To congure VOCSN and begin providing High Flow therapy:
1 Select the Therapy tab on the touchscreen, then Ventilation. To congure therapy, you must select an inactive
preset.
Note: If there is only one Preset visible on the screen, rst press Settings & Alarms and use the Preset Enable
control to enable a second Preset.
2 Press Settings & Alarms to congure the Preset you intend to use to deliver therapy. To change a control
setting, highlight the relevant control and then select Edit. Enter the new setting value, and then select ACCEPT.
Congure the inactive Preset as follows:
Control Setting
Circuit Type Valveless
Mode SIMV-Pressure or SIMV-Volume
High Flow On
Patient Type <desired setting>
Flow <desired setting>
3 Press Exit to return to the Ventilation screen, and then begin using the Preset you congured for High Flow. Press
the Preset to highlight it, and then press START beneath it.
4 Run a Pre-Use Test. If not prompted to begin one automatically, press the Menu tab, and then the PRE-USE
TEST button on the touchscreen. Carefully follow the on-screen instructions. Ensure the Pre-Use Test passes
successfully.
5 Next, begin Oxygen therapy. Select the Therapy tab on the touchscreen, then Oxygen. Select the Preset you
would like to congure, and then press Settings & Alarms. Set the Oxygen therapy controls as follows:
Control Setting
Oxygen Source External High-Pressure
Oxygen Delivery Mode FiO2
FiO2 <desired setting>
Low and High FiO2 Alarms <desired settings>
6 Press Exit to return to the Oxygen screen, and then begin using the Preset you congured for Oxygen therapy.
Press the Preset to highlight it, and then press START beneath it.
7 You are now ready to connect the patient and deliver High Flow therapy.