TysonBio SuperB User manual

Blood Glucose Monitoring System
For in vitro diagnostic use only
User’s Manual
Do not reuseKeep dryKeep away from sunlight
In Vitro Diagnostics Medical Device
Caution, consult accompanying documents Read Instructions for use
Store at +4ºC to +30ºC
ID:B03G106873
Revision:01
TysonBio

User’s Manual
Customer Service
If you have any questions or concerns regarding this product or its operation, or any
attempt to correct a problem fails, please call our Customer Service. Our trained
specialists will be happy to assist you, train you, or even reaffirm your results
Customer Service Hotline:
1-800-948-3894 (U.S.) (24 hours a day)
+886-37-585998
our office hours are from
8:30 am (GMT0030) - 5:30 pm (GMT0930)
or E-Mail us at
website:www.tysonbio.com
When you call our Customer Service, please have your TysonBio SUPERB Meter,
TysonBio SUPERB Test Strip and all other system supplies available. This will
allow us to answer any of your questions with speed and efficiency.

Coding Your TysonBio SUPERB Meter ............................................................23
The Control Solution Test .................................................................................25
Testing Your Blood Glucose Level ....................................................................29
Preparing the Test Strip ..............................................................................30
Obtaining a Blood Sample ..........................................................................31
Applying Blood Sample to the TysonBio SUPERB Test Strip ....................32
Memory Features ................................................................................................34
Transfer Test Results to a Computer .................................................................36
Care and Storage ................................................................................................37
Display Messages ...............................................................................................38
Specifications .....................................................................................................45
Limitations of the Procedure ......................................................................47
Service and Warranty .........................................................................................47
Customer Service
Important Information .........................................................................................3
Introduction ..........................................................................................................7
About the Product ................................................................................................8
About the TysonBio SUPERB Meter ..........................................................9
About the TysonBio SUPERB Meter Display ............................................11
About the TysonBio SUPERB Test Strips .................................................13
About the TysonBio SUPERB Test Strip vial ............................................14
About the Glucode Chip ............................................................................15
Installing/Replacing the Battery .................................................................16
Setting Up the Meter ..........................................................................................17
Setting Up the Date, Time .........................................................................17
Setting Up the Alarm .................................................................................18
Setting Up the HI and LO Alarm ................................................................20
Enter PC Link Mode ...................................................................................21
Measure Unit and Code Setting .................................................................21
12
Table of Contents

Important Information
34
Forearm
Palm
Available Alternate Sites Testing : Palm and Forearm
TysonBio SUPERB Blood Glucose Monitoring System provides you alternate sites
testing (AST). This system provides you to test on the palm and the forearm with
the equivalent results to fingertip testing.
Alternative site results differ from fingertip results when glucose levels are chang-
ing rapidly such as after a meal, after taking insulin, during exercise, or when you
are ill or under stress.
We strongly recommend you doAST ONLY in the following intervals:
● In a pre-meal or fasting state (more than 2 hours since the last meal).
● Two hours or more after taking insulin.
● Two hours or more after exercise.
DO NOT use AST if:
● You think your blood glucose is low.
● You are unaware of hypoglycemia.
● Your AST results do not match the way you feel.
● You are testing for hyperglycemia.
● Your routine glucose results are often fluctuating.
● If you are pregnant.
There are limitations for doing AST. Please
consult your healthcare professional before
you do AST.

5 6
Physiologic differences in the circulation between the finger and other test sites
like the forearm and palm may result in differences in blood glucose measure-
ments from the other test sites and your fingertips. Changes in blood glucose
may be observed in finger blood samples sooner than blood samples from the
forearm and other alternate sites. Rub the alternate test sites about 20 seconds
before lancing. If you are testing for hypoglycemia (low blood glucose), or if you
suffer from hypoglycemia unawareness, we recommend that you test on your
fingertips.
Talk to your doctor to see if alternate site testing is right for you. With a little bit
of education, you can give your fingertips a rest and maybe test more often than
you do now. For people with diabetes, more frequent testing is a good thing.
Just remember: any time you want to be sure of an accurate, up-to-date blood
glucose reading, test on your fingertip.
1.
2.
Low or high blood glucose readings can indicate a potentially serious medical
condition. If your blood glucose reading is unusually low or high, or if you do not
feel the way your reading indicates, repeat the test with a new test strip. If your
reading is not consistent with your symptoms or if your blood glucose result is
less than 60 mg/dL (3.3 mmol/L) or higher than 240 mg/dL (13.3 mmol/L) you
should contact your healthcare professional and follow his or her treatment
advice.
Any change or administer of medication based on the TysonBio SUPERB blood
glucose test results without the consent and advice of a physician or healthcare
professional is not recommended.
Severe dehydration and excessive water loss may cause false low results. If
you believe you are suffering from severe dehydration, consult your physician
immediately.
●
●
●
CAUTION:

