Step 2: Straight Arm
Insert the straight arm into the joint on
the Elite Rail Clamp, position, and flip
the cam joint handle to lock.
Step 1: Rail Clamp
Place Elite Rail Clamp onto the rail
over the sterile drape and turn top
knob clockwise to tighten.
Step 5: Retract
If additional retraction is necessary,
utilize the micro-adjustable handle by
turning the knob to retract further.
Press pawl to release retraction.
Step 4: Hook
Use the T-handle at proximal end of
the femur hook to position hook in
place. Insert ball into elevator arm
handle to hold in place.
Step 3: Elevator Handle
Position Elevator Handle and lock in
place by closing the joint.
Femur Elevator Set Up Instructions
© 2022 Thompson Surgical Instruments, Inc.
® The Thompson “T Circle” logomark is a Registered Trademark of Thompson Surgical Instruments, Inc.
Patents: trpat.com
Rev B
011123
dc hipfeum
10341 East Cherry Bend Road
Traverse City, Michigan 49684
phone: 231.922.0177
fax: 231.922.0174
thompsonsurgical.com
Emergo Europe
Westervoortsedijk 60
6827 AT Arnhem
THE NETHERLANDS
EC REP
If Using a Hana®Table, or Leg
positioning system, such as Arch
Place rail extender over the sterile drape
and turn top knob clockwise to tighten.
CAUTION: Avoid compressing the patient’s body with
frame components to prevent nerve damage.
The Thompson Retractor is provided in a non-sterile
condition. Refer to Thompson Retractor IFU for cleaning,
sterilization, and care instructions.
Reference the Thompson Retractor IFU for additional
warnings and cautions.
For other Thompson Retractor Systems, reference
corresponding User Manual for each.