Trotec T3000 User manual

TRT-BA-T3000-TC2014-04-007-EN
T3000
EN
OPERATING MANUAL
MULTIFUNCTION MEASURING
METER

2 EN
multifunction measuring meter T3000
Table of contents
Notes regarding the operating manual.................................2
Safety .....................................................................................2
Information about the device................................................4
Transport and storage...........................................................6
Operation ...............................................................................6
PC software..........................................................................22
Maintenance and repair ......................................................23
Errors and faults..................................................................23
Using the trace gas sensor TS800SDI (optional)..............24
Using the trace gas sensor TS810SDI (optional)..............28
Available SDI sensors..........................................................32
Available BNC electrodes ....................................................33
Further accessories (optional)............................................35
Disposal ...............................................................................35
Notes regarding the operating manual
Symbols
Warning of electrical voltage
This symbol indicates dangers to the life and health of
persons due to electrical voltage.
Warning
This signal word indicates a hazard with an average
risk level which, if not avoided, can result in serious
injury or death.
Caution
This signal word indicates a hazard with a low risk
level which, if not avoided, can result in minor or
moderate injury.
Note
This signal word indicates important information (e.g.
material damage), but does not indicate hazards.
Info
Information marked with this symbol helps you to carry
out your tasks quickly and safely.
Follow the manual
Information marked with this symbol indicates that the
operating manual must be observed.
You can download the current version of the operating manual
and the EU declaration of conformity via the following link:
T3000
https://hub.trotec.com/?id=44146
Safety
Read this manual carefully before starting or using the
device. Always store the manual in the immediate vicinity
of the device or its site of use!
Warning
Read all safety warnings and all instructions.
Failure to follow the warnings and instructions may
result in electric shock, fire and/ or serious injury.
Save all warnings and instructions for future
reference.
This appliance can be used by children aged from
8years and above and persons with reduced physical,
sensory or mental capabilities or lack of experience
and knowledge if they have been given supervision or
instruction concerning use of the appliance in a safe
way and understand the hazards involved.
Children shall not play with the appliance. Cleaning and
user maintenance shall not be made by children
without supervision.
• Do not use the device in potentially explosive rooms.
• Do not use the device in aggressive atmosphere.
• Protect the device from permanent direct sunlight.
• Do not remove any safety signs, stickers or labels from the
device. Keep all safety signs, stickers and labels in legible
condition.
• Do not open the device.

EN 3
multifunction measuring meter T3000
Intended use
Only use the multifunction measuring meter for preventive
maintenance and risk analysis in the fields of material,
construction and climate diagnostics. Observe and comply with
the storage and operating conditions according to the Technical
data chapter.
Intended use comprises:
• the analysis of:
– supply and exhaust air flows
– fluctuations in relative humidity
– condensate formation
– poor machine cooling
– porous seals
– heat build-up and temperature fluctuations
– excessively dry or damp materials
• leak detection with trace gas
To use the device for its intended use, only use accessories and
spare parts which have been approved by Trotec.
Improper use
Do not use the device in potentially explosive atmospheres,
when whet or for measurements in liquids.
Any unauthorised changes, modifications or alterations to the
device are forbidden.
Personnel qualifications
People who use this device must:
• take measures to protect themselves from direct contact
with live parts.
• have read and understood the operating manual, especially
the Safety chapter.
Residual risks
Warning of electrical voltage
There is a risk of a short-circuit due to liquids
penetrating the housing!
Do not immerse the device and the accessories in
water. Make sure that no water or other liquids can
enter the housing.
Warning of electrical voltage
Work on the electrical components must only be
carried out by an authorised specialist company!
Warning
Risk of suffocation!
Do not leave the packaging lying around. Children may
use it as a dangerous toy.
Warning
The device is not a toy and does not belong in the
hands of children.
Warning
Dangers can occur at the device when it is used by
untrained people in an unprofessional or improper way!
Observe the personnel qualifications!
Caution
Keep a sufficient distance from heat sources.
Note
To prevent damages to the device, do not expose it to
extreme temperatures, extreme humidity or moisture.
Note
Do not use abrasive cleaners or solvents to clean the
device.

