vacuubrand DVR 2 User manual

page 1 of 17
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully
the validity of this document with respect to his product. manual-no.: 999001 / 04/04/2012
DVR 2
Analog/Digital Rough Vacuum Gauge
Technology for Vacuum Systems
Instructions for use

page 2 of 17
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully
the validity of this document with respect to his product. manual-no.: 999001 / 04/04/2012
Dear customer,
Your VACUUBRAND vacuum gauge shall support you at your work for a long time without any trouble
and with full load output. Thanks to our large practical experience we attained much information how you
could add to an efcient application and to personal safety. Please read these instructions for use prior
to the initial start-up of your controller.
VACUUBRAND vacuum gauges are the result of many years of experience in construction and practi-
cal operation of these vacuum gauges combined with the latest results in material and manufacturing
technology.
Our quality maxim is the „zero fault principle“: Every vacuum gauge leaving our company, is tested in-
tensively including an endurance run. Therefore faults, even those which occur rarely, are identied and
can be eliminated immediately. The achievement of the specications after the endurance run is tested
for every device.
Every VACUUBRAND device achieves the specications.
We are committed to providing our customers with this high quality standard.
We know that the device cannot replace all of your real work and hope that our products contribute to
an effective and trouble-free realisation of your work.
Yours
VACUUBRAND GMBH + CO KG
After sales service:
Contact your local dealer or call +49 9342 808-5500.
The document ”Safety information for vacuum equipment” is part of this
manual! Read the ”Safety information for vacuum equipment” and observe
the instructions contained therein!
Trademark index:
VACUU•LAN® (US-Reg.No 3,704,401), VACUU•BUS®, VACUU•CONTROLTM, chemistry-HYBRIDTM, Peltronic®,
TURBO•MODETM, VARIO®, VARIO-SPTM, VACUUBRAND®(US-Reg.No 3,733,388) and also the shown com-
pany logos are trademarks of VACUUBRAND GMBH + CO KG in Germany and/or other countries.

page 3 of 17
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully
the validity of this document with respect to his product. manual-no.: 999001 / 04/04/2012
Contents
Safety information! .....................................................................................................4
General information............................................................................................................................4
Intended use.......................................................................................................................................4
Connecting the device .......................................................................................................................4
Operating conditions .........................................................................................................................5
Safety during operation .....................................................................................................................5
Maintenance and repair .....................................................................................................................5
Technical data .............................................................................................................7
Wetted parts ......................................................................................................................................7
Use and operation ......................................................................................................8
Changing the pressure unit ...............................................................................................................8
Pressure measurement .....................................................................................................................9
Clock symbol .....................................................................................................................................9
Adjusting the operating time and the measuring cycle ......................................................................9
Display of the battery‘s discharge status.........................................................................................10
Troubleshooting .......................................................................................................10
Readjustment of the vacuum gauge DVR 2 ...........................................................11
Adjustment at atmospheric pressure ............................................................................................... 11
Adjustment under vacuum............................................................................................................... 11
Adjustment to a reference pressure ................................................................................................12
Calibration in the factory .........................................................................................12
Replacing the battery ...............................................................................................13
Notes on return to the factory .................................................................................14
Health and safety clearance form ...........................................................................15
EC Declaration of Conformity .................................................................................16
➨DANGER indicates a hazardous situation which, if not avoided, will re-
sult in death or serious injury.
☞WARNING indicates a hazardous situation which, if not avoided, could
result in death or serious injury.
• CAUTION indicates a hazardous situation which, if not avoided, could
result in minor or moderate injury.
NOTICE is used to address practices not related to personal injury.
Note: The device contains a battery!
Remove battery before disposal (see section ”Replacing the bat-
tery”) and dispose of battery and device separately and according
to regulations.
NOTICE

