WELL.D MEDICAL WED-3000V User manual

WED-3000V
Veterinary Ultrasound Scanner
User Manual
SHENZHEN WELL.D MEDICAL ELECTRONICS CO. LTD.

- I -
Preface...........................................................................................................................................................1
Statement.................................................................................................................................................1
Manufacturer's warranty..........................................................................................................................1
Matters needAttention............................................................................................................................2
Safety labels ............................................................................................................................................2
General tips for equipment operation......................................................................................................3
General Safety Message..........................................................................................................................3
1 Summary....................................................................................................................................................6
1.1 Brief Introduction..............................................................................................................................6
1.2 Range of application..........................................................................................................................6
1.3 Appearance........................................................................................................................................7
1.4 Technical Specificatio .......................................................................................................................7
1.5 Electric principle block diagram .......................................................................................................8
1.6 Standard configurations.....................................................................................................................9
1.7 Optional pieces..................................................................................................................................9
2 Installation................................................................................................................................................10
2.1 Operation environmental requirements...........................................................................................10
2.2 Installing and disassembling battery ...............................................................................................10
2.3 Installing and disassembling probes................................................................................................11
2.4 Power Supply ..................................................................................................................................12
2.4.1 Power supply with adapter..............................................................................................12
2.4.2 Battery Operation..............................................................................................................12
2.5 Battery charging..............................................................................................................................13
2.5.1 Charging through main unit.............................................................................................13
2.5.2 Charging through adaptive line.......................................................................................13
2.5.3 Charging through auto-charger......................................................................................14
3 Operating procedures.............................................................................................................................15
3.1 Introduction of keyboard.................................................................................................................15
3.2 Display interface .............................................................................................................................16
3.3 Operating procedures ......................................................................................................................16
4 Image parameter adjustment.................................................................................................................17
4.1 Display mode switching..................................................................................................................17
4.2 Image freezing.................................................................................................................................18
4.3 Gain adjustment...............................................................................................................................18
4.4 Image brightness adjustment...........................................................................................................18
4.5 Image smoothing.............................................................................................................................18
4.6 Gray correction................................................................................................................................18
4.7 Dynamic range adjustment..............................................................................................................19
4.8 Depth adjustment.............................................................................................................................19
4.9 Scanning angel adjustment..............................................................................................................19
4.10 Focus adjustment...........................................................................................................................20
4.11 probe work frequency....................................................................................................................20
4.12 Frame-related adjustment..............................................................................................................21
4.13 M rate adjustment..........................................................................................................................21
4.14 Screen cleaning .............................................................................................................................21
4.15 Restore factory settings.................................................................................................................21

- II -
5 Image processing....................................................................................................................................22
5.1 Overview.........................................................................................................................................22
5.2 Cine loop.........................................................................................................................................22
5.3 Static image storage (SAVE)...........................................................................................................23
5.4 Image export (SVLOAD)................................................................................................................23
5.5 ZOOM.............................................................................................................................................24
5.6 Upside down....................................................................................................................................24
5.7 Turning around................................................................................................................................24
5.8 Pseudo color procedure...................................................................................................................25
5.9 Bar chart..........................................................................................................................................25
5.10 Image cleaning ..............................................................................................................................25
5.11 Report............................................................................................................................................26
6 Explanation...............................................................................................................................................27
6.1 Overview.........................................................................................................................................27
6.2 Type the basic information of the patient and hospital....................................................................27
6.3 Image explanation...........................................................................................................................28
6.4 date and clock adjustment ...............................................................................................................29
6.5 Language switching.........................................................................................................................29
6.6 Body mark.......................................................................................................................................29
7 Measurement...........................................................................................................................................31
7.1 Ordinary measurement....................................................................................................................31
7.1.1 Distance measurement....................................................................................................31
7.1.2 perimeter/area measurement .........................................................................................31
7.1.3 Volume measurement......................................................................................................33
7.1.4 M mode measurement.....................................................................................................33
7.2 Obstetrics measurement ..................................................................................................................33
8 Check and Maintenance ........................................................................................................................41
8.1 Service life.......................................................................................................................................41
8.2 Main unit maintenance....................................................................................................................41
8.3 Probe maintenance ..........................................................................................................................42
8.4 Correct usage of probe ....................................................................................................................43
8.5 Battery information .........................................................................................................................43
9 Simple Trouble Shooting........................................................................................................................45
9.1 Check...............................................................................................................................................45
9.2 Simple trouble shooting...................................................................................................................45
10 Transportation and Storage.................................................................................................................46
10.1 Environment requirements on transportation and storage.............................................................46
10.2 Transportation................................................................................................................................46
10.3 Storage...........................................................................................................................................46
Appendix AAcoustic output reporting table............................................................................................47
Appendix B Obstetrics...............................................................................................................................51