Your TysonBio SUPERB System is made up of several components. This booklet
is designed to guide you through the operation of the system with detailed direc-
tions and pictures. Your TysonBio SUPERB System contains:
● TysonBio SUPERB Meter
● One vial of TysonBio SUPERB Test Strips (10 strips)
● One Glucode Chip (outside the strip vial)
● One Lancing Device (Option)
● Ten Lancets (Option)
● One Control Solution
● One carrying case
● User’s Manual
● Log book (Option)
● Quick Reference Guide
Note: 25, 50 test strips vial can be purchased individually.
Thank you for choosing the TysonBio SUPERB Blood Glucose Monitoring System,
the latest advancement in biosensor technology. As you already know, Self moni-
toring of blood glucose (SMBG) is a necessary part of the treatment plan for people
with diabetes mellitus. The Diabetes Control and Complications Trial (DCCT) has
confirmed the significant benefits from SMBG when practiced as part of a larger
intensive strategy to tightly control blood glucose concentrations.
Adapted for its ease of use and quick response time, the TysonBio SUPERB
System has the ability to process accurate results utilizing only a small volume of
blood for in vitro diagnostics. The TysonBio SUPERB Blood Glucose Monitoring
System is intended for use in the home and in professional settings to monitor
whole blood glucose levels obtained from the fingertip, palm and forearm. This is
an over the counter (OTC) product.
Please read the entire User’s Manual carefully before using this product.
Introduction About the Product
7 8

Display Your test results are displayed here. The large, easy to read screen
guides you through the test by using numbers and symbols.
M Button The M(mode) button is used to set the year, month, day and time,
enters the memory mode to show the consecutive 7/14/30-day average results,
and turns the Meter on and off.
S Button The S(select/set) button is used to select year, month, day, time,
memory of results and control solution mode.
Test port This is where you insert the Test Strip into the Meter.
Battery Cover
Label
Reset Button Press this button to re-set the system.
Glucode Chip Insert the Glucode Chip for coding.
Data port Connect interface cable to transfer data.
Unit Switch Switch for mmol/L or mg/dL.1 mmol/L =18 mg/dL
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
M
S
RESET
FRONT BACK
2
4
5
7
8
3
16
9
10
About the TysonBio SUPERB Meter
910
mmol/L mg/dL
S/N:M05B13286
0537
For in vitro diagnostic use only
Manufacturer:
Tyson Bioresearch, Inc.
Chu-Nan,Taiwan
Customer Service
Call:+886-37-585998
Tyson ID:T502
Blood Glucose Meter
TysonBio

Memory Symbol Indicates in memory mode.
Temperature Message Appears if your meter is out of operating tempera-
ture range.
Battery Symbol Appears when the battery is low or must be replaced.
Blood Drop Symbol Flashes when it is ready to apply sample to Test Strip.
Warning Symbol Appears when result is out of the range of your setting
reference value.
Alarm Function Symbol Appears when the clock alarm is set.
Test Strip Symbol Appears to tell you when the meter is ready for test.
Error Symbol Appears when the strip or meter error.
Day Average Symbol Shows the average value of test results from the last
7-, 14- or 30-days.
Test Result Area Display test results.
6.
7.
8.
9.
10.
11.
12.
13.
14.
15.
Date
Time
Control Solution Test
Appears when doing a
control solution test and
marking the result as a
control solution test.
Units of Measurement
Unit of mmol/L or mg/dL will
appear with the test result.
Code Symbol Appears
with the code number of the
Test Strips currently in use.
1.
2.
3.
4.
5.
4
6
8
7
13
15
1
9
3
2
5
10
14
12
11
DAY AM
PM
AVG
Mem
mmol/l mg/dL
Code
H
L
About the TysonBio SUPERB Meter Display
11 12