page 4 of 17
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully
the validity of this document with respect to his product. manual-no.: 999001 / 04/04/2012
Safety information!
General information
☞ Read and comply with this manual before installing or operating the equip-
ment.
☞ Do not use the equipment if it is damaged.
Remove all packing material, remove the product from its packing-box, remove the
protective covers and keep, inspect the equipment.
If the equipment is damaged, notify the supplier and the carrier in writing within three
days; state the item number of the product together with the order number and the
supplier’s invoice number. Retain all packing material for inspection.
If the equipment is not used immediately, replace the protective covers. Store the
equipment in suitable conditions.
Intended use
☞The device is designed for ambient and gas temperatures at the pressure trans-
ducer connection of +10°C to +40°C at continuous operation, up to +80°C for
short periods (< 5 minutes) at the pressure transducer. If installing the device into
a cabinet or a housing check maximum temperature. Ensure that the maximum
permitted gas temperature at the pressure transducer (see ”Technical data”) is
not exceeded.
Use the equipment and all system parts for the intended use only, i.e. for measure-
ment of vacuum in vessels designed for that purpose.
Connecting the device
☞Avoid uncontrolled overpressure (e. g. when connecting to a locked or blocked
tubing system). Risk of bursting.
☞Comply with max. permitted ambient and gas temperature (see ”Technical
data”) and make sure ventilation is adequate if the equipment is installed in a
housing or if ambient temperature is elevated.
☞ Power supply by an internal 9V battery.
Do not charge the battery. Do not open the battery. Do not throw the battery into
re. Do not heat over 100°C and do not bring in contact with water.
• Comply with maximum permissible pressures at the pressure transducer. See
section ”Technical data”.
• Connect hoses gas tight at the pressure transducer connection.
• Ensure the stability of hose connections.
• In case of residues or aggressive or condensable media install a gas washing
bottle if necessary.
Avoid high heat supply (e.g., due to hot process gases).
Connect the vacuum gauge to the vacuum application.
Position the vacuum gauge and its vacuum line in such a way that condensate can-
not ow towards the internal pressure transducer.
NOTICE
NOTICE
NOTICE

page 5 of 17
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully
the validity of this document with respect to his product. manual-no.: 999001 / 04/04/2012
Allow the equipment to equilibrate to ambient temperature if you bring it from cold
environment into a room prior to operation. Notice if there is water condensation on
cold surfaces.
Comply with all applicable and relevant safety requirements (regulations and guide-
lines). Implement the required actions and adopt suitable safety measures.
The battery is free from mercury and cadmium.
Recommended type of battery see section ”Technical data”.
When replacing the battery, always mind its correct orientation when inserting it into
the battery compartment.
When using a battery of a different chemical system (e. g. alkaline), lifetime may be
considerably shorter.
Because of their lower capacity, the use of rechargeable batteries is not recom-
mended.
Operating conditions
➨The devices are not suitable for applications which involve dangerous or ex-
plosive gases or explosive or ammable mixtures.
• Ensure that the materials of the equipment’s wetted parts are compatible with the
substances in the vacuum system, see section ”Technical data”.
Safety during operation
➨Adopt suitable measures to prevent the release of dangerous, toxic, explosive,
corrosive, noxious or polluting uids, vapours and gases.
☞ Prevent any part of the human body from coming in contact with the vacuum.
☞Attention: If the pressure is higher than approximately 1060 mbar the pressure
reading becomes incorrect (saturation of the pressure transducer). The display
ashes. Immediate pressure relief necessary! Risk of bursting!
☞Use only genuine spare parts and accessories. Otherwise safety and perfor-
mance of the equipment as well as the electromagnetic compatibility of the equip-
ment might be reduced.
Possibly the CE mark becomes void if not using OEM spare parts.
Electronic equipment is never 100% fail-safe. This may lead to an ill-dened sta-
tus of the equipment or of other connected devices. Provide protective measures
against malfunction and failure. Ensure that in case of failure the device and the
vacuum system always will turn into a safe status.
Maintenance and repair
➨ Attention: Due to the operation the equipment might be contaminated by harmful
or dangerous substances, clean or decontaminate prior to maintenance.
NOTICE

page 6 of 17
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully
the validity of this document with respect to his product. manual-no.: 999001 / 04/04/2012
Ensure that maintenance is done only by suitably trained and supervised techni-
cians.
Interior components of the device can only be repaired at the factory.
Returned products will not be repaired or calibrated until the completed health and
safety clearance form has been received. In order to comply with law (occupational,
health and safety regulations, safety at work law and regulations for environmental
protection) vacuum pumps, components and measuring instruments returned to the
manufacturer can be repaired only when certain procedures (see section ”Notes on
return to the factory”) are followed.
NOTICE

page 7 of 17
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully
the validity of this document with respect to his product. manual-no.: 999001 / 04/04/2012
Technical data
Type DVR 2
Measuring principle capacitive, absolute pressure,
gas type independent
Pressure transducer ceramic diaphragm (alumina),
Pressure reading analogue and digital LCD display
Pressure units mbar, Torr or hPa (selectable)
Measuring cycle automatic, once in 3s, once in 1s, three times in
1s (factory-set: automatic)
Automatic sleep mode operation time: 1-1000 minutes or continuous
operation (”On”), (factory-set: 15 minutes)
Measuring range (absolute) 1080 mbar - 1 mbar (810 Torr - 1 Torr)
Maximum permissible pressure at pressure
transducer 1.5 bar absolute
Measurement uncertainty (absolute) after careful
adjustment and at constant temperature
<±1 mbar (0.75 Torr) /
±1 digit
Temperature coefcient <±0.07 mbar/K (0.05 Torr/K)
Maximum permissible temperature of gaseous
media at pressure transducer
continuous operation: 40°C,
for short periods (< 5 minutes): up to 80°C
Ambient temperature range (operation) 10°C to +40°C
Ambient temperature range (storage) -10°C to +60°C
Permitted relative atmospheric moisture during
operation (no condensation) 30% to 85%
Degree of protection according to IEC 529 IP 40
Vacuum connection small ange KF DN 16 and screw-in hose nozzle
DN 6/10 mm
Power supply battery 9V Lithium, 1.2 Ah, Ultralife U 9VL
Internal volume of measurement chamber 4.23 cm3(without hose nozzle)
Weight approx. 375 g
Dimensions L x W x H 115 mm x 115 mm x 66 mm
Wetted parts
Components Wetted materials
Sensor Aluminium oxide ceramic
Sensor housing stainless steel
Seals chemically resistant uoroelastomer
Vacuum connection / hose nozzle PBT / PPS glass bre reinforced
We reserve the right for technical modication without prior notice!

page 8 of 17
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully
the validity of this document with respect to his product. manual-no.: 999001 / 04/04/2012
Use and operation
Changing the pressure unit
➨ Select the desired pressure unit with the ” UP / DOWN” key.
➨ Switch the vacuum gauge off.
➨ Press ”MODE” key and keep it pressed while switching the
DVR 2 on (”ON/OFF” key).
☞The pressure unit as of the last operation (Torr, mbar or
hPa) is displayed.
➨ Press ”ON/OFF” key to conrm the selected pressure unit
and to terminate the mode.
Display
pressure units, selectable
adjustment mode/
warning symbol automatic sleep mode
replace battery/
display of battery capacity
ON/OFF key
UP/DOWN key
MODE key
Keys
The keys are on the rear side
of the device underneath the
symbols at the front side.
Torr
+
mbar

page 9 of 17
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully
the validity of this document with respect to his product. manual-no.: 999001 / 04/04/2012
Adjusting the operating time and the measuring cycle
➨ Press ”MODE” key to select the time setting mode.
☞The clock symbol and the currently set operating time in
minutes are displayed.
To decrease the operating time:
➨ Press ”MODE” key to change the arrow to the left.
➨ Press ”UP/DOWN” key to decrease the operating time.
➨ Press ”ON/OFF” key to conrm.
☞The device switches to the mode ”Adjusting the measur-
ing cycle”.
☞The measuring cycle can be set to Automatic, 1 meas-
urement in 3s, 1 measurement in 1s and 3 measurements
in 1s.
☞The arrow to the right indicates that the time can be in-
creased using ”UP/DOWN” key.
➨ Press ”UP/DOWN” key to increase operating time.
☞The measuring cycle as of the last operation (factory-set:
”A”) is displayed.
➨ Press ”UP/DOWN” key to select the measuring cycle: A
= Automatic (automatic determination of the measuring
cycle; frequent measurement in case of great pressure
change), S1 = 1 measurement in 3s, S2 = 1 measurement
in 1s, S3 = 3 measurements in 1s.
☞ The ashing of the arrow (only ”S1”, ”S2” or ”S3”) symbol-
izes the frequency of the pressure measurement).
☞ Press ”MODE” key to change the direction of the adjust-
ment.
Adjusting the measuring cycle:
☞The vacuum gauge is equipped with an automatic sleep
mode to save battery power when the instrument is not
used. When the clock symbol appears on the display, the
vacuum gauge will switch off automatically after approx. 1
minute.
➨ Press ”ON/OFF” key. The clock symbol will disappear and
the preselected operation time will start again.
Clock symbol
☞The time until automatic switching off (operating time) can be set by the user within a range from
1 to 1000 minutes (factory-set: 15 minutes) or continuous operation ”On” can be selected.
Upon switching on, the absolute pressure (total pressure, in-
dependent of gas type) ranging from 1 mbar to 1080 mbar is
displayed in the selected pressure unit (Torr, mbar or hPa).
☞The pressure transducer was adjusted using factory stand-
ards. In general, there is no need for readjustment by the
user because of the excellent longtime stability (see sec-
tion ”Readjustment”).
Pressure measurement

page 10 of 17
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully
the validity of this document with respect to his product. manual-no.: 999001 / 04/04/2012
➨ Press ”ON/OFF” key to conrm the measuring cycle and to
terminate the mode.
When switching off the device, the remaining battery capac-
ity is displayed qualitatively by the number of arrows in the
display.
☞If the battery symbol appears in the display during meas-
urement, the battery should be replaced.
☞The correct operation of the device is no longer ensured.
Display of the battery‘s discharge status
Troubleshooting
Fault Possible cause Remedy
❑No display or display
disappears.
➨Automatic sleep mode? ✔Switch on, increase opera-
ting time if necessary.
➨Battery at? ✔Replace battery.
➨Other cause? ✔Contact local distributor.
❑Incorrect pressure read-
ing.
➨Pressure transducer not cor-
rectly adjusted?
✔Readjust pressure trans-
ducer.
➨Moisture in the pressure
transducer?
✔Dry or evacuate the pressure
transducer and readjust if
necessary.
➨Battery at, battery symbol is
displayed?
✔Replace battery.
❑Display is ickering and/
or battery symbol ap-
pears.
➨Battery at? ✔Replace battery.
❑Adjustment mode cannot
be activated.
➨At the applied pressure, read-
justment is not possible.
✔Admit air to atmospheric
pressure or evacuate to a
pressure below 20 mbar.
❑All segments of the LCD
are displayed or no dis-
play although battery has
been replaced.
➨Pressure transducer or meas-
uring electronics defective?
✔Contact local distributor.

page 11 of 17
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully
the validity of this document with respect to his product. manual-no.: 999001 / 04/04/2012
Readjustment of the vacuum gauge DVR 2
➨ Press ”UP/DOWN” key and keep it pressed
while switching on.
☞The vacuum gauge switches into the adjust-
ment mode (indicated by a warning triangle).
☞The arrow at the right indicates, that the pres-
sure reading can be increased using ”UP/
DOWN” key.
➨ Press ”UP/DOWN” key to increase the read-
ing to the actual atmospheric pressure.
The equipment has been adjusted at the factory. In general, there is no need for adjustment by the user
due to the excellent longtime stability.
Depending on operation conditions, type of application and accuracy requirements, an inspection and
readjustment may become necessary.
Admit air to the vacuum gauge. Make sure that the vacuum connection at the vacuum gauge is at at-
mospheric pressure.
Attention: Determine the exact actual atmospheric pressure, e. g. by using an accurate barometer or
get accurate reading from the weather service, e. g. at the next airport, etc. (take into account the differ-
ence in altitude between e. g. airport and laboratory).
To reduce the reading:
➨ Press ”MODE” key to change the direction
of adjustment. The arrow moves to the left
indicating that the pressure reading can be
reduced.
➨ Press ”UP/DOWN” key to reduce the reading.
Evacuate the vacuum gauge to a pressure <0.5 mbar (<0.4 Torr) e. g. by applying a good rotary vane
pump.
➨ Press ”UP/DOWN” key and keep it pressed
while switching on.
☞The vacuum gauge switches into the adjust-
ment mode (indicated by a warning triangle).
☞The reading is adjusted automatically to
”zero”.
Attention: Adjustment under vacuum with a pressure higher than 0.5 mbar (0.4 Torr) reduces the ac-
curacy of measurement. If the pressure is signicantly higher than 0.5 mbar (0.4 Torr), adjustment to a
reference pressure is recommended.
➨ Press ”ON/OFF” key to conrm the adjust-
ment and to terminate the adjustment mode.
Adjustment at atmospheric pressure
+
➨ Press ”ON/OFF” key to conrm the adjust-
ment and to terminate the adjustment mode.
Adjustment under vacuum
+