- 1 -
Preface
Statement
This publication, including pictures and illustrations, is property of Shenzhen Well.D Medical
Electronics Co., Ltd. and under protection of international copyright law.
Information in this document is not annotated to change. The manufacture shall not state nor
observe any warranty basing on this point, and definitely give up any implied warranty basing on
any special purpose of selling or making benefit.
Without previous written permission from the producer, this document must not be photocopied,
reproduced or translated into other languages.
We preserve the right of revision on this document without still further notice.
Some pictures in this manual, which are schematic diagrams for indication only, may disaccord
with the real object, and then the real object should be regarded as the final.
is trade marks of Shenzhen Well.D Medical Electronics Co., Ltd. Any abuse of these trade
marks without permission will be sued to assume legal responsibility according to laws.
Manufacturer's warranty
Shenzhen Well.D Medical Electronics Co., Ltd. assumes the responsibility for equipment security,
reliability and performance only under the preconditions that the disassembly, assembly and
maintenance of the equipment are all performed by its assigned professional and the equipment is
used strictly in compliance with the operation manual.
Shenzhen Well.D Medical Electronics Co., Ltd. ensures a guarantee period within a year and a
half since the delivery day and promises there is no problem with the new equipment in material
and technology. Within the guarantee period, Shenzhen Well.D Medical Electronics Co., Ltd. will
maintain the equipment and replace the parts of non-man-made damages free of charge. But will
not repair or replace the equipment surface if it is damaged.
This guarantee is only available for failures occurred when the equipment is operated in
compliance with the operation manual. And the guaranteed equipment can only be used in the
prescribed range given in manual.
This guarantee excludes losses or damages caused by external reasons such as thunder struck,
earthquake, theft, unsuitable use or abuse and refitting the equipment.
Shenzhen Well.D Medical Electronics Co., Ltd. shall not be responsible for damages caused by
other devices or arbitrary connection to other devices.

- 2 -
Shenzhen Well.D Medical Electronics Co., Ltd. shall not be responsible for losses, damages or
injuries caused by delayed service request.
When there is problem with the products, please contact Shenzhen Well.D Medical Electronics
Co., Ltd. and explain the equipment model, serial number, date of purchase and the problem.
Matters need Attention
To ensure operational safety and long-term stable equipment performance, please read this
operation manual closely and understand the equipment functions, operation and maintenance at
all points before operating the equipment, especially contents of "Warning", "Caution" and
"Note".
Maloperation or inobservance of the instructions given by manufacturer or its agents may result
in equipment damage or personal injury.
The following convention works through this manual to lay special emphasis on some
information.
"Warning": Stands for neglect of it will cause severe personal injury, death or realized
property loss.
"Caution": Stands for neglect of it will cause slight personal injury or property damage.
"Note": to remind user of installation, operation or maintenance information. These
information is very significant but with no risk. Any warning against dangers shall not be
contained in NOTE.
Safety labels
Equipment labels explanation:
Note! Refer to accompanying documents
Signal output
USB2.0
IPX7
Protection against dripping water
ClassⅡequipment
Mouse
Probe socket
Packing and transportation labels explanation:

