TPS WP-80D User manual

Congratulations !
You have purchased the latest in Handheld pH-ORP-Temperature
instrumentation. We trust that your new WP-80D will give you many
years of reliable service.
The WP-80D is a breeze to operate. This manual has been designed to
help you get started, and also contains some handy application tips. If at
any stage you require assistance, please contact either your local TPS
representative or the TPS factory in Brisbane.
The manual is divided into the following sections:
1. Table of Contents
Each major section of the handbook is clearly listed. Sub-sections
have also been included to enable you to find the information you
need at a glance.
2. Introduction
The introduction has a diagram and explanation of the display and
controls of the WP-80D. It also contains a full listing of all of the items
that you should have received with your WP-80D. Please take the
time to read this section, as it explains some of items that are
mentioned in subsequent sections.
3. Main Section
The main section of the handbook provides complete details of the
WP-80D, including operating modes, calibration, troubleshooting,
specifications, and warranty terms.
4. Appendices
Appendices containing background information and application notes
are provided at the back of this manual.
Model WP
-
80D
pH, ORP, Temp Meter
Version: 2.0
Date: 1-Jul-2012

Page 2
TPS Pty Ltd
ABN 30 009 773 371
4 Jamberoo Street
Springwood, Brisbane,
Australia, 4127
Phone : (07) 32 900 400
International : 61 7 32 900 400
Fax : (07) 3808 4871
International : 61 7 3808 4871
Web : www.tps.com.au

Page
3
Contents
1. Introduction .....................................................................................5
1.1
WP-80D Display and Controls ............................................................... 5
1.2
Unpacking Information ........................................................................... 7
1.3
Specifications......................................................................................... 8
2. WP-80D Menu Structure................................................................10
3. Operating Modes...........................................................................11
3.1
Selecting pH, mV/ORP or Relative mV/ORP Mode............................. 11
4. pH Calibration................................................................................13
4.1
Calibration Procedure .......................................................................... 13
4.2
Calibration Notes ................................................................................. 15
4.3
Calibration Messages .......................................................................... 15
5. mV/ORP Calibration ......................................................................16
6. Relative mV/ORP Calibration........................................................16
6.1
Notes.................................................................................................... 17
7. Temperature Calibration...............................................................18
7.1
Calibration Procedure .......................................................................... 18
7.2
Calibration Notes ................................................................................. 18
7.3
Calibration Messages .......................................................................... 19
7.4
Manual Temperature Setting ............................................................... 19
8. Good Laboratory Practices (GLP)................................................20
8.1
To recall GLP information on the display............................................. 20
8.2
Failed Calibration ................................................................................. 21
8.3
Printing GLP Information to the RS232 Port........................................ 21
8.4
Instrument Serial Number.................................................................... 22
8.5
Additional GLP Features...................................................................... 22
9. Notepad Function..........................................................................23
9.1
Recording Readings into the Notepad................................................. 23
9.2
Recalling Records from the Notepad................................................... 23
9.3
Erasing Records from the Notepad ..................................................... 24
9.4
Printing Records from the Notepad to the RS232 Port........................ 24
10. Automatic Data logging.............................................................25
11. RS232 Port .................................................................................27
11.1
Setting the Baud Rate...................................................................... 27
11.2
Sending Readings to the RS232 Port.............................................. 27
11.3
RS232 Configuration........................................................................ 27
11.4
Communication and Statistical Software ......................................... 27
11.5
Commands....................................................................................... 28
11.6
Data Format ..................................................................................... 29

Page 4
11.7
GLP Data Format ............................................................................. 30
11.8
Importing Data into Microsoft Excel ................................................. 31
12. Battery Saver Function ............................................................. 33
13. Recharging the Battery............................................................. 34
14. Clock Function .......................................................................... 35
14.1
Setting the Clock .............................................................................. 35
14.2
Displaying or Hiding the Clock ......................................................... 35
15. Selecting Buffers for Auto Buffer Recognition ....................... 36
16. Initialising the WP-80 ................................................................ 37
17. Instrument firmware version number. ..................................... 37
18. Troubleshooting........................................................................ 38
18.1
General Errors.................................................................................. 38
18.2
pH and mV/ORP Troubleshooting ................................................... 39
18.3
Temperature Troubleshooting.......................................................... 40
19. Appendices: pH Sensor Theory .............................................. 41
19.1
Asymmetry of a pH Sensor .............................................................. 41
19.2
Slope of a pH Sensor ....................................................................... 42
19.3
pH Temperature Compensation....................................................... 43
20. Warranty .................................................................................... 44

