IAQ IAQ Setting
(Normal IAQ or Wide Range)
Operation Instructions
1. Power on and Measure:
Connect a PC power or adapter within the USB cable to start up the device. The LCD will display “---” during warm-up.
After warm-up, the LCD shows the CO2concentration for Main Display.
2. Automatic Background Calibration:
“OFF” is the default.
2.1 Press Mode Key(○
2) for more than 3 sec. to show ” A on/A oFF ”, then press Select Key (○
3) for ON/OFF.
2.2 ON: Turn on automatic background calibration to calibrate the sensor every 24- hour per day.
2.3 OFF: Turn off automatic background calibration.
Note: Automatic background calibration should be used in a well-ventilated environment. It is not recommended to use
the function in a greenhouse, wine cellar, narrow or poorly ventilate space, otherwise it will affect the accuracy.
3. Buzzer ON/OFF:
“ON“ is the default.
3.1 Press Mode Key(○
2) for more than 3 sec. to show “A on/A oFF”, then press Mode Key(○
2) to show “b on”.
3.2 Press Select Key (○
3) for ON/ OFF(b on/b oFF).
4. Auto-Switch Display (option):
“OFF“ is the default. If select “ON”, the CO2 / Relative Humidity / Temperature readings will be displayed in sequence.
4.1 Press Mode Key(○
2) for more than 3 sec. to show “A on/A oFF “, then press Mode Key(○
2) to show “b on”, press Mode Key(○
2)
again to show “S oFF ”.
4.2 Press Select Key (○
3) for ON/OFF(S on/S oFF).
4.3 When the LCD shows the ambient temperature and relative humidity readings, the LED with CO2concentration will be lit off
automatically.
5. Manually Display Temperature and Relative Humidity (option):
5.1 Short press Mode Key(○
2) to show relative humidity.
5.2 Then short press Mode Key(○
2) again to show ambient temperature.
6. IAQ Setting:
“OFF“ is the default, whose LED light is white. Please select the “Normal IAQ” or “Wide Range” setting.
6.1 Press Mode Key(○
2) for more than 3 sec. to show “A on/A oFF “, then press Mode Key(○
2) to show “b on”, press Mode Key(○
2)
again to show “S oFF”, and then press Mode Key(○
2) to show “I on/--”.
6.2 Press Select Key (○
3) for “I on” / “I --”.
6.3 I on : 800~1,400ppm for “Normal IAQ” setting with 800ppm(Green LED light)~1,400ppm(Red LED light).
6.4 I - - : 800~3,000ppm for “Wide Range” with 800ppm (Green LED light)~3,000ppm(Red LED light).
7. Switch°C/°F(option):
On main screen, short press Select Key (○
3) to show°C/°F.
8. Outside calibration and clear
8.1 Outside calibration
(1) Calibrate the device under 400ppm CO2concentration in outdoor environment.
(2) Press and hold Mode Key (○
2), then press Select Key (○
3) for more than 3 sec. to show “CAL”.
(3) Press Select Key (○
3) for 600 sec. to start the countdown for calibration, and then return back to the main screen automatically.