Page 18 21501-D02
Electronic, Self-levelling, Benchtop Isolation System
4.5 Further Tuning and PID Parameter Adjustment
The control box contains a number of hall effect sensors which use a position control
loop to determine the command output. The purpose of the position loop is to match
the actual leg position and the demanded position. This is achieved by comparing the
demanded position with the actual leg position to create a position error, which is then
passed through a digital PID-type filter. The filtered value is the valve command
output.
Proportional term - Increasing the proportional term will increase the amount of
effective pressure used to correct a given position error. Typically this is used to
minimise the amount of position error when an impulse event or disturbance affects
the current target position. If the proportional term is too high this can lead to
overshoot and general instability. If this is too low it can result in a sloppy response.
It accepts values in the range 0 to 327.
Integral term - Increasing the integral term minimises following error and final position
error. If Integral is too high this will typically lead to overshoot of the demanded
position. If the integral term is too low final position may take a long time to reach, if
at all. It accepts values in the range 0 to 100.
Derivative term - Increasing the derivative term decreases the rate of change of
controller output. Typically this is used to reduce the overshoot from a given motion.
If derivative is too high it can become sensitive to noise from the measured position
error. If derivative is too low, fluctuations may arise during motion. It accepts values
in the range 0 to 327.
If the PID parameters loaded in the Weight Selection panel do not provide satisfactory
performance, the values can be tuned to your particular application. The cautions
above must be read and understood before changing PID parameters.
Caution
The PID parameters must be set according to the load being supported. Default
values for specific loads have already been optimized and stored within the
controller. These are loaded in the ‘Weight Selection’ set up screen (see
Section 4.3.) and should not need adjustment under normal operating
conditions.
If problems are encountered (e.g. instability/hunting of the system, or
incomplete positional adjustments) these PID parameters should be adjusted
further to tune the system for the given application. If instability is experienced,
the proportional parameter in particular should be reduced in conjunction with
the error band (see previous section). Normally, only minor adjustment of the
Proportional parameter should be necessary, though some trial and error will
be required before the ideal settings for a specific application are achieved.
USE EXTREME CAUTION WHEN CHANGING PID PARAMETER VALUES.
PERFORM ONLY SMALL CHANGES. INCORRECT PID SETTINGS CAN
CAUSE MAJOR INSTABILITY WHICH COULD RESULT IN DAMAGE TO
DELICATE EXPERIMENTS OR APPLICATIONS.