4x4 ECO-SERIES user manual

Body Armor 4x4

Body Armor 4x4 ECO-SERIES user manual

ROBLIN DROP Dimensions

ROBLIN

ROBLIN DROP Dimensions

EMG EMG-ACB BARREL Installation information

EMG

EMG EMG-ACB BARREL Installation information

MiniSolar MT-150 Assembling Guide

MiniSolar

MiniSolar MT-150 Assembling Guide

Parkside PSZ 250 B2 Operation and safety notes

Parkside

Parkside PSZ 250 B2 Operation and safety notes

ABB REM 610 Technical reference manual

ABB

ABB REM 610 Technical reference manual

Tyent ACE-13 user manual

Tyent

Tyent ACE-13 user manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.

testo 316 - 1
Bedienungsanleitung
Überwachungsgerät zum Aufspüren von Gaslecks.
Das testo 316-1 findet kleinste Leckagen von brennbaren Gasen an Rohrleitungen, Flanschen,
Verschraubungen usw. Das biegbare Sondenrohr erreicht selbst schwer zugängliche Stellen.
Vielseitige Einsatzmöglichkeiten machen das Gasleck-Überwachungsgerät unentbehrlich für jeden
Fachmann.
Nicht an spannungsführenden Teilen messen!
Gerät nicht in Umgebungen über 95 %rF (kondensierend) betreiben.
Zulässige Lager- und Transporttemperatur sowie die zulässige Betriebstemperatur beachten
(z. B. Meßgerät vor direkter Sonneneinstrahlung schützen)!
testo 316-1 nicht einsetzen in geschlossenen Räumen in denen sich Gase zu einem explosiven
Gemisch angesammelt haben.
Sicherstellen, dass die Gaskonzentration 20% UEG nicht übersteigt.
An Frischluft Empfindlichkeit des Sensors einstellen.
Vor der Gaslecksuche grundsätzlich eine Funktionsprüfung durchführen.
Einstellen der Empfindlichkeit des Sensors in gasbelasteter Umgebung verschiebt die
Alarmschwellen nach unten.
Bei unsachgemäßer Behandlung oder Gewaltanwendung erlöschen die Gewährleistungsansprüche!
Einschalten
- Schaltknopf drehen
- Aufheizphase des Sensors beginnt
- LED leuchtet rot
- Alarmsignal ertönt
- Aufheizphase max. 30 sec.
- LED leuchtet grün
- Alarmsignal aus
Betriebsbereitschaft
- Kein Alarmsignal
- LED leuchtet grün
Empfindlichkeit einstellen an Frischluft
- Schaltknopf nach rechts drehen bis akustisches
Signal ertönt.
- Schaltknopf leicht nach links drehen bis Signal
verstummt.
Lecksuche durchführen
- Akustisches Signal bei ausströmenden Gas
- Signal wird mit zunehmender Konzentration
schneller
- Alarm > 200 - 10000 ppm
- LED wechselt von grün auf gelb
- Signalton
- Alarm > 10000 ppm
- LED wechselt von gelb auf rot
- Signalton
Funktionsprüfung durchführen
- Sensor mit niedrg konzentriertem Gas beauf-
schlagen (max. 10 sec.) oder Sensor anhau-
chen.
- Spricht der Sensor nicht an (kein Alarm), ist
das Gerät defekt und darf nicht mehr verwendet
werden. Gerät muss zur Instandsetzung
in den Service.
Grüne LED
Einschalten
Betriebsbereitschaft
Empfindlichkeit einstellen
Messen
Changing the battery
Technical data
Ordering data Part no.
Measuring range: 100 to 10,000 ppm (1 Vol%) CH4
1. Alarm threshold: at max. 200 ppm LED is yellow
2. Alarm threshold: at max. 10,000 ppm LED is red
Battery lifetime: > 5 h with alkali manganese IEC 6LR61
Operating temperature: 4 to 45 °C
Storage/transport
temperature: -20 to +50 °C
Dimensions: 90 x 57 x 42 mm
Weight: Approx. 300 g
testo 316-1, gas leak monitoring instrument with Instruction manual and battery 0632.0316
Case 0516.0182
TopSafe, indestructible protective case with stand 0516.0189
Plastic case to transport and store measuring instrument and accessories 0516.3120
Low Batt Signal
If during use the LED switches off and a one sec cyclic audible signal is obtained, replace the
battery within 15 min for continued working.
- Open housing screw (rear of instrument).
- Insert 9 V block battery. Observe polarisation.
- Close housing.
