Veridian Healthcare 11-516 Piggy User manual

INSTRUCTION MANUAL • ENGLISH & ESPAÑOL
Pediatric Nebulizer Systems
Model 11-515 Puppy
Model 11-516 Piggy
Please read this instruction manual
completely before operating this unit.
3-YEAR 2-YEAR
3-YEAR
2-YEAR
1-YEAR
1-YEAR

2 • ENGLISH
Pediatric
Character
Compressor Body
Disposable
Nebulizer
Air Tubing
Child Mask
Angled
Mouthpiece
(5) Air Filters
Instruction Manual
FAQ Information Sheet
Replacement Parts
Order Form
Stop!
Please ensure you have all of the following
components before using your Nebulizer
IF YOU ARE MISSING ANY PARTS, INCLUDING INSERTS OR
INSTRUCTION MANUALS, DO NOT RETURN TO PLACE OF
PURCHASE. CONTACT CUSTOMER CARE AT 866-326-1313.

ENGLISH • 3
Care and Safety Information.....................................................................................4
Introduction..................................................................................................................5
Product Features..........................................................................................................6
Nebulizer Mask/Mouthpiece Selection..................................................................7
Using Your Nebulizer ........................................................................................... 8-11
Cleaning and Disinfection................................................................................12-13
Changing the Filter .................................................................................................. 14
Troubleshooting ....................................................................................................... 15
Warnings and Cautions........................................................................................... 16
Product Specications............................................................................................. 16
Warranty ..................................................................................................................... 17
Instrucciones en Español.................................................................................. 19-35
INDEX
Manufactured for:
Veridian Healthcare
1175 Lakeside Drive
Gurnee, IL 60031
www.veridianhealthcare.com
Made in China
#93-1270 12/14
©2014 Veridian Healthcare, LLC
Toll-Free Customer Care Help Line: 1-866-326-1313
Monday – Friday 8:30 a.m. – 4:30 p.m. CST

4 • ENGLISH
CARE & SAFETY INFORMATION
NOTE: Read all instructions carefully before use. The following basic precautions are needed
when using an electrical product. Failure to read and observe all precautions could result in
personal injury or equipment damage.
Improper care or use of your nebulizer may result in injury, damage to the unit or
ineective treatment. Following these instructions will ensure the blood pressure
monitor’s ecacy and long life.
• Never allow the nebulizer to be used without adult supervision
• Do not use the nebulizer for anything except medical nebulization treatment as directed
by a physician. Do not use the nebulizer in a manner inconsistent with these instructions
• Never use any liquid in the nebulizer other than those prescribed by your doctor. Liquids
such as cough medicines or essential oils can harm both the machine and the patient
• The nebulizer is an electronic device, keep away from water to avoid electrical shock. Do
not immerse the unit in liquid and do not use while bathing. If the unit should fall into
water, do not reach for it without rst unplugging the unit
• Unplug the unit before cleaning it, when lling the nebulizer, and after each use. Do not
leave the unit plugged in when not in use
• Always ensure that the air vents are not blocked or obstructed in any way
• Do not use the nebulizer in a very dusty area
• Dropping the nebulizer or causing other heavy impact may damage it
• Do not expose the unit to direct sunlight, humid environments, extreme temperatures,
strong static electricity or electromagnetic waves
• If the nebulizer is not functioning correctly, if it makes unusual sounds, or if during use
you feel pain or discomfort, immediately stop use
• Do not allow the nebulized medicine to come in contact with your eyes
• The mouthpiece or mask should be replaced at least once per year
• Do not use the unit if it has been dropped to the ground, exposed to extreme
temperatures or high humidity, or damaged in any way
• Do not disassemble the nebulizer or attempt to repair it

ENGLISH • 5
INTRODUCTION
Thank you for purchasing a Pediatric Compressor Nebulizer. Please read this
instruction manual completely before using your unit to ensure safe and eective use,
as well as long life of the product.
Compressor nebulizers are used to eciently deliver medication in the form of a mist to the
lungs. Prescribed medicine is placed in the nebulizer and the compressor blasts air through
the liquid medication, converting it to an aerosol mist which is then inhaled. This method
of delivery is benecial in the treatment of asthma and other respiratory ailments, as the
medicine is administered directly to the bronchial passages of the lungs to provide quick and
eective relief. With proper use and care, your new nebulizer will provide you with many years
of safe, eective breathing therapy treatments.
Federal law allows nebulizers to be sold by a physician’s prescription only. These
devices should be used only at the direction of a doctor or healthcare professional and
should be used only as intended and described in this manual.

