Ugin e.ON 200 NEO User manual

Rev 141119
INSTRUCTIONS FOR USE
Ceramic firing and pressing furnace
25 rue de la Tuilerie 38170 SEYSSINET-PARISET –France
Tél. National : 04.76.84.45.45 - Tél. International : +33.4.76.84.45.43
Email : info@ugin-dentaire.fr - http : www.ugindentaire.fr


Rev 141119
e.ON 200 NEO / e.ON 250 NEO
UGIN DENTAIRE
3
e.ON 200 NEO / e.ON 250 NEO
A. Equipment description................................................................................................................ 4
B. INSTRUCTIONS FOR USE.............................................................................................................. 5
1. SAFETY INSTRUCTIONS.................................................................................................................... 5
1.1. USE................................................................................................................................................... 5
1.2. SAFETY INSTRUCTIONS.................................................................................................................... 5
1.3. WASTE DISPOSAL............................................................................................................................. 6
2. EC DECLARATION OF CONFORMITY ................................................................................................ 7
3. TECHNICAL INFORMATION.............................................................................................................. 8
3.1. TECHNICAL DATA............................................................................................................................. 8
3.2. CONDITIONS OF USE, TRANSPORT AND STORAGE ......................................................................... 9
3.3. ACCESSORIES ................................................................................................................................... 9
4. INSTALLATION AND START-UP ...................................................................................................... 10
4.1. UNPACK....................................................................................................................................... 10
4.2. INSTALLATION............................................................................................................................. 10
4.3. START-UP .................................................................................................................................... 10
5. USE................................................................................................................................................. 12
5.1. START SCREEN............................................................................................................................. 12
5.2. CONFIGURATION ........................................................................................................................ 13
5.3. DOCUMENTS............................................................................................................................... 18
5.4. CERAMIC FIRING PROGRAMMES................................................................................................ 19
5.4.1. PROGRAM PARAMETERS (ZONE 3)............................................................................................. 20
5.4.2. COPY, START AND GRAPHIC VIEW KEYS (ZONE 4)...................................................................... 23
5.4.3. FAVORITE CERAMIC .................................................................................................................... 23
5.5. SPECIAL CERAMIC FIRING PROGRAMMES (SP)........................................................................... 24
5.5.1. FAVORITE CERAMIC SPECIAL ...................................................................................................... 24
5.6. PRESSING PROGRAMMES........................................................................................................... 25
5.6.1. PRESSING CYCLE PHASES ............................................................................................................ 26
5.6.2. CONSEILS DE PRESSAGE.............................................................................................................. 27
5.6.3. FAVORITE PRESSING ................................................................................................................... 27
5.7. SAVING & PROGRAMMES’ TRACEABILITY .................................................................................. 28
5.8. PERIPHERALS............................................................................................................................... 28
6. MAINTENANCE AND TROUBLE-SHOOTING................................................................................... 29
6.1. MAINTENANCE .............................................................................................................................. 29
6.2. PROTECTION.................................................................................................................................. 30
6.3. TROUBLE-SHOOTING .................................................................................................................. 31
Non-contractual images
Translation into English from original in French

Rev 141119
e.ON 200 NEO / e.ON 250 NEO
UGIN DENTAIRE
4
A. Equipment description
N°
Description
1
Power switch ON/OFF
2
Power socket (furnace)
3
Connector for vacuum pump
4
Vacuum pump power socket
5
Rating plate
6
Quick coupling Ø6
(only for eON 250 NEO)

