Zeizz SteREO Discovery.V12 Reference guide

B 46-0008 e 12/06 1 www.zeiss.de/stemi
SteREO Discovery.V12 Brief instructions for use
For your safety
Observe the
following
instructions:
−The stereomicroscope SteREO Discovery.V12 including the original accessories may only be used for
the microscopy procedures described in these instructions and in the detailed Operating Manual.
The manufacturer accepts no liability if used for any other procedures.
−Read the Operating Manual (B 46-0009) carefully; follow the safety instructions described here.
−Connect the unit to the mains only at a correctly installed earthed socket.
−Unplug the unit from the mains before opening it, and before changing a bulb or fuse.
−Read the related user instructions before using any additional accessories.
−When the motorised focussing is in motion there is a crush hazard!
During movement, keep your hands away from the movement area.
−When using the microscope with transmitted-light there is a dazzle hazard.
Use the dazzle protector provided with the transmitted-light unit.
Controls and functional elements
NOTE: The microscope configuration shown here may vary from the existing configuration!
1Binocular Ergo phototube S 5°-45° for mounting eyepieces and a camera using an adapter
2Microscope body SteREO Discovery.V12
3 Dazzle protector, swing forward for use; when using the microscope with transmitted-light illumination
4Transmitted-light unit S for use only in connection with cold-light source KL 1500 LCD / KL 2500 LCD for transmitted-light illumination
5Support for glass plate with diameter 120 mm for objects or support of tables using adapter plate 84/120
6Support for opal glass for transmitted-light illumination (diameter 84 mm)
7Coded nosepiece 3 fold, coded or carrier S (not shown) for supporting the microscope body (support diameter 76 mm)
8Oblique reflected-light illumination with flexible light guide and focussing attachment.
9Illumination control for transmitted-light microscopy: see section Adjusting the illumination
10 System control panel SYCOP; for operation see section SY CO P Touch screen controls
11 Foot pedal for focussing / zoom function
12 Control unit EMS-1 (component of SY C OP ); connection of various functional units
13 Cold-light source KL 2500 LCD for reflected-light; see user instructions for setting and operating
14 Cold-light source KL 2500 LCD for transmitted-light; see user instructions; only in connection with transmitted-light unit S
15 Focussing control unit HIP (Human Interface Panel); power supply unit for operation without SYC O P included in scope of supply
16 Motor focussing on the column; caution! During movement, keep your hands out of the movement area - crush hazard!
17 Zoom control unit HI P (Human Interface Panel); power supply unit for operation without SY C OP included in scope of supply
Switching the device on
Switch-on procedure varies according to the existing configuration.
•Check that all cables are correctly plugged in.
•SY C OP : Switch the rocker switch at the rear of the control unit
EMS-1 to Iand press button on the front (1).
•HI P : Press button (2).
NOTE: Press button
on the control units EMS-1 (1) again to
switch into standby mode.

www.zeiss.de/stemi 2 B 46-0008 e 12/06
HI P controls
The HI P (Human Interface Panel) allows operation without
SY C OP . The first HI P is used for motorised focussing, the
second HI P for zoom operation. If a SY C OP is connected, the
HI P is not necessary and is only of limited function.
There is a basic mode and a setting mode.
Zoom control unit HI P in basic mode
The display shows the magnification (Mag) and the field diameter
(Field). Briefly press the Mode button repeatedly to display
successively field diameter, resolution and depth of focus.
Zoom value is set by the thumb wheel.
Focussing control unit HI P in basic mode
The header line of the display shows the focus position Z. Briefly
press the Mode button to reset the focus value to zero.
Set the focus position by moving the thumb wheel. Press the
thumb wheel to switch between FocusSpeed and FineFocus.
Parameters in the Setting mode (selected by Mode button)
applies to Zoom control unit only:
Set Lens to change objective variables
Set Eyepiece to change eyepiece variables
Confirmation to switch starting enquiries on / off
applies to both control units:
Zoom / Focus Speed to set the speed of movement
Backlight to adjust background illumination
Beep Level to switch acknowledgement beep on / off
Reset Param? to set on default variable
Thumb wheel ©ª
•©moves to higher position
•ªmoves to lower position
•In setting mode: select list entry
Buttons Mem1 / Mem2
•Briefly press the buttons to call up saved values
•Press for two seconds duration to save the current value
Mode button
•Press for two seconds duration to switch between the modes
•In setting mode: briefly press to select Parameters
NOTE: Changes to the parameter variables are only saved when
you switch back into basic mode.
Button
•HI P switch on / switch off
SYCOP controls
Joystick:
•Shift to left: reduce zoom value
•Shift to right: increase zoom value
•Shift upwards: move the focus plane upwards
•Move downwards: move the focus plane downwards;
CAUTION: crush hazard! Avoid risk of damaging the specimen by
using the specimen protection function or limit switches.
4 buttons for illumination intensity;
the two left hand buttons are for transmitted-light (TL)
the two right hand buttons are for reflected-light (RL)
•– Reduce brightness / double click switches the illumination off
•+ Increase brightness or switch on again
See also section SY CO P Touch screen controls.
SY CO P Touch screen controls
The touch screen is a touch-sensitive screen. Press with the finger on
the relevant button to call up the function.
After switching on, the main menu shows the following values:
−Date and time; four short status display windows
−Visual overall magnification (Mag), objective (Obj), eyepiece (Eyep)
−Field diameter (Field), resolution (Resol), depth of focus (Depth)
−Illumination intensity, colour temperature for reflected-light (RL)
and transmitted-light (TL)
•Press button Zto reset the current focus position to zero
•Press button POWER OFF to switch the system into standby mode
•Use the menu buttons to call up the FUNCTION, MEMORY, KEY
SETTINGS or SETUP submenus