6 • ENGLISH
PRODUCT FEATURES
Nebulizer
Holder
Built-In
Handle
Built-In
Handle
Power
Cord
Power
Cord
Pediatric nebulizers include a complete
nebulizer kit – child size mask, angled
mouthpiece, nebulizer, 3.3’ air tubing and
ve air lters.
Air Filter
Cover
Air Filter
Cover
Air Tubing
Connector
Port
Air Tubing
Connector
Port
Air Vents
Air Vents
Power
Switch
Power
Switch

ENGLISH • 7
NEBULIZER MASKS/MOUTHPIECE SELECTION
Angled mouthpiece
Tee adapter mouthpiece
Face mask
Pacier mouthpiece
This unit includes one standard angled mouthpiece and one child mask. Additional styles
of mouthpieces and masks are available for order on the enclosed order form, by calling
Customer Care at 866-326-1313 or at your local medical supply dealer.
Proper use of the mouthpiece is
important in helping to ensure eective
and ecient medication delivery.
Angled mouthpiece (included): put the
mouthpiece into the mouth, holding
gently between the teeth and close the
lips around the mouthpiece; hold the
nebulizer in place with a free hand.
Face mask (child size included): select
the appropriate size for the patient
- child or adult - and place over the
patients nose and mouth. Secure in
place with the elastic strap around
the patient’s head and hold the
nebulizer in place with a free hand.
Tee adapter mouthpiece (not
included): put the mouthpiece end
into the mouth, holding gently
between the teeth and close the
lips around the mouthpiece; hold
the nebulizer in place with a free
hand.
Pacier mouthpiece (not
included): put the pacier nipple
into the baby’s mouth; an adult
should hold the nebulizer, being certain
the medicine cup remains upright.

8 • ENGLISH
USING YOUR NEBULIZER
Before the rst use of the nebulizer or if the unit has not been used for an extended period of
time, it should be cleaned thoroughly as described in the Cleaning and Disinfection section
of this manual.
The patient receiving treatment should be seated or slightly reclined in a comfortable position
as treatments take several minutes. Keep in mind that the nebulizer cup must remain vertical
during treatment.
1. Place the compressor unit on a at, stable surface within operating range of the patient
and an electrical outlet. Ensure that the air vents on the side of
the unit housing are not blocked or obstructed.
2. Plug the power cord into the electrical outlet.
3. Attach one end of the air tube to the tube connector at the back
of the unit, Fig. 1. Turn on the power briey to ensure that the
unit is functioning properly and test that air is coming through
the tubing, then turn the power o. Fig. 1
Disposable Nebulizer (INCLUDED)
Depending upon the individual treatment program, the disposable
nebulizer should last approximately 2 weeks. Worn out or
cracked nebulizers can aect the particle size, compromising the
eectiveness of the medication delivery.
Reusable Nebulizer (NOT INCLUDED)
Depending upon the individual treatment program, a reusable
nebulizer should last approximately 6 months of once-daily use.
A reusable nebulizer must still be cleaned after each use and
inspected for cracks or other imperfections before each use.

ENGLISH • 9
USING YOUR NEBULIZER
4. Fill and assemble the
nebulize.
Step 1: Pour the medication
into the nebulizer cup
following the physician
prescribed dosage, Fig 2.
Step 2: Replace the top of the
nebulizer and twist to close;
gently pull the top to ensure
the top is secure, Fig 3.
5. Attach the selected
mouthpiece to the top of the
nebulizer, taking care not to
spill the medication in the
cup, Fig 4.
6. Attach the nebulizer to the
air tubing, Fig 5.
7. Position the mouthpiece at
the patient’s mouth.
Treatment time varies depending on the type of medication and
volume of medication being used. It is important to complete the entire
prescribed dose of medication for maximum eectiveness. Consult with
your prescribing physician or healthcare professional for information on
your specic medication therapy treatment plan.
Fig. 4 Fig. 5
Fig. 3Fig. 2