page 12 of 17
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully
the validity of this document with respect to his product. manual-no.: 999001 / 04/04/2012
Control of measuring equipment
The VACUUBRAND DAkkS calibration laboratory is accredited by the Deutsche Akkreditierungsstelle
GmbH (national accreditation body of the Federal Republic of Germany) for the measurable variable
pressure in the pressure range from 7.5*10-4 Torr to 975 Torr (10-3 mbar to 1300 mbar) in accordance with
the general criteria for the operation of testing laboratories dened in the DIN EN ISO/IEC 17025:2000
series of standards (accreditation number D-K-15154-01). The DAkkS is signatory to the multilateral
agreements of the European cooperation for Accreditation (EA) and of the International Laboratory Ac-
creditation Cooperation (ILAC) for the mutual recognition of calibration certicates.
Rely on calibration in the VACUUBRAND calibration laboratory:
- To meet the requirements of the DIN ISO 9000ff and 10012 series of standards regarding the calibra-
tion of inspection, measuring and test equipment at specied intervals.
- To document that the vacuum gauges calibrated are traceable to national standards of the PTB
(Physikalisch-Technische Bundesanstalt; German national institute for science and technology and
the highest technical authority of the Federal Republic of Germany for the eld of metrology and
certain sectors of safety engineering).
Evacuate the vacuum gauge to an exactly known reference pressure within the range of 0 .... 20 mbar
(0 .... 15 Torr). Switch the vacuum gauge to the adjustment mode (see ”Adjustment under vacuum”).
Attention: The accuracy of the value of the reference pressure will directly affect the accuracy of the
adjustment. If the nominal ultimate vacuum of a diaphragm pump is used as ”reference” vacuum, the ac-
curacy of the adjustment of the vacuum gauge might be doubtful. The diaphragm pump may not achieve
the specied value due to condensate, poor state, failure of the valves or the diaphragm.
➨ Press ”UP/DOWN” key to adjust the reading
from ”0” (”zero”) to the actual reference pres-
sure in the vacuum line in the range of 0 to 20
mbar (0 to 15 Torr).
➨ Press ”ON/OFF” key to conrm the adjust-
ment and to terminate the adjustment mode.
Adjustment to a reference pressure
Calibration in the factory
DAkkS calibration DVR 2 ............................................................................................................900217

page 13 of 17
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully
the validity of this document with respect to his product. manual-no.: 999001 / 04/04/2012
Note: Some of the components on the printed circuit board are sensitive to electrostatic discharge.
Do not touch components. If necessary, ground the printed circuit board and the person replacing the
battery suitably.
Attention: Use only a battery of the recommended type (see ”Technical data”).
The battery must not be charged or connected to an external voltage.
➨Remove the battery.
➨Insert the new battery into the battery compartment cor-
rectly matching the positive and negative polarity (as
shown by the engraving inside the compartment).
☞When reassembling the vacuum gauge, make sure that
the switch pad and the printed circuit board are positioned
correctly.
➨Carefully screw both parts of the housing together.
battery switch pad
printed circuit board
Replacing the battery
➨Unscrew the four countersunk head screws at the rear
side of the housing with a Phillips screw driver.
➨Disassemble the housing parts carefully.
Attention: The two housing parts are connected by the
sensor cable.