Instruction for use (rev 141119)
e.ON 200 NEO / e.ON 250 NEO
UGIN DENTAIRE
5
B. INSTRUCTIONS FOR USE
1. SAFETY INSTRUCTIONS
1.1. USE
e.ON 200 NEO is a ceramic firing furnace and e.ON 250 NEO is a ceramic firing and pressing furnace. These
furnace are intended to firing the ceramic dental prostheses with or without pressing (according to
furnace model).
These furnaces have to be operated only by dental professionals.
e.ON 200 NEO and e.ON 250 NEO have to use in the conditions and for the uses described in this manual.
Any other use that will be made of it (such as firing other materials, heating food products, etc.) is
considered inappropriate. In this case, the manufacturer declines all liability and immediately cancels the
equipment warranty.
According to the product’s instructions and functionalities, it is imperative to read carefully the
instructions for use before furnace installation and use. Retain the manual for future reference.
1.2. SAFETY INSTRUCTIONS
Pictograms’ meaning used in this instructions for use:
General warning
Electrical hazard
Burn hazard - Hot surface
Observe following safety instructions:
-During operation and after use, the furnace head surface, tray, support pressing, firing base
may reach high temperatures: do not touch
-Wear suitable gloves and the tweezer (delivered in accessories) for handling prosthetic
objects and support pressing/firing base
-The unit has to be connected to an electrical installation in accordance with the applicable
standard in the country in which it is being used. This installation has to provide protection
against overcurrent, overload and ground faults
-This furnace requires its own power line and has to be connected to a circuit breaker
-Connect it directly to the mains socket. Do not use multiple sockets
-Ensure that the connection cables are in perfect condition to prevent short circuits
-Before any technical intervention, the unit has to be switched off (with the breaker) and
disconnected from the mains power supply
-Do not spill liquid on the ventilation openings or inside the furnace

Instruction for use (rev 141119)
e.ON 200 NEO / e.ON 250 NEO
UGIN DENTAIRE
6
-Ensure that the pneumatic hoses (power socket and vacuum pump) are in perfect condition
-Ensure that the movement of the tray is not blocked by an object, this may result in
malfunction of the unit and damage to certain components
-The furnace must be placed on a flat surface, keeping sufficient clearance around the unit to
ensure proper ventilation
-Do not block the ventilation openings to avoid overheating
-Firing base and support pressing are made up ceramic fibres. Handle carefully
-They can release dust: remove with a vacuum cleaner - do not blow or use compressed air
-For moving, take hold the furnace by its base. Never lift by upper part of furnace: risk
material damage
-Never use the furnace without firing base / support pressing. Use only those provided by
UGIN DENTAIRE. Before use, check it (no dirt or damage). If damage: do not use. Place only
one firing base or one support pressing on the tray (do not stack, never simultaneously)
-Use only genuine spare parts. The use of non-original spare parts voids any warranty for your
device
-Do not introduce stranger objects inside the unit during the maintenance operations. It is
forbidden to modify the material without authorization
-Do not execute different operations of maintenance from those brought back in the manual.
Whichever operation not included in this manual, can involve risks
-For whichever information concerning installation, maintenance and use, contact the
customer’s service UGIN DENTAIRE
1.3. WASTE DISPOSAL
Observe the waste disposal regulations for electrical and electronic equipment: do not dispose of with
household waste.
The unit is subject to Directive 2012/19/EU on waste electrical and electronic equipment and to the laws of
the country in which it is being used, and must be disposed of in accordance with applicable regulations.
Firing base and support pressing made up refractory ceramic fibres. They must be disposed in accordance
with applicable regulations in force.

Instruction for use (rev 141119)
e.ON 200 NEO / e.ON 250 NEO
UGIN DENTAIRE
7
2. EC DECLARATION OF CONFORMITY
The equipment’s EC declaration of conformity is provided as appendix of instructions for use.
Content of the declaration EC:
Name and address of the manufacturer:
UGIN DENTAIRE
25 rue de la Tuilerie
38170 SEYSSINET-PARISET –FRANCE
UGIN DENTAIRE declares, on its own exclusive responsibility, that the product:
- Ceramic firing furnace e.ON 200 NEO
- Ceramic firing and pressing furnace e.ON 250 NEO
- Serial number: see the equipment’s EC declaration of conformity
Fulfills the essential requirements of health and safety of the following directives:
- Machinery directive 2006/42/EC
- Electromagnetic compatibility directive 2014/30/EU