10 • ENGLISH
USING YOUR NEBULIZER
IF THE VAPOR BEGINS TO SLOW
AND THERE IS STILL MEDICATION
IN THE CUP, GENTLY TAP THE SIDE
OF THE CUP TO SETTLE THE LIQUID
AT THE BOTTOM OF THE CUP FOR
NEBULIZATION.
Mouthpiece position Mask position
Remember that masks should
cover the patients mouth and nose,
mouthpieces should be in the
mouth with the lips closed around
the end of the mouthpiece.
8. Turn the Power switch on to begin
treatment. Inhale deeply and exhale
normally. Ensure the nebulizer is
vertical during treatment for proper
nebulization of the medication and
to prevent spilling of the liquid.
9. If the vapor output becomes
sporadic or if the noise the unit
makes changes during use, turn
o the unit and consult the
TROUBLESHOOTING section of this
instruction manual.
10. When the treatment is complete,
turn the unit o and unplug it from
the outlet.
11. Disassemble, clean and store
your nebulizer. Be certain that the
nebulizer pieces are completely dry
before storage.

ENGLISH • 11
USING YOUR NEBULIZER
Compressor nebulizers contain a powerful compressor motor which may heat up the
unit if it is left running too long or if the vents are blocked while operating. The unit
will automatically shut o if the temperature of the unit becomes too high.
If this occurs, follow these steps:
1. Turn the power o and disconnect the power cord.
2. Wait at least 30 minutes until the unit has cooled down completely.
3. Turn the unit back on.
4. If the unit heats up too quickly or emits any unusual noises beyond the normal sound of
the compressor motor, contact Customer Care at 866-326-1313.

12 • ENGLISH
The nebulizer and mouthpiece should be thoroughly cleaned after each use and disinfected
each day if on a regular therapy program unless a dierent cleaning regimen is specied by
your healthcare professional; additional hygienic precautions may be required for use with
high-risk patients.
The masks, mouthpieces, tubing and nebulizer are not dishwasher safe. The below instructions
are for manual cleaning and disinfecting of your nebulizer and mask pieces.
Cleaning:
1. Ensure that the power cord is
disconnected from the power outlet.
2. Disconnect the air tube, nebulizer and
mouthpiece. Do not wash or submerge
the air tubing.
3. Disassemble the nebulizer by twisting
the top and bottom and separating the
pieces, Fig 1-2.
4. The nebulizer and mouthpiece or mask
should be washed in warm water,
using a mild detergent. Do not wash or
submerge the air tubing; if necessary,
wipe the outside of the tubing with a damp cloth taking care not to get water inside the
tubing.
5. Allow the nebulizer and mouthpiece to air dry completely, or dry them thoroughly with
a clean, soft towel. Do not store any parts wet as remaining moisture creates a risk of
bacterial growth.
6. Occasionally wipe the body of the unit with a damp cloth and allow it to air dry. Only
clean the body of the unit when it is unplugged and completely cool.
Never use soap pads or powder cleanser to clean any part of the unit. Never use any cleaning
products that may be harmful when swallowed or inhaled.
CLEANING AND DISINFECTION
Fig. 2Fig. 1