4 EN
multifunction measuring meter T3000
Information about the device
Device description
The T3000 multifunction measuring meter is used for carrying
out detailed analysis in the fields of material and construction
diagnostics. To do so, the device has a 24 bit analogue / digital
converter which delivers precise results and is permanently
stable compared with analogue measuring instruments.
The operating elements are found on the front and sides of the
robust housing. A scratch proof colour display with touch
function(1) allows entering and selecting values and functions
and also displaying detected results. You can also navigate the
device software for selecting measuring functions by using the
cross control(3) and the "OK" key(5).
Pressing the "Back" key(2) returns you to the previous screen
of the device software. Pressing the "Main menu" key(6) opens
the main menu which is also shown when you start the device.
Pressing the "Illumination on/off" key(4) switches the
background illumination for the colour display and the keys on
and off and also serves for dimming.
The sensor connections (one SDI connection(7) and one BNC
connection(8)) can be found above the colour display(1). At the
SDI connection(7), connect the appropriate digital SDI sensor
for the corresponding situation. The universal interface of the
SDI sensors allows the multifunction measuring meter to
automatically detect the sensor construction and/or indicates
when the wrong type of sensor is connected. Depending on the
sensor construction, it is calibrated to the prevailing surrounding
conditions after connection. Any sensor settings are saved
directly to the corresponding SDI sensor. The various measured
values, such as relative humidity, dew point or flow rate, are
automatically calculated by the SDI sensors and transmitted to
the multifunction measuring meter.
At the BNC connection (8), you connect passive wood or building
moisture sensors.
The measured results of the sensor which is connected to the
SDI or BNC connection are shown in the device's colour display
(1).
The "Power" button(10) and a USB connection(11) can be
found on the sides of the device.
You can connect the device to a computer by using the supplied
USB connection cable. Then you can extract and analyse your
measured results with the optional MultiMeasure Studio
Standard or MultiMeasure Studio Professional software. Contact
your Trotec customer service.
Device depiction
1
2
3
4
5
6
7
8
9
10
11
No. Operating element
1 Scratch proof colour display with touch function
2 "Back" key
3 Cross control with "Up", "Down", "Left", "Right" keys
4 "Illumination on/off" key
5 "OK" key
6 "Main menu" key
7 SDI connection (5-pin)
8 BNC connection
9 Battery compartment with cover
10 "Power" button
11 USB port

EN 5
multifunction measuring meter T3000
Technical data
Parameter Value
Model T3000
Dimensions (lengthx widthx
height)
approx. 170 x 62 x 34 mm
Operating range
Operating temperature 0°C to +50°C
Relative humidity <90% or <20g/m3 (non-condensing)
Height above sea level (SL) max. 3000m
Storage
Relative humidity < 95% (non-condensing)
Ambient temperature -20°C to +60°C
Energy
Battery 4x Alkaline LR6AA, 1.5 V batteries
Optional power supply 5 V USB
Power input, active Approx. 400mW
Battery lifespan, passive approx. 1 year
Battery lifespan, active at least 24 h
Sensor supply 5.5 V ±10 % DC, 200 mA max.
Compatibility
Sensors* SDI sensors (digital) Air temperature and humidity
Material temperature
Material moisture
Air current
H2 trace gas detection
BNC electrodes Material moisture (construction materials and wood)
according to the resistance measuring method
Measured value memory for approx.200 measuring projects consisting of up to 3x3,600 (=10,800) measured
values, that is a total of 200x10,800 = 2,160,000 measured values
Amount of tiles in the grid max. 50 x 40 (max. amount of grid fields in one measurement)
Real time clock with calender programmed to 2099
Trace gas sensor (optional) TS 800 SDI TS 810 SDI
Response sensitivity 1 ppm H21 ppm H2
Measuring range 0to1000ppmH210to1000ppm H2
Response time <1s <1s
Pore size at stainless steel
sinter filter
>50μm >50μm
Power supply 11 – 14 V 5 – 5.5 V
Battery ("BatteryPack") 12 V / 2Ah lead battery -
Current consumption
active
passive
approx.400 mA
approx.20mA (sensor
inactive)
-
-
Ambient temperature
Operation
Storage
0 °C to +50 °C
-20 °C to +50 °C
10 °C to +60 °C
-20 °C to +50 °C
Relative humidity <95 % (non-condensing) <95 % (non-condensing)
* Data changes reserved.

6 EN
multifunction measuring meter T3000
Scope of delivery
• 1x Multifunction measuring meter T3000
• 1 x USB connection cable
• 4x Alkaline LR6 AA, 1.5 V batteries
• 1 x Display protection film
• 1 x Factory test certificate
• MultiMeasure Studio Standard PC software (to be
downloaded from www.trotec.de under the "Downloads"
menu)
• 1x Manual
Transport and storage
Note
If you store or transport the device improperly, the
device may be damaged.
Note the information regarding transport and storage of
the device.
Transport
To safely transport the multifunction measuring meter and
accessories, use the optional MultiMeasure transport case(12)
recommended by Trotec. Otherwise, protect the device during
use and transport with an optional device bag (13). Contact your
Trotec customer service.
12
13
Storage
When the device is not being used, observe the following
storage conditions:
• dry and protected from frost and heat
• protected from dust and direct sunlight
• The storage temperature is the same as the range given in
the Technical data chapter.
• Remove the batteries from the device.
• To store the device, use the optional transport case (see
chapter "Transport and storage") wherever possible.
Operation
Inserting the batteries
Insert the supplied batteries before first use.
Caution
Make sure that the surface of the device is dry and the
device is switched off.
9
1. Remove the cover of the battery compartment(9).
2. Insert the batteries with correct polarity.
3. Insert the cover of the battery compartment(9).
ðThe device can now be switched on.
Switching the device on
1. Press and hold the Power button(10).
ðAn acoustic signal is emitted.
2. Hold the button until a second beep is emitted. Then,
release the "Power" button(10).
ðAfter approx. 3 seconds, the colour display will be
switched on. The device is ready to go as soon as the
main menu is shown.