Confirmation Window This is where you can
confirm that enough blood has been applied to the
target area.
Target Area This is where blood is drawn into the
Test Strip.
Contact Bar This end of the Test Strip is inserted
into the Test Port to activate the TysonBio SUPERB
Meter. (white side face up, Front view).
1.
2.
3.
The TysonBio SUPERB Test Strip is used to perform
the blood glucose test in conjunction with the TysonBio
SUPERB Meter.
Each strip can be used only once.
Code Number
Control Solution Range
Lot Number
Expiration Date
VIAL LABEL TEST STRIP VIAL
About the TysonBio SUPERB Test Strip About the TysonBio SUPERB Test Strip Vial
13 14
3
1
2
5-5D
Front Back
0537
Manufacturer:
Tyson Bioresearch, Inc.
5F, #22, Ke E. Road III Science-
Based Industrial Park, Chu Nan,
Taiwan Tel : +886-37-585998
Store at 4-30°C ( 39-86°F). Do not refrigerate.
Do not store test strips outside this vial.
Re-cap immediately.
WARNING:Always close lid tightly.
For testing of glucose in capillary whole blood.
For in vitro diagnostic use.
Use within 90 days after first opening.
Use
TysonBio
SUPERB
meter only !
TysonBio
AUG.2011

The Glucode Chip is used to ensure that your TysonBio SUPERB Meter provides
you with accurate results. Before performing the blood glucose test or starting a
new strip vial, please check your TysonBio SUPERB Meter with the Glucode Chip
to ensure proper performance (further instructions for use of the Glucode Chip is
provided on page 22-24).
Note:
● Always use the Glucode Chip to ensure your TysonBio SUPERB Meter is
performing properly.
● DO NOT misplace your Glucode Chip.
966 Code Number
When the battery is low, the Meter will show a warning signal.
When this display appears, it is time to replace a new battery.
When the low battery warning signal is shown, the TysonBio
SUPERB Meter will provide accurate results for approximately
50 more measurements.
From the back of the Meter, gently slide and remove the
battery cover.
Insert a 3-Volt Lithium battery (CR-2032) with the “+” side
facing up.
Slide battery cover back into place.
RWSET
1.
2.
3.
About the Glucode Chip Installing/Replacing the Battery
How to Replace the Battery
15 16
S/N:M05B13286
0537
Forin vitro diagnostic use only
Manufacturer:
TysonBioresear ch, Inc.
Chu-Nan,Taiwan
CustomerSer vice
Call:+886-37-585998
TysonID:T502
Blood Glucose Meter
TysonBio

Press the Mbutton for 3 seconds to enter setting mode.
Press and release Sbutton to choose “24Hr” or “12Hr”
display. With your favorite clock mode, press Mbutton and
the Year setting will be appeared. Press and release the S
button until the correct year appears.
To enter the year and move on to choose the month, press
the Mbutton. To change the month, press the Sbutton.
To enter the month and move on to choose the day, press
the Mbutton. To change the day, press the Sbutton.
1.
2.
3.
4.
5.
6.
7.
To turn the Meter on, first insert a battery, then press the reset button located at the
back of the Meter and proceed with the following instructions.
M
S
To enter the day and move on to choose the hour setting,
press the Mbutton.
To change the hour, press the Sbutton. To enter the hour
and move on to change the minutes, press the Mbutton.
To change the minutes, press the Sbutton. To enter the
minutes and move on to choose the alarm setting, press
the Mbutton.
To set the first alarm, press the M button (y). If you press
S button (n), it will move to health management alarm
setting (step 11). To change the hour, press Sbutton. To
enter the hour and move on to change the minutes,
press Mbutton. To change the minutes, press Sbutton.
To enter the minutes and move on to the second alarm
setting, press the Mbutton. (Fig.1)
AM
Fig.1
Setting Up the Alarm
Setting Up the Date, Time
Setting Up the Meter
17 18

