York HFG02 User manual

Products User Manual Template – Issue 2
Document number Y0321/DOC/024 Rev. 3
Author: ...............................................................
D Cullen BEng (Hons), MSc, CEng, CMgr, MIET, MIEEE, MCMI
Test Instrumentation Manager
Checked: ..... ................................................................................
M S Anslow MEng, CEng, MIET, MIEEE
Senior Engineer
Approved: ................................................................
D Cullen BEng (Hons), MSc, CEng, CMgr, MIET, MIEEE, MCMI
Test Instrumentation Manager
HFG02
Operation
Manual

HFG02 Operation Manual
York EMC Services Page 2 of 25 Doc. No. Y0321/DOC/024 Rev. 3
HFG02
Operation Manual
Document Number
Y0321/DOC/024
CHANGE HISTORY SHEET
Date Issue Pag/Para Description
17/02/2017 1.0 - 1st issue
20/03/2017 2.0 p19 EUT graphs added
07/02/2018 3.0 Safety instructions updated
UNCONTROLLED ELECTRONIC COPY
Users should satisfy themselves that they are in possession of the most up to date
version of this manual by calling +44 (0) 330 4303456 or by emailing

HFG02 Operation Manual
York EMC Services Page 3 of 25 Doc. No. Y0321/DOC/024 Rev. 3
Limited Warranty
York EMC Services warrants each HFG02 to be free
from defects in material and workmanship for a period of
one year from the date of shipment to the purchaser.
This warranty extends to the original purchaser only. It
does not apply to any product or parts subject to misuse,
neglect, accident, unauthorised service or abnormal
conditions of operation.
In the event a product covered by this warranty fails,
York EMC Services, will, without charge, repair and
recalibrate the instrument if returned to their test
laboratory within one year of the original purchase –
provided that York EMC Services examination discloses,
to its satisfaction, that the product is defective. York EMC
Services may, at its option, replace the product in lieu of
repair. If the defect was caused by misuse, neglect,
accident, unauthorised service or abnormal conditions of
operation, repairs will be billed at a nominal cost. In such
cases, an estimate will be provided before work is
started, if requested by the purchaser.
For warranty service, contact York EMC Services.
Provide the serial number of the equipment and
complete details regarding the failure. You will then be
given shipping instructions. Return the instrument to the
factory. Repairs will be made and the equipment will be
returned to you. York EMC Services assumes no
responsibility for loss of, or damage to, products in
transit.

HFG02 Operation Manual
York EMC Services Page 4 of 25 Doc. No. Y0321/DOC/024 Rev. 3
DISCLAIMER
THE HFG02 HAS BEEN DESIGNED TO PRODUCE HARMONIC AND FLICKER
DISTURBANCES TO THE POWER SUPPLY TO WHICH THE TEST PORT HAS
BEEN CONNECTED, AND AS SUCH IT IS ONLY INTENDED TO BE
CONNECTED TO SUITABLE TEST EQUIPMENT. CONNECTION TO THE
PUBLIC MAINS DISTRIBUTION NETWORK IS NOT PERMITTED.
YORK EMC SERVICES LTD ASSUMES NO LIABILITY FOR ANY DAMAGE
THAT MAY RESULT FROM THE USE OF THIS EQUIPMENT OUTSIDE ITS
RECOMMENDED OPERATING CONDITIONS.
PRECAUTIONS
THE HFG02 MUST NOT BE OPERATED WITH THE TEST SUPPLY INPUT
PLUGGED DIRECTLY INTO A PUBLIC MAINS SUPPLY SYSTEM.
WHEN OPERATED, THE HFG02 WILL CAUSE HARMONIC AND FLICKER
EMISSIONS TO THE SUPPLY CONNECTED TO THE TEST SUPPLY INPUT.

