2
Figure 5. The Feed option appears on the Home screen when NanoDrop Lite
is docked on the printer
Troubleshooting Paper Feed
If the paper does not feed properly (no paper comes out from the top
of the printer), remove the paper from the printer and repeat steps 2
through 6.
To remove paper from printer
1. Unplug the power supply to printer.
2. Remove panel from back of printer by pushing the two release
buttons on either side of the paper holders while simultaneously
pulling the panel up.
Figure 6. Release button location
Once the panel has been removed, you will see a green rocker
switch on the right side of the roller.
3. Press switch down to release paper (see Figure 7).
4. Pull paper gently out of the printer.
Figure 7. Rocker switch location
5. Push the green switch up to the horizontal position once paper
has been removed.
6. Snap back panel into place.
7. Plug in the power supply to printer, allow instrument to initialize
and repeat the steps in “To load paper into printer” section.
Printing Labels
When the instrument is connected to the printer, there are three ways
to print sample labels:
•IfAutoprint is ON, a label will print automatically after each
sample measurement.
To change Autoprint navigate to Home > Tools & Settings >
Settings. Press Change to toggle to Autoprint On or Autoprint
Off.
• When Autoprint is turned off, press Print after measuring your
sample.
If you need to print more than one label for the same sample
measurement, press Print again.
• Press Print when viewing results from previously measured
samples in the Sample History screen (Home > Tools & Settings
> Sample History).
This allows you to print labels from previously measured samples
that are stored in the instrument’s memory. Use the arrow keys to
display the sample data you want to print.
Tip Helpful tips for installing the paper
• Keep the paper centered as it enters the slot on the back of
the printer.
• Grab the paper as close to the paper feed slot as possible when
feeding into the printer.
• You will hear a noise as the paper engages in the printer. At
this time, firmly push the paper into the slot to get it started.
• If none of the above works, clip a small corner from the ends
of the paper and reinsert the paper.
Tip Ensure you return switch to horizontal position before
you put the panel on.
Note The printer labels are cryogenic, and per manufacturing
specifications they perform well under a wide temperature
range: 70° C to −196° C.