vita VACUMAT 6000 MP User manual

VITA VACUMAT®6000 MP
VITA – perfect match.
Operating manual
VITA Farbkommunikation
VITA Farbkommunikation
VITA shade control
VITA shade reproduction
VITA shade communicationVITA shade determination
Date of issue: 09.18


3
VITA VACUMAT®6000 MP
Table of contents
1 Introduction 4
2 Scope of delivery 5
2.1 Control unit 5
2.2 Accessories (can be purchased separately) 5
3 Technical information 6
3.1 General description of the furnace VITA VACUMAT 6000 MP 6
4 Technical data 6
4.1 Dimensions/weights 6
4.1.1 Furnace VITA VACUMAT 6000 MP 6
4.2 Electrical data 6
4.2.1 Furnace 6
4.2.2 Vacuum pump 6
5 Intended use 7
6 Safety information 7
6.1 Pictograms 7
7 Ambient conditions 8
8 Safety functions 8
9 Installation and first use 9
9.1 Installation location 9
9.2 Device connections 9
9.3 Pressurized air connection 10
9.4 Status display 10
9.5 Fuses 10
9.6 Information about the identification plates 10
9.7 Connecting the device to the mains voltage 11
9.8 Switching off the device or stopping its operation 11
10 Cleaning the furnace 12
10.1 Cleaning firing for the firing chamber 12
10.2 Firing chamber insulation 12
11 CE mark 12
12 Fan 12
13 Mains power supply failure 13
14 Warranty and liability 13
14.1 Spare parts 13
14.2 Service 13
15 Alphabetical table of contents 14

4
VITA VACUMAT®6000 MP
1 Introduction
Dear Customer.
Thank you for choosing to purchase a VITA VACUMAT 6000 MP.
This well-designed and proven heating system for ceramic furnaces uses a quartz/
Kanthal firing muffle and newly developed insulating material in the firing chamber,
which ensures many years of uniform firing results with all ceramic materials.
The high-quality temperature control and automatic temperature adjustment
ensures a precision of plus/minus 1°C.
This device has been designed according to the the latest technological
benchmarks and complies with all international safety standards. However,
incorrect usage can impact the health and safety of the user, as well as damage
the device. Please read this operating manual carefully and follow the information
provided.
Reading and understanding this operating manual will help you maintain safety,
reduce expenses as a result of repairs and downtime and increase the reliability
and service life of the device. All of the illustrations and drawings in this operating
manual are intended for general explanatory purposes and are not authoritative for
the detailed construction of the device.
The operating manual must always be kept close to the device. It must be read
and followed by all persons responsible for performing tasks such as operation,
troubleshooting during operation and cleaning and servicing (maintenance,
inspection, repairs), either with or on the device.
We hope that you will find using the VITA VACUMAT 6000 MP to be an enjoyable
and successful experience.
Copyright
These operating instructions must be treated as confidential. They should only be
used by authorized persons. They may only be transmitted to third parties with the
written consent of VITA Zahnfabrik H. Rauter GmbH & Co. KG.
All documentation is protected by copyright.
The transmission and reproduction of documents, even excerpts, and the use
and communication of their content are not permitted without express written
authority.
Violations are an offense and liable to compensation.
We reserve all rights to exercise intellectual property rights.