- 3 -
Handle carefully
Temperature limit
Upward
Piling layer limit
Keep dry
Protect against heat
General tips for equipment operation
Note:
The equipment is use only for veterinary.
◆In operation
1. Heat radiation holes are strictly prohibited to be covered.
2. After closedown, do not switch on the equipment within 2 - 3 minutes.
3. On scanning, if any abnormal case is found, stop scanning immediately and shut down
the equipment.
4. The patient is prohibited to touch any non-applied part of the equipment.
5. When operating, do not press the keyboard panel too much, otherwise the equipment
might be damaged.
◆After operation
1. Power off the equipment.
2. Pull out the plug from power supply socket instead of pulling the cable.
3. Clean off the couplant on the probe with soft medical sterilized cotton ball.
4. Put the probe into the special case.
General Safety Message
Safety of the operator and patients and reliability of the equipment are taken into consideration
during designing and producing, the following safety precaution must be implemented:

- 4 -
1. The equipment shall be operated by qualified operating staff or under their instructions.
2. Do not open the equipment and change the parameters without permission. If necessary,
please turn to for Shenzhen Well.D Medical Electronics Co., Ltd. or its authorized agent
for service.
3. The equipment has already been regulated into its optimal performance. Do not adjust
any preset control or switch unless operate as per instructions in the manual.
4. If there is equipment failure, please shut down the equipment at once and contact for
Shenzhen Well.D Medical Electronics Co., Ltd. or their authorized agent.
5. If it is needed to connect the equipment with other company’s' electronic or mechanical
devices,please contact Shenzhen Well. D Medical Electronics Co,. Ltd. before
connection.
6. Equipment operation, storage and transportation environment
Environmental requirements on normal operation:
a) Environment temperature range: 0℃~ +40℃
b) Relative humidity range: ≤80%
c) Atmosphere pressure range: 70KPa~106Kpa
Environment requirements on equipment storage and transportation:
a) Environment temperature range: -20℃~+55℃
b) Relative humidity range: <80%(20℃)
7. Do not hit the fragile TFT-LCD display. If it cracks, deal carefully with it in case the liquid
crystal gets into eyes or mouths.
8. Must not hit the inner chargeable Li-ion battery nor throw it into fire in case it trigger an
explosion;Do not short circuit the battery output electrodes in case the battery be
damaged; and please use the original binding charger to charge the battery. More over,
because used battery will cause environment pollution, please handle the battery
correctly for recovery processing.
9. Must not disassemble the power supply adapter. If failures happen, it should be handled
by the professional; the charging output can only be used for charging the battery of the
equipment, any improper use on other battery may cause explosion, fire and other
unexpected hazards.
10. Must not short circuit the output of the adapter,a long term short circuit shall result in
adapter damage.
11. Please use standard power cord as the input line of the network power supply for the

- 5 -
adapter to reduce risk.
12. Shenzhen Well.D Medical Electronics Co., Ltd. shall not take any responsibility for any
risk resulted from unauthorized re-fitment by the users.
13. To disconnect the equipment from the power supply network by unplug the adapter from
the power supply network.
14. Refer to appendix B for sound output parameters.
Warning:
When working in an intensive electromagnetic environment beyond its pre-stated one,
WED-3000V's image may be interfered thus influence the diagnose. At this time, stop
the device and reuse it until the electromagnetic interference is removed.
Warning:
When WED-3000V works with other devices in reduplication or in close parallel,
unforeseen EMC problem may appear.
If they have to work closely, observe all the devices and verify if they are influenced by
unexpected EM coupling.