HFG02 Operation Manual
York EMC Services Page 5 of 25 Doc. No. Y0321/DOC/024 Rev. 3
1 INTRODUCTION 6
2 DESCRIPTION 7
4 DESCRIPTION OF SYMBOLS 8
5 SPECIFICATION 9
5.1 OPERATING MODES 9
5.2 HARMONICS AND FLICKER MEASUREMENT SYSTEM TEST SUPPLY 9
5.3 OPERATOR INDICATORS 9
5.4 POWER REQUIREMENTS 10
5.5 ENCLOSURE 10
5.6 COMPLIANCE 10
5.7 ENVIRONMENT 10
6 OPERATION 11
6.1 CONTROLS AND INDICATORS 11
6.2 SUPPLYING POWER TO THE HFG02 12
6.3 OPERATING MODES 12
6.4 ENABLING THE TEST 13
6.5 TEST SUPPLY INDICATORS 13
6.6 PERFORMING A TEST. 14
6.7 TROUBLESHOOTING A TEST 15
7 OUTPUT PERFORMANCE 17
7.1 FLICKER LOAD OPERATION 17
7.2 HARMONICS LOAD OPERATION 17
7.3 EUT LOAD OPERATION 19
7.4 CALIBRATION 21
8 SAFETY 22
9 ELECTROMAGNETIC COMPATIBILITY 23
9.1 GENERAL 23
9.2 EMISSIONS 23
9.3 IMMUNITY 23
9.4 NON-APPLICABILITY 23
10 MAINTENANCE AND CALIBRATION 24
10.1 DISPOSAL 24

HFG02 Operation Manual
York EMC Services Page 6 of 25 Doc. No. Y0321/DOC/024 Rev. 3
1 INTRODUCTION
The Harmonics & Flicker Generator 02 (HFG02) has been designed by York EMC
Services for the purpose of providing an easy and reliable way to externally check
the performance of measurement systems used for testing against the EN/IEC
61000-3-2 harmonics and interharmonics and EN/IEC 61000-3-3 flicker standards.
The HFG02 allows the user to periodically verify their test equipment and its
operation together, helping to ensure compliance with standards and laboratory
quality procedures.
The HFG02 provides a series of harmonic and flicker disturbances of a nominal
but stable level. The HFG02 may therefore be used to verify the stability of a
measurement system. Additionally, due to its own stability it may be used as a
transfer standard from a calibrated system, or be used as a repeatable source of
interference for diagnostic and/or development design work.

HFG02 Operation Manual
York EMC Services Page 7 of 25 Doc. No. Y0321/DOC/024 Rev. 3
2 DESCRIPTION
The HFG02 simulates equipment under test (EUT), generating known, repeatable
levels of harmonic and flicker disturbance in one of ten modes of operation. These
modes are divided into flicker, harmonics and EUT profiles.
The flicker test modes generate a fixed level of mains disturbance at a rate of 1 Hz
and 8.33/8.6 Hz, depending on the frequency of the Test Supply under test. These
modes support verification of the test systems used for EUT flicker disturbance
measurements as described in the EN/IEC 61000-3-3 test standards.
The harmonics test modes are designed to produce a variety of harmonic and/or
interharmonic-rich current waveforms. These current waveforms support
verification of the test systems used for mains-powered equipment load current
testing as described in the EN/IEC 61000-3-2 test standards.
A selection of EUT modes is also provided. These modes generate disturbances
which simulate certain types of real-world EUT not covered by the generalised
harmonics and flicker modes.
The HFG02 can be used with test systems operating at both 110 Vac and 230 Vac
(nominal) supply voltages, and with either 50 Hz or 60 Hz supply frequencies. The
HFG02 constantly monitors the Test Supply under test and adjusts itself
automatically, indicating the detected voltage and frequency to the operator.
Connection to the test system is by a captive lead pre-fitted with a CEE 7/7
European style plug that fits directly with power outlet used by a wide range of
harmonics and flicker equipment.
The HFG02 itself is powered by a wide-input range AC/DC power supply unit that
is powered separately from the Test Supply under test. This ensures that the
disturbance signals generated by the HFG02 are not affected by the need to derive
the HFG02’s own supply needs from the Test Supply under test.
For ease of access all connections, user controls, and indicators are located on
the front panel of the HFG02. The HFG02 is supplied with the external power unit
with mains cable, manufacturer’s test certificate, and a CDROM containing digital
copies of the manual, test certificate, and files of the tabulated certificate data.