Page
5
1. Introduction
1.1 WP-80D Display and Controls

Page 6
Press to record readings into memory. See section 9.1.
Also used to select primary and secondary buffers for automatic
buffer recognition. See section 15.
Press to show or hide the date and time. See section 14.2.
Press to start or stop automatic logging. See section 10.
Alternatively, press to transmit current reading plus date and time to
the RS232 port. See section 11.2.
Press to zero relative mV/ORP, when relative mV/ORP mode is
selected. See section 6.
Press to access the user-friendly menu system which makes the
WP-80D a breeze to operate.
and
The and keys are used when calibrating temperature readout
(section 7), setting manual temperature compensation (section 7.4),
setting the clock (section 14.1), setting the automatic logging period
(section 10), and displaying GLP information (section 8.1).
The key is also used to initialise the WP-80D at turn-on. See
section 16.
Switches the WP-80D on and off.
Display
32 character alpha-numeric display with user-friendly menu and
prompting system. Shows pH/mV/ORP and temperature
simultaneously. Date and time can also be displayed.

Page
7
1.2 Unpacking Information
Part No
1. WP-80D dual Channel pH-ORP-Temperature Instrument...... 121110/1
................................................................................................... 121110/3
................................................................................................... 121110/5
2. pH Sensor:
(see cable label for part No)
1, 3 or 5m cable
3. Temperature Sensor:
(see cable label for part No)
1, 3 or 5m cable
4. pH7.00 Buffer, 200mL............................................................ 121387
5. pH4.01 Buffer, 200mL............................................................ 121381
6. Right angle BNC adapter....................................................... JSBNCR
7. Battery charger ...................................................................... 130037
8. Manual
Options that may have been ordered with your WP-80D:
1. Extended cable...................................................................... 130040
2. RS232 Serial Interface Cable................................................. 130041
3. Communication software for Windows 95 and later ............... 130086
4. USB to Serial Adaptor (requires 130041 also) ....................... 130087
5. Hard Carry Case.................................................................... 130059
6. Battery charger lead for 12V cigarette lighter socket.............. 130046
7. Battery charger lead for 12V DC, with battery clips ................ 130052
8. Solar Panel............................................................................ 130012

Page 8
1.3 Specifications
pH
Range ..................................0 to 14.00 pH
Resolution ............................0.01 pH
Accuracy ..............................±0.01 pH
mV/ORP
Range ..................................0 to ±500.0 and 0 to ±1500 mV
(auto-ranging)
Resolution ............................0.15 and 1 mV
Accuracy ..............................±0.3 and ±1 mV
Temperature
Range ..................................-10.0 to +120.0
o
C
Resolution ............................0.1
o
C
Accuracy ..............................±0.2
o
C
General Specifications
Temperature Compensation:
pH
(only – not for mV/ORP)
.....Automatic 0 to 100
o
C
Manual 0 to 100
o
C
Calibration:
pH ........................................Auto Standard Recognition in pH4.01,
pH6.86, pH7.00, pH9.18 or pH10.01
buffers
pH Sensor Asymmetry Range
-1.00 to +1.00 pH
pH Sensor Slope Range.......85.0 to 105%
Input Impedance ..................>3 x 10
12
Ω
mV/ORP...............................Sensor calibration not available.
Temperature.........................Use a reference thermometer.
Temperature Sensor Offset ..-10.0
o
C to +10.0
o
C

Page
9
Memory.................................... 3600 readings including date and time
Automatic Logging....................
User-set for one reading every 1 to 90
seconds, 1 to 90 minutes, or 1 to 24 hours.
RS232 Output ......................... 300, 1200, 9600 & 19200 baud. 8 bits, no
parity, 1 stop bit, XON/XOFF Protocol.
Clock........................................ Calendar clock displays date, month,
hours, minutes & seconds.
Year is recorded in memory and
transmitted to the RS232 port, but is not
displayed.
Battery Saver ........................... On : Auto switch-off after 5 minutes
Off : Continuous use
Bar Graph display of battery charge level.
Readout of battery voltage available for
troubleshooting.
Good Laboratory Practices....... Date, Time and Value of last pH and
Temperature calibration are stored, and
can be recalled or sent to the RS232 port
at any time.
Power....................................... 6V NiMH Rechargeable Battery for
approx 50 hours operation.
Dimensions .............................. 195 x 110 x 55 mm
Mass ........................................ Instrument only : Approx 520g
Full Kit : Approx 2.5kg
Environment............................. Temperature : 0 to 45
o
C
Humidity : 0 to 95 % R.H.

