manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Unit
  6. •
  7. Multimeter
  8. •
  9. Unit UT136A User manual

Unit UT136A User manual

Measurement Operation
First check on 9V battery, then turn rotary switch to
the measuring position. If the low battery, “ ” sign.
will be displayed on LCD panel. Nearly to “ ” sign
on the meter front panel terminal input which alarm
not exceed the testing voltage and current input value
limitation.
A. DC/ AC Voltage Measurement (See figure 1)
figure 1
1. In each range, the meter has an input impedance
of 10MΩ. This loading effect can cause measurement
errors in high impedance circuits. If the circuit
impedance is less than or equal to 10kΩ, the error is
negligible (0.1% or less).
2. Press SELECT to switch AC or DC voltage
measurement.
3. The display value of AC measurement is effective
value of sine wave( average value).
4. The 400mV AC range can obtain only under
manual mode.
Note:
To avoid possibly damages to the meter, please do
not attempt to input higher than 500V.
To avoid electrical shock, please pay attention during
the high voltage measurement.
B. Capacitor Measurement(UT136B only, figure 2).
figure 2
1. The Meter will display a fixed value as below which
is the Meter internal fixed distributed capacitance
value. To ensure accuracy when measuring a small
value of capacitance, the tested value must subtract
this value, REL mode can help on that.
2. It is normal to take several seconds to obtain a
stable reading when testing a high value capacitor.
3. To avoid possibly damages to the meter, please do
not attempt to input higher than 60V DC and 33V rms
AC.
7. Battery in meter: 9V 6F22 or NEDA 1604 or
IEC6LR61
8. Deficiency of Built in Battery: LCD display “ ”.
9. Product size: 72mm×137mm×35mm
10. Net Weight: About 200g (battery include)
11. Conform: CE
Button function and auto power off
1. HOLD button
Press HOLD to enter and exit hold mode in any mode
except frequency measurement.
2. SELECT button
Press SELECT button to select the alternate
functions including AC/DC voltage, AC/DC current,
auto and manual ranging, frequency and duty cycle.
Used as REL function button under resistance
measurement.
3. AUTO POWER OFF
To preserve battery life, the Meter automatically goes
into a “sleep” mode if you do not press any button for
around 15 minutes. The Meter can be activated by
pressing any button, then returns to the display for
the function selected previously. To disable the sleep
mode function, press SELECT button while turning on
the meter.
4. BUZZER
The buzzer phonate go with every time button be
effectual pressed. When the meter will auto power off
in 1minutes the buzzer beeps 5 times. Before power
off there will be a long time buzzer beeps.
UT136A/B/C/D Operating Manual
UT136A/B/C/D
Handheld Digital Multimeter
Operating Manual
Safety Information
Meter UT136A/B/C/D complies with the standard
IEC/EN61010-1, in pollution degree 2, overvoltage
category (CAT II 600V)and double insulation. Use
the meter only as specified in this operating manual,
otherwise the protection provided by the meter may
be impaired.
1. Before using the meter inspect the case. Do not
use the meter if it is damaged or the case (or part
of the case) is removed. Look for cracks or missing
plastic. Pay attention to the insulation around the
connections.
2. Inspect the test leads for damaged insulation or
exposed metal. Replace damaged test leads with
identical model number or electrical specifications
before using the meter.
3. Replace the battery as soon as the battery
indicator “ ” Appears. With a low battery, the
meter might produce false readings that can lead to
electric shock and personal injury.
4. When measurement is on / off against the correct
testing position.
5. Do not apply more that the rated voltage, as
marked on the meter in order to avoid possible
electric shock or personal injury and avoid possible
damage to the meter.
6. Do not change the measuring range during the
testing as it causes to damage the meter.
7. When each measurement has been completed,
disconnect between the test leads and the circuit
under test connection, then turn the meter power
off and remove the test leads away form the input
terminals of the meter.
8. When the meter working at an effective voltage
over 60V in DC and 30Vrms in AC, special care
should be taken.
9. Do not use or store the meter in an environment of
high temperature and humidity. The performance of
the meter may deteriorate after dampened.
10. The internal circuit of the meter shall not be
altered at will to avoid damage of the meter and any
accident.
11. Soft cloth and mild detergent should be used to
