1. General Introduction
Indoor Air Quality is used to measure for the quality of air in interiors; the
comfortable indoor environmental quality can make people feel fresh, work
effectively and it’s good for the health. The CO2concentration is the important
factor of good indoor air quality. People breath in oxygen and breath out CO2,
nowadays people often close the windows to avoid noise and enjoy the
comfortable living and working environment provided by air-conditioning
systems, which results in the fact that the indoor concentration of CO2is far
higher than the outdoor average’s. With the high CO2concentration and non-
proper ventilation, people will feel headaches, dull, fatigued, lose of
concentration and correspond to the high levels of dust, chemicals and
bacteria in the air.
2. ZGkb201 ------ Multi IAQ Monitor (CO2 / Temperature / Humidity)
ZGkb201 Multi IAQ monitor is a cost-effective, air quality indicator, the large
single digit LED display with three most critical parameters of indoor air quality,
the current CO2concentration, the ambient temperature and the relative
humidity.
CO2-----CO2concentration indicates the amount of fresh air ventilation being
introduced to a space.
Ambient temperature----The suitable ambient temperature brings the
comfortable space.
Relative humidity----The relative humidity in air indicates the amount of
moisture.
Features:
NDIR (Non-Dispersive-Infrared) technology is being used to measure CO2
concentration up to 3000 ppm (parts per million)
The Large single digit LED clearly displays the CO2concentration,
Relative humidity and ambient temperature
The various (blue, yellow and red) colors of LED’s show the indoor air
quality situation
The Reliable CO2sensor provides the long-term calibration stability
Dimmer Function can automatically adjust the display of single digit LED
to reach the power saving
Benefits:
Enjoy the more comfortable, hygienic and healthier Indoor Air Quality
Save energy, get the proper ventilation, the over-ventilation of the building
can be reduced.
Improve the IAQ satisfaction in public.
Application:
ZGkb201 has the widely application for classroom, kindergarten, meeting room,
office building, shopping mall, theater, exhibition, hospital, hotel, sports center,
gymnasium, railway station, bus station, airport and other public place where
the comfortable and healthy indoor air quality is important.
3. LED Display
There are three modes: CO2, RH, Temperature measure mode.
CO2, RH, Temperature measure mode:
When CO2concentration is below
800ppm, the green LED will light.
When CO2 concentration is between
800ppm and 1200ppm, the yellow
LED will light.
When CO2 concentration exceeds
1200 ppm, the red LED will light.
When CO2concentration exceeds
the CO2measurement range
(0-3,000ppm), the CO2display
area will display with “Hi”. It will
disappear when the CO2
concentration returns to the
measurement range.
When RH exceeds the RH
measurement range 20%, the
humidity display area will display
with “Lo”. It will disappear when the
RH returns to the measurement
range.
When RH exceeds the RH
measurement range 90%, the
humidity display area will display
with “Hi”. It will disappear when the
RH returns to the measurement
range.
Ref.No.:122011
4. SPECIFICATIONS
Method - NDIR
Display –Single Digit LED
Sample Method - Diffusion or flow through (50 ~200 ml/min)
CO2& Temperature & RH Specification:
CO2
Measurement Range
0-3,000 ppm display
Display Resolution
1ppm at 0~1,000ppm;
5ppm at 1,001~2,000ppm;
10ppm at 2,001~3,000ppm
Accuracy
0~2,000ppm: ±70 ppm or ±5%
of reading, whichever is
greater; over 2000ppm: +/-7%
Repeatability
±20 ppm
Temperature
Dependence
Typ.±0.2% of reading per °C
or ±2 ppm per °C, whichever is
greater, referenced to 25°C
Pressure Dependence
0.13% of reading per mm Hg
Response Time
About 2min for 63% of step
change
Warm-Up Time
<60 seconds at 22°C
Temperature
Temperature Range
0°C ~40°C display
Display Resolution
0.1°C
Display Unit
°C
Accuracy
±1°C
Response Time
20-30 minutes (case must
equalize with environment)
RH
Measurement Range
20%-90% RH
Display Resolution:
1%RH
Accuracy
±5%RH@23°C
Response time
<5 min for 63% of step change
Operating &
Storage
condition
Operating
Temperature
0°C~ 40°C
Humidity Range
0 ~ 95% RH non-condensing
Storage Temperature
-20°C~ 60°C
Power
Supply
Power Source
Universal 100~240VAC,
50/60Hz , External Power Brick
(included in package)
Power Consumption
7 Watts (Max.)
5. Installation Instruction
Please take out the ZGkb201 and AC adaptor from the package carefully.
There have two options to install ZGkb201.
Installation Option
1. Hang up the ZGkb201 with wire rope.
2 .Choose a suitable location to install ZGkb201, fix the screw(diameter of
wallboard nail is bigger than 4mm)to the wall.
6. Weight & Dimensions
Weight: 2.4 kg
Dimension:
EMC/RFI
Readings may be affected if the unit is operated within radio frequency
electromagnetic field strength of approximately 3 volts per meter, but the
performance of the instrument will not be permanently affected.
Radiant Innovation Inc. Http://www.ZyAura.com
1F, No.3, Industrial East 9th Road, Science-Based
Industrial Park, HsinChu, Taiwan 300.
Inch(mm)

