VistaLab ali-Q 2000-0005 User manual

Operating Manual
ali-Q™: Aliquoting Pipet Controller
Operating Manual
or click here
or scan QR code

Table of Contents
Safety Instructions........................................................................................................................................... 3
Getting Started ................................................................................................................................................ 4
Component Diagram + Materials................................................................................................................. 4
Package Contents......................................................................................................................................... 5
Terms and Definitions .................................................................................................................................. 5
Operation and Use .......................................................................................................................................... 6
Shipping Mode ............................................................................................................................................. 6
Sleep Mode .................................................................................................................................................. 6
Pipetting speed control................................................................................................................................ 6
Wall Mount .................................................................................................................................................. 6
Conventional Pipetting Mode........................... ...........................................................................................7
Aliquot Mode ............................................................................................................................................... 8
Intelligent Measuring System ...................................................................................................................... 9
LED Indicators .............................................................................................................................................. 9
Technical Specifications ................................................................................................................................ 10
Troubleshooting ............................................................................................................................................ 11
Maintenance.................................................................................................................................................. 12
Cleaning / Decontaminating ...................................................................................................................... 12
Autoclaving ................................................................................................................................................ 12
Replacing Membrane Filter........................................................................................................................ 12
Replacing Nozzle Insert .............................................................................................................................. 12
Recharging Battery .................................................................................................................................... 13
Replacing Battery ....................................................................................................................................... 13
Calibration .................................................................................................................................................... 14
Verification................................................................................................................................................. 14
In-lab Adjustment ...................................................................................................................................... 14
Warranty........................................................................................................................................................ 15
Send-in Service ........................................................................................................................................... 15
Ordering Information.................................................................................................................................... 16
Transport, Storage and Disposal................................................................................................................... 17
Declaration of Conformity ............................................................................................................................ 18

Page | 3
Safety Instructions
Read this operating manual completely before using the device for the first time. Also, observe the instructions for use ofthe
accessories. This instrument may sometimes be used with hazardous materials, operations, and equipment. Itisbeyond the scope
of this manual toaddress all the potential safety risks associated with its use in such applications. Itisthe responsibility ofthe
user ofthis instrument toconsult and establish appropriate safety and health practices and determine the applicability of
regulatory limitations prior to use.
ali-Q is an aliquoting pipet controller designed for aspirating and dispensing aqueous solutions with plastic orglass serological
pipets in 1to 100 mL volumes. It is intended for laboratory use. Any use of this instrument in amedical orIVD setting isthe sole
responsibility of the user.
! Please read the following carefully!
1. Every user must read and understand this operating manual before operation.
2. Follow general instructions for hazard prevention and safety instructions; e.g., wear protective clothing, eye protection and
gloves.
3. Observe all specifications provided byreagent manufacturers.
4. Never use the instrument in an atmosphere with a danger ofexplosion. Highly flammable liquids must notbe pipetted.
5. Only use the instrument for pipetting liquids, with strict regard to thedefined limitations ofuse. If in doubt, contact VistaLab.
6. Always use the instrument in such a way that neither the user nor any other person is endangered. Avoid splashes. Only use
suitable vessels.
7. Never use force onthe instrument.
8. Use only VistaLab accessories and spare parts. See Ordering Information on pg 18. Do not attempt tomakeanytechnical
alterations. Do not dismantle theinstrument any further than is described in the operating manual!
9. Always check the instrument for visible damage before use. Ifthere is a sign ofa potential malfunction, immediately stop
pipetting. Consult the 'Troubleshooting' section of this manual or contact VistaLab if needed.
10. Only the original Power supply can be used for recharging ofthe Lithium Ion battery. Do not replace with non-rechargeable
batteries or rechargeable batteries of other manufacturers.
11. The Power supply must be protected against moisture and must beused only for this instrument.
12. Only authorized service personnel may service the instrument. Unauthorized repair or service may void theunits’ warranty
13. DO NOT AUTOCLAVE ali-Q.Only approved parts and accessories may be removed from unitfor autoclaving. See pg. 12,
Cleaning / Decontaminating.
14. WARNING! Improper use ofthe instrument orthe battery (short-circuit, mechanical damage, overheating etc.) may cause
the explosion of the battery.
Operating Conditions:
oOperating Temp: 15°C to 40°C (59°F to 104°F)
oStorage Temp: -20°C to + 50°C (-4°F to +122°F)
oAltitude up to 2000 m
oRelative Humidity: 10% – 95%, non-condensing
oAtmospheric Pressure: 70 – 106 kPa (525 – 795 mmHg)
oVapor Pressure up to max. 500 mbar. Aspirate slowly above 300 mbar, to prevent liquid boiling
oDensity up to 9.0 g/cm3
ali-Q Body Chemical Compatibility for cleaning and decontaminating:
Water, diluted ethanol or isopropanol. For more information on chemical compatibility, see the support area of our website,
www.vistalab.com/support
Safety Compliance:
See Declaration of Conformity at end of manual
EN 61010-1 General Safety Requirements CSA C22.2, No. 61010.1
EN 61326 EMC Requirements UL 31111-1
<- Back to T.O.C.

Page | 4
Getting Started
Description Material
1 Body/ Main Housing Nylon
2 Aliquot volume-set dial Nylon
3 Charging port, Micro USB N/A
4 Aliquot button Nylon
5 Aspirate button Nylon
6 Dispense button Nylon
7 Battery, Lithium ion Lithium Polymer
8 Battery compartment cover PC/ ABS
9 Nozzle Cover Nylon
10 Nozzle Insert Silicone
11 Membrane Filter, 0.45 µm Polypropylene
12 LED Indicator N/A
<- Back to T.O.C.
Materials Chart

Page | 5
Package Contents:
ali-Q pipet controller
Power supply (w/ universal outlet adapters)
Micro USB charging cord
Wobble-notTM serological pipet samples (1 per size:5mL, 10 mL, 25 mL, 50 mL)
1Wall Mount
1Replacement membrane filter
Quick Reference Guide + Safety Sheet
Certificate of Calibration
Terms and Definitions:
Aliquot – (noun) A portion of a total amount of a solution or suspension; (verb) To separate a volume of
solution or suspension into aliquots.
Aspirate - to fill with liquid
Dispense - to eject liquid
Serological Pipet (pipet) - disposable glass or plastic tube with volume graduations used to aspirate and
dispense liquid. Sometimes spelled ‘pipette’.
Accuracy - How close the measured volume is to the target volume
Accuracy (%)
=( − )
100
Precision (CV%) - How close measurements (aliquots) are toeach other; independent of the target volume
Performance - Another way of referring to the unit’s accuracy and precision.
<- Back to T.O.C.
Coefficient of Variation (%)
=
Standard Deviation
Mean
100

Page | 6
Operation and Use
Shipping Mode:
ali-Q isshipped from the factory in Shipping mode for safety and to preserve battery life. When you first
take your ali-Q out of the box, you must wake it out of shipping mode bypressing the aliquot button with
the unit inthe upright position.
To put the ali-Qback into Shipping mode,turn the unit completely upside down and press and hold the
aliquot button. Shipping mode should only beused when shipping or transporting ali-Q.
Sleep Mode:
ali-Q automatically goes into Sleep mode to preserve battery life after being motionless for 10 seconds.
The LED will blink RED whenentering sleep mode. Once the ali-Q is picked up or moved, it will
automatically wake itself up, andLED will blink GREEN
Pipetting speed control:In Conventional Pipetting Mode (see next page), the speed that the ali-Q
aspirates and dispenses liquid is controlled by the touch-sensitive aspirate and dispense buttons. The
amount of force used to push on the buttons determines the speed that the liquid enters or exits the
pipet.
↑ BUTTON FORCE = ↑ SPEED
Wall Mount:ali-Q may be stored in the wall mount to avoid damage from accidental drops andchemical
spills (and save precious bench space!). ali-Q may also be placed in wall mount with sero pipets up to
50mL installed. Wall mount will firmlyadhere to any smooth, clean, non-painted surface, such asthe
inside of a hood, using the 3M Command strips included.Clean the surface with alcoholbefore applying
the wall mount.Remove the stickers on the adhesive backs and immediately press onto surface and keep
applying moderate pressure for about 30 seconds. DO NOT place ali-Q into wall mount for about 30
minutes,to allowadhesive to set.
<- Back to T.O.C.