clean the surface of the meter when servicing. No
abrasive and solvent should be used to prevent the
surface of the meter from corrosion, damage and
accident.
International Electrical Symbols
Double Insulated
Grounding
Deficiency of Built in Battery
Warning: Refer to the Operating Manual
AC (Alternating Current)
DC (Direct Current)
Conforms to Standards of European Union
Overall Specification
1. Maximum voltage between any terminals and
grounding. Refer to different range input protection
voltage.
2. 10A terminal: 10A H 250V Fast Type Ceramic Fuse
Φ5x20mm.
3. mA terminal: 0.5 A H 250V Fast Type Ceramic
Fuse Φ5x20mm.
4. Maximum display 4000.
Operating temperature: 0℃~40℃(32℉~104℉)
Relative Humidity:0℃~30℃ below≤75%,
30℃~40℃≤50%
Storing temperature:-10℃~50℃ (14℉~122℉)
5. Electromagnetism:
Under 1V/m emission: Best Total Accuracy= Specific
Accuracy + Measurement 5%.
Over 1V/m emission do not have any specific index
6. Elevation: 0~2000m
C. DC/AC Current Measurement (see figure 3)
figure 3
1.Press SELECT to switch AC or DC current measurement
2.Turn off power to the circuit before the connection
between the test leads across with the object being
measured.
3. Selecting the correct terminal input and turn the
rotary switch to select the measuring function. In case
of no any idea on the value input of the current, just
simply test from the high value to low one.
4. Fuses are located on mA and10A terminal input.
Never attempt the test lead connect to any circuits
especially on the power supply terminal and may be
hurt.
5. For the safety purpose, less than 10 seconds is
for each measurement duration and keep 15 minutes
duration for next measurement during the current
measurement over 5A.
D. Resistance Measurement (see figure 4)
figure 4
1. The LCD displays “OL” indicating open-circuit for
the tested resistor or the resistor value is higher than
the maximum range of the meter.
2. To maintain the resistance measurement accuracy,
discount circuit power and discharge all the high
voltage capacitors during the measuring resistance.
3. The test leads cause 0.1Ω~0.2Ωresistance
variation during the measurement, In order to obtain
precision readings in low-resistance measurement.
Need to make the short circuit on the test leads and
mark the measurement value which show on LCD
display. Then deduct this variation value on the
measurement value which come the meter.
4. If Ωreading with shortened test leads is not less
than or equal to 0.5Ω, check for loose test leads,
incorrect function selection or others.
5. For high-resistance measurement greater than
1MΩ, it is normal to take several seconds to obtain a
stable reading with short test leads for measurement.
6. Do not input higher than DC 60V and AC 30V rms
voltage to prevent any damage and accident.
E. Diodes Measurement (see figure 5)
figure 5
1. Disconnect circuit power and discharge all the high
voltage capacitors before measuring resistance.
2. Place the red test lead on the component’s anode
and place the back test lead on the component’s
cathode, a good diode should still produce a forward
voltage drop reading of 500~800mV. The LCD
displays “OL” indicating open-circuit for the tested
diodes or the testing the diodes with polarity.
3. The open circuit for diodes is 1.5V.
4. Do not input higher than DC 60V and AC 30V rms
voltage to prevent any damage and accident.
F. Continuity Measurement (see figure 6)
Range Resolution Accuracy
400Ω 0.1Ω ±(1.2% + 2)
4KΩ 0.001KΩ
40kΩ 0.01kΩ ±(1% + 2)
400kΩ 0.1kΩ
4MΩ 0.001MΩ ±(1.2% + 2)
40MΩ 0.01MΩ ±(1.5% + 2)
Overload Protection: 500V AC/ DC
7. Diode, Continuity Test
Function Resolution Remark
0.001V Display positive voltage decline
<10Ω buzzer, >10Ω may or
may not buzzer
Overload Protection: 500V DC / AC.
8. Frequency and duty cycle
Range Resolution Accuracy
10Hz 0.01Hz±
100Hz 0.1Hz
1KHz 0.001KHz
10KHz 0.01KHz (0.5%+3)
100KHz 0.1KHz
1MHz 0.001MHz
10MHz 0.01MHz
0.1~99.9% 0.1% For refer only
9. Temperature Measurement(For UT136C only)
Range Resolution Accuracy
-40~0℃-(8% +5)
0~400℃1℃±(2.5%+3)
400~1000℃±(3% +3)
Overload Protection: 500V DC /AC.
Enclosed point contact K type thermocouple can
onl y be used on less than 230℃temperature
measurement.
10. NCV (For UT136D only)
Function Range Remarks
<10mm: buzzer beeps
NCV 220V/50Hz 10~50mm: may and may not beeps
>50mm: Buzzer not sound
Maintenance And Service
Warning
Make sure the test leads take off the terminal and the
circuit as well as power off the meter if want to open
the meter cover.
1.General Service and Maintenance
Periodically wipe the case with a damp cloth and mild
detergent. Do not use abrasives or solvents.