HFG02 Operation Manual
York EMC Services Page 8 of 25 Doc. No. Y0321/DOC/024 Rev. 3
3 DESCRIPTION OF SYMBOLS
Symbols displayed on the product are listed below:
Caution
Protective Earth Terminal
Alternating current
HFG02 power supply on
HFG02 power supply off
Direct current
DC connector polarity
ESD handling precaution
Separate collection for WEEE-
Waste Electrical and Electronic
Equipment

HFG02 Operation Manual
York EMC Services Page 9 of 25 Doc. No. Y0321/DOC/024 Rev. 3
4 SPECIFICATION
4.1 Operating modes
(See Section 5. Output Performance for details of typical measurement results)
4.1.1 Current harmonics
Frequency range 5 Hz to 2.0 kHz (50 Hz Test Supply)
5 Hz to 2.4 kHz (60 Hz Test Supply)
Depending upon mode selection
Stability < 5% of the test limit in EN 61000-3-2 standard
4.1.2 Flicker disturbance
Flicker rate 1 Hz (50 Hz and 60 Hz Test Supply)
8.3 Hz (50 Hz Test Supply only)
8.6 Hz (60 Hz Test Supply only)
Stability < 5% of the reading
4.1.3 EUT profiles
Half-wave rectifier Load current is drawn every alternate half-cycle
Dimmer Load current is drawn at every cycle between 90º-180º
and 270º-360º phase angles.
Intermittent flicker Load current is drawn for varying length periods over a
20 minute recurring cycle
4.2 Harmonics and flicker measurement system test supply
4.2.1 Test Supply ratings
Voltage 80 Vac to 130 Vac (110 Vac nominal input)
200 Vac to 253 Vac (230 Vac nominal input)
Current 3 A (peak)
Frequency 50 Hz or 60 Hz
Power rating 500 W max
Connector CEE 7/7 European style plug for connecting the HFG02
to the Test Supply under test.
Insulation rating 3000 Vrms
4.3 Operator indicators
(See Section 4. Operation for a full description of the control and indicator functions)
4.3.1 Test Supply indicators Measured voltage < 80 Vac to >253 Vac
Measured frequency < 40 Hz to > 61 Hz

HFG02 Operation Manual
York EMC Services Page 10 of 25 Doc. No. Y0321/DOC/024 Rev. 3
4.3.2 Operation indicators Test enabled / running
System alarm
4.4 Power requirements
HFG02
Input connector 2.1 mm DC power jack (inner conductor +ve)
Voltage input 24 Vdc
Power consumption 24 W max
Power supply unit
Input connector IEC plug
Voltage input 100 Vac to 240 Vac (nominal)
Power consumption 50 W max
4.5 Enclosure
Overall dimensions (not including captive leads):
Length 330 mm
Width 320 mm
Height 170 mm
Weight 8 kg
4.6 Compliance
Safety Low Voltage Directive, 2014/35/EU
EMC EMC Directive, 2014/30/EU
4.7 Environment
4.7.1 Operating Temperature Range
Recommended Ambient 10ºC to 30ºC
Maximum Ambient 5ºC to 40ºC
4.7.2 Humidity
Relative humidity: 20% to 80% (non-condensing)
4.7.3 Protection
IP rating IP10
NOTE. Do not allow any obstructions to interfere with the operation of the fans, or
site the HFG02 such that airflow is restricted. Doing so will cause the equipment to
run at an increased temperature and in so doing may affect its operation.

HFG02 Operation Manual
York EMC Services Page 11 of 25 Doc. No. Y0321/DOC/024 Rev. 3
5 OPERATION
5.1 Controls and indicators
Identifier Description
Start Test Pressing this switch enables the test mode, and applies the
chosen load profile to the measurement system, as appropriate
for the test supply detected.
Pressing this switch again whilst the test is running disables the
test mode.
Test enabled indicator Indicates that the test mode is running.
Test Mode selector This switch selects one of ten load profile test modes available to
be used. Changing this setting whilst a test is running will
automatically disable the current test mode.
Alarm indicator This indicator flashes if the temperature within the enclosure
exceeds the preset limit. In this state the test mode will be
disabled and the HFG02 will wait until the temperature has
dropped sufficiently before the test can be resumed.
A steady indicator warns of a system failure, which will need to
be cleared by fully restarting the HFG02 (refer to Section 4.7).
Test Supply indicators These indicators display the frequency and RMS voltage of the
Test Supply. Section 5.5. gives details of the ranges indicated.
HFG02 power switch Main On/Off switch for the HFG02.
HFG02 power indicator Indicates that the HFG02 is on.
HFG02 power connector
HFG02 power supply input socket (24 Vdc).
Test supply fuse Protective fuse for the Test Supply input
Test supply cable Captive lead connector for the Test Supply input
Chassis earth Additional earthing point for the HFG02 enclosure
Figure 1: HFG02 front panel and controls