page 14 of 17
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully
the validity of this document with respect to his product. manual-no.: 999001 / 04/04/2012
If the measures listed in section ”Troubleshooting” do not work: Return the equip-
ment to the factory for inspection and repair.
Interference with the equipment (except replacement of the battery) voids all war-
ranties.
Safety and health of our staff, laws and regulations regarding the handling of dan-
gerous goods, occupational health and safety regulations and regulations regarding
safe disposal of waste require that for all pumps and other products the “Health and
safety clearance form“ must be sent to our ofce duly completed and signed before
any equipment is dispatched to our premises.
Fax or post a completed copy of the health and safety clearance form to us in ad-
vance. The declaration must arrive before the equipment. Enclose a second com-
pleted copy with the product. If the equipment is contaminated you must notify the
carrier.
No repair / DAkkS calibration is possible unless the correctly completed form
is returned. Inevitably, there will be a delay in processing the equipment if in-
formation is missing or if this procedure is not obeyed. Eventually, acceptance
of the returned equipment will be refused.
If the product has come in contact with chemicals, radioactive substances or other
substances dangerous to health or environment, the product must be decontami-
nated prior to sending it back to the factory.
To expedite repair and to reduce costs, please enclose a detailed description of
the problem and the product’s operation conditions with every product returned for
repair.
We submit quotations only on request and always at the customer’s expense. If
an order is given, the costs incurred are offset from the costs for repair or from the
purchase price, if the customer prefers to buy a new product instead of repairing the
defective one.
☞ If you do not wish a repair on the basis of our quotation, the equipment
might be returned to you disassembled and at your charge!
We replace parts due to optical aspects upon your request (at your expense).
Before returning the equipment ensure that (if applicable):
☞Equipment has been cleaned and/or decontaminated.
☞All ports have been sealed.
☞Equipment has been properly packed, if necessary, please order an original pack-
aging (costs will be charged), marked as appropriate and the carrier has been
notied.
☞Ensure that the completed health and safety clearance form is enclosed.
We hope for your understanding for these measures, which are beyond our control.
Scrapping and waste disposal:
Dispose of the equipment and any components removed from it safely in accord-
ance with all local and national safety and environmental requirements. Particular
care must be taken with components which have been contaminated with dangerous
substances from the process. Do not incinerate uoroelastomer seals and O-rings.
☞You may authorize us to dispose of the equipment at your expense.
Notes on return to the factory
Repair - return - DAkkS calibration
NOTICE
NOTICE

page 15 of 17
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully
the validity of this document with respect to his product. manual-no.: 999001 / 04/04/2012
Health and safety clearance form
1. Device (Model): ................................................................... 2. Serial no.: .........................................
3. Reason for return / malfunction:
..............................................................................................................................................................
4. Has the device been used in a copper process step (e.g., semiconductor production).
☐yes ☐no
5. Substances (gases, liquids, solids) in contact with the device / which have been pumped:
..............................................................................................................................................................
..............................................................................................................................................................
..............................................................................................................................................................
..............................................................................................................................................................
6. Prior to return to the factory the device has been decontaminated. ☐yes ☐no
Description of the decontamination method and the test / verication procedure:
..............................................................................................................................................................
..............................................................................................................................................................
7. The device is free of hazardous, harmful substances. ☐yes ☐no
8. Protective measures required for VACUUBRAND employees:
..............................................................................................................................................................
9. If the paint is damaged, we wish a repaint or a replacement of parts for reason of appearance (repaint
and replacement at customer’s expense). ☐yes ☐no
10. Legally binding declaration
We assure for the returned device that all substances, which have been in contact with the device are
listed in section 5 and that the information is complete and that we have not withheld any information. We
declare that all measures - where applicable - have been taken listed in section “Return to the factory”.
By our signature below, we acknowledge that we accept liability for any damage caused by providing in-
complete or incorrect information and that we shall indemnify VACUUBRAND from any claims as regards
damages from third parties. We are aware that as expressed in § 823 BGB (Public Law Code of Ger-
many) we are directly liable for injuries or damages suffered by third parties, particularly VACUUBRAND
employees occupied with handling/repairing the product.
Shipping of the device must take place according to regulations.
Name: ....................................................................... Signature: ..............................................................
Job title: .................................................................... Company’s seal:
Date: .......................................................................
Release for repair grant by VACUUBRAND (date / signature): ..............................................................
Devices will not be accepted for any handling before we have received this declaration.
Please read and comply with ”Notes on return to the factory”.
Oil lled pumps: Drain oil prior to shipping absolutely!
VACUUBRAND GMBH + CO KG
Alfred-Zippe-Straße 4
97877 Wertheim, Germany
Tel.: +49 9342 808-5660
Fax: +49 9342 808-5666
E-Mail: [email protected]
www.vacuubrand.com