Instruction for use (rev 141119)
e.ON 200 NEO / e.ON 250 NEO
UGIN DENTAIRE
8
3. TECHNICAL INFORMATION
3.1. TECHNICAL DATA
The ceramic furnaces e.ON 200 NEO and e.ON 250 NEO are intended for firing the ceramic dental
prostheses. Then, e.ON 250 NEO model enable to firing with pressing. These furnaces have to be operated
only by dental professionals and in the conditions described in this manual (see chapter 1.1 and 3.2).
e.ON 200 NEO
e.ON 250 NEO
Dimensions (mm)
-Width
320
320
-Height
572
719
-Depth
360
360
Weight
22 Kg
27 Kg
Power consumption
1 850 W
Supply voltage
230 V
Frequency
50/60 Hz
Pneumatic system
(pressure level)
minimum 1.7 bar –maximum 6 bar
Noise level
< 70 dB(A)
e.ON 200 NEO
e.ON 250 NEO
-100 ceramic firing programmes
100 corresponding “Favorite”programmes
-100 special ceramic firing programmes (SP)
100 corresponding “Favorite”programmes
-50 pressing programmes
50 corresponding “Favorite”programmes
X
-Touch screen 7’’ with ANDROID navigation system
-USB port for data transfer and storage, and for use of a mouse
for navigation and parameter settings
-Magnetic storage tray for the prosthetic restorations
-7 languages (French, English, German, Italian, Spanish, Russian and
Mandarin Chinese)
Legend:
included
X
not included

Instruction for use (rev 141119)
e.ON 200 NEO / e.ON 250 NEO
UGIN DENTAIRE
9
3.2. CONDITIONS OF USE, TRANSPORT AND STORAGE
Authorised conditions of use
-Ambient temperature: +5°C to +40°C
-Humidity range: maximum relative humidity 80% for temperatures until 31°C, without condensation
and decrease linear until 50% to 40°C, without condensation
-Ambient pressure: up to an altitude of 2000 m above sea level
Authorised conditions of storage
-Ambient temperature: -20°C to +65°C
-Humidity range: maximum relative humidity 80%
-Ambient pressure: up to an altitude of 2000 m above sea level
Conditions to transport
-Wait the complete cooling before packing/transporting
-Use the original packaging only, with the protective elements (polystyrene...)
-Take hold the furnace by its base, never lift by upper part of furnace
-Install the furnace in the packaging with the protective elements. Also protect the accessories for
transportation
3.3. ACCESSORIES
Accessories delivered with the furnace (included in the packaging):
eON 250 NEO
UGIN code
eON 200 NEO
UGIN code
-1 support pressing
FCE05THE0058
-1 firing base
FCE05THE0057
-1 firing base
FCE05THE0057
-1 tweezers
INS03DBT0026
-3 alumina pistons Ø 13
FCE05THE0061
-1 USB-stick
FCE05ELQ0070
-1 cylinder set Ø 35 (100g)
CYL03ACH0003
-1 power cable
FCE05ELC0014
-1 cylinder set Ø 50 (200g)
CYL03ACH0004
-1 casting ring set No.2 (with alumina
piston D13mm)
FCE06BAG0004
-1 casting ring set No.3 (with alumina
piston D13mm)
FCE06BAG0005
-1 casting ring (Ø 81)
FCE05DIV0052
-1 silicone cylinder (Ø 80)
CYL03FCE0002
-1 cylinder base E max Ø 35 (100g)
FCE05DIV0083
-1 cylinder base E max Ø 50 (200g)
FCE05DIV0084
-1 hook kit
FCE05TOL0117
-1 firing tongs
FCE05DIV0080
-1 tweezers
INS03DBT0026
-1 USB-stick
FCE05ELQ0070
-1 power cable
FCE05ELC0014
-1 PU hose Ø4/6
FCE05PNE0015
Accessory recommended (not delivered with the furnace –to order separately)
-Vacuum pump - UGIN DENTAIRE code: FCE01PPE0001

