Unit UT801 User manual

LCD Display
1. Manual Range Indicator of manual range
2. Warning ! Indicator for Warning signal
3. The battery is low
4. Indicator for high voltage signal
5. Indicator for Negative reading
display
6. AC Indicator for AC voltage or current
( DC indicator do not display)
7. Data hold is active
8. Test of diode
9. The continuity buzzer is on
10. Number Indicates testing reading
11. Units of measurement:
General Specifications
1. Maximum Voltage between terminal input and
COM: 1000V( except 200mV, 250V)
2. μA mA terminal input protection: (CE)250mA 265V
auto recovery fuse
3. 10A terminal input protection: (CE)F1 (10A H
1000V) Fast type melted fuse Φ10.3x38mm
4. Resistance input protection: PTC/1000V
5. Capacitance input protection: (CE)F2, F3 (0.5A H
1000V) Fast type melted fuse Φ6.35x31.8mm
6. Frequency input protection: PTC/1000V
7. Temperature input protection: (CE)0.5A 1000V
fuse
8. terminal input protection: PTC/1000V
9. hFE input protection: (CE)250mA 265V auto
recovery fuse, F3 (0.5A H 1000V) Fast type melted
fuse Φ6.35x31.8mm
10. Display: LCD full func tion signal display,
maximum reading is 1999(UT801) , 19999(UT802)
Updates 2-3 times / second
11. Range: Manual
12. Polarity Display: Auto
13. Overload indication: 1
14. Battery Deficiency:
15. Operating Temperature: 0~40℃(32℉~104℉)
16. Storing Temperature: -10~50℃(14℉~122℉)
17. Relative Humidity: 0℃~30℃ below ≤75%
30℃~40℃ ≤50%
18. Electromagnetic Field: Under 1V/m the influence
of radiated radio-frequency electromagnetic field
phenomenon, Total accuracy= specific accuracy+
measuremen t 5%, Over 1V/m radiated radio-
frequency electromagnetic which do not have any
reference data on this topic.
19. Power: AC(external power adapter AC220V/
DC9V-200mA) or DC(internal battery type 2 R14/1.5V
6 pieces)
20. Product size: (300x245x105)mm
21. Product Net Weight : About1500g(without the
accessories)
22. Safety Compaliances : IEC 61010: CATⅡ1000V
6. Do not overload voltage or current on EITHER
between terminal and terminal OR between terminal
and grounding which indicate on meter limitation.
7. The rotary switch should be placed in the right
position and no any changeover of range shall be
made during measurement is conducted to prevent
damage of the Meter.
8. Do not use or store the meter in an environment
of high temperature, humidity, flammable and
electromagnetic environment. The performance of
the meter may deteriorate after dampened.
9. The internal circuit of the meter shall not be
altered at will to avoid damage of the meter and any
accident
10. Replace the battery as soon as the battery
indicator “ ”Appears. With a low battery, the
meter might produce false readings that can lead to
electric shock and personal injury.
11. Turn the meter off when it is not is use and take
out the battery when not using for a long time.
UT801/802 Operating Manual
Bench Type Digital Multimeters
UT801/802
Operating Manual
Overview
Digital Bench-Type Multimeter Model UT801 is the
maximum reading 1999 and 3 1/2 digits and UT802
is the maximum reading 19999 and 4 1/2 digits,
both models are in manual range, DC / AC current
type digital multimeter, This is also the extra large
characters in LCD display backlight with full function,
full measurement and full overload protection as well
as a good product design outlook, In addition to all
the conventional features include DC/AC voltage,
DC/AC current, resistance, frequency, capacitance,
temper at ur e ℃, Tr an si sto r hF E、diode an d
continuity buzzer.
This operating manual covers information on
safety and cautions. Please read the relevant
information carefully and observe all the Warnings
and Notes strictly.
Unpacking Inspection
Open the package case and take out the Meter.
Check the following items carefully to see any missing
or damaged part. If you find any missing or damage,
please contact your dealer in your country.
● Operating Manual 1 piece
● Test Lead 1 pair
● Alligator Clip 1 pair
● K Type Temperature Probe 1 piece
( For the temperature under 230℃testing)
● Multi-Purpose Socket 1 piece
● Power Cord 1 piece
(AC220V 50Hz DC9V/200mA)
Safety Information
This M et er co mp li es with t he sta nd ar d s
IEC61010-1 in pollution degree 2, overvoltage
category (CAT II 1000V) and double insulation. If you
can not follow up this operating instruction to use the
meter and it reduces the chance to have an using
protection.
1. Before using the Meter and Test Leads inspect
both items. Do not use the Meter and Test Leads
if it is damaged or the case (or part of the case) is
removed or no reaction on LCD display. Prohibited
to use the meter without housing or housing without
screw fix up in order to avoid possible electric shock
or to avoid possible damage to the meter or to the
equipment under test.
2. If the damage of test leads, use only the same
model number or identical electrical specifications
replacement parts.
3. Do not use your finger to touch on any testing
cable, connector, unused terminal input or circuit
during the testing stage.
4. When the meter working at an effective voltage
over 60V in DC or 30V rms in AC, special care should
be taken for there is danger of electric shock.
5. Selecting the correct terminal input and turn the
rotary switch to select the measuring function. In
case of no any idea on the value input of the current,
just simply test from the high value to low one.
Unit of Voltage: The millivolt, volt
Unit of current: Microampere, milliampere,
ampere
Unit of electrical resistance: Ohm, thousand
ohms, trillion ohm
Unit of electrical capacity: Accepts the farad,
the microfarad
Unit of Frequency: Kilohertz
Unit of Temperature: Degree Celsius Factor
Unit of Triode enlargement: Times
V ,Vm
μA, mA, A
Ω, kΩ, MΩ
nF/μF
kHz
℃
β
Operational MeasurementGuide (see Diagram 1, 2,
3)
Diagram 1
Diagram 2
Diagram 3
Functions
Symbol Terminal Input Explanation
V V ←→ COM DC Voltage Measurement
V V ←→ COM AC Voltage Measurement
Ω V ←→ COM Resistance Measurement
V ←→ COM Diode / Continuity Buzzer
Measurement
kHz V ←→ COM Frequency Measurement
A mA μA ←→ COM mA/μA DC Current Measurement
10A ←→ COM ADC Current Measurement
A mA μA ←→ COM mA/μA AC Current Measurement
10A ←→ COM A AC Current Measurement
F V←→ mA μA Capacitance Measurement
(Use Multi-Purpose Socket)
V ←→ mA μA Temperature Measurement
(Use Multi-Purpose Socket)
hFE V ←→mA μA Triode Enlargement Factor
(Use Multi-Purpose Socket) Measurement
℃
Functional Buttons
Turn the power on and off Turn the display
backlight on and off(suitable for battery
supply, backlight turn on about 10 seconds
after auto shut down.)
Press Hold to enter and exit the hold mode
in any mode, the meter beeps.
LIGHT
HOLD
P/N:110401104405X DATE:2018.06.26 REV.1
序号 项目 内容
1
2
3
4
5
6
7
尺寸
材质
颜色
外观要求
装订方式
表面处理
其它
DWH
设 计 MODEL
机型:
Part NO.
物料编号:
CHK
审 核
APPRO.
批 准
说明书菲林做货要求:
版本
/
/
UT801,UT802 110401104405X
外形尺寸:(210x285)±1mm,折叠成形尺寸(70x142.5)±1mm:
用60g书纸
黑色
印刷完整清晰,版面整洁.无分层.残损.毛边等缺陷.
风琴折(三折),再中间对折,封面图在外面
REV.1 修改成股份有限公司
黄荣伟2018.06.25

