18 | Sesamoid®Plasty User Guide
Technical Specifications (cont.)
Guidance and Manufacturer’s Declaration
Electromagnetic Emissions
The Sesamoid Plasty Computer is intended for use in the electromagnetic environment specified below. The
customer or the user of the Sesamoid Plasty Computer should assure that it is used in such an environment.
Emissions test Compliance Electromagnetic environment – guidance
RF emissions CISPR 11 Group 1 The Sesamoid Plasty uses RF energy only for its internal
function. Therefore, its RF emissions are very low
and are not likely to cause any interference in nearby
electronic equipment.
RF emissionsCISPR 11 Class B The Sesamoid Plasty Computer is suitable for use in
all establishments, including domestic establishments
and those directly connected to the public low-voltage
power supply network that supplies building used for
domestic purposes.
Harmonic emissions IEC 61000-3-2 Class D
Voltage fluctuations/ flicker emissions IEC
61000-3-3
Complies
Electromagnetic Immunity
The Sesamoid Plasty Computer is intended for use in the electromagnetic environment specified below. The
customer or the user of the system controller computer should assure that it is used in such an environment.
Immunity test IEC 60601 Test level Compliance level
Electromagnetic
environment – guidance
Electrostatic discharge
(ESD) IEC 61000-4-2
± 8 kV contact
± 2, ± 4, ± 8, ± 15 kV air
± 8 kV contact
± 2, ± 4, ± 8, ± 15 kV air
Floors should be wood, concrete
or ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30 %.
Electrical fast transient/
burst IEC 61000-4-4
± 2 kV for power supply lines
± 1 kV for input/output lines
± 2 kV line(s) to line(s)
± 1 kV line(s) to earth
Mains power quality should be that
of a typical commercial or hospital
environment.
Surge IEC 61000-4-5 ± 0.5, ± 1 kV line(s) to line(s)
± 0.5, ± 1, ± 2 kV line(s) to ground
for power supply lines
± 2 kV line(s) to ground for input/
output lines
± 0.5, ± 1 kV line(s) to line(s)
± 0.5, ± 1, ± 2 kV line(s) to ground
for power supply lines
± 2 kV line(s) to ground for input/
output lines
Mains power quality should be that
of a typical commercial or hospital
environment.
Voltage dips, short
interruptions and voltage
variations on power
supply input lines IEC
61000-4-11
0 % UT; 0.5 cycle
At 0°, 45°, 90°, 135°, 180°, 225°,
270° and 315°
0 % UT; 1 cycle and 70 % UT; 25/30
cycles
Single phase: at 0°
0 % UT; 250/300 cycle
0 % UT; 0.5 cycle
At 0°, 45°, 90°, 135°, 180°, 225°,
270° and 315°
0 % UT; 1 cycle and 70 % UT; 25/30
cycles
Single phase: at 0°
0 % UT; 250/300 cycle
Mains power quality should be that
of a typical commercial or hospital
environment.
Power frequency (50/60
Hz) magnetic field IEC
61000-4-8
30 A/m 30 A/m Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital environment.
NOTE: UTis the AC mains voltage prior to application of the test level.