Instruction for use (rev 141119)
e.ON 200 NEO / e.ON 250 NEO
UGIN DENTAIRE
10
4. INSTALLATION AND START-UP
4.1. UNPACK
Unpack e.ON 200 NEO / e.ON 250 NEO and check that it is in perfect order (furnace and accessories)
Any defects can be reported to the carrier
Remember to get the carrier to sign the delivery note
NOTE: keep the original packaging to transport the furnace (see chapter 3.2)
4.2. INSTALLATION
Install in a room aired. Observe the conditions indicated in chapter 3.2
Place the furnace on a level and aired surface. Keep a distance from other objects around not less than
10 or 15 cm. Keep it away from heat sources (for example, radiators and/or other equipment that
release heat). Avoid vibrations and shocks
Do not install in a place with explosion hazard
Its forbidden placing inflammables, toxic, volatile or explosives around the furnace
Place the vacuum pump near to the furnace
This unit requires its own power line and has to be connected to a circuit breaker.
Grounding is mandatory to avoid accidents and the dangers of overheating the power line.
Due to the power of furnace: connect it directly to the mains, do not use multiple sockets (risk
of electrical interference)
4.3. START-UP
NOTE: for marks, see chapter A –equipment description
1) Connect the vacuum tube to the vacuum connector at the back of the furnace (No. 3)
2) Connect the vacuum pump power cord to the vacuum pump power socket (No. 4)
3) Connect the compressed-air tube to the quick coupling Ø6 (No. 6) (only for e.ON 250 NEO model)
4) Connect the power cord to the power socket (No. 2)
5) Connect the furnace to the laboratory’s pneumatic system (only for e.ON 250 NEO model)
IMPORTANT: the minimum pressure required to ensure proper functioning is 1.7 bar. An error
message will appear in case of insufficient pressure. The maximum pressure is 6 bar
6) Make sure that the mains voltage complies with the voltage indicated on the rating plate (No. 5),
connect the power cord to a 10/16 A grounded power socket
7) Activate the ON/OFF switch (No. 1): the application start to loading and the tray go down

Instruction for use (rev 141119)
e.ON 200 NEO / e.ON 250 NEO
UGIN DENTAIRE
11
8) When the tray has reached the lower position, the firing or pressing base can be placed on top
9) After approximately 40 seconds, the start screen of the furnace appears (Fig 1). The installation is
completed
Fig 1: start screen
IMPORTANT: even if the use of this furnace is very intuitive, it is recommended to read the
instructions for use carefully prior to its first use. These instructions are directly accessible by
pressing the “document” key

Instruction for use (rev 141119)
e.ON 200 NEO / e.ON 250 NEO
UGIN DENTAIRE
12
5. USE
5.1. START SCREEN
There are three display zones:
Display in zone 1:
-Temperature inside the firing chamber
-UGIN DENTAIRE’s logo
-Pressure inside the firing chamber
-Date
Display in zone 2:
Display pictograms (e.ON 250 NEO: 10; e.ON 200 NEO: 8)
-Equipment configuration (Config.): see chapter 5.2
-Ceramic, Favorite ceramic, Ceramic SP, Favorite ceramic SP, Pressing, Favorite pressing: access to
parameters of each program (see chapters 5.4, 5.5, 5.6)
-Opening, Closure: to raise or lower tray in stand-by mode
-Documents : display the instructions for use
Display in zone 3:
-Return key (back arrow) : present in each screen, to return to the previous screen. It is not
displayed in the start screen
-Hour (update in Configuration)

