vacuubrand DCP 3000 User manual

page 1 of 36
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully
the validity of this document with respect to his product. Manual-no.: 999162 // 19/01/2011
Vacuum gauge
Technology for Vacuum Systems
Instructions for use
DCP 3000

page 2 of 36
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully
the validity of this document with respect to his product. Manual-no.: 999162 // 19/01/2011
➨DANGER indicates a hazardous situation which, if not avoided, will re-
sult in death or serious injury.
☞WARNING indicates a hazardous situation which, if not avoided, could
result in death or serious injury.
• CAUTIONindicatesahazardoussituationwhich,ifnotavoided,could
result in minor or moderate injury.
NOTICEisusedtoaddresspracticesnotrelatedtopersonalinjury.
DisconnectequipmentfromACpower.
Dear customer,
YourVACUUBRAND vacuum gauges should support you for many years without trouble and with op-
timal performance. Thanks to our long practical experience we have much information and advice on
how you can achieve powerful application usage and personal safety through our products. Please read
theseinstructionsforusebeforetheinitialoperationofyourdevice.VACUUBRANDvacuumgauges
are the result of many years of experience in design and construction and practical operation of these
gauges combined with the latest developments in material and manufacturing technology.
Ourqualitymaximisthe„zerofaultprinciple“:Everyvacuumgaugeleavingourcompany,istestedin-
tensivelyincludinganendurancerun.Thereforefaults,eventhosewhichoccurrarely,areidentiedand
canbeeliminatedimmediately.Theachievementofthespecicationsaftertheendurancerunistested
for every device.
Every VACUUBRAND device achieves the specications.
We are committed to providing our customers with this high quality standard.
We know that the device cannot replace all of your real work and hope that our products contribute to
an effective and trouble-free realisation of your work.
YoursVACUUBRANDGMBH+COKG
After sales service: Contact your local dealer or call +49 9342 808-193.
NOTICE

page 3 of 36
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully
the validity of this document with respect to his product. Manual-no.: 999162 // 19/01/2011
Portuguê
Ρyccкий
Polski
Nederl.
Suomi
日本語
中文
한국어
DCP 3000 V2.0
Deutsch
English
Français
Italiano
Español
Türkçe
Reset / Language selection
1 switch off 2 press both
3 turn
4 press
Portuguê
Ρyccкий
Polski
Nederl.
Suomi
日本語
中文
한국어
DCP 3000 V2.0
Deutsch
English
Français
Italiano
Español
Türkçe

page 4 of 36
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully
the validity of this document with respect to his product. Manual-no.: 999162 // 19/01/2011
Contents
Reset / Language selection .....................................................................................3
Safety information ......................................................................................................5
General information............................................................................................................................5
Intended use ......................................................................................................................................5
Ambient conditions ............................................................................................................................6
Operatingconditions .........................................................................................................................6
Safety during operation .....................................................................................................................6
Technical data .............................................................................................................8
Technical data of display unit ............................................................................................................8
Technical data of power supply .........................................................................................................8
TechnicaldataofpressuretransducersVSK3000andVSP3000 ..................................................9
Wetted parts ....................................................................................................................................10
Use and operation ....................................................................................................11
Assemblingthecountryspecicmainsplug.................................................................................... 11
Display and symbols ....................................................................................................................... 11
ChangingthevacuumconnectoratpressuretransducerVSK3000 ..............................................13
Notes on operation ..........................................................................................................................13
Menu ..........................................................................................................................16
Conguration ............................................................................................................17
Connecting several pressure transducers.............................................................19
Manualswitchingbetweenpressuretransducers ............................................................................19
AutomaticswitchingbetweenpressuretransducersVSK3000andVSP3000...............................19
Measuringdifferentialpressure ........................................................................................................20
Reconguringthepressuretransducer ............................................................................................20
Setpoints ...................................................................................................................21
ConguringthevalvesordigitalI/Ointerfacestosetsetpoints........................................................21
Interface parameters ................................................................................................22
Setting the interface ........................................................................................................................22
Notes concerning the function data logger (recording and sending of data)...................................23
Read commands .............................................................................................................................24
Write commands .............................................................................................................................26
Troubleshooting .......................................................................................................27
Accessories ..............................................................................................................28
Cleaning the pressure transducers ........................................................................29
Calibration in the factory .........................................................................................29
Readjustment of pressure transducer VSK 3000 ..................................................30
Readjustment of pressure transducer VSP 3000 ..................................................31
Notes on return to the factory .................................................................................32
Health and safety clearance form ...........................................................................33
Trademark index:
VACUU•LAN®(US-Reg.No3,704,401),VACUU•BUSTM,VACUU•CONTROLTM,chemistry-HYBRIDTM, Peltron-
icTM,TURBO•MODETM,VARIO®,VARIO-SPTM,VACUUBRAND®(US-Reg.No3,733,388)and alsotheshown
companylogosaretrademarksof VACUUBRANDGMBH+CO KG in Germanyand/orothercoun-
tries.

page 5 of 36
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully
the validity of this document with respect to his product. Manual-no.: 999162 // 19/01/2011
General information
To operate the vacuum gauge DCP 3000, at least one external
pressure transducer VSK 3000 or VSP 3000 or MPT 100 is required!
Read and comply with this manual before installing or operating the equip-
ment.
Remove all packing material, remove the product from its packing-box, remove the
protective covers and keep, inspect the equipment.
If the equipment is damaged, notify the supplier and the carrier in writing within three
days; state the item number of the product together with the order number and the
supplier’s invoice number. Retain all packing material for inspection.
Do not use the equipment if it is damaged.
If the equipment is not used immediately, replace the protective covers. Store the
equipment in suitable conditions.
Intended use
☞ OperatethedeviceonlyincombinationwithVACUUBRANDgenuineaccesso-
ries.Makesure thattheindividual componentsareonlyconnected, combined
and operated according to their design and as indicated in the instructions for
use.
☞ Comply with notes on correct vacuum and electrical connections, see section
”Useandoperation”.
• Thedeviceisdesignedforambient and gas temperatures at the pressure trans-
ducerconnectionof+10°Cto+40°Catcontinuousoperation,upto+80°Cfor
short periods (< 5 minutes) at the pressure transducer. If installing the device into
a cabinet or a housing check maximum temperature. Ensure that the maximum
permitted gas temperature at the pressure transducer (see ”Technical data”) is
not exceeded.
Usetheequipmentfortheintendeduseonly,i.e.formeasurementofvacuuminves-
sels designed for that purpose.
Connecting the device
☞Avoid uncontrolled overpressure (e.g., when connecting to a locked or blocked
tubing system). Risk of bursting.
• Comply with max. permitted pressure at the pressure transducer, see section
”Technical data”.
•Connecthosesatthepressuretransducergastight.
• The device is equipped with a short circuit proof wide range power supply
with an integrated overload protection.
• Checkthatmains voltageandcurrent conformwiththe equipment(seerating
plate).
• Prior to using the wall power supply, assemble and lock the suitable mains plug
(included in delivery) to the power supply.
Safety information
NOTICE
NOTICE

page 6 of 36
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully
the validity of this document with respect to his product. Manual-no.: 999162 // 19/01/2011
• Comply with max. permitted ambient and gas temperature (see ”Technical
data”) and make sure ventilation is adequate if the equipment is installed in a
housing or if ambient temperature is elevated.
• Avoidhighheatsupply(e.g.,duetohotprocessgases).
• Incaseofresiduesoraggressiveorcondensablemediainstallagaswashing
bottle if necessary.
Position the pressure transducer and its vacuum connection lines so that conden-
satecannotowtowardsthepressuretransducer.
Ensurestabilityofthehoseconnection.Complywithallrelevantsafety require-
ments.
When the device is brought from cold environment into a warm room for operation
bedewing may occur. Allow the device to acclimatise.
Complywith national safety regulations and safety requirements concerning the
useofvacuumandelectricalequipment.Complywithallapplicable and relevant
safety requirements (regulations and guidelines), implement the required ac-
tions and adopt suitable safety measures.
Ambient conditions
To the best of our knowledge the equipment is in compliance with the requirements
of the applicable EC-directives and harmonized standards with regard to design,
typeandmodel,especiallydirectiveEN61010-1.Thisdirectivegivesindetailcon-
ditions under which the equipment can be operated safely (see also IP degree of
protection).
Adopt suitable measures in case of differences, like using the equipment outdoors,
installationinaltitudesofmorethan1000mabovemeansealevel,conductivepol-
lution or dewiness.
Operating conditions
➨The devices are not suitable for applications which involve dangerous or ex-
plosive gases or explosive or ammable mixtures.
Ensure that the materials of the wetted parts are compatible with the substances
processedinthevacuumsystem,seesection“Technicaldata”.
Safety during operation
➨Adopt suitable measures to prevent the release of dangerous, toxic, explosive,
corrosive,noxiousorpollutinguids,vapoursandgases.
➨Prevent any part of the human body from coming in contact with the vacuum.
OperationwithpressuretransducerVSK 3000:
☞Attention: Ifthepressureishigherthanapproximately1060mbarthepressure
reading becomes incorrect (saturation of the pressure transducer). The display
ashes.Immediate pressure relief necessary! Risk of bursting!
NOTICE
NOTICE
NOTICE

page7of36
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully
the validity of this document with respect to his product. Manual-no.: 999162 // 19/01/2011
OperationwithpressuretransducerVSP 3000:
☞Attention:Maximumpressurereading:1*103 mbar.Pressurevaluesabove1000
mbar can not be displayed. Danger of unnoticed overpressure! Risk of burst-
ing!
OperationwithpressuretransducerMPT 100:
☞SeeinstructionsforuseofMPT100.
• Useonlygenuine spare parts and accessories.Otherwisesafetyandperform-
ance of the equipment as well as the electromagnetic compatibility of the equip-
ment might be reduced.
PossiblytheCEmarkorthecTÜVusmarkbecomevoidifnotusingOEMspare
parts.
Electronicequipmentisnever100%fail-safe.Thismayleadtoanill-denedsta-
tus of the equipment or of other connected devices. Provide protective measures
against malfunction and failure. Ensure that in case of failure the device and the
vacuum system always will turn into a safe status.
Maintenance and repair
➨ Beforestartingmaintenanceunplugthepowersupply(wallplug).
➨Wait two minutes after isolating the equipment from mains to allow the capaci-
tors to discharge.
• Attention: Due to the operation the equipment might be contaminated by harmful
or dangerous substances, clean or decontaminate prior to maintenance.
Ensure that maintenance is done only by suitably trained and supervised techni-
cians.
Interior components of the device can only be repaired at the factory.
Returnthepressuretransducertothefactoryforrepair.Openingthehousingofa
pressure transducer will void any warranty.
Returned products will not be repaired or calibrated until the completed health and
safety clearance form has been received. In order to comply with law (occupational,
health and safety regulations, safety at work law and regulations for environmental
protection) vacuum pumps, components and measuring instruments returned to the
manufacturer can be repaired only when certain procedures (see section ”Notes on
return to the factory”) are followed.
NOTICE
NOTICE

page 8 of 36
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully
the validity of this document with respect to his product. Manual-no.: 999162 // 19/01/2011
We reserve the right for technical modication without prior notice!
Technical data
Technical data of power supply
Technical data of display unit
Display unit DCP 3000
Display illuminatedLCDgraphicdisplay
Pressureunits/scale(selectable) mbar, Torr or hPa
Ambient temperature range (operation) 10°Cto+40°C
Ambient temperature range (storage) -10°Cto+70°C
Permitted relative atmospheric moisture during
operation (no condensation) 30%to85%
Maximumpermissiblerangeofsupplyvoltage 24 V= ( ±6V )
Maximumpermissiblecurrentofconnected
valves (connected components) 4A(withsufcientlypowerfulpowersupply)
Maximumpowerdraw(internal) 3.4W(140mAat24V=)
DegreeofprotectionaccordingtoIEC529
(front side) IP 42
Interface RS232C
Cablelengthofpowersupply approx. 2 m
Weight (without power supply) 0.44kg
DimensionsLxWxH(incl.foot,withoutmains
adapter) 144 mm x 124 mm x 114 mm
Power supply (wall plug)
Inputvoltage(±10%) 100-240V~,47-63Hz
Max.currentdraw 0.8A
Ambient temperature range (operation) 0°Cto+40°C
Ambient temperature range (storage) -20°Cto+85°C
Outputvoltage 24 V=, short-circuit proof
Max.outputcurrent 1.25 A
Mainsconnection exchangeableplugEurope/UK/US/AUS
Weight 0.24kg
DimensionsLxWxH 108mmx58mmx34mm