- 6 -
1 Summary
1.1 Brief Introduction
This equipment is high resolution linear/convex Veterinary Ultrasound Scanner. It adopts
micro-computer control and digital scan converter (DSC), digital beam-forming (DBF),
real time dynamic aperture (RDA), real time dynamic receiving apodization, real time
Dynamic receiving focusing (DRF), Digital frequency Scan (DFS), frame correlation
technologies to endue its image with clarity, stability and high resolution.
There are five display modes: B, B+B, B+M, M and 4B; And 256 gray scale.
The system can process real time image display, freeze, save, load, up and down flip, left
and right flip, capacity cine loop; Multi-level scanning depth, dynamic range, Scan angle,
frame correlation factor regulation and focus number, focus position, etc.
Date, clock display; name, sex, age, hospital annotation; Distance, circumference, area,
volume and heart rate measurement; EDD Measurement for equine, bovine, sheep, dog
and cat. It offers 16 body marks. And Multiple-choice probes for the clinic diagnosis
demands.
Video output offers connection to external video image printer and large-screen display
and other equipments. High speed USB port provides real time image transfer to the PC.
Combined power supply mode of AC adapter and built-in chargeable battery, 3 battery
charging modes and the low power consumption and advanced power management
technology promise more lasting battery operation.
7 inches TFT-LCD display and field programmable gate array (FPGA) and surface
mounted technology (SMT) make this equipment compact and light in weight.
Jet molding enclosure with hand-held structure.
The equipment consists of main unit, probe and adapter.
1.2 Range of application
Suitable for diagnosis on horses, cows, sheep, pigs, cats and dogs and other animals.

- 7 -
1.3 Appearance
Front sketch
Side interfaces sketch
Top knobs sketch
1.4 Technical Specificatio
MODEL
WED-3000V
Probe
L1-5/7.5MHz
HF linear
LV2-4/6.5MHz
Rectal
C1-11/50R/3.5M
Hz convex
C1-12/20R/5.0M
Hz
micro-convex
35LV150CP
Backfat probe
Detect Depth(mm)
≥80
≥140
≥90
≥160
Resolution
(mm)
Lateral
≤1 (Depth≤
60)
≤3 (Depth≤
80)
≤5 (80<
Depth≤130)
≤3 (Depth≤
60)
≤2 (Depth≤
80)
Axial
≤1 (Depth≤
80)
≤1 (Depth≤
80)
≤1 (Depth≤
60)
≤1 (Depth≤
80)
Blind zone(mm)
≤3
≤6
≤5
≤5
Geometric
position
precision
Horizont
al
≤5
≤7.5
≤7.5
≤5
Vertical
≤5
≤5
≤5
≤5

- 8 -
Monitor size
7 Inch TFT LCD
Display mode
B、B+B、4B、B+M、M
Image gray scale
256 Scale
Image storage
64 Frame
Dynamic range
0-192dB(Adjustment range 64-192dB)
Cine loop
≥340 Frame
Scan depth
70mm~240mm
Image flip
Up/down, left/right
Focus
Focus Number, Focus position
Body mark
16
Image Process
Pseudo color, GAMA, Image Smoothen, Histogram
Probe Frequency
Adjustable(3 point each probe)
Angle
Show, Adjustable
Measure
Distance, Circumference, area, volume, Gestational age,
expected date
Character display
Date, clock, name, PID, age, Sex, hospital name, doctor,
full-screen character editor
Output report
Support
Interface
2*USB2.0,VIDEO,MOUSE
Battery Continuous work
≥3Hours
Size
L(265mm)*W(46mm)*H(153mm)
Net weight
800g
Power
45W
1.5 Electric principle block diagram

- 9 -
1.6 Standard configurations
Main unit
power adapter
Battery
35LV150CP Backfat probe
Adapter power supply cord
Coupling gel 250ml
Manual
Final examination report
Packing List
1.7 Optional pieces
LV2-4/6.5MHz Rectal probe
C1-11/50R/3.5MHz convex array probe
C1-12/20R/5.0MHz micro convex probe
L1-5/7.5MHz HF linear probe
Printer
CD
High speed USB cable
Warning:
Please use the spare parts specified above. The manufacturer shall not bear the
responsibility of safety problem, unexpected EMC performance drop caused by
arbitrary adoption of non-specified spare parts.

