Getting Started
Once you have unpacked & checked your machine, you will
need to set it up in the following order:
1. Fill your Vaporiser with Isoflurane via the correct Key
Filling Device insuring you do not over fill the vaporiser.
2. Connect the Oxygen hose to the Regulator on the
Oxygen Bottle by means of screwing on the Hand
Wheel.
3. Turn on the Oxygen Regulator & check you have
pressure to your machine by turning the knob on the
flow meter.
4. Connect the Rodent Circuit –Fresh Gas to the Vaporiser
outlet.
5. Connect the Scavenger tube to either a Waste Gas
Scavenge Filter, or to an Active Scavenge Device.
Fresh Gas
Scavenge Tube to
waste gas
Leak Testing Your Machine
1. Turn on your oxygen supply & check you have pressure in the gauge.
2. Check that the flowmeter is off & the bobbin is reading on zero.
3. Occlude the patient end of the circuit with your finger.
4. Turn on the flowmeter to around 0.3-0.4.
5. The Bobbin in the flowmeter should ever so slightly be dropping.
6. If the Bobbin is floating at a constant level, you may have a leak.
7. Once completed, turn off the flowmeter.
Some common leak sites are:
Holes in hoses
Flow meter not turned off –Do Not Over Tighten
Hose connections
Leaks can easily be located by using soapy water which will produce bubbles at the leak site.