5
VITA VACUMAT®6000 MP
2 Scope of delivery
Device supplied in a special box with:
• 1 VITA VACUMAT 6000 MP furnace, coated or in stainless steel
• 3 firing sockets and 2 sets of pressure discs
• 1 plug-in status display
• 1 cable for mains power connection
• 1 furnace tweezer
• 1 package of firing trays A + B
• 1 package of firing trays G
• 1 dental equipment DVD
• 1 cable for the control unit
• 1 pressure regulator for attachment to the vacuum pump
• 1 pressure tube, vacuum tube
2.1 Control unit
The VITA VACUMAT 6000 MP can be equipped with the following control units:
• VITA vPad comfort with 10“ TFT-Display, photo viewer,
memory for 500 firing programs for control of 1 VITA VACUMAT
6000 M / VITA VACUMAT 6000 MP and VITA ZYRCOMAT 6000 MS.
• VITA vPad excellence with 10“ TFT-Display, photo viewer,
memory for 1000 firing programs for control of 1 to 4 VITA VACUMAT
6000 M / VITA VACUMAT 6000 MP and VITA ZYRCOMAT 6000 MS.
To operate with 1 VITA VACUMAT 6000 M / 6000 MP or VITA ZYRCOMAT
6000 MS, the VITA vPad comfort can be used. An additional VITA SWITCHBOX
is required for the operation of 2 or more VITA VACUMAT 6000 M / 6000 MP /
ZYRCOMAT 6000 MS with a VITA vPad excellence.
You will need the VITA MultiPump
and the VITA vPad excellence for
operation with a VITA vacuum pump for up to 4 VITA VACUMAT 6000 M or
VITA VACUMAT 6000 MP.
Please read the information in the operating manual of the corresponding
control unit.
2.2 Accessories (can be purchased separately):
• Side panels, set of 2 pc. each
• Vacuum pump: 230/240 V, 50/60 Hz, 115 V, 50/60 Hz
or 100 V/60 Hz.
• FDS (Firing Data System) – firing data administration program for PC
• Magnetic numbers 1-4, set of 4 pc. each

6
VITA VACUMAT®6000 MP
3 Technical information
3.1 General description of the furnace VITA VACUMAT 6000 MP
• High-performance technology – extremely accurate temperature control for
optimum firing results
• Time-saving, user-friendly convenience and a small footprint
• Casing made of coated or stainless steel plate
• Optical operating status display
• Firing object storage tables, extendable
• High-quality material used to insulate firing chamber
• Quartz firing muffle
• Temperature sensor (platinum/rhodium-platinum)
• Automatic temperature calibration before every program startup
• Temperature accuracy +/- 1°C
• Pneumatic pressing mechanism
4 Technical data
4.1 Dimensions / weights
4.1.1 Furnace VITA VACUMAT 6000 MP
• Width: 230 mm
• Depth: 370 mm
• Height: 630 mm
• Housing, weight: lacquered steel 18.7 kg, stainless steel 20.1 kg.
• Effective dimensions
of the firing chamber: diameter: 90 mm, height: 55 mm
• Firing chamber temperature: max. 1200°C
4.2 Electrical data
4.2.1 Furnace
• Electrical connection: 230 V AC, 50 Hz
or 100/110 V AC, 50/60 Hz
• Power consumption: max. 1500 W
4.2.2 Vacuum pump
• Electrical connection: 230 V, 50/60 Hz
• or 100/110 V, 50/60 Hz
• Power consumption: max. 200 W
• Final vacuum: < 960 mbar
• Dimensions: 320 x 110 x 220 mm
• Weight: approx. 6.4 kg

7
VITA VACUMAT®6000 MP
5 Intended use
Basic information on the instrument design
The device is designed according to state-of-the-art technology and recognized
safety standards. However, if it is used inappropriately, hazards for the health and
safety of the user or third parties may arise as well as the risk of damaging the
device and other material property.
Unauthorized modes of operation
The operation of the device with power sources, products, etc., which are subject
to hazardous materials regulations, using equipment modified by the user or
operation that could have any negative impact on the health of the operating
personnel, is not permitted.
Authorized modes of operation
The operation of the device is only permitted if this operating manual has been
completely read and understood and the procedures described in it have been
observed. Any other or additional use, e.g., processing of products other than
those intended, as well as handling of hazardous materials or substances injurious
to health, is considered to be contrary to the recommended use. The manufacturer/
supplier will not be liable for any damage resulting from such unauthorized use.
The risk of such use is borne exclusively by the user.
6 Safety information
6.1 Pictograms
This pictogram warns the user about hazardous voltage.
Before opening the unit, always disconnect the device from the mains
current by unplugging the AC adapter.
This pictogram warns the user about hot surfaces that can cause
burns.
Dispose of electrical and electronic equipment separately, not with
household waste.
The black bar under the "wheeled bin" symbol indicates that the
device was placed on the market after August 13, 2005.
Please note that the device is subject to European Community
Directive 2002/96/EC (WEEE) and the national laws valid in your
country, and it must be disposed of accordingly. Contact your dealer if
you need to dispose of the device.
This pictogram warns of hazardous situations with the risk of personal
injury or damage to the device.
This pictogram points to useful pieces of advice, explanations and
supplements regarding the handling of the equipment.
Hazardous voltage
Hot surface
Separate disposal
Note
Information