Page | 7
Conventional Pipetting Mode:
1 Attach serological pipet - Holdthe pipet as near to its
upper end as possible, andcarefully insert itinto the
nozzle until itfits tightly. NOTE: ali-Q iscompatible with all
brands and sizes ofpipets!
WARNING!
Using excessive force when inserting pipets risksbreakage
and injury. If you are having difficulty inserting pipets fully,
try using VistaLab Wobble-notTM low insertion force pipets.
2 Aspirate liquid - Immerse the pipet’s tip into the liquid.
Press andhold the gray aspirate button and fill the pipet
so that the meniscus isabove the volume graduation
desired. Then press the bottom gray dispense button until
bottom of meniscus is at the desired volume graduation.
ATTENTION!
Do not fill the pipet over the max volume! Severe damage
can occur if any liquid enters ali-Q! Decrease pressure on
aspirate button when reaching max volume toslow down
aspiration.
3 Dispense liquid - Carefully position the tip of the full
pipet over the receiving vessel. Press and hold the gray
dispense button until meniscus reaches desired volume
graduation or when pipet is completely empty.
ATTENTION!
NEVER invert pipet orlay onbench when filled with liquid.
Scan QRcode or click to watch
ali-Q video
NOTE:
The motor will activate at seemingly random times when pressing the aspirate or dispense buttons.
Don’t worry. It’s normal!
For more information on ali-Q’s Intelligent Measuring System, see pg. 9.
<- Back to T.O.C.

Page | 8
Aliquot Mode:
1Attach pipet.
2Set the aliquot volume-set dial - Turn the dial so that the arrow points to the volume you wish to
aliquot.
3Aspirate - Immerse the pipet tip into the liquid. Press and hold the aspirate button and fill the pipet.
Be sure to aspirate enough liquid to dispense as many aliquots as you will need.
Ex: If you need to dispense 10 aliquots of 2 mL, you will need to aspirate at least 20 mL of liquid.
4Aliquot - Carefully position the tip of the full pipet over the first receiving vessel. Press and hold the
purple aliquot button and the LED turns solid GREEN. Hold button until green LED turns off and motor
turns on. You will feel andhear the motor. Then,carefully transfer to the next vessel andrepeat for
each aliquot.
NOTE: If you release the aliquot button too soon, i.e. before the full aliquot volume has dispensed, the LED
will blink RED and you will hear an “alert” sound. The aliquot did not dispense fully/properly, and will not
be accurate. Re-do the aliquot.
NOTE: Discard first aliquot for optimal precision.
<- Back to T.O.C.

Page | 9
Intelligent Measuring System:
ali-Q aliquot mode utilizes a state-of-the art measuring system that incorporates information and
feedback from multiple sensors to give real-time feedback and parameter adjustment for different
variables that affect the accuracy and precision when in aliquot mode.
LED Indicators:
When charger is disconnected:
LED Status
OFF Battery charge is 20% or more.
RED blink (slow) Max. 2 hours battery remaining. Charge recommended.
RED blink (fast) Battery completely dead. Unit will not function until battery is charged.
When charger is connected:
LED Status
ORANGE Battery is recharging.
GREEN Battery is fully charged. Remove charger.
NOTE:
-ali-Q can still be used when connected to charger
-If LED does not light up during charging, see Troubleshooting section
<- Back to T.O.C.
-Varying environmental conditions: Temperature, humidity and pressure all have significant effects
on the performance of any liquid handling device... EXCEPT for the ali-Q! ali-Q can sense changes in
ambient conditions and compensate as necessary to ensure accurate aliquoting- whether you're in
the deserts of Arizona, or the mountains of Colorado!
-Varying pipet angle:One ofthe most useful (and ergonomic!) features ofthe ali-Q isitsability to
adjust the dispensed aliquots for varying pipet angles. This is especially useful in a hood, where space
andmaneuverability is limited. Whether you hold the pipette completely vertical or tilted, you can
beconfident that performance will not beaffected.
-Varying brands/sizes serological pipets: Every brand of pipet has different sizes and geometries.
These geometric differences contribute to changes in the physical characteristics of the system. ali-Q
can sense the effect of these differences and will automatically compensate accordingly, to optimize
performance.
NOTE: Discard first aliquot for optimal precision.