page 9 of 36
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully
the validity of this document with respect to his product. Manual-no.: 999162 // 19/01/2011
Technical data of pressure transducers VSK 3000 and VSP 3000
Type VSK 3000 VSP 3000
Measuringprinciple
ceramic diaphragm (alumina),
capacitive, absolute pressure,
gas type independent
thermal conductivity according to
Pirani, dependent on gas type
Measuringrange(absolute) 1060mbar-0.1mbar
(795Torr-0.1Torr)
1*103mbar-1*10-3 mbar
(7.5*102Torr-7.5*10-4 Torr)
Resolution 0.1mbar 10%ofdisplayeddecade
Measurementuncertainty
(absolute) after careful adjust-
ment and at constant tem-
perature
<±1mbar(0.75Torr)/
±1 digit
1*102mbar-1*10-2 mbar
(7.5*101Torr-7.5*10-3Torr):
±15%ofdisplayedvalue
Maximumpermissiblepres-
sure at pressure transducer 1.5 bar (1125 Torr) absolute
Maximumpermissibletem-
perature of gaseous media at
measurement connection
continuousoperation:40°C,
forshortperiods(<5minutes):upto80°C
Temperaturecoefcient <±0.07mbar/K(0.05Torr/K) notspecied
Ambient temperature range
(operation) 10°Cto+40°C
Ambient temperature range
(storage) -10°Cto+60°C
Permitted relative atmospher-
ic moisture during operation
(no condensation)
30%to85%
Range of supply voltage
(viaVACUU•BUS) 6-30VDC/5mA 18-30VDC/65mA
DegreeofprotectionIEC529 IP 54
Communication VACUU•BUS
Weight
withsmallange
with hose nozzle
with tubing connection
185 g
180g
178g
180g
185 g
-
Housingdimensions
diameter
length with vacuum
connection
60mm
60mm(smallange)
95 mm (hose nozzle)
63 mm (tubing connection)
60mm
58 mm
97mm
-
Vacuum connection
smallangeDN16or
hosenozzle6/10mmorconnec-
tionforPTFEtubingDN8/10mm
smallangeDN16and
screw-inhosenozzle6/10mm
Internal volume of measure-
ment chamber
withsmallange:4.1cm3
withhosenozzle:4.4cm3
withtubingconnection:3.5cm3
2.9 cm3
withhosenozzle:2.5cm3
Cablelength approximately2.0m
We reserve the right for technical modication without prior notice!
Technical data MPT 100: See instructions for use of MPT 100

page10of36
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully
the validity of this document with respect to his product. Manual-no.: 999162 // 19/01/2011
Wetted parts
Components Wetted materials
VSK 3000
Sensor aluminium oxide ceramics
Sensor housing, measurement chamber PPS,glassbre
Seal at sensor chemicallyresistantuoroelastomer
Hosenozzle PP
Clampingring PA
Smallange stainless steel
VSP 3000
Sensor aluminium oxide ceramics
Sensor housing, measurement chamber, small
ange PBT,glassbre/PUR
Hosenozzle/O-ring PPS,glassbre/FPM
We reserve the right for technical modication without prior notice!

page 11 of 36
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully
the validity of this document with respect to his product. Manual-no.: 999162 // 19/01/2011
warningnotice(ifnecessaryincombinationwithothersymbols),ashing
venting valve switched on (if connected)
mbar
Torr
hPa preselected pressure unit
1013.2 actual absolute pressure at the pressure transducer
VSK 1 / VSP 1
active pressure transducer (status line)
Use and operation
WhenswitchingontheDCP3000fortheveryrsttime,amenutoselectthelanguageofthedeviceis
displayed. Select the desired language (e.g. ”English”) by turning the selection knob and press to con-
rm.Thenselectthepressureunit(”mbar”, ”Torr” or” hPa”) in the same way.
It is possible to access the language selection menu at any time by switching on the device while
keeping the selection knob pressed.
After switching on, the version number of the software is displayed.
Theconnectedcomponents(valves,externalsensor)areidentiedandconguratedautomatically.
Assembling the country specic mains plug
☞The wall power supply is delivered with mains plugs for
Europe,UK,USAandAustralia.
☞To replace the mains plug press the locking key and re-
move mains plug.
☞Assemble the suitable mains plug to the power supply and
lock.
Display and symbols
mbar
1013. 2
mbar
VSK 1
1.0
x 10 3
mbar
VSP 1

page 12 of 36
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully
the validity of this document with respect to his product. Manual-no.: 999162 // 19/01/2011
Keys
Venting: Only after connection of an external venting valve
• Shortpressingventsshort-time,controlrunson.
• Pressinglongerthan2secondsventsthesystemto
atmospheric pressure.
Rear side DCP 3000
selection knob
• Presstoreachtheset-upmenuofthefunction
• Turntoreachtheparameterset-up
• Presstoreachthesetvalue
• Turntochangethesetvalue
• Presstoconrmandtoreachfurtherparametersortoexittheset-upmenu
on/offswitch
connection power supply
rating plate
serial interface
RS232C
Attention: Do not cant when assembling or removing plug connections! Comply with correct
positioning of the plug. To connect further components use Y-adapters and extension cables
VACUU•BUS.
Further information on how to use several sensors or valves on request.
sockets for connection
of external pressure
transducerand/or
venting valve