EN 7
multifunction measuring meter T3000
Switching the device off
1. Press and hold the Power button(10).
ðAn acoustic signal is emitted.
2. Hold the button until a second beep is emitted. Then,
release the "Power" button(10).
ðThe device is switched off.
Overview and navigation
When using the device, take special note of the following
important operating elements and displays:
15
16 17 18 19
62
20
Item Function
2 opens the previous menu
6 opens the main menu
15 name of current screen
16 indicates the connected sensor type(SDI/BNC/NTC)
17 display of current time
18 locking/unlocking the touch function:
press the Power button(10) for1s
19 background illumination indication:
press the Illumination on/off key(4)
20 battery loading status indicator/plug symbol when
charged via USB
All options which can be selected via touch function can also be
selected by using the cross control(3) and the "OK"key(5). For
safety reasons, some options can only be selected and
configured by using the cross control and the "OK"key(5) (e.g.
date and time in the Settings screen).
Activating large print
The current measured values shown in the upper half of the
display can also be displayed in large print.
1. Briefly press the "On/Off"key(10) in any measuring mode
screen.
2. To exit the large print display, press the "OK" button.
Info
When the key lock is enabled, large print cannot be
activated or deactivated.
Setting the language
1. Press the Settings key in the main menu.
2. Press the language selection key(21) in the "Settings"
screen.
21

8 EN
multifunction measuring meter T3000
3. Press the key with the desired language from the following
screen.
ðThe selected screen language is activated immediately.
4. Press the "Main menu" key(6) to open the main menu.
Main menu
You can open the following submenus from the main menu:
Submenu Function
Measuring mode Carrying out measurements
Archive open archived measured values
Settings make any device settings
submenu measuring mode
The following measuring methods are available in the
Measuring mode submenu:
Measuring
method
Function
Wood moisture measure moisture in wood (with BNC
electrode)
Building moisture measure moisture in walls and floors (with
SDI or BNC electrodes)
Flowrate measure air/gas flows (with SDI sensor)
Humidity measure humidity (with SDI sensor)
Temperature measure temperature (with SDI sensor)
Hydrogen measure hydrogen content (trace gas; with
SDI sensor)
"Measuring mode – Wood moisture" screen
22
28 29 30 31
24
25
23
27
26

EN 9
multifunction measuring meter T3000
Item Function
22 Opens the menu for selecting the wood type.
To change the wood type, please proceed as follows:
1. Tap the displayed type of wood(22).
2. Press the "OK" button(5) to activate the "wood
type" input box.
3. Press the "Up and "Down" keys on the cross
control(3) to select the initial letter of the desired
type of wood.
4. Press the "Right" key to select any other letters of
the desired type of wood.
5. confirm the setting by pressing the OK button(5)
6. Use the "Down" key to navigate to the search result
list of "wood types" or use the "Left" and "Right"
keys to scan forwards or backwards through the
"wood type" list page by page.
7. Confirm the selected wood type by pressing the
"OK" button.
23 M%:
Shows the wood moisture as a mass percentage (relative
mass proportion of the humidity to the total mass).
24 Opens the menu for selecting the internal temperature
sensor(INT) or for entering a fixed temperature
value(FIX). The measured or manually specified
temperature value is shown on the right.
25 Resets the measured values shown in item27 to zero.
26 Opens the menu for selecting the indication of measured
values.
Minimum: Indicates the smallest detected measured
value from a measuring period.
Maximum: Indicates the largest detected measured
value from a measuring period.
Average: Shows the average value of all detected
measured values from a measuring period.
Hold: Pauses the current detected measured value and
shows it continually (when this option is selected).
Off: Switches off the measured values in item27.
27 Measurement value display according to selection in
item26:
Minute: Duration of the measured interval
M%: Shows the measured value (e.g. as average value)
as a mass percentage.
°C: Shows the measured (INT) or manually specified
temperature values (FIX).
28 Opens the menu for carrying out a grid measurement
(see section Measuring method wood moisture– Grid
measurement).
Item Function
29 Opens the menu for setting the measuring range in
heightx width (edge lengths) and their activation/
deactivation as well as the specified temperature value
(FIX)
(see section Measuring method wood moisture–
Measuring range).
30 Opens the menu for setting a fixed recording duration for
an automatic measurement
(see section Measuring mode wood moisture–
Automatic measurement).
31 Saves the measured values in the archive.
"Measuring mode– Wood moisture– Grid measurement"
screen
32
34 35 36 37
33
Item Function
32 digits:
dimensionless display of the measured value; the lower
the measured electric resistance, the higher the
moisture in the wood and therefore, the higher the digit
value
33 Measuring grid:
Use the grid measurement function to carry out a grid
measurement (e.g. of a wall). The grid measurement can
be carried out either with or without height and length
details, but the grid is always used. Please proceed as
follows:
1. Press the "OK"button(5) to select the grid.
2. Use the keys on the cross control to select the grid
measuring spot.
3. Press the "OK" button(5) or the button(34) to
confirm the selected grid measuring spot. => The
dimensionless measured value is recorded.
4. Use the keys of the cross control to select the next
grid measuring spot and repeat the process.