8.
9.
To set the second alarm, press the M button (y). If you press S
button (n), it will move to health management alarm setting (step
11). To change the hour, press Sbutton. To enter the hour and
move on to change the minutes, press Mbutton. To change the
minutes, press Sbutton. To enter the minutes and move on to the
third alarm setting, press the Mbutton. (Fig.2)
To set the third alarm, press the M button (y). If you press S button
(n), it will move to health management alarm setting (step 11). To
change the hour, press Sbutton. To enter the hour and move on to
change the minutes, press Mbutton. To change the minutes, press
Sbutton. To enter the minutes and move on to the fourth alarm
setting, press the Mbutton. (Fig.3)
10.
11.
12.
To set the fourth alarm, press the Mbutton (y). If you press Sbutton
(n), it will move to health management alarm setting (step 11). To
change the hour, press Sbutton. To enter the hour and move on to
change the minutes, press Mbutton. To change the minutes, press
Sbutton. To enter the minutes and move on to health management
“HI alarm setting” press the Mbutton. (Fig.4)
Note: When the alarm is ringing, press S or M button to turn off or
meter will turn off alarm automatically after 30 sec.
To change the “HI alarm setting” press the Sbutton. To enter the
‘HI alarm setting” and move on to “LO alarm setting” press M
button. (Fig.5)
To change the “LO alarm setting” press the Sbutton. To enter the
“LO alarm setting” and move on to PC link setting, press M
button. To finish the setting, press S button (no) to turn off the
meter.
Fig.4
Fig.5
PM
mg/dL
mg/dL
19 20
Fig.2
Fig.3
AM
PM
Setting Up the HI and LO Alarm

13. If you press M button (yes), "PC" link mode for transferring test
data will appear. Further instruction on transferring test data,
see page 36. Further press Mbutton to turn off the meter. (Fig.6)
1.
2.
Select unit, mg/dL or mmol/L on the back.
Note: The mg/dL is the standard unit in the United States.
Meters used in the United States must be set to read in
mg/dL.
Plug the Glucode Chip into the Glucode Chip slot for strip
coding.
Note: 1mmol/L = 18mg/dL
Fig.6
Enter PC Link Mode
Measure Unit and Code Setting
RESET
19
Checking the Performance of Your TysonBio SUPERB Meter
The Glucode Chip is used for checking and coding your TysonBio SUPERB Meter. To
Check your Meter:
1.
2.
Insert the Glucode Chip into meter’s Glucode Chip slot.
All segments of the LCD display will appear indicating
that the TysonBio SUPERB Meter is functioning
properly.
If you are a first time user or start a new strip vial,
please continue the coding procedure.
DAY AM
PM
AVG
Mem
mmol/l mg/dL
Code
H
L
21 22
mmol/L mg/dL
S/N:M05B13286
0537
For in vitro diagnostic use only
Manufacturer:
Tyson Bioresearch, Inc.
Chu-Nan,Taiwan
Customer Service
Call:+886-37-585998
Tyson ID:T502
Blood Glucose Meter
TysonBio

Each batch of TysonBio SUPERB Test Strip is encoded with a number that must
match one of the code numbers stored in the TysonBio SUPERB Meter. In order
to ensure an accurate result, the code number on the display must also match the
code number found on the TysonBio SUPERB Test Strip vial.
Insert the TysonBio SUPERB Glucode Chip, Conduc-
tive Bar’s end first, into the slot.The TysonBio SUPERB
Meter will automatically be turned on.
Note: Code the TysonBio SUPERB Meter whenever
you start a new test strip vial.
All segments of the LCD display will appear indicating
that the TysonBio SUPERB Meter is functioning prop-
erly. Then a beep will sound, followed by the code
number.
1.
2.
Note: If you are using the Meter for the first time
without inserting the Glucode Chip, a “000” will
appear on the screen and then turn off automati-
cally after 3 sec.
Match the code number on the TysonBio SUPERB
Meter display with the code number on the Test Strip
vial. If the code matches, you may press the M
button to turn it off and begin to test lately.
If the meter code and vial code do not match, please
ask the store/distributor to replace a new vial.
3.
4.
Code
Code
Coding Your TysonBio SUPERB Meter
23 24

The Control Solution is used to check and to ensure that the TysonBio SUPERB
Meter and TysonBio SUPERB Test Strip are working together properly and that you
are performing the test correctly.
When you are using your Blood Glucose Monitoring System for the first time.
When you are using a new batch of Test Strips.
Anytime you question the performance of the system, or on a regular basis to
ensure accuracy, e.g. once a week.
When you adjust your diabetic medication plans.
When your blood glucose test result is lower or higher than your normal level.
1.
2.
3.
4.
5.
Make sure the Control Solution is at room temperature
prior to testing.
Insert a TysonBio SUPERB Test Strip, white side face up,
Contact Bar’s end first, into the test port. The TysonBio
SUPERB Meter will turn on automatically.
All segments of the LCD display will appear, indicating that
the TysonBio SUPERB Meter is functioning properly.
A beep will sound and the code number will appear followed
by a flashing “ “ symbol. Be sure the code number on the
display matches the code number on the TysonBio SUPERB
Test Strip vial. If the code numbers do not match, code the
TysonBio SUPERB Meter by using Glucode Chip, accord-
ingly.
1.
2.
3.
4.
Note: Control Solutions are required but not supplied.
There are three Control Solutions at different desired range of aqueous glucose
(Low, Normal or High). Two levels of Control Solutions should be performed. For
more information about the Control Solution, please read your Control Solution
package insert.
To Perform a Control Solution test:
M
S
The Control Solution Test
When to perform a Control Solution Test
mg/dL
25 26