page 13 of 36
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully
the validity of this document with respect to his product. Manual-no.: 999162 // 19/01/2011
Notes on operation
Operation principle of the VSK 3000
TheVSK3000isequippedwithacapacitivepressuretransducerwithceramicdiaphragmtomeasure
the actual pressure independent of the gas type and depending on the vacuum, i. e. absolute.
Thepressurevalueisdisplayedwitharesolutionof0.1mbarinthepressurerangefrom0.1mbarto
1060mbar.
Operation principle VSP 3000
ThepressuretransducerVSP3000reliesonthefactthatthethermalconductivityoftheresidualgas
in the vacuum chamber is a measure for the gas pressure. The thermal conductivity of gases is propor-
tional to the pressure within a certain range and is related to the molecular mass.
The thermal conductivity of gases and vapours varies with their molecular mass. Therefore the pressure
measurement is dependent on the gas type. The pressure transducer has been adjusted for air at the
factory.
Changing the vacuum connector at pressure transducer VSK 3000
Installing the hose nozzle:
Unscrewtheange(usinga17mmopen-endwrench,ifnecessary)toexposethecompressiontting
(1). Slip the compression nut, and then the ferrule, onto the smooth end of the supplied hose nozzle (2).
SlidethesmoothendofthehosenozzleontothecompressionttingontheVSK3000gaugehead,and
tightenthecompressionnutrmlynger-tight(3).
(1)
(3)
(2)
(1) (2)
(3)
Installing the PTFE-tubing connection(PTFE-tubingID:8mm,OD:19mm):
Unscrewtheange(usinga17mmopen-endwrench,ifnecessary)toexposethecompressiontting(1).
Slip the compression nut, and then the ferrule, onto the PTFE-tubing (2). Slide the PTFE-tubing onto the
compressionttingontheVSK3000gaugehead,andtightenthecompressionnutrmlynger-tight(3).

page 14 of 36
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully
the validity of this document with respect to his product. Manual-no.: 999162 // 19/01/2011
General information on handling the VSP 3000
ThepressuretransducerVSP3000whichhasbeendevelopedparticularlyforuseinchemicallaborato-
riesisapressuresensorworkinginthenevacuumrange.Overthewholemeasuringrangethepres-
sure reading is indicated in exponential notation.
AllowtheVSP3000aperiodofapprox.20minutestopreheatandtomeetitsspecications.Evenin
caseofthecontrollerorvacuumgaugebeingswitchedoff,theVSP3000staysreadyforoperation.
Tode-energizethepressuretransducer,unplugeitheritsVACUU•BUSlineorthepowersupplyofthe
controller or vacuum gauge.
Theinteriorofthepressuretransducerishighlysensitive!Donotinsertngersor
tools into the measuring chamber.
VACUU•BUS
Readout of the pressure transducer via VACUU•BUS line by the vacuum gauge DCP 3000 using
VACUU•BUSprotocol.Maximumcablelengthinsidebuildings:30m(extensioncableVACUU•BUS2m:
order-no.612552/extensioncableVACUU•BUS10m:order-no.2618493).
Don’t use more than one DCP 3000 within a VACUU•BUS system. MorethanoneDCP3000within
thesameVACUU•BUSsystemwillinterferewitheachotherandcauseerrormessagesoftheconnected
components.
NOTICE
Diagram:GasdependenceofaVSP3000
true pressure [mbar]
displayed pressure [mbar]
air
helium
argon
☞ Pressureofgaseswithsimilarmass,suchasO2orCO,canbereadoffdirectlywithintheuncertainty
of the measurement.
☞ Withgasesoflowerorhighermass(H2,He,Ar,CO2)itisrecommendedtoreadjusttheVSP3000
using the gas to be measured.