8
VITA VACUMAT®6000 MP
7 Ambient conditions
• Use indoors
• Ambient temperature 2°C to 40°C
• Relative humidity: 80% at 31°C
• Max. altitude: 3800 m above sea level (standard elevation zero, NHN).
• Fluctuations in nominal voltage are not greater than plus/minus 10% of the
nominal voltage.
8 Safety functions
The furnace is in operation with the control unit
VITA vPad comfort or
VITA vPad excellence
and has the following safety and monitoring features:
• Temperature sensor monitoring
• Temperature monitoring
• Vacuum monitoring
• Power failure protection
• Lift monitoring
• Press stroke monitoring

9
VITA VACUMAT®6000 MP
9 Installation and first use
9.1 Installation location
• Install the device in a dry, heated room. The distance to the closest wall
should be at least 25 cm (see also Section 7. Ambient conditions).
• When the temperature is below 15°C (e.g., after transport), leave the device
to stand for approx. 30 minutes at room temperature before using it for the
first time.
• Ensure that the surface where the device is installed is heat-resistant.
The radiation and heating of the device is within a non-hazardous range.
Nevertheless, heat-sensitive surfaces of furniture and veneers could become
somewhat discolored over time, due to the constant influence of heat.
• Prevent direct sunlight from coming into contact with the device.
• Do not place any flammable objects in the vicinity of the device.
Do not place the control unit directly in the heat radiation area of the firing
chamber.
• Do not set up the device in such a way that makes it difficult to press the main
switch and pull out the power supply cord.
Before using the device for the first time, read the corresponding operating
manual for the control unit.
9.2 Device connections
Status display
(see sect. 9.4)
Pressure connection
max. 5 bar
Connection for control unit
VITA Switchbox
Connection for control unit
VITA vPad excellence/VITA vPad comfort
Vacuum tube connection
Device fuses (see Section 9.5)
Connection for vacuum pump
(see sect. 4.2.2)
Pressure regulator 3 bar
Mains power supply connection
Main switch

10
VITA VACUMAT®6000 MP
9.3 Pressurized air connection
The vacuum pump is equipped with a pressure regulator.
The input pressure from the laboratory pressurized air network should be a
maximum of 6 bar.
Set the pressure output of the device to a maximum of 5 bar.
See the operating manual regarding the settings for the pressing pressure
VITA vPad excellence /comfort (see Section 19.3)
9.4 Status display
The status display shows the following operating modes:
• Green – device in standby operation
• Blue – program active
• Red – error
Read the corresponding operating manual for the control unit.
9.5 Fuses
On the back of the device (see Section 9.2) there are 2 fuses for the device.
The identification plates show information about the fuses used in the device.
Fuses with other ratings must not be used.
230 V model
T 8 H 250 V
100/110 V model:
T 15 H 250 V
The manufacturer is not liable for accidents to the user occurring when
the device is open!
9.6 Information about the identification plates
This pictogram warns the user about hazardous voltage. Before
opening the unit, always disconnect the device from the mains
current by unplugging the AC adapter.
After the back plate has been removed, a residual voltage of up to
400 V may still be present in components in the area of the power
supply unit on the circuit board, even when the device is switched off.
Do not place any objects in the area of the lift plate.
When the device is switched on, the lift will move down to the lower
position. For setting down firing objects, use the laterally extendable
storage plates.
Hazardous voltage
Residual voltage
Note