10 EN
multifunction measuring meter T3000
Item Function
34 Enters the current measured value to the marked grid
measuring spot. Afterwards, the grid measuring spot
below will be selected.
35 Skips the grid measuring spot below the currently
selected grid measuring spot and selects the grid
measuring spot that follows afterwards. Use this function
if no measurement can be made at this position because
of a power cabinet or a window etc.
36 Skips to the start of the next column.
37 Saves the measured values in the archive.
"Measuring mode – Wood moisture – Measuring range"
screen
38
40
41
42
39
Item Function
38 Specifies the total height of the grid area to be
measured. The value is used during grid measurement.
Use the "OK"button(5) to go to the settings. Afterwards,
you can adjust the values with the control keys and
move on to the next field.
39 Specifies the total width of the grid area to be measured.
The value is used during grid measurement.
Use the "OK"button(5) to go to the settings. Afterwards,
you can adjust the values with the control keys and
move on to the next field.
40 Activates ("On") or deactivates ("Off") the use of
specified edge lengths (38, 39) during grid
measurement. For deactivation, you must specify the
corresponding edge lengths of the grid area in the PC
software after grid measurement.
41 Specifies the temperature value(FIX) for the temperature
compensation which is necessary for accurate wood
moisture calculation if the integrated sensor is not being
used.
42 Saves the settings.
Info
The settings saved here are also used for all following
measurements, unless they are deactivated (in the
corresponding menu item)!
"Measuring mode– Wood moisture– Automatic
measurement" screen
44
45
43
46 47 48
Item Function
43 M%:
Shows the measured value as a mass percentage
(relative mass proportion of the humidity to the total
mass).
44 Selection for temperature compensation:
- internal sensor(INT)
- specified, fixed temperature value(FIX)
45 Opens a screen for selecting the recording duration for
measuring over a long period.
46 Starts recording.
The key turns red once recording has started.
47 Stops the current recording. The detected values are
automatically saved to the current measuring project.
48 Pauses the current recording.
Key 46 flashes.
Press this key or key46 again to continue recording.

EN 11
multifunction measuring meter T3000
"Measuring mode– Building moist" screen
52
54
50
55 56 57
49
51
53
Note
When using BNC electrodes, the measured value is
shown as the dimensionless unit digits.
When using dielectric measuring SDI sensors, the
measured value is shown as digits, M% or CM%.
Item Function
49 Opens the menu for selecting the construction material:
No material: a non-destructive, indicative (digit)
measurement is carried out
Anhydrite floor: A building moisture measurement for
anhydrite screed is carried out (with SDI sensor
TS660SDI only). The measurement results (indicative)
in M% and CM% are calculated with an equation which
is suitable for this material.
Cement floor: building moisture measurement for
cementitious screed (with SDI sensor TS660SDI only).
the measurement results (indicative) in M% and CM%
are calculated with an equation which is suitable for this
material
50 Indicates the building moisture:
digits: dimensionless display of the measured value (with
BNC electrode):
M%: indicates the measured value as a mass
percentage (with SDI sensor TS660SDI only)
CM%: indicates the measured value according to the
calcium carbide method
51 Resets the measured values shown in item53 to zero.
Item Function
52 Opens the menu for selecting the indication of measured
values.
Minimum: Indicates the smallest detected measured
value from a measuring period.
Maximum: Indicates the largest detected measured
value from a measuring period.
Average: Shows the average value of all detected
measured values from a measuring period.
Hold: Pauses the current detected measured value and
shows it continually (when this option is selected).
Off: Switches off the measured values in item53.
53 Measurement value display according to selection in
item52:
Minute: Duration of the measured interval
M%: Shows the measured value (e.g. as average value)
as a mass percentage.
CM%: Shows the measured value (e.g. as average value)
according to the calcium carbide method.
54 Opens the menu for carrying out a grid measurement
(see section Building moisture– Grid measurement).
55 Opens the menu for setting the measuring range in
heightx width (edge lengths) and their activation/
deactivation as well as the specified alarm value
(see section Building moisture– Measuring range/alarm
value).
56 Opens the menu for setting a fixed recording duration for
an automatic measurement.
(see section Measuring mode building moisture–
Automatic measurement).
57 Saves the measured values in the archive.