page 15 of 36
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully
the validity of this document with respect to his product. Manual-no.: 999162 // 19/01/2011
Max. permitted pressure at the pressure transducer:
1.5 bar (absolute).
VSK 3000:
☞ The display ashes at a pressure higher than approxi-
mately1060mbar.Immediate pressure release. Risk of
bursting!
VSP 3000:
☞Attention:Maximumpressurereading:1*103 mbar. Pres-
surevaluesabove1000mbararenotdisplayed.Risk of
unnoticed overpressure. Risk of bursting!
Prior to use
➨ Connectthegaugehead VSK 3000orVSP3000withconnecting cable (VACUU•BUS line) toa
socketattherearsideoftheDCP3000.
➨ Connectthegaugeheadtothevacuumchamberviathesmallangeconnection,thehosenozzleor
PTFEtubing.Avoidcontamination(oil/oilmist)ofthegaugeheadwhengeneratingthevacuumwith
anoil-lledvacuumpump.
☞Do not mount the gauge head directly at the oil-sealed pump. The diameter of the pipelines should
be as large as possible.
☞Inside a vacuum system where evaporation occurs (e.g. rotary evaporator) the vacuum is not uniform
(e.g., a condenser acts as pump or the vacuum in the pipeline is lower than in the system). Therefore
choose carefully the position where to connect the gauge head.
☞ Condensateanddepositsatthegaugeheadfalsifythemeasurementresult.
☞If residues occur or when working with aggressive or condensable substances, install a gas washing
bottle in front of the pressure gauge head.
☞Mount the gauge head in such a position that condensate can not ow into it.
VSP 3000:
Recommended orientation: Vertically with vacuum connection pointing downwards. If mounting the
VSP 3000 in any other orientation, a readjustment is recommended (see section ”Readjustment of
VSP3000”).
☞If necessary, clean the gauge head.
➨Switch the equipment on.
Automatic detection of connected components
Whenswitchingonthedevicetheactualcongurationoftheconnectedcomponentsischecked.
Connected components (valves, sensors, etc.) are automatically detected, used and monitored until
thedeviceisswitchedoff.RecongurationispossiblebyswitchingtheDCP3000off/on.Identicalcom-
ponentsmustbeconguredbeforehandtodifferentVACUU•BUSaddresses(e.g.,VSK1andVSK2),
seesection”Connectingseveralpressuretransducers”.
➨ Further information on how to use several sensors or valves on request.
Thecongurationmenuisstoredinthedevice,see”Conguration”withspecicsettingsadaptedauto-
matically to the connected components.
Depending on the connected components (valves, sensors) some menu items are not active!

page 16 of 36
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully
the validity of this document with respect to his product. Manual-no.: 999162 // 19/01/2011
Menu
RS-232
Baud 19200
Parity 8-N-1
HandshakeRTS-CTS
Send Off
Data Cancel
------Back-------
Conguration
Adjustment1000mbar
RS-232 . . .
Sensors . . .
Display . . .
Setpoint . . .
LogData Off
- - - - - - Graphic - - - - - - -
------Back---------
Display
Brightness 100%
Contrast 40%
Sound Off
Unit mbar
LanguageEnglish
------Back--------
Sensors
VSK1 1001mbar
VSP1 1.0E+3mbar
VSK2 780mbar
VSP2 7.8E+2mbar
------Back-------
Switching between pressure transducers while the pressure indication or graph-
ic are displayed:
➨Press the selection knob and keep it pressed. Then turn the still pressed selection knob until the
desired pressure transducer is displayed. Release selection knob.
1013 . 2
mbar
VSK 1
1.0
x
10 3
mbar
VSP 1
pressure indication
with VSK 3000
pressure indication
with VSP 3000
Setpoint
On1 50mbar
Off1 10mbar
On2 9E-2mbar
Off2 5mbar
------Back-------

