INTRODUCTION
Congratulations on your purchase of Extech’s 380926 Clamp meter + Digital MultiMeter. This meter
can measure to 2000A through the clamp and also measure to 400mA through the test leads. With
Capacitance, Diode and Duty Cycle measurements, this meter provides the user with a full featured
multimeter combined with the high amperage current clamp. This meter is shipped fully tested and
calibrated and, with proper use, will provide years of reliable service.
SAFETY
Safety Symbols
This symbol, adjacent to another symbol or terminal, indicates the user must refer to
the manual for further information.
This symbol, adjacent to a terminal, indicates that, under normal use, hazardous
voltages may be present
Double insulation
WARNING: This indicates that a potentially hazardous condition which, if not avoided, could result
in death or serious injury.
CAUTION: This indicates that a potentially hazardous condition which, if not avoided, could result
in injury or damage to the meter Reply to request.
Safety Precautions
1. WARNING: Improper use of this meter can cause damage, shock, injury or death. Read and
understand this user’s manual before operating the meter.
2. Make sure any covers or battery doors are properly closed and secured.
3. Always remove the test leads before replacing the battery or fuses.
4. Inspect the condition of the test leads and the meter itself for any damage before operating the
meter. Repair or replace any damage before use.
5. Do not exceed the maximum rated input limits.
6. Use great care when making measurements if the voltages are greater than 25VAC rms or
35VDC. These voltages are considered a shock hazard.
7. Always discharge capacitors and remove power from the device under test before performing
Capacitance, Diode, Resistance or Continuity tests.
8. Remove the battery from the meter if the meter is to be stored for long periods.
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

380926 -EN-V7.1-3/12
3
METER DESCRIPTION
Front panel
1. Current Sense Jaws
2. Function switch
3. Keypad
4. LCD display
5. Input jacks
Note: The tilt stand and battery compartment door are
located on the rear of the meter.
Symbols
AC current or Voltage
DC current or Voltage
Continuity
HDisplay hold
REL Relative
AUTO Auto range
Diode
Display Backlight
Units
mV, V millivolt, volt (voltage)
Ω, kΩ, MΩ ohm, kilohm, megohm (resistance)
% percent (duty cycle)
μA, mA, A microamps, milliamps, amps (current)
Hz, kHz hertz, kilohertz (frequency)
nF, μF nanofarad, microfarad (capacitance)
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

380926 -EN-V7.1-3/12
4
OPERATING INSTRUCTIONS
Current Measurements using the Clamp, DC/AC
WARNING: Make sure that the test leads are disconnected from the meter’s terminals before
making current measurements with the clamp jaw.
WARNING: For safety, use the rubber input terminal cover when measuring current using the
Clamp.
1. Set the Function switch to 2000A position.
2. Press the FUNC key to select AC or DC current
3. For DC current measurements, push & hold
the DCA ZERO button until the reading
indicates zero.
4. Press the trigger to open jaw. Fully enclose
the conductor to be measured (see
diagram).
5. The clamp meter will automatically select
the proper range (Auto Range). To select
the range manually, press the RANGE
button before pressing DCA ZERO button.
6. Read the measured value from the LCD
display.
NOTE: If the DCA zero button is used to zero the
meter, the clamp meter remains in the auto-range mode. If the REL button is used, the
clamp meter will be placed in the manual range mode.
Current Measurements using the Test Leads, DC/AC
1. Set the Function switch to the mA or µA position.
2. Press the FUNC key to select AC or DC.
3. Connect the black test lead to the COM terminal and the red test
lead to the mA µA input jack.
4. Break the circuit under test (put the meter in series with circuit
under test) refer to diagram.
5. Read the measured value from the LCD display.
NOTE: The maximum reading for direct input current is 240mA AC/DC.
INCORRECT
More than one
conductor in jaws
CORRECT
One conductor in jaws
µA and mA
Current
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