5.
6.
7.
After the “ “ symbol appears, press the Sbutton for 3 sec to
switch control solution mode. The “ “ symbol indicates that
the Meter will mark your current test as a Control Solution
test. Now you are ready to apply the Control Solution.
Discard the first drop of Control Solution and squeeze a small
drop of Control Solution on a clean nonabsorbent surface
(such as a clean piece of wax paper). Do not apply Control
Solution to the Test Strip directly from the bottle.
Bring the tip of the Test Strip to lightly touch the drop of
Control Solution. Control Solution is automatically pulled into
the strip through the tip. Hold until the meter beeps. The
meter will now being counting down from 8 to 1 and the
Control Solution test result will appear.
Caution: Please remember that in order to prevent
contamination, follow the above instructions for perform-
ing the Control Solution test.
mg/dL
AM
mg/dL
Comparing Control Solution Results
The TysonBio SUPERB is functioning properly if the result falls inside the speci-
fied range printed on the Test Strip vial. If the test result falls outside the specified
range, repeat the test.
Results falling outside the specified range may be caused by:
● Error in performing the test
● Expired or contaminated Control Solution
● Expired or contaminated TysonBio SUPERB Test Strips
● Improper coding of the TysonBio SUPERB Meter
● TysonBio SUPERB Meter malfunction
Note: The result will not be incorporated into the average in the Control Solution
test “ “ mode. DO NOT use the system if the problem persists.
Please call Tyson Bioresearch, Inc., Customer Service at +886-37-585998 or
website of www.tysonbio.com
M
S
27 28

Before testing blood glucose, you need the following items:
1. TysonBio SUPERB Meter
2. TysonBio SUPERB Test Strip
3. Adjustable Automatic Lancing Device
4. Sterile Lancet
5. Glucode Chip
Caution: To reduce the chances of infection:
● Never share a lancing device and lancet with another person.
● Always use a new and sterile lancet. Lancets are for single use only.
● Always use a new TysonBio SUPERB Test Strip.
Test Strips are for single use only.
● Do not get lotion, oil, dirt or debris in or on the lancet and lancing device.
1.
2.
3.
4.
Wash hands using soap and warm water. Rinse and dry
thoroughly.
Pull out a TysonBio SUPERB Test Strip from the vial and re-cap
the vial immediately.
Insert the Test Strip white side face up, into the test port.
A beep will sound and the code number will appear followed by a
flashing “ “ symbol. Now you are ready to obtain a blood sample.
Reminder: Make sure that the code number on the display
matches the code number displayed on the Test Strip vial. If
the code number on the display does not match the code
number on the Test Strip vial, code the Meter by using
Glucode Chip, accordingly.( See page 23)
M
S
mg/dL
Testing Your Blood Glucose Level Preparing the Test Strip
29 30

SAMPLE MAY BE OBTAINED FROM FINGER, PALM or FOREARM
For further instructions please see the insert provided with lancing device.
Unscrew and remove the endcap of the lancing device.
Insert the lancet into the lancet holder.
Twist and remove the protective lancet cover to expose the sterile needle tip.
Recapping and screwing the endcap to the body of the lancing device.
Pull back on the cocking mechanism until it comes to a stop.
Place the lancing device firmly against the finger tip and press the release
button.
After sampling, remove the endcap containing the used lancet carefully.
Remove the lancet from the lancing device. Discard the used lancet properly
in accordance with local, state, and federal law.
1.
2.
3.
4.
5.
6.
7.
8.
After obtaining a blood sample, place the drop onto the test well of the TysonBio
SUPERB Test Strip.
Blood is automatically drawn into the TysonBio SUPERB Test Strip.
As soon as enough blood has filled the Confirmation Window (see picture) of the
Test Strip, the TysonBio SUPERB Meter will beep and begin counting down from
8 to 1.
Your blood glucose result will appear on the LCD panel and will be stored into
the Meter memory automatically.
1.
2.
3.
4.
Note: To ensure accurate results, make sure the Confirmation Window of the Test
Strip is completely filled with your blood sample.
M
S
mg/dL
M
S
M
S
AM
mg/dL
Obtaining a Blood Sample Applying Blood Sample to the TysonBio SUPERB Test Strip
31 32

