• Calculation
Combined Chlorine = Total Chlorine - Free Chlorine
Technical Data
Free chlorine: 0.05 - 6.0 mg/l Cl2
Total chlorine: 0.05 - 6.0 mg/l Cl2
(Accuracy Chlor: 0-1mg/l±0.2;1-2mg/l±0.3;2-3mg/l±0.4;3-6mg/l±0.5)
Bromine: 0.1 - 13.5 mg/l Br
(Accuracy: 0-2mg/l±0.5;2-4mg/l±0.7;4-7mg/l±0.9;7-13mg/l±1.2mg/l)
pH value: 6.5 - 8.4 pH (Accuracy ±0.2 pH)
Cyanuric Acid: 1 - 80 mg/l Cys (Accuracy ±15 mg/l)
Alkalinity: 10 - 500 mg/l CaCO3 (Accuracy ±50 mg/l)
Light source: LED, filter (λ= 528 nm)
Battery:
9 V-
block battery
(Life 600 tests).
Auto-OFF: Automaticswitchoff5minutesafterlastkeypress
Ambient
temperature:0–40°C,relativehumidity: 30 –90%,
conditions
: non-condensing
CE:
DIN EN 55 022, 61 000-4-2, 61 000-4-8,
50 082-2, 50 081-1, DIN V ENV 50 140, 50 204
•
Maintainance
1. Keep the instrument in a clean condition.
2. Nosolventorabrasivematerialsshouldbeusedtocleantheinstrument.
3. Let the instrument dry when not used, store under ambient
environmental conditions.
4. Remove battery if instrument is longer unused.
• Trouble shooting
1. Use only tablets in white foil.
2. Touching the tablets can cause contamination.
3. Close the grey lid on top of the sample chamber before
zeroing/reading.
4. Alwaysensure thatthesampleandthelidarethoroughlyrinsed
when changing from one test to the other.
• Fitting the battery (9 volts, 6LR61)
BAT Change the batteries with this display indication.
Check
position of the black O-Ring when opening/ closing the battery
chamber lid!