Standard Frequency Series Generators
Pre-Installation
5
M-2008-7-12
3.3 LINE POWERED GENERATORS – POWER LINE REQUIREMENTS
Operation
GENERATOR MODEL TXR 325D TXR 325 M TXR 425
Maximum Power kW 40 kW 40 kW 40 kW
Maximum mA 400 mA 400 mA 400 mA
Maximum kVp 125 kVp 125 kVp 125 kVp
Power Line 230 / 240 VAC
Single Phase, 50 /60 Hertz
230 / 240 VAC
Single Phase, 50 /60 Hertz
230 / 240 VAC
Single Phase, 50 /60 Hertz
Line voltage automatic compensation: +10%.
Maximum line regulation for maximum kVA demand: 5%.
NOTES: For lines below 230 or above 240 VAC requires matching the autotransformer
to the supply line. Instructions in the installation document and labels in the x-ray console
provide instructions for Line Strap Adjustments.
GENERATOR MODEL TXR 525 SFQ
Maximum Power kW 50 kW
Maximum mA 500 mA
Maximum kVp 125 kVp
Power Line 230 / 240 VAC
Single Phase, 50 /60 Hertz
Line voltage automatic compensation: +10%.
Maximum line regulation for maximum kVA demand: 5%.
NOTES: For lines below or above 240 VAC requires matching the autotransformer to the supply line. Instructions in
the installation document and labels in the x-ray console provide instructions for Line Strap Adjustments.
• RMS line current during a X-ray exposure, minimum line power required, Generator stand-by consumption (W),
the differential sensitivity (mA) and the thermomagnetic breaker should be:
LINE VOLTAGE SINGLE - PHASE GENERATORS
40 Kw 40 Kw 40 Kw 50 Kw
240 VAC 160 A 160 A 160 A 180 A
Maximum kVA required Maximum kW x 1.25
Stand-by Consumption 240 W
Differential Sensitivity
(Earth Leakage / Ground Fault) 30 mA
Differential, Thermomagnetic (Fuses)
and Contactor
50% of the RMS line current
(RMS = momentary line current based on 100 ms X-ray exposures)
NOTES: For lines below or above 240 VAC requires matching the autotransformer to the supply line. Instructions in
the installation document and labels in the x-ray console provide instructions for Line Strap Adjustments.