ADJUSTMENT
WARNING: To adjust a 2nd stage correctly, it’s must be connected to a appropriate
and well adjusted 1st stage (with a interstage pressure between 9,3 - 9,7 bar) and a
tank with a minimal filling pressure of 180 bar.
1. Check or the adjustment knob is still screwed out completely to achieve the lightest
breathing resistance and position the V.I.V.A. flow vane in the max. position.
2. Before connecting the 2nd stage adjustment tool to the 2nd stage and without pushing
on the purge, slowly blow through the inlet tube to detect a leak. This information will
indicate or the orifice is in contact (no leak) or not (leak) with the seat."
"
If a leak is detected, screw the 2nd stage adjustment tool to the 2nd stage and slowly
screw the orifice one turn clockwise and repeat the test. If there is no leakage, the 2nd
stage can be adjusted.
3. Connect the Digi-Mag adapter mouthpiece to the G260 mouthpiece and turn on the
Digi-Mag. Zeroize the Digi-Mag pressure sensor if needed.
4. Connect the appropriate and well adjusted 1st stage to a tank with a minimal filling
pressure of 180 bar, connect the 1st stage to the 2nd stage adjustment tool with the
L.P. hose and slowly open the tank valve.
5. Push slowly on the tool adjustment knob to counteract the thrust of the interstage
pressure, turn slowly the knob to look for the groove of the orifice."
"
As soon as the groove is found, slowly unscrew the adjustable orifice - counter
clockwise - until a small leak is detected, then screw in the orifice - just enough
clockwise - to stop the leak.
6. Cycle the 2nd stage 10 times or more by pushing on the purge, then readjust as in-
dicated above.
7. Inhale slowly through the Digi-Mag and note the cracking effort. It should be between
1,0 to 1,2 inches of water (2,5 to 3,0 milibars).
8. If the cracking effort is higher then indicated, use the 4 mm
Allen key and unscrew by a ¼turn the fine adjustment
screw inside the adjustment knob."
"
Hold the adjustment knob tight, otherwise not the fine ad-
justment screw will be unscrewed, but the adjustment knob will be unscrewed inside
the inlet tube."
"
note : The adjustable orifice MUST always be readjusted as per paragraph 5 and 6
before measuring the cracking effort again."
"
Proceed by successive steps to reach the recommended values(1,0 to 1,2 inches of
water or 2,5 to 3,0 milibars).