HFG02 Operation Manual
York EMC Services Page 12 of 25 Doc. No. Y0321/DOC/024 Rev. 3
5.2 Supplying power to the HFG02
Figure 2: HFG02 power supply
The HFG02 is powered by an external 24 Vdc power supply unit which connects
to the input socket on the front panel . Pressing the power switch with the
24 Vdc unit connected will turn the HFG02 on, which will be confirmed by the
green indicator . Pressing the power switch again will turn the HFG02 off.
5.3 Operating modes
Figure 3: HFG02 mode selection
The operating mode is selected using the Test Mode switch . The operation of
each test mode is described in Table 1 below. Note that if the Test Mode switch is
changed whilst a test is running, the test is automatically disabled.
Mode Test Mode Setting Load Profile Description
Flicker
8.3/8.6 Hz Produces Pst value of > 1
(above the pass limit defined in EN/IEC61000-3-3)
1 Hz Produces Pst value of < 1
(below the pass limit defined in EN/IEC61000-3-3).
Harmonics
Fixed Single repetitive harmonic-rich current waveform
Fluctuating Repeating current profile alternating between two distinct
waveforms over a 10 second cycle
Fixed + 10 Hz Fixed harmonic waveform with additional 10 Hz
interharmonics
5 Hz Fixed interharmonics at frequency steps of 5 Hz
10 Hz Fixed interharmonics at frequency steps of 10 Hz
EUT
Half-wave rectifier Simulates the effect of a half-wave rectified mains input
Dimmer Simulates the waveform of an analogue dimmer switch
Intermittent flicker Simulates the effect of a varying duty cycle load
Table 1: HFG02 load operation test modes

HFG02 Operation Manual
York EMC Services Page 13 of 25 Doc. No. Y0321/DOC/024 Rev. 3
5.4 Enabling the test
Figure 4: Test Enable switch and indicator
The selected test mode is enabled by pressing the Start Test switch and the
indicator displays the running test status.
5.5 Test Supply Indicators
The HFG02 monitors the input from the Test Supply under test and displays the
voltage and frequency measurements on the front panel indicators as shown in
Table 2.
Test Supply monitoring is continuous and automatic, and does not require a test
mode to be enabled. Tests are automatically disabled if the Test Supply voltage or
frequency are outside their permitted ranges.
Test Supply Voltage Voltage measured Display Indication
110 VAC LED 230 VAC LED
Supply voltage not detected Below detection range
Invalid voltage – too low < 80 VAC
110 Vac range 80 to 130 VAC
Invalid voltage – mid range 130 to 200 VAC
230 Vac range 200 to 253 VAC
Invalid voltage – too high > 253 VAC
Test Supply Frequency Frequency measured Display Indication
50 Hz LED 60 Hz LED
Supply input not detected Below detection range
Invalid frequency – too low < 49 Hz
50 Hz range 49 to 51 Hz
Invalid frequency
–
mid
range 51 to 59 Hz
60 Hz range 59 to 61 Hz
Invalid frequency – too high > 61 Hz
Table 2: Test Supply measurement indicators

HFG02 Operation Manual
York EMC Services Page 14 of 25 Doc. No. Y0321/DOC/024 Rev. 3
5.6 Performing a test.
The HFG02 acts as the equipment under test in the harmonics and flicker test
system, and should be connected to the single-phase test power supply as per the
typical test setup used (Figure 5 for example).
Figure 5: Example of a typical test setup
Connect the HFG02 power supply and turn the HFG02 on by pressing the
power switch . The power supply indicator should be illuminated green to
indicate the HFG02 is operating correctly.
The Test Supply Input cable can be plugged into the Test Supply power output
socket. Note that if the Test Supply does not provide a safety earth connection as
one of the conductors (e.g. 2-pin Live and Neutral only) then the Chassis Earth
wiring stud on the front of the HFG02 must be connected to a suitable earth first
(not shown). A suitable earth point may be provided elsewhere on the Harmonics
and Flicker test system.
Configure the Harmonics and Flicker Test System to measure harmonic or flicker
disturbance as per the system instructions, with the Test Supply configured to
provide the required voltage and frequency.
Once the Test Supply has been safely connected, the Harmonics and Flicker Test
Supply can be enabled. The HFG02 Test Supply indicators should show
confirmation of the selected Test Supply output voltage and frequency.
Select the required test load operation using the Test Mode switch .

