1
1. Entering clinical setting interface: While on standby, press both the “+”and “-”buttons and hold for
three seconds. The interface will be shown as below:
Figure1 Standy Figure2 Clinical setting interfaceA
Figure3 Clinical setting interface B
2. Parameters setting
From the clinical setting interface, press the “+”and “-”button to toggle between different options and
options on the interface, press the “ok”button to choose an option or setting that is highlighted. Or
you can choose an option or setting by directly touching on the interface at corresponding area. The
interface will only show a few lines at a time. You can toggle between pages to show more options or
settings by touching the “Previous”and “Next”.
The available options and settings are shown in the chart below:
OFF/1/2/3/4/5/6/7/ Patient
*Note: “Min pressure”and “Max pressure”are only selectable inAuto CPAP mode, and will be
replaced by “Therapy mode”when therapy mode has been set into CPAP.
**Requires further operation. Will be explained in following contents.
Manual titration: