7
in sequence, according to their magnifications, into the threaded holes of the
nosepiece. Place the specimen in the center of the field of view, first use low
power objective to find the specimen image and then observe it with high
power objective, at the same time, adjust fine focusing knob until the image is
clear. When using 100X oil-immersion objective, the space between the front
of objective and specimen surface should be full of cedar oil. Drops of cedar
oil must be strictly free from air bubble, and the objective should be cleaned
immediately after using, otherwise it may solidify and make cleaning
difficult.
2、In order to get a sharp and clear image, you must adjust the illumination
system. When using objectives of different magnification, please adjust the
iris diaphragm of the condenser to bring its numerical aperture and objective’s
numerical aperture into coincidence. Generally speaking, the rate of objective
increase,the location of the condenser need to move up. Aperture is design for
adjust NA and not for adjust brightness. Usually, aperture is open to 70~80% of
the objective’s exit pupil, until can see clearly image in the view.
3、When to watch the same type of specimens, can adjust limit screw to proper
position and lock it up with nut. Then directly make coarse focusing knob up to the
position where the first speciment can focusing clearly. That can simplify the
operation and reduce the working intensity.
Note: The same type of specimen refer to the thickness of the slice 、the coverslip
and the carry slip are uniform.
7 Maintence and storage
1、When you open the carton, please be careful not to make the lens dropping
2、All the lens has been adjusted by manufacturer already, please do not
disassemble by yourself.
3、Nosepiece, coarse and fine focus is installed precisely, please do not
disassemble by yourself.
4、You should make the instrument clean, and often wipe the dusts.
5、Place the instrument in a shady, cool and dry place, when finishing the
operation, always use the dust cover for protection.
8 Troubleshooting Guide
If problems occur during use, please review the following list and take remedial
action as needed. If you cannot solve the problem after checking the entire list,
please contact the service department for assistance.
1、Optical Part