Other ZyAura Measuring Instrument manuals

ZyAura ZG106 User manual

ZyAura

ZyAura ZG106 User manual

ZyAura ZGkb201p User manual

ZyAura

ZyAura ZGkb201p User manual

ZyAura ZG1683R User manual

ZyAura

ZyAura ZG1683R User manual

ZyAura ZGW01 User manual

ZyAura

ZyAura ZGW01 User manual

ZyAura ZGw08VRC User manual

ZyAura

ZyAura ZGw08VRC User manual

ZyAura ZGW063RY User manual

ZyAura

ZyAura ZGW063RY User manual

ZyAura ZGm27 User manual

ZyAura

ZyAura ZGm27 User manual

ZyAura ZG1683RU User manual

ZyAura

ZyAura ZG1683RU User manual

ZyAura ZGw08PR User manual

ZyAura

ZyAura ZGw08PR User manual

Popular Measuring Instrument manuals by other brands

PCB Piezotronics 3503A1120KG Installation and operating manual

PCB Piezotronics

PCB Piezotronics 3503A1120KG Installation and operating manual

Bender ISOMETER iso685-D-P manual

Bender

Bender ISOMETER iso685-D-P manual

Access Laser L15 Series Operator's manual

Access Laser

Access Laser L15 Series Operator's manual

YOKOGAWA AR-SA user manual

YOKOGAWA

YOKOGAWA AR-SA user manual

Laser 2000 ADM-1000 operating instructions

Laser 2000

Laser 2000 ADM-1000 operating instructions

Bushnell NEO GHOST user manual

Bushnell

Bushnell NEO GHOST user manual

Fluke CableAnalyzer DSX-5000 Getting started guide

Fluke

Fluke CableAnalyzer DSX-5000 Getting started guide

Apera Instruments PC60 instruction manual

Apera Instruments

Apera Instruments PC60 instruction manual

Dahua Modular VTO user manual

Dahua

Dahua Modular VTO user manual

Retrotec DM32 Operation manual

Retrotec

Retrotec DM32 Operation manual

Victor DM3218 user manual

Victor

Victor DM3218 user manual

SATO SK-110TRH Calibration guide

SATO

SATO SK-110TRH Calibration guide

socomec COUNTIS AT operating instructions

socomec

socomec COUNTIS AT operating instructions

BRONKHORST MASS-STREAM D-6400 instruction manual

BRONKHORST

BRONKHORST MASS-STREAM D-6400 instruction manual

Pitney Bowes DM300c series instructions

Pitney Bowes

Pitney Bowes DM300c series instructions

GSi Protek A333 operating manual

GSi

GSi Protek A333 operating manual

Sensit HXG-3 Quick start instructions

Sensit

Sensit HXG-3 Quick start instructions

Keysight Technologies 86120B user guide

Keysight Technologies

Keysight Technologies 86120B user guide

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.