Solution 2 ensures correct
threshold monitoring of the mat's
resistance. Use Solution 2 only if
Solution 1 does not provide
satisfactory results. However, the
advantage of solution 1 is that the
mat and the instrument ground
connections are monitored which
is more meaningful than the mat
resistance.
5. Connect the system to earth
facility using a cord inserted into
the 4mm plug socket
6. Plug the DC power supply plug
(224700) into the power jack in the
rear of the unit. Route the wire
from the supply to a nearby mains
socket and plug the power supply
into the socket.
Unit Operation
Install the operator wristband and
ground cord (225200 and 225202).
The right-angled plug inserts into the
socket on the yellow remote box. The
wrist strap is monitored as soon as
the cord is connected to the remote
box socket.
Note: The remote box is fitted with
three sockets, a jack stereo socket
and two 4mm plug socket. The latter
are for guest use, and are not
monitored.
The front panel
The LED located in the centre of the
unit indicates the status of the
workstation. On the far right of the
unit is an orange LED indicating that
the unit is powered.
The workstation status is indicated
via 4 LED and a buzzer. The buzzer
and the two red LED labeled "Low"
and "High" indicate if the wrist strap
resistance, when worn, is lower than
0.75MΩor higher than 35MΩ.
These LED should not come on
when the resistance of the wrist strap
is within these limits. Assuming
Solution 1 has been chosen, the LED
labeled "Mat" will turn red when the
resistance to ground of the work
surface is higher than 100MΩ.
Limited Warranty
Vermason expressly warrants that for a period
of one (1) year from the date of purchase,
Vermason
Single Continuous Monitor For Wrist
Strap and Benchmats
will be free of defects in
material (parts) and workmanship (labour).
Within the warranty period, a unit will be
tested, repaired or replaced at Vermason’s
option, free of charge. Call Customer Service
at
0044 (0) 1462 672005
for a Return Material
Authorisation (RMA) and for proper shipping
instructions and address. Any unit under
warranty should be shipped prepaid to the
Vermason factory. You should include a copy
of your original packing slip, invoice, or other
proof of purchase date. Warranty repairs will
take approximately two weeks.
If your unit is out of warranty, Vermason will
quote repair charges necessary to bring your
unit to factory standards. Call Customer
Service at
0044 (0) 1462 672005
for a Return
Material Authorisation (RMA) and proper
shipping instructions and address.
Warranty Exclusions
THE FOREGOING EXPRESS WARRANTY IS
MADE IN LIEU OF ALL OTHER PRODUCT
WARRANTIES, EXPRESSED AND IMPLIED,
INCLUDING MERCHANTABILITY AND
FITNESS FOR A PARTICULAR PURPOSE
WHICH ARE SPECIFICALLY DISCLAIMED.
The express warranty will not apply to defects
or damage due to accidents, neglect, misuse,
alterations, operator error, or failure to properly
maintain, clean or repair products.
Limit of liability
In no event will Vermason or any seller be
responsible or liable for any injury, loss or
damage, direct or consequential, arising out of
the use of or the inability to use the product.
Before using, users shall determine the
suitability of the product for their intended use,
and users assume all risk and liability
whatsoever in connection therewith.
Should the mat resistance be below
this limit, and the worn wrist strap
resistance be between 0.75 and
35MΩlimits, the green "OK" LED
should turn on and the buzzer will
remain silent.
Parking a station
When an operator has to leave the
work area, his wrist strap monitor
should be parked by unplugging his
wrist strap ground cord from the
remote box. In this situation the
green "OK" LED will not illuminate
and the buzzer will remain silent.
Certification
Depending upon the requirements of
a facility, periodic verification of the
monitoring unit may be performed
using a calibrated decade box
Vermason product code 223000.
The calibration procedure is
described in the Instructions for Use
of this decade box.
Specifications
Test voltage: Max 5 Volts DC
Test current: Max 5µA
Resistance settings
Lower limit: 0.7-0.8MΩ
adjustable from 0.5
to 2MΩ
Upper limit: 34-36MΩadjustable
from 25 to 50MΩ
Accuracy, mat test:
± 10% (both mat connectors wired to
the monitor, wiring solution 1).
Calibration, mat test:
90-100MΩ(fully adjustable from 25
to 110MΩ)
Station status indication via 4 LED
+power-on LED
Station parking facility (unplug wrist
strap from remote box)