Page 10
2. WP-80D Menu Structure
A detailed breakdown of the menu system of the WP-80D is shown below
.
This diagram provides a quick reference for the WP-80D menu functions.
→F1:Cal. →F1:pH1 (available when Channel 1 in pH mode)
F2:pH2 (available when Channel 2 in pH mode)
F3:Temp
→F2:Notepad →F1:Recall
F2:Erase
F3:Print
F4:Prog. →F1:Min
F2:Sec
F3:Hr
↑↓:Set Period
→F3:Mode →F1:Chan1 →F1:pH.
F2:mV/ORP
F3:Rel mV/ORP
→F3:Mode →F2:Chan2 →F1:pH.
F2:mV/ORP/
F3:Rel mV/ORP
→F4:Setup →F1:Batt →F1:OFF
(Battery Saver)
F2:ON
(Battery Saver)
F2:Clock
F3:GLP →F1:Recall
F3:Print
F4:Baud →F1:300
F2:1200
F3:9600
F4:19200

Page
11
3. Operating Modes
The WP-80D has two channels. Each channel can be individually set up
for pH, mV/ORP or Relative mV/ORP mode.
3.1 Selecting pH, mV/ORP or Relative mV/ORP Mode
To select pH, mV/ORP or Relative mV/ORP mode…
1. Select the Mode menu ( , then →
→→
→F3:Mode)…
Select Channel
F1:Chan1 F2:Chan2
2. Select F1:Chan1 for Channel 1 or F2:Chan2 for Channel 2…
F1:pH
F2:mV/ORP
F3:Rel mV/ORP
F4:OFF
2. Press to select pH mode.
Press to select mV/ORP mode.
Press to select Relative mV/ORP mode.
Press to turn that channel off.
Press to quit and retain the current selection.
3. The WP-80D will now display the selected modes, along with
Temperature and the Time. . eg.
7.00pH 500.0m
25.0
12:00:00
Notes
1. If Channel 1 is set to Relative mV/ORP, then Temperature is NOT
displayed.
2. Temperature compensation does not apply in mV/ORP mode.
3. If the temperature sensor is unplugged, and one or both of the
channels is set for pH, the manual temperature setting is displayed
with 1
o
C resolution.
4. If the temperature sensor is unplugged, and neither of the channels is
set for pH, no temperature value is displayed.

Page 12
5. The decimal point is replaced by a ∗
∗∗
∗if a pH or Temperature
calibration has failed (sections 4.1, and 0) if the unit is initialised
(section 16), or if the unit has lost its factory calibration (section 18.1).

Page
13
4. pH Calibration
A “ ∗” in place of the decimal point indicates that the pH readout is not
calibrated, or a past calibration has failed. The “ ∗” will be removed once
a full two-point pH calibration has been successfully performed.
4.1 Calibration Procedure
1. Plug the pH sensor into the Channel 1 or Channel 2 socket and the
temperature sensor into the Temperature socket.
2. Switch the meter on.
3. Ensure that you are using buffers which have been selected for
automatic buffer recognition. See section 15 for a detailed
explanation.
4. Select pH Mode.
Either ( →
→→
→F3:Mode →
→→
→F1:Chan1 →
→→
→F1:pH)
or ( →
→→
→F3:Mode →
→→
→F2:Chan2 →
→→
→F1:pH)
5. Ensure that temperature has already been calibrated, or manually set
(see sections 7.1 and 7.4). NOTE: If the decimal point in the
temperature reading is shown by a "∗
∗∗
∗", then the temperature readout
is not calibrated.
6. Remove the wetting cap from the pH sensor.
7. Rinse the pH and Temperature sensors in distilled water and blot
them dry.
8. Place both electrodes into a small sample of pH7.00 (or pH6.86)
buffer, so that the bulb and reference junction are both covered, as
per the diagram over the page.
DO NOT place the electrodes directly into the buffer bottle.
9. Select pH Calibration.
Either →
→→
→F1:Cal. →
→→
→F1:pH1
or →
→→
→F1:Cal. →
→→
→F2:pH2

Page 14
10. When the reading has stabilised, press the key to calibrate. If a 1
point calibration has been performed, the "∗
∗∗
∗" will not be removed until
a full 2 point calibration has been performed.
11. Rinse the pH and Temperature electrodes in distilled water and blot
them dry.
12. Place both sensors into a small sample of pH4.01, pH9.18 or pH10.01
Buffer, so that the bulb and reference junction are both covered, as
per the diagram in step 8, above. DO NOT place the electrodes
directly into the buffer bottle.
pH9.18 and pH10.01 buffers are unstable once the bottles have
been opened. Discard immediately after use.
13. Select pH Calibration for the same channel as calibrated in step 9.
Either →
→→
→F1:Cal. →
→→
→F1:pH1
or →
→→
→F1:Cal. →
→→
→F2:pH2
14. When the reading has stabilised, press the key to calibrate. The
"∗" will now be replaced by a decimal point, if calibration was
successful.
15. The WP-80D is calibrated and ready for use in this mode. Discard the
used samples of buffer.

Page
15
4.2 Calibration Notes
1. A 1-point calibration should be performed at least weekly. In
applications where the electrode junction can become blocked, such
as dairy products, mining slurries etc, a 1-point calibration may have to
be done daily.
2. A full 2-point calibration should be performed at least monthly. Of
course, more frequent calibration will result in greater confidence in
results.
3. All calibration information is retained in memory when the WP-80D is
switched off, even when the battery is removed. This information can
be recalled or printed later using the GLP function (see section 8).
4. The WP-80D displays the value of the pH buffer to which it will attempt
to calibrate. Ensure that the buffer value displayed corresponds to the
buffer that you are using.
4.3 Calibration Messages
1. If a 1-point calibration has been successfully performed, the WP-80D
will display the following message, and the asymmetry of the
electrode. For example…
1 Point Cal.OK
Asy= 0.10pH
2. If a 1-point calibration has failed, the WP-80D will display the following
message, and the failed asymmetry value of the electrode. For
example:
1 Point Cal.Fail
Asy= 1.50pH Hi
or :
1 Point Cal.Fail
Asy=-1.50pH Lo
3. If a 2-point calibration has been successfully performed, the WP-80D
will display the following message, and the asymmetry and slope of the
electrode. For example…
2 Point Cal.OK
Asy= 0.10pH
then :
2 Point Cal.OK
Slope=100.0%
4. If a 2-point calibration has failed, the WP-80D will display the following
message, and the failed slope value of the electrode. For example…
2 Point Cal.Fail
Slope=130.0% Hi
or :
2 Point Cal.Fail
Slope= 70.0% Lo

Page 16
5. mV/ORP Calibration
The mV/ORP section is factory calibrated. There is no user-calibration
facility for this mode.
6. Relative mV/ORP Calibration
Select Relative mV/ORP mode when measurements relative to a known
standard are required. Calibration of the Relative mV/ORP mode is simply
a matter of zeroing the reading when the sensor is in the known standard.
1. Plug the Redox sensor into the Channel 1 or Channel 2 socket.
Temperature compensation is not applied in Relative mV/ORP mode,
so the temperature sensor does not need to be connected.
2. Switch the meter on.
3. Select Relative mV/ORP Mode
Either ( →
→→
→F3:Mode →
→→
→F1:Chan1 →
→→
→F3:Rel mV/ORP).
or ( →
→→
→F3:Mode →
→→
→F2:Chan2 →
→→
→F3:Rel mV/ORP).
4. The display should now be showing Relative mV/ORP on the top line
with absolute mV/ORP or date and time on the bottom line. For
example…
500.0R 7.00pH
660 F4:Zeros
↔
500.0R 7.00pH
31/12 12:00:00
Press to alternate between absolute mV/ORP or the clock on the
bottom line.
5. Remove the wetting cap from the Redox sensor.
6. Rinse the sensor in distilled water and blot dry.
7. Place the Redox sensor into a sample of the known standard. Ensure
that the platinum tip and reference junction are both covered, as per
the diagram over the page.