page17of36
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully
the validity of this document with respect to his product. Manual-no.: 999162 // 19/01/2011
Inthemenu”Conguration”thedeviceparametersareset.
Settings
☞Select the parameters via the selection knob.
Conguration
➨ SwitchtheDCP3000on.
➨Press the selection knob.
➨ Turntheselectionknobtochooseamenuandconrmbypressingtheselectionknob.
’’Conguration’’
Settings for
• Adjustment(ofthepressuretransducer)
• RS-232(serialinterface)
• Sensors(congurationandswitchingbetweenseveralpressuretransducers)
• Display(brightnessandcontrastofthedisplay,language,sound,.....)
• Setpoint(valves,digitalI/Ointerface)
• LogData(Off,1...3600seconds)
• Graphic
☞Adjustment:Adjustmentofthepressuretransducerundervacuumand/oratatmosphericpressure,
seesection”AdjustmentofpressuretransducerVSK3000/VSP3000”.
Adjustmentiscarriedoutattwoadjustmentpoints:atatmosphericpressureandundervacuum.
If no adjustment is possible in the actual pressure range, ”---- mbar” is displayed.
☞RS-232:Congurationoftheinterface,settingofparametersandcommands,seesection”Interface
parameters”.
Baudrateisselectableon19200,9600,4800or2400,parityon”8-N-1”,”7-O-1”or”7-E-1”,Hand-
shakeon”no”,”Xon-Xoff”or”RTS-CTS”andremoteon”On”or”Off”.
Automaticsendingofpressurevalues”off”(onlyondemand)orinatimeintervalof1....360sec-
onds. To switch off automatic sending select ”off” or stop via serial interface.
Send data of the recording.
☞Sensors: Selection of the gauge head to be controlled (maximum 4 sensors of each type VSK
3000orVSP3000(alternativelymax.4xMPT100insteadofVSP3000)orreferencesensorsVSK
3000).
☞Display: Selection of the device parameters
”Brightness”between0-100%and”Contrast”between0-100%ofthedisplay
”Sound””On”or”Off”incaseoffaults
”Units” (pressure units) ”mbar”, ”hPa” or ”Torr”
”Language” of the display ”German”, ”English”, ”French”, ”Italian”, ”Spanish”, ”Turkish”, ”Korean”,
”Chinese”,”Portuguese”,”Russian”,”Polish”,”Dutch”,”Japanese”,”Finnish”,
☞Setpoint:Maximumfoursetpoints“On”andfoursetpoints“Off”canbepresetafterappropriatecon-
gurationofvalvesand/ordigitalI/Ointerfaces(seesection“Setpoints”).
☞Log data:Recordingandinternalstorageoftimeandpressuredataofupto4sensors(LogData
”Off”orinterval1...3600seconds)

page 18 of 36
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully
the validity of this document with respect to his product. Manual-no.: 999162 // 19/01/2011
Conguration
Adjustment 1000mbar
RS-232 . . .
Sensors . . .
Display . . .
Setpoint . . .
Logdata Off
- - - - - - Graphic - - - - - -
------Back--------
The screen-shot shows the factory-set values.

page 19 of 36
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully
the validity of this document with respect to his product. Manual-no.: 999162 // 19/01/2011
Connecting several pressure transducers
ItispossibletoconnectseveralexternalpressuretransducerssimultaneouslytotheDCP3000.Maxi-
mumnumberofpressuretransducersofthesametype(VSK3000orVSP3000)connectedtoone
DCP3000:four.Alternativelyconnect1xMPT100insteadofoneVSP1000fromDCP3000software
version2.03or4xMPT100insteadofVSP1000fromDCP3000softwareversion2.22.Thatisatotalof
eightexternalpressuretransducersatmaximum.UptofouradditionalpressuretransducersVSK3000
maybeconnectedasreferenceifconguredaccordingly(see”Measuringdifferentialpressure”).Only
the reading of one pressure transducer is displayed at a time.
☞ TheVACUU•BUSaddressoftheactiveanddisplayedpressuretransducerisshownintheupperleft
corner (status line). The readings of all pressure transducers are shown in the menu ”Sensors”. If in
themenu“Sensors”thedisplayofasensorisashing,thisindicatesafaultatthesensor.
☞ Eachpressuretransducerneedsanindividualaddress(VSK1toVSK4/VSP1toVSP4/MPT1to
MPT4)tocommunicatewiththeDCP3000viaVACUU•BUS.Allpressuretransducersareshipped
withtheirVACUU•BUSaddressfactory-settoVSK1orVSP1orMPT1.
☞ UsetheDCP3000torecongurethepressuretransducersVSK3000orVSP3000(e.g.,changing
theVACUU•BUS addressfromVSK1toVSK2),seesection”Reconguringthepressure trans-
ducer”.
CongurationoftheMPT100:SeeinstructionsforuseoftheMPT100.
Manual switching between pressure transducers
1st possibility:
➨Select the menu item ”Sensors” in the menu ”Conguration”.
☞ ConnectedpressuretransducersaredisplayedintheorderoftheirVACUU•BUSaddress:VSK1-
VSP1-VSK2-VSP2-...-VSP4.
➨ Select the desired pressure transducer by turning and pressing the selection knob.
➨ Presstheselectionknobagaintoexitthecongurationmenu.
☞The selected pressure transducer is displayed in the status line as active pressure transducer.
2nd possibility:
☞Switching during pressure indication or in graphic display
➨Press the selection knob and keep it pressed.
➨ Turntheselectionknobuntilthedesiredpressuretransducerisindicated(seeVACUU•BUSaddress
in the status line).
➨ Release the selection knob.
☞The selected pressure transducer is displayed in the status line as active pressure transducer.
Automatic switching between pressure transducers VSK 3000 and VSP 3000
IfbothapressuretransducerVSK3000andaVSP3000areconnectedtothevacuumgaugeDCP
3000theDCP3000automaticallydisplaysthereadingsofthepressuretransducermoresuitableinthe
current pressure range.
ThisrequiresbothpressuretransducerstobeconguredtocorrespondingVACUU•BUSaddresses(i.e.
VSKxandVSPx).IfthetwoVACUU•BUSaddressesdonotcorrespond(i.e.VSKxandVSPy),anau-
tomaticswitchingbetweenpressuretransducersisnotpossible.Itthenisnecessarytorecongurethe
addressofonepressuretransducerinordertoenableautomaticswitching(seesection”Reconguring
the pressure transducer”.)
Switchover points between pressure transducers VSK 3000 and VSP 3000:
Incaseofdecreasingpressure:SwitchingfromVSK3000toVSP3000atapressureof1mbar
Incaseofincreasingpressure:SwitchingfromVSP3000toVSK3000atapressureof5mbar
Theactivepressuretransducerisshowninthestatusline.Furthermore,aVSK3000isalwaysdisplayed
indecimalnotation,aVSP3000inexponentialnotation.