11
VITA VACUMAT®6000 MP
Never operate the device in any case without the firing socket installed.
In continuous operation (max. firing temperature, max. firing time), some parts
of the firing chamber may reach high temperatures (above 70°C). Do not reach into
the open firing chamber when the device is switched on. There is a risk of touching
electrically live and hot parts.
9.7 Connecting the device to the mains voltage
Important! Before first use, please read Section 6 Safety Information!
For information about the connections, refer to Section 9.2
• Attach the connection cable to the control unit and the furnace.
• Plug in the status display.
•
Connect the vacuum pump, electrical connection and tubing connection.
• Connect the device to the mains power supply using the mains power cable
supplied.
• Install the pressure tube for the pressing mechanism.
Important!
Avoid electrical multiway socket outlets with extensions,
as there is a risk of fire with overload.
• Switch on the device at the main switch. The lift moves to the lower position.
• Clean or wipe the lift plate and the lift plate gasket (dust particles from the
insulation are deposited during transportation of the device).
• Attach the firing socket to the lift plate.
• To accommodate the investment ring, 2 firing sockets are available, each with
2 pressure discs with different diameters.
Important: For pressing, always insert both pressure discs into the
firing socket.
Important! Never operate the device in any case without the firing
socket installed!
9.8 Switching off the device or stopping its operation
When the device is not in operation, the lift should be moved into the firing
chamber and the device must be switched off at the main switch (see Section 9.2).
Closing the firing chamber protects the insulation and prevents moisture
absorption.
Please read the operating instructions for the control unit used.

12
VITA VACUMAT®6000 MP
10 Cleaning the furnace
Before each cleaning operation, remove the power supply plug!
It is not necessary to clean the inside of the firing chamber.
Cleaning the casing of the firing chamber at regular intervals with a damp cloth
contributes to its operating safety.
In all cleaning work, you must never use any detergents or flammable
liquids.
10.1 Cleaning firing for the firing chamber
Consult the operating manual of the control unit for information about the cleaning
firing.
10.2 Firing chamber insulation
The firing chamber contains ceramic mineral fibers as insulation material
(Index No. 650-017-00-08), which is classified as a CAT 2 carcinogen
(Annex VI, EC 1272/2008).
When working with the firing chamber or exchanging the firing muffle, fiber dust
may be discharged. Exposure to this dust can potentially be carcinogenic on
inhalation, as well as result in irritation of the skin, eyes and respiratory organs.
When exchanging the firing muffle, please proceed as follows:
• Wear long-sleeved protective clothing
• Wear safety goggles as well as protective gloves
• Use a dust vacuum system or wear a FFP 2 respirator.
Once work has been completed, rinse dust from unprotected skin using cold water.
Wash workwear separately from everyday clothing.
11 CE mark
With the CE mark, a legally binding declaration is made that the device meets
the essential requirements of European Community Directive 2006/95/EC (Low-
Voltage Directive) as well as of European Community Directive 2004/108/EC (EMC
Directive).
12 Fans
The device is equipped with two fans. The fans are temperature-controlled
.
Activation, deactivation and speed of the fans are controlled automatically.
The fans prevent excessive heating of the device itself and the pressing
mechanism and contribute to the overall operating safety.
If a fan fails, an error message is shown on the display (for information about this,
refer to the Error Messages section of the control unit’s operating manual).
For safety reasons, the device should not be operated without a fan. The upper
cover of the firing chamber and the openings in the rear cover must not be closed
or blocked.