Page
17
8. When the reading has stabilised, press the key to zero the Relative
mV/ORP reading. The Relative mV/ORP reading will now be zero, and
the absolute mV/ORP reading will remain unchanged. For example…
0.0R 7.00pH
660
F4:Zeros
or: 0.0R 7.00pH
31/12 12:00:00
9. The WP-80D Relative mV/ORP mode is now zeroed and is ready for
use. The readout can be re-zeroed by pressing the key whenever
required.
6.1 Notes
1. If Relative mV/ORP is selected for BOTH channels, then the WP-80D
offers you a choice between Channel 1 or Channel 2 when you press
the key to zero the Relative mV/ORP reading.
Select Channel
F1:Chan1 F2:Chan2
2. The Relative mV/ORP offset is retained in memory when the WP-80D
is switched off, even when the battery is removed.
3. The Relative mV/ORP zero is reset when leaving Relative mV/ORP
mode (see section 3)

Page 18
7. Temperature Calibration
A “ ∗” in place of the decimal point indicates that the Temperature
readout is not calibrated, or a past calibration has failed. The “ ∗” will be
removed once Temperature has been successfully calibrated.
7.1 Calibration Procedure
1. Plug the temperature sensor into the Temperature socket.
2. Switch the meter on.
3. Place the sensor into a beaker of room temperature water, alongside
a good quality mercury thermometer. Stir the probe and the
thermometer gently to ensure an even temperature throughout the
beaker.
4. Select Temperature Calibration ( →
→→
→F1:Cal. →
→→
→F3:Temp).
5. The reading from the probe is now displayed on the right of the
display, and the value you are going to set is shown on the left. For
example…
> 25.0< 20*0
↑
↑↑
↑↓
↓↓
↓:Set F1:Cal.
6. When the reading on the right has stabilised, press the and
keys until the reading on the left shows the same temperature as the
mercury thermometer.
7. Press the key to calibrate the temperature readout.
Alternatively, press the key to abort temperature calibration.
7.2 Calibration Notes
1. Temperature calibration information is retained in memory when the
WP-80D is switched off, even when the battery is removed. This
information can be recalled later using the GLP function (see section
8).
2. Temperature does not need to be recalibrated unless the
Temperature probe is replaced or the meter is initialised.

Page
19
7.3 Calibration Messages
1. If a temperature calibration has been successfully performed, the
WP-80D will display the following message and the offset value of the
probe. For example:
Calibrate OK
Offset= 1.0
2. If a temperature calibration has failed, the WP-80D will display the
following message, and the failed offset value of the probe. For
example…
Calibrate Failed
Offset= 10.5
7.4 Manual Temperature Setting
If the temperature sensor is not connected, the temperature of the sample
solution must be set manually for accurate pH measurements. A separate
thermometer will be required for this.
1. Switch the meter on.
2. Measure the temperature of the sample.
3. Select Temperature Calibration ( →
→→
→F1:Cal. →
→→
→F3:Temp).
4. The current temperature setting is now displayed. For example…
> 25.0< Man Temp
↑
↑↑
↑↓
↓↓
↓:Set F1:Save
5. Press the and keys until the display shows the temperature of
the sample.
6. Press the key to save the temperature value.
Alternatively, press the key to quit and retain the current setting.

Page 20
8. Good Laboratory Practices (GLP)
The WP-80D keeps a record of the date and time of the last pH
asymmetry, pH slope and Temperature offset calibrations as part of GLP
guidelines.
8.1 To recall GLP information on the display
1. Switch the meter on.
2. Select the GLP menu ( →
→→
→F4:Setup →
→→
→F3:GLP).
3. Select F1:Recall from the menu.
4. The instrument model, firmware version number, and instrument serial
number are displayed, along with a prompt describing how to scroll
through the GLP information. For example…
WP80D V2.0 S1234
↓
↓↓
↓:More ↑
↑↑
↑:Back
5. Press the key to sequentially scroll through the GLP information
for all parameters. Press the key to scroll back to previous data.
The sequence of information displayed is shown below. Press to
abort at any time.
WP80D V2.0 S1234
↓
↓↓
↓:More ↑
↑↑
↑:Back
:↓
↓↓
↓:↑
↑↑
↑
pH Asy 0.10pH
@ 31/12/10 11:40
:↓
↓↓
↓:↑
↑↑
↑
pH Slope 100.0%
@ 31/12/10 11:50
:↓
↓↓
↓:↑
↑↑
↑
Temp Offset 0.1
@ 31/12/10 12:00
:↓
↓↓
↓:↑
↑↑
↑
Exit