This manual suits for next models

1

Other Unit Multimeter manuals

Unit UT136A User manual

Unit

Unit UT136A User manual

Unit UT132B User manual

Unit

Unit UT132B User manual

Unit UT70C User manual

Unit

Unit UT70C User manual

Unit UT30B User manual

Unit

Unit UT30B User manual

Unit UT58D User manual

Unit

Unit UT58D User manual

Unit UT120A User manual

Unit

Unit UT120A User manual

Unit UT513 User manual

Unit

Unit UT513 User manual

Unit UT603 User manual

Unit

Unit UT603 User manual

Unit UT50E User manual

Unit

Unit UT50E User manual

Unit UT181A User manual

Unit

Unit UT181A User manual

Unit UT105 User manual

Unit

Unit UT105 User manual

Unit UT8803 User manual

Unit

Unit UT8803 User manual

Unit UT207A User manual

Unit

Unit UT207A User manual

Unit TU13BA User manual

Unit

Unit TU13BA User manual

Unit UT136A User manual

Unit

Unit UT136A User manual

Unit UT70A User manual

Unit

Unit UT70A User manual

Unit UT204A User manual

Unit

Unit UT204A User manual

Unit UT58E User manual

Unit

Unit UT58E User manual

Unit UT201 User manual

Unit

Unit UT201 User manual

Unit UT58A User manual

Unit

Unit UT58A User manual

Unit UT107 User manual

Unit

Unit UT107 User manual

Unit UT531 User manual

Unit

Unit UT531 User manual

Unit UT89XE User manual

Unit

Unit UT89XE User manual

Unit UT61B User manual

Unit

Unit UT61B User manual

Popular Multimeter manuals by other brands

Axio AX-T01 instruction manual

Axio

Axio AX-T01 instruction manual

Mastech MS8269 user manual

Mastech

Mastech MS8269 user manual

Fluke 10 Service manual

Fluke

Fluke 10 Service manual

Extech Instruments MO290 user guide

Extech Instruments

Extech Instruments MO290 user guide

Monacor DMT-1050 instruction manual

Monacor

Monacor DMT-1050 instruction manual

Keysight Technologies U1273A user guide

Keysight Technologies

Keysight Technologies U1273A user guide

Velleman DVM92 user manual

Velleman

Velleman DVM92 user manual

Uniks M32 user manual

Uniks

Uniks M32 user manual

Keithley DMM7510 user manual

Keithley

Keithley DMM7510 user manual

LaserLiner MultiMeter-Pocket manual

LaserLiner

LaserLiner MultiMeter-Pocket manual

Keithley 2110-100 Calibration manual

Keithley

Keithley 2110-100 Calibration manual

multicomp pro MP730007 manual

multicomp pro

multicomp pro MP730007 manual

Rigol DM3068 user guide

Rigol

Rigol DM3068 user guide

Truper MUT-36 manual

Truper

Truper MUT-36 manual

Profile PIN-112 quick start guide

Profile

Profile PIN-112 quick start guide

Aneng AN113B user manual

Aneng

Aneng AN113B user manual

Fluke 867B Service manual

Fluke

Fluke 867B Service manual

IDEAL 61-603 Technical manual

IDEAL

IDEAL 61-603 Technical manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.