- 10 -
2 Installation
2.1 Operation environmental requirements
Environment temperature range: 0℃~+40℃
Relative humidity range: ≤80%
Atmosphere pressure range: 70KPa~106Kpa
Power supply: a.c.100V~240V,50Hz±1Hz/60Hz±1Hz
Avoid strenuous vibration during operation; Keep away from equipments with high
electric field, high magnetic field and high voltage; avoid strong sunlight on the display;
Keep the equipment well-ventilated, damp proof and dustproof.
Note:
Check according to the "packing list" after uncasing and make sure that there is no
shipping damage, and then install the equipment according to requirements and
methods described in “Installation”.
Warning:
If there is breakage at unpacking check, it is banned to use the device to ensure
security.
The probe should be protected from felling off or crashing and the manufacturer.
2.2 Installing and disassembling battery
Install battery: Set the battery into the battery slot and move the battery release key on its side to top
till the battery is inserted completely and then release the key (refer to figure).
Disassemble battery: It is the reverse process of installation (refer to figure).

- 11 -
2.3 Installing and disassembling probes
Install probes:
The probe jack lies in the bottom of the right side of the equipment. Adjust the probe plug
mark position and insert it, turn the plug lock sleeve counter-clockwise. There is only one plug
jack which is also compatible for those optional probes (refer to figure).
Disassemble probes:
It is the reverse process of installation.
Warning:
Installation and disassembly of the probe must comply with the operation instructions,
must not twiddle the sheath around the plug (refer to figure).
Warning:
Do not by all means unplug or plug the probe connector at state of log on in case the
probe and main unit be damaged.
1
2
3

- 12 -
2000 times of efficient connection and disconnection of the probes are
guaranteed.Once the probe is connected with the main unit, do not unplug nor plug it
at discretion in case poor contact happen.
Warning:
The probe should be protected from felling off or crashing and the manufacturer
assumes no responsibility for the kind of hazard.
Warning:
Must not touch the contact pin in the probe connector.
Warning:
Please handle the equipment carefully.
2.4 Power Supply
The equipment provides two automatic switch-over modes to supply power: adapter and built-in battery.
2.4.1 Power supply with adapter
1. Check the input power cord plug of the adapter to see if it matches the EPS outlet.
2. Check the EPS to see if it is in the specified range and the power cord to see if it is connected well.
3. Check the adaptor to see if it works well: Plug the power cord into the AC input outlet, switch on the
power switch of the outlet, if the DC output indicating light turns on green, it works well.
4. Shut the power switch of the outlet.
5. Insert the DC output plug of the adapter into the DC14V/3A outlet in the rear of the equipment, switch
on the outlet.
6. Press the power switch, the main unit is power on and the power indicator light turns on, that means the
adapter operate well.
2.4.2 Battery Operation
1. Install the battery correctly on the main unit.
2. Switch on the power switch on the left side of the equipment. When the main unit is power on, the
power indicator will turn on.
3. The equipment can start operation.
Note:
At power off state, there is still certain power consumption with the main unit. If the
main unit can not be turned on when pressing down the switch, that is to say the
battery is running up and needs charging.
Note:
When the main unit charging/volume indicator light is flashing, that is the battery is
running up and it should be charged as soon as possible.

- 13 -
Warning:
It is prohibited to use any other power supply except the standard adapter as the
external power supply for the main unit.
2.5 Battery charging
There are 3 ways to charge the battery.
2.5.1 Charging through main unit
1. Install the battery correctly into the main unit.
2. Insert the plug of “ ” of the adapter into the “DC14V/3A” interface on the rear.
3. Connect power cord of “ ”of the adapter to the AC. EPS.
4. Whenever the main unit is power on or off, the “Charging/Volume” indicator light turns on, it means
the battery is charged through adapter; when the lights turn on, it means the battery is fully
charged(refer to figure).
Charging through main unit
Tip:
To prolong the battery operation time, please finish charging when the indicator lights
show that the battery is full.
2.5.2 Charging through adaptive line
1. Take out the battery from the main unit or take out the spare battery.
2. Connect the round end of the adaptive line to the " " terminal of the adapter, and the flat end to
the charging terminal of the battery.
3. Connect power cord of " " of the adapter to theAC EPS.