B 46-0008 e 12/06 3 www.zeiss.de/stemi
Adjusting the illumination
Reflected-light illumination:
•Connect the light guide for reflected-light to the cold-light source
(RL) and attach it to the stand plate or nosepiece (or carrier S).
Alternatively clamp the ring illuminator to the single objective (not
to the nosepiece).
•Switch on the cold-light source for reflected-light (RL) with rocker
switch I.
•Align the light guide on to the specimen.
•Adjust the illumination intensity on the cold-light source or use
the two right hand buttons on SY C OP .
Transmitted-light illumination with transmitted-light unit S:
•Connect the light guide for transmitted-light (TL) to the cold-light
source.
•For lower magnifications, insert the opal glass.
•To avoid dazzle from transmitted-light, swing the dazzle protector
into position.
•Switch on the cold-light source for transmitted-light (TL) with
rocker switch I.
•Adjust the illumination intensity on the cold-light source or use
the two left hand buttons on SY C OP .
•The illumination levers on the transmitted-light unit allow the
brightfield / darkfield to be adjusted as depicted alongside. To
optimise the contrast with oblique light various intermediate
positions can be employed (see explanation of symbols).
Symbols for illumination control
Light beam exits vertically (reflector at 45° inclination)
Inclining the light beam towards the observer (dazzle hazard!) Inclining the light beam away from the observer
Light beam exits centrally Translation of the light beam towards the observer
Diffuse light quality with the white reflector plate Directed light quality with the mirror reflector plate
Adjusting the stereomicroscope
The stereomicroscope should be connected and switched on.
Focus and zoom can be controlled using H I P or S Y C O P . See section
H I P controls or SYCOP controls.
•Place a specimen on the insert plate (3).
•If the dioptric eye values are known, set the eyepieces (2)
accordingly.
•Select the viewing height by twisting the eyepiece support (1) by
up to max. 120° and adjust it to the pitch of individual eyes, i. e.
when looking in the eyepieces only an uninterrupted circle of light
should be seen.
•Tilt the binocular part viewing angle and set it within the range
5° - 45°.
•Open the aperture diaphragm using thumb wheel (4).
•When searching for objects, first set the zoom (5) to the lowest
magnification.
•Focus on a small sharp detail in the centre of the image (6).
•Set the maximum zoom value (5) and adjust for exact focus on the
detail (6).
•Then go back to the lowest zoom value (5) and correct any out-of-
focus separately for each eye, using the adjustable eyepieces (2)
(compensation for variations in the eye).
Once the microscope has been compensated in this way, image focus
will be retained over the entire zoom range.
At the camera outlet of the Ergo phototube (8) a camera can be
attached by a suitable adapter if desired. To change between visual
observation and documentation, swing over the lever (7). The right
hand beam path is switched over completely.
Other Zeizz Microscope manuals
Popular Microscope manuals by other brands

Blue-Ray Biotech
Blue-Ray Biotech EzScope 101 Quick operation guide

Elenco Electronics
Elenco Electronics EDU-41008 instructions

Jeulin
Jeulin 571 220 manual

Denville Scientific
Denville Scientific MZ180 Series user manual

PCE Health and Fitness
PCE Health and Fitness PCE-VMM 100 user manual

BestScope
BestScope BS-2010E user manual