Page | 10
Technical Specifications
Accuracy and Precision:
Conventional pipetting mode:
Accuracy and precision (CV%) is determined by the graduations on the serological pipet
used (typically ±2%, see manuf. spec) as well as the user’s ability to consistently
eyeball and control the liquid meniscus.
Aliquot mode:
ali-Q is calibrated to perform within the following performance specifications, under
controlled environmental conditions with Grade 3 Purified water:
With Wobble-notTM serological pipet (any size):
With any other brand of serological pipet (any size):
Dimensions: 7.2’’ x 7.0’’ x 2.2’’ (18.3 cmx 17.8 cm x 5.6 cm)
Weight: 385g
Volume Capacity:Accepts 1to100 mL serological pipets
Pipet Compatibility:Compatible with ANY size and ANY brand serological pipet. Optimal
performance when using VistaLab Wobble-not serological pipets. Not for use with Pasteur pipets.
Mains/ power supply:
Input Voltage = 100-240 VAC
Frequency = 50-60 Hz
Input Current = 190-320 mA
Output Voltage = 5V
Output Current = 2 A
Rechargeable Battery:
Type = Lithium-Polymer
Voltage = 3.7V
Capacity = 1100 mAh
Charging Time = ~ 3 hr
Number of Dispenses = ~2000 dispenses (with a 25mL pipet)
Ambient Conditions:
Ambience = For indoor use only
Ambient Temperature = 15 - 40°C
Relative Humidity = 10% - 95%, non-condensing
Atmospheric pressure = 70 – 106 kPa (525 - 795 mmHg)
Volume Accuracy Precision (CV%)
0.5 mL 2% 2%
5.0 mL 2% 1%
Volume Accuracy Precision (CV%)
5.0 mL 3% 2%
0.5 mL 4% 3%
<- Back to T.O.C.
With Wobble-notTM serological pipet (any size):
With any other brand of serological pipet (any size):
Volume Accuracy Precision (CV%)
0.3mL 3% 3%
3.0 mL 2% 1%
Volume Accuracy Precision (CV%)
3.0 mL 4% 3%
0.3 mL 5% 4%
ali-Q 0.5-5mL ali-Q LS (Low Speed) 0.3-3mL

Page | 11
Troubleshooting
Problem
Possible Cause
Suggested Fix
Slow aspiration
-Membrane filter clogged
-Not enough force on aspirate
button
-Replace membrane filter
-Press aspirate button with more force
to increase aspirate speed
Aspirating air bubbles
-Pipet tip not fully immersed in liquid
while aspirating.
-Pipet not fully inserted into nozzle
insert
-Aspirating too fast for solution
(ex: detergents)
-Make sure pipet tip is fully immersed
in liquid throughout aspiration
-Insert pipet more
-Lightly press the aspirate button to
slow aspirate speed
Pipet drips liquid
-Pipet not fully inserted into nozzle
insert
- Crack in serological pipet
-Membrane filter clogged
-Membrane filter not properly
positioned or upside down
-Nozzle insert worn or damaged
-Valve not shutting properly
-Try Wobble-not low insertion force,
less-drip pipets.
-Get a new pipet
-Replace membrane filter
-Insert filter properly
-Replace nozzle insert
-Send ali-Q in for repair service
Pipet does not stay in ali-Q
-Nozzle insert worn or damaged
-Replace nozzle insert
NOTE: Using Wobble-not pipets will
significantly decrease damage to
the nozzle insert.
Out-of-Calibration
-Using different equipment, different
density liquid, varying environmental
conditions,
-Membrane filter clogged
-Adjust the aliquot volume-set dial up
or down depending on measured
results (p 14) to compensate
-Replace membrane filter
Motor makes alert sound
when dispensing aliquots
-Not holding down aliquot button
long enough
-HOLD aliquot button until GREEN
LED light turns off and motor starts
(you will hear and feel motor)
LED blinks or does not light
up when connected to
charger
-Battery not properly installed
-Wrong Power supply
-Battery / Power supplydamaged
-Instrument defective
-Firmly push in battery connector
-Only use Power supply
Cat. #9020-9005
-Replace battery/ replace Power
supply
-Send ali-Q in for repair service
Battery does not hold charge
-Charger not inserted fully
.
-Battery defective or drained
-Check charger connection. Replace
charging cord
-Replace battery
Instrument doesn’t work
-Dead battery
-Internal software malfunction
-Recharge battery
-Restart unit by unplugging battery for
10 seconds and plug battery back in.
If problem still occurs,contact VistaLab Technical Support at (914) 244- 4068 or [email protected]
<- Back to T.O.C