page 16 of 17
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully
the validity of this document with respect to his product. manual-no.: 999001 / 04/04/2012
Hersteller / Manufacturer / Fabricant:
VACUUBRAND GMBH + CO KG · Alfred-Zippe-Str. 4 · 97877 Wertheim · Germany
Hiermit erklärt der Hersteller, dass das Gerät konform ist mit den Bestimmungen der Richtlinie 2004/108/
EG.
Hereby the manufacturer declares that the device is in conformity with the directive 2004/108/EC.
Par la présente, le fabricant déclare, que le dispositif est conforme à la directive 2004/108/CE.
Vakuummessgerät / Vacuum gauge / Vacuomètre
Typ / Type / Type: DVR 2
Artikelnummer / Order number / Numéro d‘article: 682902
Seriennummer / Serial number / Numéro de série: Siehe Typenschild / See rating plate / Voir plaque
signalétique
Angewandte harmonisierte Normen / Harmonized standards applied / Normes harmonisées utilisées:
DIN EN 61326-1
Bevollmächtigter für die Zusammenstellung der technischen Unterlagen / Person authorised to compile
the technical le / Personne autorisée à constituer le dossier technique:
Dr. J. Dirscherl · VACUUBRAND GMBH + CO KG · Alfred-Zippe-Str. 4 · 97877 Wertheim · Germany
. . . . . . . . . . . . . . . . . . . . . . . . . . . .
(Dr. F. Gitmans)
Geschäftsführer / Managing director / Gérant
. . . . . . . . . . . . . . . . . . . . . . . . . . . .
Ort, Datum / place, date / lieu, date
Wertheim, 04.04.2012
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
(Dr. J. Dirscherl)
Technischer Leiter / Technical Director / Directeur technique
ppa.
EG-Konformitätserklärung
EC Declaration of Conformity
Déclaration CE de conformité
VACUUBRAND GMBH + CO KG
Alfred-Zippe-Str. 4 · 97877 Wertheim
T +49 9342 808-0 · F +49 9342 808-5555
www.vacuubrand.com

page 17 of 17
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully
the validity of this document with respect to his product. manual-no.: 999001 / 04/04/2012
Disclaimer: Our technical literature is only intended to inform our customer. The validity of general em-
pirical values and results obtained under test conditions for specic applications depend on a number
of factors beyond our control. It is therefore strictly the users´ responsibility to very carefully check the
validity of application to their specic requirements. No claims arising from the information provided in
this literature will, consequently, be entertained.
VACUUBRAND GMBH + CO KG
-Technology for Vacuum Systems-
© 2012 VACUUBRAND GMBH + CO KG Printed in Germany
Alfred-Zippe-Str. 4 · 97877 Wertheim / Germany
T +49 9342 808-0 · F +49 9342 808-5555
Other manuals for DVR 2
1
Table of contents
Other vacuubrand Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

Knick
Knick Stratos 2201 pH instruction manual

Trace2O
Trace2O HMUV1825P manual

TSI Instruments
TSI Instruments Alnor RVA801 Operation and service manual

Rohde & Schwarz
Rohde & Schwarz RPG HATPRO Installation and maintenance guide

Knauer
Knauer K-7400 user manual

Tempo Communications
Tempo Communications CableScout CS90B user manual