ENGLISH • 13
CLEANING AND DISINFECTION
Disinfection:
The nebulizer should be disinfected once daily, after the last use of the day, unless a dierent
disinfection regimen is specied by a healthcare professional. Disinfect the nebulizer as
follows:
1. Ensure that the power cord is disconnected from the power outlet.
2. Disconnect the air tube, nebulizer and mouthpiece. Do not wash or submerge the
tubing.
3. Disassemble the nebulizer by twisting the top and bottom and separating the pieces.
4. Rinse the nebulizer and mouthpiece or mask with hot water. Mix one part white vinegar
with three parts distilled water, in a quantity sucient to fully immerse the nebulizer and
mouthpiece or mask.
5. Submerge the nebulizer and mouthpiece or mask in the water and vinegar mixture for
thirty minutes.
6. Rinse the nebulizer and mouthpiece or mask thoroughly with water.
7. Allow the nebulizer and mouthpiece to air dry completely, or dry them thoroughly with
a clean, soft towel. Do not store any parts wet as remaining moisture creates a risk of
bacterial growth.
NOTE: Condensation may collect in the air tubing. If this occurs, the moisture must be
completely removed, following these steps:
1. Disconnect the air tubing from the nebulizer, but keep it plugged into the air tubing
connector port on the unit.
2. Turn the power on and allow the compressor to run until all moisture has dried.
Nebulizers and tubing should be replaced as needed depending on the frequency of therapy
treatments or when damaged, cracked or holes appear in the tubing. To order replacement
parts, use the included order form, call Customer Care at 866-326-1313 or contact your local
dealer.

14 • ENGLISH
CHANGING THE FILTER
• Never operate the unit without a lter.
• Replace the lter after every ten uses, or sooner if it turns gray, appears contaminated is
blocked or becomes wet.
• Use only Veridian Healthcare lters; additional lters are available for order using the
included order form, calling Customer Care at 866-326-1313 or contacting your local
dealer.
• Do not substitute cotton or any other materials for the lters.
• Do not attempt to clean the lters or reuse a lter.
To replace the lter, follow these steps:
1. Change the lter only when the unit is o,
unplugged and completely cool.
2. Gently pull the lter cap out of the
compressor.
3. Remove the lter from the lter cap.
4. Place a new lter in the lter cap.
5. Return the lter cap to its place, ensuring
that it is rmly in place.

ENGLISH • 15
TROUBLESHOOTING
The nebulizer produces no aerosol, or a small amount of aerosol.
• There may be too much or too little medicine in the nebulizer. The amount of liquid
should be between 1.0 ml and 6.0 ml.
• The hose connectors may be blocked and need to be cleaned
• The nebulizer may not be held upright. Always hold the nebulizer vertically during use.
• The medication may not be suitable for nebulization. Check with your doctor.
• The lter may be in need of replacement
• The air hose may be kinked
Some medicine remains in the nebulizer at the end of my treatment.
This is normal. Once you hear a signicant dierence in the sound of the nebulizer, you may
stop inhaling.
What medications can be used in the nebulizer?
Only medications prescribed by your doctor should be used in the nebulizer.
Should any special steps be taken for pediatric users?
Children should never use the nebulizer without adult supervision. For use with babies
and children, the mask should cover the mouth and nose completely to ensure eective
inhalation.
Treatments with the mask takes longer than treatments with the mouthpiece.
This is normal. Less medication is delivered per breath with the mask because the aerosol is
combined with ambient air through the holes in the mask.
Can I share the nebulizer with another person?
No. Each user must have their own nebulizer to ensure proper hygiene. The compressor unit
may be shared but the nebulizer mouthpiece, and mask pieces may not.
Why must the nebulizer be replaced yearly?
Regular wear on the nebulizer can aect the composition and delivery of the medicine.
Regular replacement of the nebulizer is also hygienic.

16 • ENGLISH
WARNING AND CAUTIONS
PRODUCT SPECIFICATIONS
Nebulizer therapy is intended for use with physician prescribed medication only.
Over the counter medication, essential/aromatherapy oils or alcohol are not suitable for
inhaling using a nebulizer. Highly viscous substances may damage the unit and make it
unsuitable for use. Never use nebulizer therapy for anything other than physician prescribed
respiratory care therapy.
Use common sense electrical caution when operating the device. This nebulizer runs on
AC power. Always turn o the power and unplug the device before cleaning or if the unit is
not working properly.
Name & Model Number Puppy Pediatric Compressor Nebulizer, 11-515
Piggy Pediatric Compressor Nebulizer, 11-516
Medication Capacity Maximum 5 ml to 6 ml
Particle Size 0.5 to 5 µm
Sound Pressure Maximum 60 dBA at 1 meter
Average Nebulization Rate Approximately 0.25-0.35 ml / minute
Compressor Pressure Range 35 to 45 Psi
Operating Pressure Range 10 to 13 Psi
Liter Flow Range 6 ~ 9 lpm
Operating Conditions Temperature: 50ºF – 104ºF (10ºC – 40ºC); Humidity: 30% – 75%
Storage Environment Temperature: -13ºF – 158ºF (-25ºC – 70ºC); Humidity: 13% – 90%
Unit Dimensions (W x H x D) 11-515 (Puppy): 7-1/4”x 9-1/4”x 6-1/2”
11-516 (Piggy): 8-1/2”x 10” x 8-3/8”
Unit Weight 11-515 (Puppy) 3 lbs. 3.5 oz.
11-516 (Piggy) - 3 lbs. 1 oz.
Power Connection 120 V ~, 60 Hz; 68 VA
Included Accessories Angled mouthpiece, child mask, nebulizer,
air tubing, ve air lters
Specications are subject to change without notice.