Page | 12
Maintenance
www.vistalab.com/services
Routine Maintenance: ali-Q will require minimal, if any, maintenance, if used as intended. VistaLab
recommends performing routine maintenance and calibration verification at least once a year. Routine
maintenance includes:
Cleaning and disinfecting outer body,nozzle cover, nozzle insert and filter
Replacing the membrane filter
Replacing nozzle insert (as-needed)
Aliquot mode calibration verification
WARNING!
ALWAYS use protective gloves and eye protection. Avoid therisk of contamination!
Cleaning / Decontaminating:
-Chemical - Occasionally wipe outer housing of the ali-Q and the nozzle cover with a damp cloth of
soapy water or diluted alcohol solutions (e.g.70% isopropanol).
-UV Sanitizing - The complete ali-Q unit (outer housing) can withstand typical output (254 nm
wavelength) from UV sterilization lamps.
NOTE: The effects of the UV exposure will likely cause discoloration over time!
Autoclaving :
NEVER AUTOCLAVE ALI-Q!! Only the nozzle cover, nozzle insert and membrane filter can
be detached from the unit and separately autoclaved at:
o121°C (250°F); 30 psi (2 bar absolute) with a holding time of at least 15
minutes.
Replacing the Membrane Filter:
1Unscrew the purple nozzle cover in the counterclockwise direction.
2Remove the membrane filter.
3Place new membrane filter with the thicker end facing downwards into the
silicone nozzle insert and thinner end into ali-Q.
4Screw the nozzle cover assembly back onto the ali-Q.
Replacing the Nozzle Insert:
1Unscrew the purple nozzle cover in the counterclockwise direction.
2Remove the membrane filter and silicone nozzle insert from nozzle cover.
3Insert new silicone nozzle insert into nozzle cover. Place membrane filter with the
thicker end inserted into the silicone nozzle insert and thinner end into ali-Q.
4Screw the nozzle cover assembly back into the ali-Q.
<- Back to T.O.C.

Page | 13
<- Back to T.O.C.
Recharging Baery Instrucons:
rechargable or rechargeable bateries from other manufacturers. WARNING:
Charge time approx. 3 hours (from 0% to 100% bry)
life approx. 40 hours (dependent on frequency and intensity of use)
Condis for recharging 0°C to 40°C (32°F to 104°F)
Recharging Procedure:
1ach the proper outlet adapter to the power supply pack and insert into wall outlet.
2h the wide end of the Micro USB charging cord to the power supply pack.
3Plug the Micro USB end of the charging cord into the charging port of the ali-Q (under the aliquot
volume dial). ali-Q can sll be used when plugged into charger.
Replacing the Baery/ Removing Battery (Hard reset):
1 Remove battery cover: On the bo�om of the ali-Q,
A) Remove screw, press latch and slide black cover forward
B) Remove the white sticker on the battery cover to expose battery cover latch hole. Insert flat, narrow
tool into latch hole and press on latch while pushing cover forward.
2 Remove baery: Grip the baery in one hand and the white connector in the other hand a nd gently pull
apart. Do not pull by the leads!
3 Replace baery: Plug (new) baery into white connector unl it snaps-in. Insert baery into compartment.
Slide battery cover back into place.
AB