13
VITA VACUMAT®6000 MP
13 Mains power supply failure
The instrument is equipped with power supply failure protection. This component
prevents a program interruption and any incorrect firing in the event of a brief
failure of the mains power supply. The power failure protection is activated as
soon as the mains power supply fails when a firing program is running.
Mains voltage failure time less than approx. 15 sec.
The program continues to run and is not interrupted. The display is out of
order during this period. Once mains power is supplied again and the program
interruption is over, the running program reappears in the display.
Mains voltage failure time longer than approx. 15 sec.
The program is interrupted and the display is out of order. Once mains power is
supplied again, the display indicates that there was a power failure.
Important! Once mains power is supplied again, the time required
for the control unit to switch back on again is approx. 20 seconds.
14 Warranty and liability
The warranty and liability are based on the terms and conditions stipulated in the
contract.
In the event of software modifications without the knowledge and approval
of VITA Zahnfabrik H. Rauter GmbH & Co. KG, all liability and warranty claims are
invalidated.
14.1 Spare parts
Spare parts must comply with the technical requirements specified by the
manufacturer. This is always ensured when using original VITA spare parts.
14.2 Service
Additional information on this device is available on our homepage:
http://www.vita-zahnfabrik.com
Software updates are available for download under Documents & Media /
Download Center / Product Information / Software Updates.
An option to register is also provided via Service / Update Messenger
so that the latest information on the device is automatically emailed to you.
In case of technical queries regarding the device or regarding repair services
and warranty provisions, contact us at:
Email: instruments-service@vita-zahnfabrik.com
Tel. +49 (0) 7761 / 562-105, -106, -101

14
VITA VACUMAT®6000 MP
P
Pictograms 7
Power failure protection 13
Pressure connection 10
S
Safety functions 8
Safety information 7
Scope of delivery 5
Service 13
Spare parts 13
Startup 9
Status display 9, 10
Switching off the device 11
T
Technical data 6
Technical information 6
U
Unauthorized modes of operation 7
V
Vacuum pump 6
VITA vPad comfort 5
VITA vPad excellence 5
A
Accessories 5
Ambient conditions 8
Authorized modes of operation 7
C
CE mark 12
Cleaning the device 12
Cleaning work 12
Connecting the device to the
mains voltage 11
Copyright 4
D
Device connections 9
Dimensions/weights 6
Disposal 7
E
Electrical data 6
F
Fan 12
Firing chamber insulation 12
Fuses 10
G
General description 6
Guarantee 13
I
Identification plates 10
Installation and first use 9
Installation location 9
Installation plates 10
Intended use 7
L
Liability 10, 13
M
Mains power supply failure 13
Mains voltage failure time 13
15 Alphabetical table of contents


VITA Zahnfabrik H.Rauter GmbH & Co.KG
Spitalgasse 3 · D-79713 Bad Säckingen · Germany
Tel. +49 (0) 7761/ 562-0 · Fax +49 (0)7761/ 562-299
Hotline: Tel. +49 (0) 7761/ 562-222 · Fax +49 (0) 7761/ 562-446
facebook.com/vita.zahnfabrik
1665_1E – 0918 (.1) S – Version (06)
With the unique VITA SYSTEM 3D-MASTER, all natural
tooth shades can be systematically determined and per-
fectly reproduced.
Please note: Our products must be used in accordance with the instructions for use.
We accept no liability for any damage resulting from incorrect handling or usage. The
user is furthermore obliged to check the product before use with regard to its suitability
for the intended area of applications. We cannot accept any liability if the product is
used in conjunction with materials and equipment from other manufacturers that are
not compatible or not authorized for use with our product and this results in damage.
The VITA Modulbox is not necessarily a component of the product. Date of issue of
this information: 09.18
After the publication of this information for use any previous versions become obsolete.
The current version can be found at www.vita-zahnfabrik.com
VITA VACUMAT®6000 MP is CE marked within the meaning of EC Guideline 2006/95/
EC, 2004/108/EC and 2011/65/EC.
Other manuals for VACUMAT 6000 MP
4
Table of contents
Other vita Dental Equipment manuals

vita
vita V60 i-Line User manual

vita
vita Easyshade Compact User manual

vita
vita Easyshade Advance 4.0 User manual

vita
vita VITA Easyshade V User manual

vita
vita Easyshade LITE User manual

vita
vita Easyshade Advance 4.0 User manual

vita
vita VITA SMART.FIRE User manual

vita
vita VITA Easyshade V User manual

vita
vita VITA FDS User manual

vita
vita YZ Series User manual