12 EN
multifunction measuring meter T3000
"Measuring mode – Building moisture – Grid
measurement" screen
60 61 62 63
58
59
Item Function
58 digits:
dimensionless display of the measured value; the lower
the measured electric resistance, the higher the
moisture in the material and therefore, the higher the
digit value
59 Measuring grid:
1. Press the "OK"button(5) to select the grid.
2. Use the keys on the cross control to select the grid
measuring spot.
3. Press the "OK" button(5) or the button(34) to
confirm the selected grid measuring spot.
ðThe dimensionless measured value is recorded.
4. Use the keys of the cross control to select the next
grid measuring spot and repeat the process.
60 Enters the current measured value to the marked grid
measuring spot. Afterwards, the grid measuring spot
below will be selected.
61 Skips the grid measuring spot below the currently
selected grid measuring spot and selects the grid
measuring spot that follows afterwards. Use this function
if no measurement can be made at this position because
of a power cabinet or a window etc.
62 Skips to the start of the next column.
63 Saves the measured values in the archive.
"Measuring mode – Building moisture – Measuring range/
alarm value" screen
66
67
68
64
65
Item Function
64 Specifies the total height of the grid area to be
measured. The value is used during grid measurement.
Use the "OK"button(5) to go to the settings. Afterwards,
you can adjust the values with the control keys and
move on to the next field.
65 Specifies the total width of the grid area to be measured.
The value is used during grid measurement.
Use the "OK"button(5) to go to the settings. Afterwards,
you can adjust the values with the control keys and
move on to the next field.
66 Activates ("On") or deactivates ("Off") the use of
specified edge lengths (64, 65) during grid
measurement. For deactivation, you must specify the
corresponding edge lengths of the grid area in the PC
software after grid measurement.
67 Only when using an SDIsensor:
Specifies the limit value for the alarm signal of the
dimensionless measuring unit "digits". Once the limit
value is reached, an alarm signal is emitted.
If no SDI sensor is used, the field's background colour
will be grey and the field cannot be selected.
68 Saves the settings.
Info
The settings saved here are also used for all following
measurements, unless they are deactivated (in the
corresponding menu item)!

EN 13
multifunction measuring meter T3000
"Measuring mode – Building moisture – Automatic
measurement" screen
70
71
69
72 73 74
Item Function
69 Building moisture indication according to the selected
material.
No material: digits
Anhydrite screed or cement screed: M%, CM%
70 Shows the remaining time until the automatic
measurement finishes.
71 Opens a screen for selecting the recording duration for
measuring over a long period.
Once you confirmed your selection, the selected
remaining time will be adopted in the display(item70).
72 Starts recording.
The key turns red once recording has started.
73 Stops the current recording. The detected values are
automatically saved to the current measuring project.
74 Pauses the current recording.
Key 72 flashes.
Press this key or key72 again to continue recording.
"Measuring mode – Flowrate" screen
77
81
76
82 83
78
75
80
79
Item Function
75 Shows the measured flow rate as a numerical value in the
selected unit (e.g. m/s).
Select the unit in the "Settings" screen which you can
access from the main menu.
76 Shows the measured flow rate as visual bars.
77 Shows the measured temperature value.
78 Resets the measured values shown in item80 to zero.
79 Opens the menu for selecting the indication of measured
values.
Minimum: Indicates the smallest detected measured value
from a measuring period.
Maximum: Indicates the largest detected measured value
from a measuring period.
Average: Shows the average value of all detected
measured values from a measuring period.
Hold: Pauses the current detected measured value and
shows it continually (when this option is selected).
Off: Switches off the measured values in item80.
80 Measurement value display according to selection in
item79:
Minutes: Duration of the measured interval
m/s: Shows the measured flow rate as a numerical value in
the selected unit (e.g. m/s). Select the unit in the "Settings"
screen which you can access from the main menu.
°C: Shows the measured temperature value.
81 Opens the menu for setting a fixed recording duration for
an automatic measurement
(see section Measuring mode flowrate– Automatic
measurement).
82 Opens the menu for setting the measuring range in heightx
width (edge lengths) as well as the form of the object to be
measured (e.g. pipe or shaft) for the volumetric flow
measurement
(see section Measuring mode flowrate – Measuring range).
83 Saves the measured values in the archive.