Page | 14
Calibration
ali-Q’s aliquot mode volume delivery is factory calibrated to aliquot the volume set on the aliquot volume-
set dial to the factory specs (see Technical Specifications, next page) in a state-of-the-art ISO 17025
accredited calibration lab under the following conditions:
-Temperature: 20°C - 22°C (68°F - 71.6°F)
-Relative humidity: 45% - 75%
- Using Wobble-not 25mL serological pipet (Cat. #4090-0025)
- Using ISO 3696 Grade 3 purified water
NOTE: Using ali-Qwith different liquids more/ less dense than water, under varying environmental
conditions will slightly affect the accuracy of the dispensed aliquots. However, precision is always
maintained with the ali-Q thanks to its Intelligent Measuring System.
Calibration Verification:
VistaLab recommends sending ali-Q to VistaLab forcalibration,or performing verification,at least once
ayear,or whenever conditions and equipment change.
In-lab Verification /Adjustment:
Users can verify the accuracy of the dispensed aliquots in four ways:
1Gravimetric - Usea digital scale andwater, determine the weight (in grams) of the dispensed
aliquot. See the Performance Verification document for gravimetric procedure at
www.vistalab.com/support
2Colorimetric - Use a spectrophotometer to measure the absorbance value of diluted dye ina
solvent ofcontrolled volume.Absorbance value then converted into a volume (µL).
3Visual - Use an approved graduated cylinder (Class A, To contain, calibration certified).
4 Visual (least reliable) - Use the volume graduations on the serological pipet as a reference of
how much liquid is being dispensed.
NOTE: Most serological pipets have ±2% accuracy (at full scale) of their volume graduations, so
this is the least accurate way of verifying.
To compensate for errors in dispensed aliquots, adjust the aliquot volume-set dial up ordown to
compensate for differences.
<- Back to T.O.C.

Page | 15
Warranty
VistaLab guarantees the functionality of ali-Q for 1 year from the Date of Manufacture (original proof of
purchase required).
VistaLab shall not be liable for the consequences of improper handling, use, servicing, operating or
unauthorized repairs of the instrument or the consequences of normal wear and tear especially of wearing
parts such as pistons, seals, valves and the breakage of glass as well as the failure to follow the
instructions of the operating manual. We are not liable for damage resulting from any actions not
described in the operating manual or if non-original spare parts or components have been used.
For warranty claims, please contact Technical Support to receive Return Authorization. Unit will be
inspected for defects and repaired/replaced as necessary at the sole discretion of VistaLab. Customer will
be responsible for service charges for non-warranty damage.
To validate warranty, register your ali-Q at www.vistalab.com/register-your-pipette
Send-in Service:
To send your ali-Q sent in for routine calibration and maintenance or repairs, go to
www.vistalab.com/service to create a service order and pre-pay using Purchase Order or Credit Card.
Follow instructions on website and email confirmation to send-in for service. Be sure to decontaminate
the unit before packaging. Put ali-Q into Shipping mode for safety by holding it completely upside-down
and pressing purple aliquot button (red light will blink 2X).
<- Back to T.O.C.

Page | 16
Ordering Information
Catalog Number
Description
2000-0005
ali-Q aliquoting pipet controller
Including power supply, wall mount,1 spare 0.45 µm membrane filter (non-
sterile), Wobble-not pipet samples (1 each; 5mL, 10 mL, 25 mL, 50 mL)
Accessories and Parts:
Catalog Number Description Quantity
9020-4001 Replacement Rechargeable Lithium ion Battery 1/pack
9020-4002 Replacement Battery Cover 1/pack
9020-4005 Replacement Membrane Filter, 0.45 µm, non-sterile 5/pack
9020-4008 Replacement Silicone Nozzle Insert 2/pack
9020-4010 Replacement Nozzle Cover 1/pack
9020-9005 Replacement Power Supply w/Universal outlet adapters 1/pack
9020-9008 Wall Mount 1/pack
Wobble- not Serological Pipets:
Catalog Number Description Quantity
4090-0005 Wobble-not serological pipets, 5 mL 200/case, Indiv. wrapped
4090-0010 Wobble-not serological pipets, 10 mL 200/case, Indiv. wrapped
4090-0025 Wobble-not serological pipets, 25 mL 200/case, Indiv. wrapped
4090-0050 Wobble-not serological pipets, 50 mL 100/case, Indiv. wrapped
Distributors:
Go to www.vistalab.com/distributors to find your local distributor.
<- Back to T.O.C.
2000-0003ali-Q LS (Low Speed) aliquoting pipet controller
Including power supply, wall mount,1 spare 0.45 µm membrane filter (non-
sterile), Wobble-not pipet samples (1 each; 5mL, 10 mL, 25 mL, 50 mL)