page20of36
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully
the validity of this document with respect to his product. Manual-no.: 999162 // 19/01/2011
Measuring differential pressure
A differential pressure between two pressure transducers can only be determined and indicated be-
tweentwopressuretransducersVSK3000connectedtotheDCP3000.
☞ Thisrequiresoneofthetwopressuretransducerstobedenedasreference(e.g.,VACUU•BUSad-
dress ”Ref. 1”).
☞ In addition, both pressure transducers have to be congured to corresponding VACUU•BUS ad-
dresses(VSKxandRef.x).Ifthetwoaddressesdonotcorrespond(VSKxandRef.y),nodifferen-
tialpressurewillbedeterminedordisplayed.Thenitisnecessarytoreconguretheaddressofone
pressuretransducer toenable differentialpressuremeasurement(seesection”Reconguringthe
pressure transducer”.)
➨ Thedisplayedpressurevalueis:(”pressurevalueofRef.x”minus”pressurevalueofVSKx”)mbar
Reconguring the pressure transducer VSP 3000 or VSK 3000
Conguration of the MPT 100: See instructions for use of MPT 100.
ReconguringtheVACUU•BUSaddressofapressuretransducer:
1. SwitchthevacuumgaugeDCP3000off.
2. PreferablyconnectonlythepressuretransducertobereconguredtotheDCP3000.
3. Keepkey”VENT”pressedandswitchtheDCP3000on(key”ON/OFF”).
4. Select the menu item ”Vacuubus” by turning and pressing the selection knob.
5. TherstcomponentdetectedbytheVACUU•BUSisdisplayed(e.g.,VSK1).Ifnecessary,selectthe
pressuretransducertobereconguredbyturningandpressingtheselectionknob.
6. Keepkey”VENT”pressedandpressselectionknobadditionally.Theinversedisplayofthecompo-
nent is replaced by a frame.
7. Turntheselectionknobuntiltheaddress,towhichyouwanttorecongurethepressuretransducer
(e.g.,VSK2orRef.1),isdisplayed.
8. Presstheselectionknob.Ifthecongurationwassuccessful,themenuisquitandtheDCP3000
restarts.
AcongurationofmorethanonepressuretransducertothesameVACUU•BUSaddresswillresultin
errorconditionsattheDCP3000andmustthereforebeavoided!
DonotcongureaVSK3000asVSP3000.
DonotcongureaVSP3000asVSK3000.
Furtherinformationconcerningthecongurationofothercomponents(e.g.,valves)ondemand.
Other manuals for DCP 3000
1
Table of contents
Other vacuubrand Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

Anritsu
Anritsu MS2687B Brochure & specs

Test Equipment Depot
Test Equipment Depot IDEAL 700 Series quick start guide

Kikusui
Kikusui KMP1000 user manual

Afriso EURO-INDEX
Afriso EURO-INDEX DTA 10 instruction manual

Apera
Apera ORP60 instruction manual

Orient Detectors
Orient Detectors Titan Ger 1000 user manual

TPI
TPI SP620 user manual

Endress+Hauser
Endress+Hauser Proline Promag 50 operating instructions

Tektronix
Tektronix TLA7000 Series installation manual

Renishaw
Renishaw OMM-S SPRINT quick start guide

Seafloor
Seafloor HyDrone-G2 RCV Quick start guide & user manual

Pfeiffer Vacuum
Pfeiffer Vacuum TPG 500 installation instructions