Note on disposal: Dispose of empty batteries responsibly. Batteries should be placed in plastic
bags to avoid short-circuits.
Measuring instrument
conforms with:
2014/30/EU
0973 3161 de en 04
testo 316 - 1
Instruction manual
Monitoring instrument to detect gas leaks.
The testo 316-1 detects even the smallest leaks of combustible gases in pipelines, flanges and
screw connections etc. The bendable probe stem has no problem reaching even inaccessible
points. Its wide range of applications makes the gas leak monitoring instrument indispensable for
every professional.
Do not measure on live parts.
This Instrument is not designed to operate while wet or in an environment of
condensing humidity (> 95 %RH).
Observe maximum storage and transport temperature and maximum operation temperature
(e.g. protect measuring instrument from direct sunlight)!
Do not use testo 316-1 in enclosed rooms where gases have gathered to form an explosive
mixture.
Ensure that the gas concentration does not exceed the lower explosion limit of gas mixtures by
20%.
The alarm limits are adjusted downward when the sensitivity is set in a gas environment.
Always carry out a function test prior to gas leak detection.
Set sensitivity of sensor in fresh air.
Warranty no longer applies in the case of inexpert handling or if force is used.
Switching on
- Turn control button
- Warm-up phase of sensor begins
- LED is red
- Alarm signal sounds
- Warm-up phase max 30 seconds
- LED is green
- Alarm signal off
Ready to operate
- No alarm signal
- LED is green
Setting sensitivity in fresh air
- Turn control button to the right until the sound
signal is heard
- Turn the control button to the left until the
sound signal stops.
Detecting leaks
- Sound signal when gas is escaping
- Signal becomes quicker as concentration
increases
- Alarm > 200 - 10000 ppm
- LED changes from green to yellow
- Warning sound
- Alarm > 10000 ppm
- LED changes from yellow to red
- Warning sound
Function test
- Charge the sensor with a low concentrated gas
(max. 10 sec.) or breathe on the sensor.
- If the sensor does not react (no alarm), the
instrument should be considered as defective
and must not be used. Return the unit imme-
diatly to the nearest Test service point for repair.
Green LED
Switching on
Ready to operate
Setting sensitivity
Measuring
Batteriewechsel
Technische Daten
Abgleich-Zertifikat/ Justage certificate
Messbereich: 100...10.000 ppm (1 Vol%) CH4
1. Alarmschwelle: spätestens bei 200 ppm LED leuchtet gelb
2. Alarmschwelle: spätestens bei 10.000 ppm LED leuchtet rot
Batteriestandzeit: > 5 h bei Typ Alkali-Mangan IEC 6LR61
Betriebstemperatur: 4...45 °C
Lager-/Transport-
temperatur: -20...+50 °C
Abmessungen: 90 x 57 x 42 mm
Gewicht: ca. 300 g
Serien-Nr. Optischer Signalgeber ook
Serial no. Optical emitter
Akustischer Signalgeber ook
Accustic emitter
Einstellung der Alarmschwellen
Setting alarm thresholds
Sollwert/Reference Geprüft bei/Tested at
Alarm 1 200 ppm ________ ppm ook
Alarm 2 8800 ppm ________ ppm ook
Bestelldaten Bestell-Nr.
testo 316-1, Gasleck-Überwachungsgerät inkl. Bedienungsanleitung und Batterie 0632.0316
Bereitschaftstasche 0516.0182
TopSafe, unverwüstliche Schutzhülle inkl. Aufsteller 0516.0189
Kunststoff-Koffer, für Transport und Aufbewahrung von Messgerät und Zubehör 0516.3120
Low Batt Signal
Wenn während des Einsatzes die LED ausschaltet und ein Signalton im Sekundentakt hörbar ist,
Batterie innerhalb von 15 Minuten wechseln.
- Gehäuseschraube öffnen (Rückseite).
- 9 V Blockbatterie einlegen. Polung beachten!
- Gehäuse schließen.
Entsorgungshinweis: Entladene Batterien in Sammelbox werfen. Um Kurzschlüssen vorzubeugen,
Batterien einzeln in Plastikbeutel geben!
Messgerät konform zu:
2014/30/EU