Page | 17
Transport, Storage and Disposal
Transport:
Before sending the ali-Q in to an authorized service center, you must decontaminate the controller per
instructions on pg 12. Please note the following:
Air Temperature
Relative Humidity
Atmospheric Pressure
General Transport
-20°C – +50°C
10% - 95%
30 kPa – 106 kPa
Air Freight
-5°C – +45°C
10% - 95%
30 kPa – 106 kPa
Storage:
Air Temperature
Relative Humidity
Atmospheric Pressure
General Transport
-20°C – +50°C
10% - 95%
70 kPa – 106 kPa
Air Freight
-5°C – +45°C
10% - 95%
70 kPa – 106 kPa
Disposal:
The adjoining symbol means that storage batteries and electronic devices must be disposed
of separately from household trash (mixed municipal waste) at the end of their service life.
-According to the Directive 2002/96/EC of the European parliament and of the Council on
Waste Electrical Equipment (WEEE) of 27 January 2003, electronic equipment requires
disposal according to the relevant national disposal regulations.
-Batteries contain substances that can have harmful effects on the environment and
human health. Therefore, according to the Directive 2006/66/EC of the European Parliament
and the Council on Waste Batteries of 6 September 2006 batteries require disposal according
to the relevant national disposal regulations. Dispose of batteries only when completely
discharged.
WARNING!
Do not short circuit battery to discharge it!
Risk of explosion and fire due to overheated rechargeable batteries. Do not heat to over 80°C and do
not throw into fire.
VistaLab Technologies, Inc
2 Geneva Rd
Brewster NY, 10509
(888)652-6520
www.vistalab.com
<- Back to T.O.C.

Declaration of Conformity
VistaLab Technologies Inc. declares on its own responsibility that the devices
ali-Q aliquoting pipet controller, Model 2000-0005
comply with:
Green/Ecodesign Regulations
Meet CEC, DoE Level VI, Energy Star, ErP Stage 2
NRCan & GEMS
No Load Power Consumption Less 0.075W
Standards for EU
Safety requirements for electrical equipment for measurement, control and
laboratory use- General Requirements
EN 61010-1:2010
Electrical equipment for measurement, control and laboratory use- EMC
requirements
EN 61326-1:2013
Standards for Canada and USA
Safety requirements for electrical equipment for measurement, control and
laboratory use- General Requirements
CAN/CSA-C22.2
No. 61010-1
Safety requirements for electrical equipment for measurement, control and
laboratory use- General Requirements
UL 61010-1
Operation is subject to the following two conditions: (1) this device may not
cause harmful interference, and (2) this device must accept any interference
received, including interference that may cause undesired operation.
Part 15 of the FCC
Rules Class A
<- Back to T.O.C.
Copyright© 2017 VistaLab Technologies, Inc. USA. All rights reserved, including graphics and images.
Nopart ofthis publication may be reproduced without prior permission of the copyright owner.
VistaLab Technologies and the VistaLab Technologies logo are trademarks of VistaLab Technologies, Inc. USA.
ali-Q is a trademark ofVistaLab Technologies, Inc.
Registered trademarks and protected trademarks are not marked in all cases with ® or™ inthis manual.
U.S Patents are listed on www.vistalab.com/about/quality-management/
Lit # :L080-0060-001 Rev C