Quick Start Procedures Air-Cooled Models
Installation
Position the unit so the intake and discharge are not
impeded.Inadequateventilationwillcausea
reduction in cooling capacity and, in extreme cases,
compressorfailure.
Avoid excessively dusty areas and institute a
periodiccleaningschedule.Forproperoperation,
the unit needs to pull substantial amounts of air
through a condenser. A build up of dust or debris on
the fins of the condenser will lead to a loss of
cooling capacity.
The unit will retain its full rated capacity in ambient
temperaturesuptoapproximately+24°C.
Ensurethevoltageofthe powersourcemeetsthe
specifiedvoltage,±10%
.
Theplumbingconnectionsarelocatedontheright
side of the unit and are labelled SUPPLY and
RETURN.Theseconnectionsare¾inchFPT,
1 inch FPT for units with a CP-75 or TU-9 pump.
Removetheplasticprotectiveplugsfromboth
plumbingconnections.Connect the SUPPLYfitting
to the inlet of your application. Connect the RE-
TURNfitting to theoutlet of yourapplication.
Tofill thereservoir open theaccess panelon the left
rearcornerofthecasetopandremovethereservoir
coverby unscrewing thethumbscrews. Fill the
reservoirto withinoneinch ofthetop. Ifthe fluid
capacity of your application and recirculation lines is
significant,haveextrafluidonhand.
ThermoFisherrecommendsusingdistilled/
deionized water with a 0.05 to 0.1 MOhmcm
reading. If you do not have access to distilled/
deionizedwaterwerecommendusingfilteredfacility
water.
Operation
Before starting the unit, double-check all electrical
and plumbing connections. Make sure the circulat-
ing system has been filled with cooling fluid.
OnmodelsHX 300 throughHX 750,theunitmust be
connected to the power source for at least 12 hours
to allow the compressor oil to heat and separated
fromtherefrigerant.
To start the unit first connect it to main power. On
units equipped with a circuit breaker, place it to the
on position. The MAIN POWER LED will illuminate.
Place the ON/OFF switch to the ON position. The
COOL and IDLELEDs onthe front panelindicate
the status of the refrigeration system. COOL is on
whentheunitis removing heat fromthe coolingfluid,
IDLE is on when heat is being added to the fluid
reservoir.Astheoperatingtemperatureapproaches
the setpoint, the LEDs cycle.
When the unit is shut off, wait five minutes before
restartingtoallowtimefortherefrigeration
pressuresto equalize.If the pressuresare not
allowedto equalize, thecompressor will short-
cycle and no cooling will occur.
Temperature Adjustment
Todisplay thetemperaturesetpoint,while
displayingthereservoirtemperaturepressthe
NEXTENTERkey.Toadjustthetemperature
setpoint use the YES/NO keys as described in
this manual.
Flow Control
TheRECIRCULATINGFLOWCONTROLhandle
controls the flow rate to your application. In the full
verticalpositionyoureceivefullflow,thefull
horizontal position is no flow.
Periodic Maintenance
Periodicallyinspectthereservoirfluid.Ifcleaningis
necessary,flushthereservoirwithacleaningfluid
compatible with the circulating system and the
coolingfluid.
Thecoolingfluidshouldbereplacedperiodically.
Whenoperatingatlowtemperatures,theconcen-
tration of water in the cooling fluid will increase
over time, leading to a loss of cooling capacity.
Periodicvacuumingofthecondenserfinsis
necessary.Thefrequencyofcleaningdependson
theoperatingenvironment.Werecommendmaking
amonthlyvisual inspectionof thecondenser after
initialinstallation. After severalmonths, the
cleaningfrequencywillbeestablished.
UnitswithPD and TUpumpshave a strainer. If
debris is in the system, the strainer will prevent the
materialfrombeingdrawnintothepumpand
damagingthepumpvanes.
After initial installation, the strainer may become
clogged.Thestrainermustbecleanedafterthe
first week of installation. After this first cleaning, a
monthlyvisualins-pectionisrecommended.After
severalmonths,thefrequencyofcleaningwillbe
established.
Beforecleaningthestrainer,disconnectthepower
cordfromthepowersourceanddrainthereservoir.