- 14 -
4. When the " " indicator light on the adapter turns into red, the battery is in charging; when the
" " indicator light turns into green, the battery is fully charged. (Refer to figure).
Charging through adaptive line
2.5.3 Charging through auto-charger
1. Take out the battery from the main unit or take out the spare battery.
2. Connect the flat end marked with an arrow of the auto-charger to the charging terminal of the battery.
3. Plug the other end of the auto-charger into the cigar lighter socket.
4. When the "Charging" indicator light on the adapter turns into red, the battery is in charging; when the
"Charging" indicator light turns into green, the battery is fully charged. (Refer to figure).
Charging through auto-charger
Tips:
The input voltage of the auto-charger is DC9~14V/1.5A.
The output voltage of the auto-charger is DC12.6V/1A.
The operations and storage environment are the same as those of the main unit.

- 15 -
3 Operating procedures
3.1 Introduction of keyboard
Mouse
As a supplement of computer operation, mouse can make the measurement faster and more convenient. PC
mouse of which the left button, middle button and right button have specific functions is applied to the
equipment.
Left button
Freeze the image or display the measuring mark.
Middle button
End the measurement and display the result of volume measurement.
Right button
● Determine the start and end position of the cursor, and switch the start cursor and end cursor in the
distance measurement mode.
● Determine the start position of the cursor and complete the measurement of perimeters and areas in
the free arm type measurement of perimeters and areas.
● Be equal to the role of button in the elliptical type measurement of perimeters and areas.
Note:
"FUNC / 3" reserved function keys for the device.
Power switch
Freezes button
Display area
Power indicator
Probe socket
Function button

- 16 -
3.2 Display interface
3.3 Operating procedures
1. Switch on the power button“ ”.
2. Press any button to enter the B-type ultrasonic scan interface.
3. Type the basic information of the patient.
Press “ ” to type the information including name, PID, age, gender, hospital, doctor.
4. Spread ultrasonic coupling agent on the probe acoustic shadow surface, and close to the scanning area,
then the real-time image can be observed on the screen.
5. Freeze the image, and conduct measurement on the image.
6. Create a report, print the report and image by the connected video printer.
Hospital name
probe type
Focous, focal distance
Probe raster
Depth display
System date, time
Patient age
Patient gender
Doctor name
Dynamic range, frame frequency
gray scale band
Scanning direction mark
Patient name
Status indicator of cineloop
overall gain
Medical record number
Probe frequency
Current scanning angle
Frame correlation value
Near field, far field
Brightness, image smoothing, grey correction
Battery capacity indicator

- 17 -
4 Image parameter adjustment
4.1 Display mode switching
B mode switching
Press “B/0”button in the real-time scanning mode to switch current image to real-time single B mode. (The
default mode when the equipment is power on)
BB mode switching
Press “BB/1” button to switch the current image to double B mode. Real-time image and freeze image are
displayed on the screen. Press “BB”button continuously to switch the real-time and freeze states of two
images. (Attention: Backfat probe does not support)
B+M/M mode switching
Press “BM/2” button to switch the current image to B+M mode. B type and M type real-time images can be
displayed on the screen at the same time. The sample line of B type image can be moved around by mouse
and arrow button. Press this button again to switch the current image to M mode, and then an M type image
is displayed on the screen. (Attention: Backfat probe does not support)
4B mode switching
Press “FUNC/3”to switch the current image to 4B mode in the B type, BB type real-time scanning mode.
Four B type images are displayed, one of which is in a real-time state. Press“FUNC/3”button continuously
to switch the real-time and freeze states of four images. (Attention: Backfat probe does not support)
B BB B+M
M 4B
Attention:
When you call up the menu, please exit the menu before switching the display mode.
Other manuals for WED-3000V
1
Table of contents
Other WELL.D MEDICAL Medical Equipment manuals
Popular Medical Equipment manuals by other brands

NSK
NSK Osseo 100+ Operation manual

St. Jude Medical
St. Jude Medical Wi-Box AO Transmitter Instructions for use

Integra
Integra MAYFIELD Pediatric Radiolucent Horseshoe... instruction manual

AGFA
AGFA Classic E.O.S. Installation planning

MedaCure
MedaCure Comfort Zone LRM36 user manual

Weinmann
Weinmann MEDUMAT Standard2 instructions