Instruction for use (rev 141119)
e.ON 200 NEO / e.ON 250 NEO
UGIN DENTAIRE
13
5.2. CONFIGURATION
When press key “Config.”, the drop-down menu appears to customize the unit and to access a series
of specific functions.
Configuration menu contains 14 parameters.
NOTE: Change the parameters’ numeric field:
-press the active zone (white) and select the value (keys ±)
-“Accept” or “Cancel” to validate or not the choice, which allows to return to the previous screen
5.2.1. Labo name
When selecting the white field (on the right), the alphanumeric keypad appears at the bottom of the
screen. The user can now enter the name of the laboratory which is then displayed in the start screen
(instead of UGIN DENTAIRE’s logo)
5.2.2. Date
5.2.3. Hour
5.2.4. Language
The furnace default language is French. This language can be changed in the drop-down menu. Press the
button ▼ and confirms your choice. The change is instantly accepted

Instruction for use (rev 141119)
e.ON 200 NEO / e.ON 250 NEO
UGIN DENTAIRE
14
5.2.5. Temperature unit
There are two possibilities: Celsius or Fahrenheit degrees.
The default temperature reading is set to Celsius. Press °F to change to Fahrenheit.
5.2.6. Pressure unit
There are two possibilities: millibar (mbar) or millimeters of mercury (mmHg).
The default values are displayed in mbar. Select mmHg to change to millimeters of mercury.
5.2.7. Sound
The furnace is set by default to “Bip 1”. This setting can be changed in the drop-down menu. Press the ▼
button and confirms your choice. The new acoustic signal is now activated
5.2.8. Sound level
When selecting this field, the alphanumeric keypad appears at the bottom of the screen. The sound
volume set by default to 100 (maximum) can now be changed. To mute the sound, set the value to 0 (zero)
5.2.9. Screen saver (min)
The indicated value shows the time that will elapse before a screen saver is activated. This setting can be
changed from 15 to 30 minutes.
In stand-by mode, the pictograms in the start screen will disappear and the UGIN DENTAIRE’s logo will start
moving randomly on the screen. Only the RETURN key and the time indication are displayed.
You can leave this mode by touch the screen.

Instruction for use (rev 141119)
e.ON 200 NEO / e.ON 250 NEO
UGIN DENTAIRE
15
5.2.10. USB copy
When press the key ►, a menu with 5 lines appears:
-Copy program (furnace to USB stick): to copy the programs setting inside the furnace and the files
contents in folder Documents
-Copy program (USB stick to furnace): to copy the programs contents in USB stick
-Traceability of the cycles (furnace to USB stick): to copy the completed cycles
-Background: replace the background’s image
-Image display: to image modification
5.2.10.1. Copy Prog (Furnace → USB)
When a USB stick is inserted into the furnace (on the right) and it is recognized , this function allows to
copy all furnace programmes and to transfer the files in the “Documents”folder (see chapter 5.3).
To save the data, select the desired line and wait for completion indicated by the logo . This can take
some tens of seconds. The data are copied and the stick can now be removed.
IMPORTANT: the data saved on the USB stick can now be displayed and copied on a computer. 2 folders
named “PROG”and “DOCUMENTS”should appear
“PROG” folder contains several files that correspond with the various programmes (eON 250 NEO: 2 x 3
files; eON 200 NEO: 2 x 2 files):
-3 files in “BIN”format (ProgCeram, ProgCeramSP and ProgPress) are locked. They contain
all programme settings and are saved in a specific format
-3other files (ProgCeram, ProgCeramSP and ProgPress) in “TXT”format can be displayed
and printed
NOTE: only 2 programmes for the model witout pressing eON 200 NEO: ProgCeram, ProgCeramSP
“DOCUMENTS”folder contains the PDF files of the instructions for use in: they can be displayed and
printed
5.2.10.2. Copy Prog (USB → Furnace)
When the USB stick is recognized , this function allows to transfer the programmes on the USB stick to
the furnace. They will replace the programmes in the unit.
The same procedure as described in previous chapter Copy Prog (Furnace→USB) applies.
NOTE: for safety reasons, only the “BIN” files (locked) can be transferred