14 EN
multifunction measuring meter T3000
"Measuring mode – Flowrate – Automatic measurement"
screen
86
88
85
89 90
84
87
91
Item Function
84 Shows the measured flow rate as a numerical value in
the selected unit (e.g. m/s).
Select the unit in the "Settings" screen which you can
access from the main menu.
85 Shows the measured flow rate as visual bars.
86 Shows the measured temperature value.
87 Shows the remaining time until the automatic
measurement finishes.
88 Opens a screen for selecting the recording duration for
measuring over a long period.
89 Starts recording.
The key turns red once recording has started.
90 Stops the current recording. The detected values are
automatically saved to the current measuring project.
91 Pauses the current recording.
Key 89 flashes.
Press this key or key89 again to continue recording.
"Measuring mode – Flowrate – Measuring range" screen
94
95
96
92
93
Item Function
92 Selects the form of the object to be measured. The
following options are available:
1. rectangular
2. round
3. off
Depending on the selected form, a different equation is
used to calculate the measured values.
93 Specifies the diameter of the object to be measured.
Use the "OK"button(5) to go to the settings. Afterwards,
you can adjust the values with the control keys and
move on to the next field.
94 Specifies the height of the object to be measured.
Use the "OK"button(5) to go to the settings. Afterwards,
you can adjust the values with the control keys and
move on to the next field.
95 Specifies the width of the object to be measured.
Use the "OK"button(5) to go to the settings. Afterwards,
you can adjust the values with the control keys and
move on to the next field.
96 Saves the settings.
Info
The settings saved here are also used for all following
measurements, unless they are deactivated (in the
corresponding menu item)!

EN 15
multifunction measuring meter T3000
"Measuring mode – Humidity" screen
98
101 102
99
97
100
Item Function
97 Shows the following measured values:
Temperature: Temperature value in °C humidity:
Humidity in %RH (relative humidity)
Dew point: Dew point in °C
Critical dew point: Critical dew point in °C at 70%, 75%
or 80% RH (user setting)
98 Specifies how measured values are displayed in
item100:
Minimum: Always shows the smallest detected
measured value from a measuring period.
Maximum: Always shows the largest detected measured
value from a measuring period.
Average: Shows the average value of all measured
values which have been detected so far from a
measuring period.
Hold: Pauses the current detected measured value and
shows it continually (when this option is selected).
Off: switches off the display in item100
99 Resets the measured values shown in item100 to zero.
100 Shows the measured values according to the
specifications in item98.
101 Opens the menu for setting a fixed recording duration for
an automatic measurement
(see section Measuring mode humidity– Automatic
measurement).
102 Saves the measured values in the archive.
"Measuring mode – Humidity – Automatic measurement"
screen
104
105 106 107
103
108
Item Function
103 Shows the following measured values:
Temperature: Temperature value in °C
Humidity: Humidity in %RH (relative humidity)
Dew point: Dew point in °C
Critical dew point: Critical dew point in °C at 70%, 75%
or 80% RH (user setting)
104 Shows the remaining time until the automatic
measurement finishes.
105 Opens a screen for selecting the recording duration for
measuring over a long period.
106 Starts recording.
The key turns red once recording has started.
107 Stops the current recording. The detected values are
automatically saved to the current measuring project.
108 Pauses the current recording.
Key 106 flashes.
Press this key or key106 again to continue recording.

16 EN
multifunction measuring meter T3000
"Measuring mode – Temperature" screen
110
113 114
111
109
112
Item Function
109 Shows the measured temperature in °C.
110 Specifies how measured values are displayed in
item112:
Minimum: Always shows the smallest detected
measured value from a measuring period.
Maximum: Always shows the largest detected measured
value from a measuring period.
Average: Shows the average value of all measured
values which have been detected so far from a
measuring period.
Hold: Pauses the current detected measured value and
shows it continually (when this option is selected).
Off: switches off the display in item112
111 Resets the measured values shown in item112 to zero.
112 Shows the measured values according to the
specifications in item110.
113 Opens the menu for setting a fixed recording duration for
an automatic measurement
(see section Measuring mode temperature– Automatic
measurement).
114 Saves the measured values in the archive.
"Measuring mode – Temperature – Automatic
measurement" screen
116
117 118 119
115
120
Item Function
115 Shows the measured temperature in °C.
116 Shows the remaining time until the automatic
measurement finishes.
117 Opens a screen for selecting the recording duration for
measuring over a long period.
118 Starts recording.
The key turns red once recording has started.
119 Stops the current recording. The detected values are
automatically saved to the current measuring project.
120 Pauses the current recording.
Key 118 flashes.
Press this key or key118 again to continue recording.

