Ref.No.:032011-1
Using the MUTE function:
1. Press MODE, the MUTE icon flashes simultaneously.
2. Press ENTER, use up/down to select the on/off.
3. Press ENTER again to save the data.
Note: The factory setting with alarm is buzzer off, user can set the alarm on/off according to MUTE function instruction.
Setting theALTI mode:
Setting the ALARM 1 level:
1. Press MODE, the ALARM 1 icon flashes.
2. Press ENTER. Using the Up/Down to set the parameter. the default alarm 1 is 400~2900 ppm
3. Press ENTER again to save the data.
Note: 1) When the CO2 Alarm level is ≧1,000 ppm, the interval is ±100ppm, when the CO2 Alarm level is<
1,000 ppm, the interval is ±50 ppm.
2) After setting with new alarm level parameter, the green LED will light when CO2 concentration below
the alarm 1 level.
Setting the ALARM 2 level:
1. Press MODE, the ALARM 2 icon flashes.
2. Press ENTER. Using the Up/Down to set the parameter. the default alarm 2 is 500~3000 ppm
3. Press ENTER again to save the data.
*Note: 1) When CO2 Alarm level is 1,000 ppm, the interval is ±100ppm, when the CO2 Alarm level is≧<
1,000 ppm, the interval is ±50 ppm.
2) After setting with new alarm level parameter, the green LED will light when CO2 concentration below
the alarm1 level. The red LED will light when CO2 concentration exceeds the alarm 2 level, the yellow
LED will light when CO2 concentration is between the alarm1 level and the alarm 2 level.
Using the CALI (calibration) Mode:
1. Press MODE, The CALI flashes.
2. Press ENTER, the CALI icon on the display.Adjust the lower display to the ambient CO2 reading by
Up/Down button.
3. Press MODE for more than 10sec, CALIBRATION flashes. Calibration will be done automatically after 3
minute and the LCD will display “Pass” or “Fail.” If it shows “Fail,” please try again.
Setting the Datalogger Mode:
1. Press MODE, DATALOGGER icon flashes.
2. Press ENTER, the reading of CO2, RH and Temperature show on the LCD.
3. Press Up/Down to page up/down the reading. The numbers from 1 to 48 will display with CO2 reading
alternatively. If the LCD display with number “1”, it indicates the first datalogger with CO2,RH and
temperature reading,. The datalogger function can record 48 sets reading.
Remark: With the Built-in Datalogger,ZGw063RY can provide the past CO2,RH and temperature reading within
the past 24 hours. The log interval is 30 minutes per data. ZGw063RY is connected with power and
used for the first time, if the working time is more than 30 minutes, ZGw063RY will have CO2, RH and
temperature reading in datalogger, if the working time is less than 30 minutes, the LCD will display
“NULL” while using the datalogger function.
Using the MAX MIN Mode:
1. Press MODE, the MAX MIN flashes simultaneously.
2. Press ENTER, the reading of CO2, RH and Temperature show
on the LCD alternatively.
3. Press Up/Down and the “CLR” will flash on the LCD. Press Enter
to CLEAR the MAX and MIN record.
Note: With the Built-in MAX MIN mode, ZGw063RY can provide the Maximum and Minimum CO2, RH and
temperature reading record after the device is powered on. If users press up/down to clear the MAX and
MIN CO2 reading record, ZGw063RY will provide the new MAX and MIN CO2, RH and temperature
reading from that time.
Using the “RcFS” Mode:
1. Press MODE, The RcFS flashes simultaneously.
2. Press ENTER, the icon “no” shows on the LCD, Press the Up/Down to select the no/yes.
3.After selecting, press ENTER to save the changes.
Note: If the users set the data or calibrate the sensor wrongly. You can use the RcFS (Recover the factory
Setting) to come back to factory setting data, the datalogger with 48 sets of CO2,RH and temperature will
bewillbecleared.
Specification
Method – NDIR Sample Method -Diffusion or flow through (50 ~200 ml/min)
■CO2Specifications:
■Temperature specification:
Measurement
Range 0-3,000 ppm display
Temperature
Range Display 0°C to 50°C
Display
Resolution 0.1°C
Display
Resolution
1ppm at 0~1,000ppm; 5ppm at
1,001~2,000ppm; 10ppm at
2,001~3,000ppm
Display Options °C
Accuracy 0~2,000ppm, ± 70 ppm or ±5%
of reading whichever is greater;
over 2000ppm: +/-7%
Accuracy
Relay (no action and
be under the alarm
level)
±1°C when the fan blows
to the device directly, the
accuracy of temperature is
+/- 1.5 °C
Repeatability ±20 ppm @400ppm
Accuracy
Relay (action and
exceeds the alarm
level)
±2.5°C when the fan
blows to the device
directly, the accuracy of
temperature is +/- 1.5 °C
Temperature
Dependence
Typ.±0.2% of reading per °C or
±2 ppm per °C, whichever is
greater, referenced to 25°C
Response Time 20-30 minutes (case must
equalize with environment)
■RH Specification:
Pressure
Dependence
0.13% of reading per mm Hg
(Corrected via user input for
altitude)
Measurement
Range 20%-90% RH
Response
Time About 2min for 90% of step
change
Display
Resolution 1%RH
Warm-Up
Time About 60 seconds at 22°C
Accuracy ±5%RH@23°C
Zone LED
Display
Green:<800ppm
Yellow:800~1200ppm Red:
>1200ppm
Response Time <5 min for 63% of step
change
Power Supply
6VDC AC adapter (no plug)
■Operating Conditions:
Relay Output
30VDC or 250VAC, max 2A .,
SPST. Normal Open
Operating
Temperature 0°C to 50°C
Storage
Temperature -20°C to +60°C
Humidity Range 0 ~ 95% RH
non-condensing
Calibration
Note: Before calibrating, you need to obtain standard gas or semi-standard gas, there are 3 ways to get
standard gas.
■Method A: use CO2 in office/building
-Use two meters (One is the device for calibration. The other one is a calibrated (new) one.
-Use ambient room gas for calibration in office, waiting at least 10min, until the CO2 reading doesn’t change
(Notice: usermust not breathe toward the ZGw063RY, CO2 from the userwill affect the reading of
ZGw063RY)
-Use the reading of the new device as the standard reading.
-Calibrate device by the Cali Mode instruction.
■Method B: use CO2 outsides
-Use ambient room gas for calibration outside. Wait at least 10min, until the CO2 reading stabilizes.
(Notice: usermust not breathe toward the ZGw063RY, CO2 from the userwill affect the reading of
ZGw063RY)
-Use 380~420ppm as the standard reading.
-Calibrate the device by the Cali Mode instruction.
■Method C: use standard CO2 in the bottle
-Pump the standard CO2 gas (0~1000ppm, flux = 0.1~0.2 liter/min) into the ZGw063RY from the Gas Entry
Hole, waiting about 2~3min.
- Calibrate the device by the Cali Mode instruction.
1. Press MODE, theALTI icon flashes.
2. Press ENTER,ALTI show on the display And press the mode button to alter the m(meter)
3.Adjust altitude (step=100m) by up/down button.
4. Press ENTER again to save the data.