Error Codes
7
Wile 200 is equipped with error messages to guide the user to get the
most reliable results, as well as to ensure that the meter is in good
condition.
Error code Cause Solution
Critical Error The meter must be submitted to an
authorized service center
7Weight not detected Repeat test / Do weight test
8Taring result is too high Do weight test and check the scale
9Weight measurement uncertain /
meter shook during the measure-
ment
Keep the meter stable,
measurement uncertainty has in-
creased
10 Moisture measurement result
stabilize slowly Happens to very wet samples
11 Weight measurement too un-
certain / meter shook during the
measurement
Keep the meter stable, measurement
uncertainty has increased
12 Temperature of the sample lower
than 2 C / Measurement uncer-
tainty is high
The tester is most accurate when grain
and tester temperature is between 16C
and 32C
13 Temperature of the sample above
50 C / Measurement uncertainty
is high
The tester is most accurate when grain
and tester temperature is between 16C
and 32C
14 Large temp. diff. between the grain
and test cell / Measurement uncer-
tainty has increased
For better accuracy, grain and meter
test cell should be at similar tempera-
tures
15 Not on a suffciently level surface,
meter warns with sound alarm Keep meter stable during measure-
ment
16 Weight result is over 330 g Check that leveling handle did not stay
above the test cell/ Do weight test
17 Weigh result is under 50 g Check that leveling handle did not stay
above the test cell/ Do weight test
18 Hectoliter weight is 20% higher
than average Measurement uncertainty has in-
creased, Do weight test
19 Hectoliter weight is 20% lower
than average Measurement uncertainty has in-
creased, Do weight test
20 Test cell is not cleared before the
measurement Clear test cell