Instruction for use (rev 141119)
e.ON 200 NEO / e.ON 250 NEO
UGIN DENTAIRE
16
5.2.10.3. Traceability of the cycles (Furnace→USB): - -Ceram, - - Ceram SP, - - Press
This line only contains the fully completed cycles.
This function ensures the traceability of the work performed by saving the settings that were used for each
one of them. The storage capacity of the programmes is:
-60 ceramic firing cycles
-60 SP ceramic firing cycles
-20 pressing cycles
When these limits are reached, each new cycle will overwrite the oldest on the list. They are transferred in
an identical way to the previous chapters.
Example of traceability transfer:
10 Ceram, 3 Ceram SP, 5 Press (except version without pressing)
After selecting the function, the transfer will start:
1/10 Ceram, 0/3 Ceram SP, 0/5 Press (except version without pressing)
Final display: 10/10 Ceram, 3/3 Ceram SP, 5/5 Press followed by the logo indicating that the transfer
is complete. The stick can now be safely removed.
This operation can take some tens of seconds (depending on the amount of files). After this manipulation,
the counters are reset to 0.
5.2.10.4. Background .jpg (USB → Furnace)
The screen background can be replaced with any other image.
However, this image must be saved on the USB stick under the following name: “ImgFond.jpg”
Only this file will be recognized by the furnace. It is imperative that the stick contains only one file like this
when modifying the background image.
When the stick is recognized , simply select the desired line and wait for the logo to appear.
To verify that the screen background has been modified, press the RETURN key (at the bottom left) to
return to the “Config”menu and display the new background. The stick can now be safely removed.
5.2.10.5. Image display (USB/Visu/ImgVisu.jpg)
This function allows to display the image of your choice, even within a running cycle. To this end, you need
to create (on a computer) a file named “Visu”and save the selected image in this file under “ImgVisu.jpg”,
before copying them to the USB stick.
The desired document can be accessed as described earlier.
You can zoom in to enlarge specific areas. The principle is identical to that of smartphones.
To «leave» the image, press the RETURN key (at the bottom left) to return to the previous screen. The stick
can now be safely removed.

Instruction for use (rev 141119)
e.ON 200 NEO / e.ON 250 NEO
UGIN DENTAIRE
17
5.2.11. Silver test
This programme has been pre-defined and allows to calibrate the temperature of the unit in case of
unsatisfactory firing. These settings cannot be modified.
Open the firing chamber, place a 30 mm silver wire on the firing base and press the START button.
At the end of the programme, the wire should be melted only at the tip.
If not, the calibration offset may need to be modified in each programme (see chapter 5.4.1).
Repeat this process, if necessary.
5.2.12. Decontamination
This programme is similar to the silver wire test and cannot be modified. It allows to clean the firing
chamber.
Open the furnace, place a carbon rod CARBONET on the firing base and press the START button to let the
cycle run.
5.2.13. Auto-test
In case of a malfunction, the defective component is indicated.
ATTENTION: do not interrupt the test before completion
At the end of each phase, a logo appears indicating the result of the test.
At the end of test, the key can be used to return to the previous screen
Exemple :
Auto-test: problem with the vacuum in the firing chamber
5.2.14. Technician parameters

Instruction for use (rev 141119)
e.ON 200 NEO / e.ON 250 NEO
UGIN DENTAIRE
18
This function allows to update the software and process versions, and the documents.
These parameters cannot be accessed without a valid code. The code will be provided to the user along
with the required updates. Another code is reserved to the service technician for maintenance purposes.
5.2.15. Version Application
These indications refer to the version of the controller software of the furnace.
This may be useful in communications between the user and the supplier of the unit.
5.3. DOCUMENTS
Documents
This function allows to view the instructions for use directly on the furnace’s screen. The .pdf files can be
printed via a computer. This is simply done by transferring them to the USB stick supplied with the furnace
(see chapter 5.2.10.1: Copy Prog (Four → USB).
When press the “Documents” key, a menu of 6 lines appears. Select the desired file to access it. The key
arrow allows to return to the start screen. You can zoom in on specific areas. The principle is identical
to that of Smartphones.
IMPORTANT: instructions for use have a specific name. The unit will only recognize the following names:
-EON-FR.pdf
-EON-EN.pdf
-EON-DE.pdf
-EON-IT.pdf
-EON-ES.pdf
-EON-RU.pdf
-EON-CH.pdf
Two other types of files can be displayed on the furnace:
-1 video file “Video.mp4”(size 1 Go maxi)
-1 image file “Image.jpg“
The instructions for use are subject to change. They are updated via the USB stick as described in the
section “Technician parameters” (see chapter 5.2.14)
The same procedure applies to the Video and Image files.

Instruction for use (rev 141119)
e.ON 200 NEO / e.ON 250 NEO
UGIN DENTAIRE
19
5.4. CERAMIC FIRING PROGRAMMES
Pictogram
start screen
Program
Programmes name
Ceramic
C001 to C100
Favorite Ceramic
C001 to C100
The setting screen display after select the program Ceramic or Favorite Ceramic. This screen is composed
by four zones display:
Zone 1
Zone 2 Zone 3 Zone 4
Display zone 1
-date
-temperature inside the firing chamber
-program type: CERAMIC
-pressure inside the firing chamber
Display zone 2
A drop-down from C001 to C100 allow to select
the desired program. Each cycle is identified by:
-program’s number (C and 3 numbers)
-program’s name
-star (activated: “Favorite”program)
Display zone 3
This drop-down menu allow to change the
settings of selected program:
-Number, program’s name, “Favorite”star
-its parameters are shown below
Display zone 4 :
Three active keys to:
-copy a program
-start the cycle
-display the cycle graph

Instruction for use (rev 141119)
e.ON 200 NEO / e.ON 250 NEO
UGIN DENTAIRE
20
5.4.1. PROGRAM PARAMETERS (ZONE 3)
IMPORTANT: if the numeric value of the parameter exceeds the minimum/maximum limits, a message
will indicate the limits to be observed. If these parameters are not modified, the indicated limits will be
taken into account when proceeding to the next parameter
Changes the field numerical:
-Press the field. The value is displayed in green. Carry out the changes with the keypad numerical
-Confirm and press key : pass to following parameter
Program’s name
Select the corresponding field (white) to enter or change the name of the programme. This can be done
via the alphanumeric keypad that appears. To confirm or cancel, use the button in the lower left of the
screen
Drying rate (20°C to 150°C/min)
Indicates the rate of increase between the standby temperature and the preheating temperature.
Value: 20°C/min to 150°C/min
Preheating temperature (20°C to 1.000°C)
Always ≥ stand-by temperature and ≤ final temperature. Value: 20°C to 1 000°C
Close furnace time (0 to 60 minutes)
-Set at 0: the furnace will close in one continuous movement when the preheating temperature is
reached
-If the value is > 0: the tray will move in 6 equal stages during the selected time
Press for close the furnace before the end of each stage. When press , the tray stop in position
chosen by user. Press again to movement will continue
Preheating time (0 to 60 minutes)
Within a running cycle and with the furnace closed, it indicates the holding time at the preheating
temperature.Value: 0 to 60 minutes
Rate (20 to 150°C/min)
Indicates the rate of increase between the preheating temperature and the final temperature
Value: 20°C/min to 150°C/min
This manual suits for next models
1
Table of contents
Other Ugin Dental Equipment manuals
Popular Dental Equipment manuals by other brands

Ultradent Products
Ultradent Products Ultrapro Tx Air Maintenance and troubleshooting guide

Waterpik
Waterpik ION Water Flosser quick start guide

KaVo
KaVo SONICflex 1.008.6383 Instructions for use

Durr Dental
Durr Dental VSA 300 S Installation and operating instructions

Omec
Omec PL.06.00 Instruction, use and maintenance manual

KaVo
KaVo ELECTROmatic TM Instructions for use