  9. Tunix EO-250 User manual

Tunix EO-250 User manual

Other Tunix Measuring Instrument manuals

Tunix TDP-01B User manual

Tunix

Tunix TDP-01B User manual

Tunix HygroCAL User manual

Tunix

Tunix HygroCAL User manual

Tunix EO-25X User manual

Tunix

Tunix EO-25X User manual

Tunix TRPG-01B User manual

Tunix

Tunix TRPG-01B User manual

Popular Measuring Instrument manuals by other brands

Powerfix Profi 278296 Operation and safety notes

Powerfix Profi

Powerfix Profi 278296 Operation and safety notes

Test Equipment Depot GVT-427B user manual

Test Equipment Depot

Test Equipment Depot GVT-427B user manual

Fieldpiece ACH Operator's manual

Fieldpiece

Fieldpiece ACH Operator's manual

FLYSURFER VIRON3 user manual

FLYSURFER

FLYSURFER VIRON3 user manual

GMW TG uni 1 operating manual

GMW

GMW TG uni 1 operating manual

Downeaster Wind & Weather Medallion Series instruction manual

Downeaster

Downeaster Wind & Weather Medallion Series instruction manual

Hanna Instruments HI96725C instruction manual

Hanna Instruments

Hanna Instruments HI96725C instruction manual

Nokeval KMR260 quick guide

Nokeval

Nokeval KMR260 quick guide

HOKUYO AUTOMATIC UBG-05LN instruction manual

HOKUYO AUTOMATIC

HOKUYO AUTOMATIC UBG-05LN instruction manual

Fluke 96000 Series Operator's manual

Fluke

Fluke 96000 Series Operator's manual

Test Products International SP565 user manual

Test Products International

Test Products International SP565 user manual

General Sleep Zmachine Insight+ DT-200 Service manual

General Sleep

General Sleep Zmachine Insight+ DT-200 Service manual

Sensa Core Lacto Spark user manual

Sensa Core

Sensa Core Lacto Spark user manual

VOLTCRAFT UTS-1980 operating instructions

VOLTCRAFT

VOLTCRAFT UTS-1980 operating instructions

EBCHQ 94915 Operation manual

EBCHQ

EBCHQ 94915 Operation manual

Balluff BIP LD2-T017-01-EP-S4 Series user guide

Balluff

Balluff BIP LD2-T017-01-EP-S4 Series user guide

Pulsar IMP Lite Series instruction manual

Pulsar

Pulsar IMP Lite Series instruction manual

ETCR 6800 user manual

ETCR

ETCR 6800 user manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.

unix
Corporation
EO-250
Instruction
Manual
Oil/Liquid Temperature Calibrator
Temperature Range: Ambient to 250
˚
C
Table of Contents
1. Preface.....................................................................................1
2.Technical Specification............................................................. 2
3. Operating Principle.................................................................. 3
4.Safety Measures.................................................................... .. 5
5. Warranty Certificate..................................................................6
6.Calibration Report......................................................................9
Preface
Congratulation on purchase of
“
Tunix
”
make Oil/Liquid Temperature Calibrator Model : EO-
250. This instruments is one of the best available
in its class.
We have taken enough care in designing and manufacturing to give you trouble free
performance for longer period. Before starting the instrument, we suggest you to
go through the instruction manual.
Please check for following Standard accessories
1)

Product is not physically damaged.
2)

Operating manual With Warranty certificate
3)

Power Chord.
4) Silicon Oil 50 CST. (optional to be ordered separately. )
5) 17025 Accredited Calibration Certified( optinal to be ordered separately).
On Arrival
Technical Specification
1) Temperature Range: ambient temperature to 250
˚
C
2) Display Resolution: 0.1 Deg C.
3) Control accuracy: Bettter than
±
0.1
˚
C
4) Thermal in-stability:
±
0.05
˚
C @100 (after stabilization time of 15 minutes).
±
0.06
˚
C @250
˚
C (after stabilization time of 15 minutes).
5) Thermal Non-uniformity (Radial):
±
0.06
˚
C ( after stabilization time of 15 minutes)@100
˚
C.
6) Stabilization Time/settling time: 15 minutes after set point is achieved.
7) Time to reach to 250
˚
C from ambient temperature: 45 minutes.
8) Power supply: 230 VAC @50-60 Hz.
9) Current : 6A.
10) Power Consumption: single phase ,1000 Watt maximum.
11) Oil vessel capacity: 4 Liter
50 Cst.(Silicon Oil).
12) Sound: Less than 30 dB
13) Enclosure Metal(MS) Powder Coated
14) Weight: 14 Kg without oil.
Note:
1) Stability is calculated as the best data for the period of two minutes after stablization, the reported
value is Sigma(
σ
), 120 samples taken for the calculation.
2)

Temperatures above 120 deg C causes fumes which gradually increases with respect to temperature
so a exhaust blower is highly recommended to install. Fumes causes serious burning and irritation to
eyes
Operating Principle
You require a stable known temperature source.
The certainty of the calibration depends on
1)

Stability of the source.
2)

Uniformity to which the stable temperature is known.
Designed block diagram of EO-250 is given below.
230VAC
@50
Hz
PID Controller
Control
Output
SSR
heating
element/
Oil/Liquid Reservior
Block diagram of temperature controling systems
Operating Principle
IMPORTANT INSTRUCTION FOR OPTIMUM PERFORMANCE
1)

Use 10 Amp glass fuse as supplied with instrument.
2)

3 Pin Plug used should have 15 amp capacity.
3)

Cooling fan at the backside of controlling unit should be always on .
PID Controller
This is required to cool the electronics. Ensure the air passage is not blocked.
Never start the system without liquid/ fluid, It will permanently damage the instrument's heating coil
Always maintain the liquid at the level marked in the vessel at indicated below, neither above the level
indicated nor the level below indicated. Filling liquid level high above the indicated level will make
the spillage. Hot oil in such case will cause a serious burn injury.Filling the liquid below the indicated
level will damage the heating coils permanently
Always use correct platform to keep the instrument, Incorrect level may increase the vibrations.
Never carry/transport the product with filled oil vessel.
Never carry/transport the product when it is hot. Allow it to cool naturally by setting the minimum set
point.
Always use silicon oil(50 CST),
never use any other oil.
Safety Measures & Precautions
Please check for following before you switch on the product
Safety Measures & Precautions
A Glass fuse is used in the supply line to prevent any problem due to failure of heating
element. In case of fuse being fused to frequently kindly consult factory.
In the event of SSR failure temperature gets run away above set value. If it cross the
set point by more than 255°C. Switch off the power immediately and consult factory.
Do not temper wiring as it may be safety hazard.
Trouble Shooting
1)

Calibrator is not reaching set value.

Check mains for full voltage.

2)

Calibrator temperature is running over.

Check terminal 1 &2 of SSR for getting short for ever. If it is short replace it.

Check for correct polarity in your mains plug. Phase should go through SSR as per design. If
phase is not routed through SSR It may not control the temperature.

3)

Chamber is not not getting on.

Check mains.

Check

fuse.

Check tightness of all terminals on controller & main terminal strip.
4)

Chamber temperature not getting stable.

Check fan at the bottom cover it should be running.

Need turning off controller.
Warranty Certificate
This is to certify that,Temperature Oil/Liquid Temperature Calibrator Model no EO-250
having Sr. No_______________________Date__________________is properly Tested for
workmanship. We Certify our Calibrator for satisfactory performance for the period of one
year against any manufacturing defect.
Name:
Date:
Signature