The TysonBio SUPERB results range between 20 to 600 mg/dL (1.1 to 33.3
mmol/L). If a “HI” message appears on your display, your TysonBio SUPERB
Meter has detected that your blood glucose level is higher than 600 mg/dL(33.3
mmol/L). It is suggested that you review your testing procedure and test again
with a new TysonBio SUPERB Test Strip to confirm the result. If the same result
occurs, consult your healthcare professional immediately.
(Fig.7)
If a “LO” message appears on your display, your TysonBio SUPERB Meter has
detected that your blood glucose level is lower than 20 mg/dL (1.1 mmol/L). It is
suggested that you review your testing procedure and test again with a new
TysonBio SUPERB Test Strip to confirm the result. If the same result occurs,
consult your healthcare professional immediately.
(Fig.8)
If no further tests are performed, the TysonBio SUPERB Meter will shut off
automatically after 4 minutes.
After finishing the test, we recommend using a tissue paper to remove TysonBio
SUPERB Test Strip from the TysonBio SUPERB Meter for proper disposal.
5.
6.
7.
M
S
The TysonBio SUPERB Meter automatically stores the 300 most recent blood
glucose results with date and time in its memory. It also provides you with a
consecutive 7/14/30-day average of your blood glucose test results.
Press the Mbutton to turn on the TysonBio SUPERB
Meter. All segments of the LCD display will appear. The
date and time will appear first.
Press Sbutton, the result of control solution will appear
on display.*
Press Sbutton, the consecutive 7-day average will be
shown.*
Press Sbutton, the consecutive 14-day average will be
shown.*
Press S button, the consecutive 30-day average will be
shown.*
1.
2.
3.
4.
5.
Note : * If there is no test results during this stage, it will
directly appear the most recent test result.
M
S
Memory Features
33 34
(Fig.7)
(Fig.8)

Press Sbutton, the most recent test result will appear. Each
time you press the Sbutton, the TysonBio SUPERB Meter will
recall the last 300 test results accordingly. When the memory is
full, the oldest result is removed and replaced with the newest
result.
Press and release the Sbutton to the end of the memory, “000”
will show. Then press the Sbutton again or Mbutton to turn
the meter off.
6.
7.
To Record Your Results
We have provided a log book for your recording convenience. To learn more about how to record your
results, please refer to the log book for further instructions.
Exiting the Memory Mode
1. Press the Mbutton to turn off the meter.
2. Meter will shut down automatically after 4 minutes with no further action.
Note: The consecutive 7/14/30-day average is calculated from the blood glucose results obtained
during the last consecutive 7/14/30-day.
AM
Mem
mg/dL
Mem
You can use your meter with TysonBio Link Health Management Software to transfer test results to your
personal computer.
Obtain the required software and cable
For order information please call Tyson Bioresearch, Inc., Customer Service at +886-37-585998 or
website of www.tysonbio.com
Install the software on a computer
Follow the instructions provided with software to install the software.
Transfer data
Follow the instructions provided in the software to download the results from the meter. After
finishing the data transfer, press M button to turn off the meter.
Note: While in the communication mode, you will be unable to perform a blood glucose test. If the
meter is not in the PC link mode, it will not respond to computer commands.
1.
2.
3.
4.
Transfer Test Results to a Computer
Get ready to transfer test results
Setting the meter to the PC link mode, see page.21. “PC” will appear on the
display. Connect the interface cable to a serial port on your computer. Then
connect the other end of interface cable to the data port located on
the side of the meter with the meter turned off.
35 36
Table of contents
Other TysonBio Blood Glucose Meter manuals

TysonBio
TysonBio EZ Smart-608 User manual

TysonBio
TysonBio AC500 User manual

TysonBio
TysonBio HS100-B User manual

TysonBio
TysonBio HT100-B User manual

TysonBio
TysonBio AC100 User manual

TysonBio
TysonBio TB 100 User manual

TysonBio
TysonBio AC100 User manual

TysonBio
TysonBio TB200 User manual

TysonBio
TysonBio EVOLVE User manual

TysonBio
TysonBio HS200-B User manual