380926 -EN-V7.1-3/12
5
Voltage Measurements, DC/AC
1. Set the Function Switch to the Vposition.
2. Press the FUNC key to select AC or DC.
3. Insert the black test lead to the COM input jack
and the red test lead into the Vinput jack.
4. Connect the test leads in PARALLEL with the
circuit to be measured (see diagrams).
5. Read the measured value from the LCD display.
Resistance Measurements
CAUTION: Before taking any in-circuit resistance measurements remove
power from the circuit under test and discharge all capacitors.
1. Set the Function switch to the Ωposition.
2. Press the FUNC key until the ohms symbol appears on LCD.
3. Insert the black test lead to the COM input jack and the red test lead to
the Ωinput jack.
4. Connect test leads to the device under test (see diagram).
5. Read the measured value from the LCD display.
Continuity Test
1. Set the Function switch to the position.
2. Press the FUNC key until the "Ω" and " " symbols appear on the display.
3. Insert the black test lead to the COM input jack and the red test lead to the Ωinput jack.
4. Connect the test lead tips to the device to be measured (refer to diagram for resistance
measurements above).
5. Read the measured value from the LCD display.
6. If the resistance is < 10Ωapprox. an audible signal will be heard.
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

380926 -EN-V7.1-3/12
6
Diode Test
1. Set the Function switch to the diode position.
2. Press the FUNC key until the symbol appears on the LCD.
3. Insert the black test lead to the COM input jack and the red test lead to the input jack.
4. Connect the test lead tips to the diode. A typical diode forward voltage drop will indicate 0.3
to 0.7V. The reverse drop will indicate "OL", indicating high impedance. Proper diode check
should include both forward and reverse tests.
Frequency Measurement
1. Set the Function switch to the Hz position.
2. Press the FUNC key until the "Hz" symbol appears on the LCD.
3. Insert the black test lead to the COM input jack and the red test lead to the
Hz input jack.
4. Connect the tips of the test leads to the device to be measured (see
diagram).
5. Read the frequency value in Hz on the LCD.
Duty Cycle Measurements
Follow the steps as for Frequency measurements with the following exception: Press the “Hz/%”
key until the “%” symbol appears on the display.
Capacitance Measurement
1. Set the Function switch to the position.
2. Press the FUNC key until the "nF" symbol appears on the LCD.
3. Insert the black test lead to the COM input jack and the red test lead to the
red input terminal.
4. Connect the tips of the test leads to the device to be measured (see
diagram).
5. Read the capacitance value on the display.
NOTE: When making very low capacitance measurements, use the REL function
to zero any stray capacitance.
RED REDBLACK BLACK
FORWARD TEST REVERSE TEST
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

380926 -EN-V7.1-3/12
7
Automatic / Manual Range Selections
The meter defaults to the autoranging mode when turned on. "AUTO" will appear in the display.
To select manual ranging, press the RANGE key. Momentary presses of the RANGE key will
step through the ranges. To return to the Auto Range mode, press and hold the RANGE key for 2
seconds.
Relative Reading Measurements
Press the REL key to enter the relative mode. "REL" will appear in the display. In the relative
mode, the meter stores the reading that was on the display at the time of the REL key is pressed
and displays the difference between the measured value and the stored value
Press the REL key to return to normal mode.
Data Hold / Backlight key
Press the HOLD key momentarily to freeze the present reading on the LCD."H" will appear in the
display. Press HOLD again to return to normal operation.
Backlight
Press and hold the Hold key for 2 seconds to activate the backlighting. The backlight will
automatically turn off to conserve energy after approx 12 seconds.
MAINTENANCE
Battery Replacement
When the low battery symbol appears on the LCD, replace the meter’s 9V battery.
1. Remove power to the meter and remove test leads from meter
2. Remove the Phillips head screw (back of meter) and open the battery compartment.
3. Remove and replace the 9V battery.
4. Replace the compartment cover and rear screw.
You, as the end user, are legally bound (EU Battery ordinance) to return all used
batteries, disposal in the household garbage is prohibited! You can hand over your
used batteries / accumulators at collection points in your community or wherever
batteries / accumulators are sold!
Disposal: Follow the valid legal stipulations in respect of the disposal of the device at
the end of its lifecycle
Fuse Replacement
NOTE: Fuse Rating: 500mA (5mm x 20mm diameter)
1. The meter is provided with one overload protect 500mA fuse for current measurements
(direct input NOT clamp measurements).
2. To replace the fuse, open the meter case by removing the battery cover and battery and then
the four screws holding the rear case.
3. The fuse is located on the Main PCB.
4. Replace fuse and secure meter case.
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