HFG02 Operation Manual
York EMC Services Page 15 of 25 Doc. No. Y0321/DOC/024 Rev. 3
The chosen test load operation can be enabled before or after the Harmonics and
Flicker Test System has started the test routine. When ready to enable the load
operation, press the Start Test switch . The Test Enable display should
indicate that the test is running.
Pressing the Start Test switch again will disable the running test. Changing the
Test Mode selection switch will also disable the running test. The new selected
test will need to be started by pressing the Start Test switch again.
Note:
If the HFG02 test has been running for more than 2 hours continuously it is
recommended that it is allowed to cool down before power is finally removed, to
reduce any latent heat build-up. To do this, leave the HFG02 turned on but with no
running test for approximately 10 to 15 minutes before fully powering down.
5.7 Troubleshooting a test
The HFG02 monitors the Test Supply input from the Harmonics and Flicker Test
System and reports the measured voltage and frequency on the front panel display
(see Table 2). If an invalid voltage or frequency is detected, the test load operation
will be disabled (Test Enable display will be OFF).
The test load operation will also be disabled if the HFG02 detects too many
dropped voltage cycles or interruptions from the Test Supply.
If the HFG02 load exceeds its recommended operating temperature, the test load
operation will be disabled (Test Enable display will be OFF) and the Alarm
indicator will flash at a steady rate. Once the temperature has returned to
normal the Alarm indicator will turn off and the test can be resumed.
If the HFG02 detects a system error, the Alarm indicator will fully turn on without
flashing. The system error will need to be cleared by removing all power from the
HFG02 before fully restarting. If fully restarting the device fails to clear the error,
the HFG02 may require servicing.
The Alarm indicator will also fully turn on if a serious error has been detected with
the Test Supply voltage e.g. surge or transient overvoltages sufficient to cause
damage to the HFG02 itself. This error can be cleared by disconnecting the Test
Supply and restarting the HFG02.
A flowchart for troubleshooting if the test cannot be enabled is shown in Figure 6.
In the case of other problems, please contact your supplier for advice.

HFG02 Operation Manual
York EMC Services Page 16 of 25 Doc. No. Y0321/DOC/024 Rev. 3
Figure 6: Troubleshooting flowchart

HFG02 Operation Manual
York EMC Services Page 17 of 25 Doc. No. Y0321/DOC/024 Rev. 3
6 OUTPUT PERFORMANCE
Please refer to the test results supplied with the unit for its unique performance.
Typical performance results are shown in the sections below.
6.1 Flicker Load Operation
The typical flicker disturbance generated by the HFG02 in each of the steady flicker
modes when tested according to EN/IEC 61000-3-3 is shown in Table 3.
Flicker Mode
Setting Load Operation Pst Plt
8.3/8.6 Hz
8.33 Hz flicker (50 Hz Test Supply detected)
8.6 Hz flicker (60 Hz Test Supply detected)
1.10
1.06
1.10
1.06
1 Hz
1 Hz flicker (50 Hz or 60 Hz Test Supply detected) 0.45 0.45
Table 3: Typical Pst results measured to EN/IEC 61000-3-3
6.2 Harmonics Load Operation
The typical current harmonics disturbance generated by the HFG02 in each of the
harmonics modes when tested according to EN/IEC 61000-3-2 is shown below:
Figure 7: Disturbances generated in Fixed Harmonic mode

HFG02 Operation Manual
York EMC Services Page 18 of 25 Doc. No. Y0321/DOC/024 Rev. 3
Figure 8: Disturbances generated in Fluctuating Harmonics mode
Figure 9: Disturbance generated in Fixed + 10 Hz Interharmonics mode

HFG02 Operation Manual
York EMC Services Page 19 of 25 Doc. No. Y0321/DOC/024 Rev. 3
Figure 10: Disturbances generated in 5 Hz and 10 Hz Interharmonics modes
6.3 EUT Load Operation
The typical disturbances generated by the HFG02 in each of the EUT modes is
shown below:
Figure 11: Disturbances generated in Dimmer and ½ wave rectifier modes

HFG02 Operation Manual
York EMC Services Page 20 of 25 Doc. No. Y0321/DOC/024 Rev. 3
Figure 12: Half-wave rectifier current profile
Figure 13: Dimmer current profile
Figure 14: Intermittent flicker voltage profile
Table of contents
Other York Inverter manuals