Any abnormal on the meter, stop use the meter and
return to service center.
When need to have calibration on the meter, please
allow certified engineer or specific service center for
the service maintenance.
2. LCD display “ ” sign indicating battery will be
run out and need to replace a new battery, if fail to do
that. It causes the variance of the measuring result.
Battery Specification: 9V 6F22 or NEDA 1604/
IEC6LR61
Operating Steps:
(1) Turn the power in “ off "situation, then remove the
test lead out of the meter.
(2) Use screwdriver to take off the screw on the
battery cover. Then you can take off old battery for
replacing.
(3) User screwdriver to take off two pieces of screws,
then you can replacing the burned fuse(s) as
replacement.
Fuse specification:
F1 Fuse φ5×20mm, F 0.5 A H 250V
F2 Fuse φ5×20mm, F 10 A H 250V
figure 6
1. Place the test leads to the circuit which be tested,
if between both terminals show resistance >100Ω. it
is a short-circuit, no buzzer. if between both terminals
show resistance≤10Ω, it is a good connection with
continually buzzer. if between both terminals show
resistance> 10Ω, may or may not buzzer.
2. Resistance value on tested circuit display on LCD.
Unit is Ω.
Note:
1. To maintain the diodes measurement accuracy
disconnect circuit power and discharge all the high
voltage capacitors before the measuring.
2. For the continuity testing, the open circuit for
voltage is 0.5V.
3. Do not input higher than DC 60V and AC 30V rms
voltage to prevent any damage and accident.
G. Frequency and Duty Cycle Measurement (see
figure 7)
figure 7
1. Connect the test leads across with the object being
measured. The measured value shows on the display.
2.If you need to measure duty cycle, press SELECT
button to select % measurement mode.
Note:
Input Amplitude: a
When ≤100K Hz: a > 300 mV rms
When >100K Hz: a > 600 mV rms
Never input higher than 10V rms for test frequency
and duty cycle.
H. Temperature Measurement ( only UT136C, see
figure 8)
figure 8
1. Turn rotary switch to“℃”.
2. Put the multiple purpose socket into the correct
terminal. K-Type point contact temperature probe
can only be used in the measurement below 230℃,
if you want to measure over 230℃, you need to
separately buy another temperature probe
3. The LCD displays “OL” indicating point contact
temperature probe which take off. When short circuit
between ℃terminal and COM terminal, the meter
show the value of room temperature.
I. NCV Function ( UT136D only, see figure 9)
figure 9
1. Set the rotary switch to NCV and remove the test
lead from the input terminals.
2. Place the housing front part with marking towards
the 220V/50Hzbeing measured. Distance <10mm,
the meter buzzer; 10mm<Distance <50mm, the meter
may or may not buzzer; Distance >50mm, the meter
not buzzer and the LCD displays “OL”.
UT136A/B/C/D Operating Manual
Accuracy Specifications
Accuracy: ±(a% reading + b digits),guarantee for 1 year
Operating temperature: 23℃±5℃
Relative Humidity: <75%
1. DC Voltage
Range Resolution Accuracy
400mV 0.1mV ±(0.8% + 3)
4V 0.001V
40V 0.01V ±(0.8% + 1)
400V 0.1V
500V 1V
Input impedance: all range 10MΩ
Maximum input voltage: 500V AC/DC.
2. AC Voltage
Range Resolution Accuracy
400mV 0.1mV ±(1.2% + 5)
4V 0.001V
40V 0.01V ±(1.2% + 3)
400V 0.1V
500V 1V
Input impedance: about 10MΩ; Maximum input
voltage: 500V AC/DC
Frequency:40Hz~400Hz
Display effective value of sine wave( average value)
each measurement is applicable from 5% of range as
reference.
3. Capacitance (UT136B only)
Range Resolution Accuracy
4nF 0.001nF For refer only
40nF 0.01nF
400nF 0.1nF ±(4% + 3)
4μF 0.001μF
40μF 0.01μF
100μF 0.1μF ±(5% + 10)
Overload Protection: 500V AC / DC
4. DC Current
Range Resolution Accuracy
400μA 0.1μA
4000μA 1μA
40mA 0.01mA
400mA 0.1mA
4A 0.001A
10A 0.01A
Overload Protection:
mA range: F1 fuse φ5×20mm, F 0.5 A H 250V
10A range: F2 fuse φ5×20mm, F 10A H 250V
When ≤5A Continuous measurement is allowed,
When <5A Continuous measurement less than 10
seconds at an interval more than 15 minutes.
5. AC Current
Range Resolution Accuracy
400μA 0.1μA
4000μA 1μA
40mA 0.01mA
400mA 0.1mA
4A 0.001A
10A 0.01A
Overload Protection:
mA range: F1 fuse φ5×20mm, F 0.5 A H 250V
10A range: F2 fuse φ5×20mm, F 10A H 250V
When ≤5A Continuous measurement is allowed,
When > 5A Continuous measurement less than 10
seconds at an interval more than 15 minutes.
Frequency:40Hz~400Hz
Display effective value of sine wave( average value)
each measurement is applicable from 5% of range as
reference.
6. Resistance
±(1% + 2)
±(1.2% + 3)
±(1.5% + 5)
±(1.2% + 5)
±(1.5% + 3)
±(2% + 3)
0.1Ω
P/N: 110401103485X
MAY.2018 REV. 1

This manual suits for next models

3

Other Unit Multimeter manuals

Unit UT61A User manual

Unit

Unit UT61A User manual

Unit UT33B User manual

Unit

Unit UT33B User manual

Unit UT60F User manual

Unit

Unit UT60F User manual

Unit UT50D User manual

Unit

Unit UT50D User manual

Unit UT131B User manual

Unit

Unit UT131B User manual

Unit TU13BA User manual

Unit

Unit TU13BA User manual

Unit UT60B User manual

Unit

Unit UT60B User manual

Unit UT136A User manual

Unit

Unit UT136A User manual

Unit UT60F User manual

Unit

Unit UT60F User manual

Unit UT695D Series User manual

Unit

Unit UT695D Series User manual

Unit UT71A User manual

Unit

Unit UT71A User manual

Unit UT90A User manual

Unit

Unit UT90A User manual

Unit UT61B User manual

Unit

Unit UT61B User manual

Unit UT58E User manual

Unit

Unit UT58E User manual

Unit UT71A User manual

Unit

Unit UT71A User manual

Unit UT60C User manual

Unit

Unit UT60C User manual

Unit UT805A User manual

Unit

Unit UT805A User manual

Unit UT511 User manual

Unit

Unit UT511 User manual

Unit UT61D User manual

Unit

Unit UT61D User manual

Unit UT8806E User manual

Unit

Unit UT8806E User manual

Unit UT804 User manual

Unit

Unit UT804 User manual

Unit UT801 User manual

Unit

Unit UT801 User manual

Unit UT81B User manual

Unit

Unit UT81B User manual

Unit UT89XE User manual

Unit

Unit UT89XE User manual

Popular Multimeter manuals by other brands

YATO YT-73080 operating manual

YATO

YATO YT-73080 operating manual

Elma Instruments 6700 user manual

Elma Instruments

Elma Instruments 6700 user manual

Vichy VC97 user manual

Vichy

Vichy VC97 user manual

REED R5007 instruction manual

REED

REED R5007 instruction manual

alfatest ADM-128 manual

alfatest

alfatest ADM-128 manual

Fluke 8840A instruction manual

Fluke

Fluke 8840A instruction manual

Keysight U3606A Programmer's reference

Keysight

Keysight U3606A Programmer's reference

AEMC instruments 5233 user manual

AEMC instruments

AEMC instruments 5233 user manual

Aktakom AMM-1081 Operation manual

Aktakom

Aktakom AMM-1081 Operation manual

Chauvin Arnoux F 15 user manual

Chauvin Arnoux

Chauvin Arnoux F 15 user manual

Keithley DMM6500 quick start guide

Keithley

Keithley DMM6500 quick start guide

Klein Tools MM1000 instruction manual

Klein Tools

Klein Tools MM1000 instruction manual

Amprobe ac68C user manual

Amprobe

Amprobe ac68C user manual

Gossen MetraWatt METRACLIP 63 operating instructions

Gossen MetraWatt

Gossen MetraWatt METRACLIP 63 operating instructions

Maxwell Digital Multimeters MX 25 312 user manual

Maxwell Digital Multimeters

Maxwell Digital Multimeters MX 25 312 user manual

UEi UTL33T instruction manual

UEi

UEi UTL33T instruction manual

Gossen MetraWatt MetraHit 28s manual

Gossen MetraWatt

Gossen MetraWatt MetraHit 28s manual

Agilent Technologies 34410A user guide

Agilent Technologies

Agilent Technologies 34410A user guide

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.