380926 -EN-V7.1-3/12
8
SPECIFICATIONS
Function Range Resolution Accuracy Remarks
400.0A 0.1A ±(2.0% + 5d)AC/DC Current
(Clamp on) 2000A 1A ±(2.0% + 8d)
45Hz to 1KHz
4.000V 1mV
40.00V 10mV
400.0V 0.1V
AC Voltage
True RMS
1000V 1V
±(1.2% + 5d) 45Hz to 1KHz
Input Impedance:
10Mohms
400.0mV 0.1mV ±(0.5% + 2d)
4.000V 1mV
40.00V 10mV
400.0V 0.1V
DC Voltage
1000V 1V
±(1.0% + 2d)
Input Impedance:
10Mohms
400.0 Ω0.1Ω
4.000 kΩ1Ω
40.00 kΩ10Ω
400.0 kΩ100Ω
±(1.0% + 5d)
4.000 MΩ1kΩ±(2.0% + 2d)
Resistance
40.00 MΩ10kΩ±(3.5% + 5d)
400.0µA 0.1µA
4000µA 1µA
40.00mA 0.01mA
AC/DC Current
(Direct Input)
400.0mA 0.1mA
±(1.2% + 5d) 45Hz to 1KHz
5Hz 0.001Hz
50Hz 0.01Hz
500Hz 0.1Hz
5kHz 1Hz
50kHz 10Hz
Frequency
100kHz 100Hz
±(1% + 5d)
50nF 10pF
500nF 100pF
5µF 0.001µF
Capacitance
50µF 0.01µF
±(3% + 5d) Accuracy stated
after a relative "REL"
zero performed
Duty Cycle 1 to 99% 0.1% ±(1% + 5d)
Continuity Audible tone; <10 ohms approximately
Open circuit voltage; 0.5V approximately
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

380926 -EN-V7.1-3/12
9
Conductor Size 2.0”(50mm) maximum
Measurement parameters ACA, DCA, ACV, DCV, Resistance, Diode, Frequency,
Capacitance, Duty Cycle, Continuity
Current Sensor Hall Effect
Zero adjust Automatic except for DCA (Push-button)
Diode Test Test current of 0.6mA typical;
Open circuit voltage <1.6V DC typical.
Battery type 9V NEDA 1604
Range Selection Auto or manual
Display 0.6” (15mm) 5000 Count Backlit LCD
Overload Indication "OL"
Power Consumption 5mA approx.
Low Battery Indication Battery icon appears on LCD
Sampling rate 1 reading every 0.35 seconds approx.
Standards CE, IEC 1010-1 CAT III 1000V and CAT IV 600V
Operating Temperature/Humidity 32 to 122oF (0oC to 50oC) / <80%
Dimensions 10 x 2.9 x 1.5” (255 x 73 x 38mm)
Weight 0.85 lbs. (380g)
Maximum Rated Input Limits
Function Maximum Rated Input Limits
AC/DC Current 500mA AC/DC (fused)
AC Voltage 1000 VAC/DC
DC Voltage 1000 VAC/DC
Resistance 400 VAC/DC
Frequency 1000 VAC/DC
Capacitance 400 VAC/DC
Duty Cycle 1000 VAC/DC
Diode Test 400 VAC/DC
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com
Table of contents
Other FLIR Measuring Instrument manuals

FLIR
FLIR Extech HDV-TX2 User manual

FLIR
FLIR EXTECH HDV-WTX2 User manual

FLIR
FLIR Si124 User manual

FLIR
FLIR identiFINDER R300 User manual

FLIR
FLIR CM72 User manual

FLIR
FLIR MR160 User manual

FLIR
FLIR Extech Instruments MA410 User manual

FLIR
FLIR MR160 User manual

FLIR
FLIR EM54 User manual

FLIR
FLIR VP50 User manual
Popular Measuring Instrument manuals by other brands

Hanna Instruments
Hanna Instruments HI6221 instruction manual

Felix Instruments
Felix Instruments F-751 manual

Micromeritics
Micromeritics FLEX Series Pre-installation instructions

Pyle
Pyle PLMT56 instruction manual

Omicron Lab
Omicron Lab CT Analyzer user manual

Sea-Bird Scientific
Sea-Bird Scientific SBE 11plus V2 user manual