EN 17
multifunction measuring meter T3000
"Measuring mode – Hydrogen" screen
122
123
121
Item Function
121 Shows the measured hydrogen concentration in digits
(dimensionless measured value). This measured value is
shown relative to the set base value (see chapter
"Preparing for measurement with trace gas sensor
(heating phase)"). Negative measured values cannot be
shown.
122 Shows the measured hydrogen concentration in digits
(dimensionless measured value). This measured value is
shown relative to the set reference value (see chapter
"Differentiated indication measurement"). Negative
measured values can also be shown.
123 Exits the screen.
Note!
When using the differentiated indication measurement, new H2
base concentrations always occur which the displayed
measured values are based on. An evaluation and comparison
of the dimensionless measured values is then not worthwhile,
because the measurements may relate to constantly changing
H2 concentrations. Thus, the measured values are not saved.
So, only use the Measuring mode – Hydrogen for locating leaks.
"Archive" screen
125 126
124
In the "Archive" screen, you can view archived measuring
projects or open them for further processing. You can access
this screen from the main menu.
Item Function
124 Shows a list of the measuring projects which are saved
in the archive. The currently selected archive entry is
highlighted.
Navigate in the archive as follows:
1. Press the "Down" key on the cross control until the
desired archive entry is selected.
2. You may need to turn the page by pressing the
"Left" or "Right" keys on the cross control.
3. Press the "OK" key on the cross control to confirm
the selected archive entry. You can also press
"OK"(124) on the screen key.
ðThe basic data of the archive entry are displayed
on the screen (see section "Archive– Basic data
menu").
125 Selects the currently selected archive entry. Afterwards,
the basic data of the archive entry are displayed (see
section "Archive– Basic data menu").
126 Deletes all entries from the archive. Then, a safety
prompt opens. Confirm it by pressing the "OK" key if you
want to delete all entries. Otherwise, touch the "Cancel"
key.

18 EN
multifunction measuring meter T3000
The icons next to the saved measuring projects show each
measuring mode. The meaning is as follows:
Building moisture measurement
Wood moisture measurement
Flowrate measurement
Temperature measurement
Relative humidity measurement
Grid measurement
Spot measurement
Time measurement
"Archive– Basic data menu" screen
127
128
129
Item Function
127 Opens the selected archive entry. Depending on the
measuring mode, the logged values of the archive entry
will be displayed.
Use the "OKbutton(5) to return to the list view of the
archive entries.
If the archive entry represents a grid measurement, a
grid symbol button will appear. After pressing this
button, you will be able to edit the grid measurement.
Single measurements or measurements over a long
period cannot be edited in the device.
128 Deletes the selected archive entry. Then, a safety prompt
opens. Confirm it by pressing the "OK"key(5) if you
want to delete the entry. Otherwise, touch the "Cancel"
key.
129 Closes this screen and returns to the list view of the
archive entries.
"Settings" screen
You can configure the device as follows in the "Settings"
screen:
"Settings" screen– screen1
130
133
134
131
132
Item Function
130 Opens the submenu for selecting the screen language.
131 Opens the submenu for selecting the unit system.
Metric: Activates the metric unit system (e.g. for use in
continental Europe).
Imperial: Activates the imperial unit system (e.g. for use
in the USA).
132 Sets the screen brightness. This option can only be
selected by using the cross control below the colour
display.
1. Press the "Down" key on the cross control until the
scale is selected.
2. Press the "Left" and "Right" keys on the cross
control to increase or reduce the display brightness.
133 Sets the date and time. This option can only be selected
by using the cross control below the colour display.
1. Press the "Down" key on the cross control until the
date is selected.
2. Press the "OK"key(5) on the cross control.
ðThe entire row is selected.
3. Press the "Left" or "Right" keys on the cross
control to select the value to be configured.
4. Press the "Up" or "Down" keys on the cross control
to increase or reduce the value to be configured.
5. Press the "OK"key(5) on the cross control.
ðThe row is deselected.
134 Opens the next screen.