ENGLISH • 17
3-YEAR LIMITED WARRANTY
The warrantor guarantees that its Compressor Nebulizer will be free from manufacturing
defects under normal use for a period of three-years. This warranty covers only normal use,
and does not apply to use in any clinical or commercial applications. This warranty does not
cover batteries or other power sources that may be provided with or used with the nebulizer
unit. This warranty is voided if the nebulizer unit is misused or abused in any manner.
If the nebulizer fails to operate within the warranty period while under the original purchaser
owns the product, return it postage prepaid to: Attn: Repair Department, 1175 Lakeside
Drive, Gurnee, IL, 60031; include your name, address, phone number and brief description
of the problem with your nebulizer. Take care to carefully package the product to avoid
any damage that may occur while in transit; shipping insurance with returned receipt is
recommended. The warrantor will repair or replace the defective unit at our option. Repair
or replacement of the defective unit, at the warrantor’s option, is the sole remedy under this
warranty.
ANY IMPLIED WARRANTIES WHICH THE PURCHASER MAY HAVE ARE LIMITED IN
DURATION TO THE TIME THAT THE ORIGINAL CONSUMER PURCHASER OWNS THE
PRODUCT. Some states do not allow limitations or how long an implied warranty lasts, so the
above limitation may not apply to you.
This warranty constitutes the warrantor’s only reasonability and obligation to repair and/
or replace materials or components, or refund the purchase price. The warrantor will not
be responsible for any indirect, incidental, special consequential or punitive damages or
other loss, including, but not limited to damage to or loss of other property or liable to the
purchaser for any amount in excess of the cost of repair and/or replacement of the unit, or the
purchaser for any amount in excess of the cost of repair and/or replacement of the unit, or the
purchase price of the unit. Some states do not allow the exclusive or limitation of incidental or
consequential damages, so the above limitation or exclusion may not apply to you.
This warranty gives you specic legal rights, and you may also have other rights which vary
from state to state.


MANUAL DE INSTRUCCIONES • ESPAÑOL
Nebulizador Pediátricos
Modelo 11-515 Puppy
Modelo 11-516 Piggy
LEA ATENTAMENTE ESTE MANUAL DE
INSTRUCCIONES ANTES DEL USO
3-YEAR 2-YEAR
3-YEAR
2-YEAR
1-YEAR
1-YEAR

20 • ESPAÑOL
Cuerpo pediátrico
del compresor
del carácter
Nebulizador
Desechable
Tubo de aire
Máscara Pediátrico
Pipeta
(5) Filtros de aire
Manual Detallado
Preguntas con
Frecuencia Hechas
Forma de Orden
¡Alto!
Asegúrese de tener la totalidad de los siguientes
componentes de siguiente antes de usar su nebulizador
SI FALTA ALGUNO DE LOS ELEMENTOS, INCLUIDOS LOS INSERTOS
O LOS MANUALES DE INSTRUCCIONES, NO HAGA LA DEVOLUCIÓN
EN EL LUGAR DONDE EFECTUÓ LA COMPRA. COMUNÍQUESE CON
ATENCIÓN AL CLIENTE AL TELÉFONO 866-326-1313.
This manual suits for next models
1
Table of contents
Languages:
Other Veridian Healthcare Respiratory Product manuals