EN 19
multifunction measuring meter T3000
"Settings" screen– screen2
135
139
136
137
138
Item Function
135 Either specifies the period for automatic dimming of the
colour display or deactivates the function:
30sec, 1min, 5min, off
136 Either specifies the period for automatic switch-off of the
colour display or deactivates the function:
10min, 30min, 1hour, off
137 Switches signal tones/key tones on or off.
138 Selects the base value to be shown in measuring mode:
Dewpoint: Displays the value as dew point.
Absolute humidity: Displays the value as absolute
humidity in g/m3.
Mixing Ratio: Displays the value as mixing ratio (g/kg) of
the mass of steam to the mass of the dry gas.
139 Opens the next screen.
"Settings" screen– screen3
140
141
142
Item Function
140 Specifies the unit for volumetric flow:
m3/s, m3/min, m3/h, l/min
141 Calibrates the connected sensor. For more detailed
information please refer to the “Calibrate sensor”
section.
142 Opens the submenu for the temperature indication of the
critical dew point. You can choose from the following
options:
• 100%: The temperature of the calculated dew
point will be displayed (100% relative humidity).
• 80 %: The temperature at80% relative humidity
will be displayed as critical dew point.
• 75 %: The temperature at75% relative humidity
will be displayed as critical dew point.
• 70 %: The temperature at70 % relative humidity
will be displayed as critical dew point.
Info
The term "critical dew point" specifies the temperature
that is already critical for mould formation with regard
to an associated relative humidity of approx. 70%,
75% or 80%. Within these ranges, mould formation is
already possible.
Therefore, the critical dew point is often several °C
higher than the actual dew point temperature with
100%RH, where condensation actually occurs!
Knowing the critical dew point is particularly important
when carrying out measurements e.g. behind picture
frames, closets or wardrobes in environments with
otherwise inconspicuous room climate values.

20 EN
multifunction measuring meter T3000
Calibrating the sensor
The calibration process for the sensor will be carried out in the
submenu "Settings" after pressing the "Sensor" button
(Settings menu, screen3). From this menu, a single-point
calibration (zero offset) can be carried out for the selected
sensor indications. All sensors are already factory-calibrated
and have a corresponding characteristic calibration curve. By
stating a calibration value (offset) a global shift of the calibration
curve, which has an effect on the entire measuring range, is
performed for the single-point calibration!
The offset value to be entered is that value by which the
calibration curve will be shifted.
Example:
The displayed value is always 5 increments too high.
=> Change the offset value for this measurement channel by
-5.
The default setting of most sensor types' offset value is 0.0.
Changing the offset value brings about an automatic reset of the
measured values.
The offset value range depends on the respective sensor type
and the unit to be measured. When entering values outside of
the respective value range, a fault message stating the setting
limits will be emitted. The offset will be saved by most SDI
sensors and therefore also be considered for future
measurements.
Please observe that in case of flow sensors the adjustment is
effected by means of a factor. The default setting of the CAL
value here is 1.0, which corresponds to 100 %.
Example:
A setting of 0.9 would mean an adjustment to 90 % and
therefore a reduction by 10%. (measured value before: 10 m/
s2, after: 9 m/s).
A setting of 1.2 would mean an adjustment to 120% and
therefore an increase by 20%. (measured value before: 10 m/
s2, after: 12 m/s)
An exception constitutes the contact temperature measurement
using an NTC-based surface probe (e.g. TS131SDI). Here the
offset value is set via the settings menu of the sensor. In this
settings menu, you can enter the offset value for the
NTCsensor. The set offset value will be accordingly saved in
the device, not in the sensor. This means that all connected
sensors of the same type also work with this adjusted
offset!
Carrying out a measurement
Info
Please note that moving from a cold area to a warm
area can lead to condensation forming on the device's
circuit board. This physical and unavoidable effect can
falsify the measurement. In this case, the display
shows either no measured values or they are incorrect.
Wait a few minutes until the device has become
adjusted to the changed conditions before carrying out
a measurement.
After connecting a sensor and selecting the measuring mode,
the measurement starts automatically.
Carrying out a measurement with SDI sensor (e.g. with TS
660 SDI humidity sensor)
1.
2.
3.
Other manuals for T3000
1
Table of contents
Other Trotec Measuring Instrument manuals

Trotec
Trotec BR16 User manual

Trotec
Trotec BP17 User manual

Trotec
Trotec BI15 User manual

Trotec
Trotec BX11 User manual

Trotec
Trotec bz30 User manual

Trotec
Trotec MultiMeasure BASIC User manual

Trotec
Trotec BM12 User manual

Trotec
Trotec BF06 User manual

Trotec
Trotec VSC206 User manual

Trotec
Trotec PD200 User manual
Popular Measuring Instrument manuals by other brands

AMALGAMATED INSTRUMENT
AMALGAMATED INSTRUMENT 05305L manual

Teledyne Lecroy
Teledyne Lecroy M41h quick start guide

Rimik
Rimik DN200 Operation manual

THORLABS
THORLABS LFE1220W-US installation manual

Lutron Electronics
Lutron Electronics EMC-9400SD Operation manual

Ericsson
Ericsson R310s troubleshooting guide