3M Bair Hugger 775 User manual

To t a l Te m p e r a t u r e M a n a g e m e n t™ S y s t e m
3
Bair Hugger™
Temperature Management Unit
Model 775
Operator’s Manual
1 Temperature management unit
Model 775 Operator's manual
2 Unité de contrôle de température,
Modèle 775 Manuel de l’utilisateur
3 Temperaturmanagementgerät
Modell 775 Benutzerhandbuch
4 Unità per la gestione della
temperatura Modello 775
Manuale d’uso
5 Unidad de control de temperatura,
Modelo 775
Manual del operador
6 Model 775 temperatuurregelunit
Gebruikershandleiding
7 Temperaturregleringsenhet,
Modell 775 Bruksanvisning
8 Temperaturstyringsenhed Model
775 Brugervejledning
9 Temeraturadministrasjonsenhet
Modell 775 Brukerhåndbok
0 Lämpötilan valvontayksikkö edestä
Malli 775 Käyttöopas
- Unidade de Gestão
de Temperatura Modelo 775
Manual do Utilizador
=
Μονάδα Διαχείρισης
Θερμοκρασίας Μοντέλο 775
Εγχειρίδιο Χρήσης
H Dmuchawa Model 775 Instrukcja
obsługi
J Hőmérséklet szabályozó egység
775-ös Modell Használati utasítás
F Termoregulační jednotka, Model
775 Návod k obsluze
Q Temperatūros kontrolės įrenginys,
Modelis 775 Naudotojo vadovas
M
Устройство управления
температурой, модель 775
Руководство по эксплуатации
: Çalıştırma El Kitabı
Vücut Isısı Yönetim Ünitesi
Model 775
E ベアーハガー 体温管理装置
モデル775 取扱説明書
I 775 型温度管理仪 操作员手
册
U


English 1
Français 15
Deutsch 31
Italiano 47
Español 63
Nederlands 79
Svenska 95
Dansk 111
Norsk 127
Suomi 143
Português 159
Ελληνικά
175
Polski 191
Magyar 207
Česky 223
Lietuvių Kalba 239
Русский
255
Turkish 271
日本語
287
英语
303
319
3
Bair Hugger
Temperature Management Unit
Model 775
Operator’s Manual


Table of Contents
Introduction ��������������������������������������������������������������� 1
Description of the Total Temperature ManagementTM System ������������� 1
Indications ������������������������������������������������������������ 1
Contraindications ����������������������������������������������������� 1
Warnings ������������������������������������������������������������� 1
Cautions �������������������������������������������������������������� 2
Notices ���������������������������������������������������������������� 3
Proper Use and Maintenance ������������������������������������������� 3
Read Before Servicing Equipment ��������������������������������������� 3
Overview and Operation ���������������������������������������������������� 4
Unit Power-On-Reset �������������������������������������������������� 4
Overview of Control Panel ��������������������������������������������� 5
Mounting the Temperature Management Unit to an IV Pole �������������� 7
Instructions for Use ���������������������������������������������������� 8
Viewing the Temperature Mode Timer ���������������������������������� 9
What to Do in Case of an Over-Temperature Condition ������������������ 9
What to Do in Case of a Fault Condition �������������������������������� 9
General Maintenance ������������������������������������������������������ 10
Cleaning the Cabinet and Hose ���������������������������������������� 10
Technical Service and Order Placement ������������������������������������� 11
Technical Service ����������������������������������������������������� 11
When You Call for Technical Support ���������������������������������� 11
In-Warranty Repair and Exchange ������������������������������������� 11
Specications ������������������������������������������������������������� 12
Denition of Symbols ������������������������������������������������������ 14


1
Bair Hugger Model 775 Temperature Management Unit Operator’s Manual
English 771022D
Introduction
Description of the Total Temperature Management™ System
The 3MTM Bair HuggerTM brand Total Temperature Management system consists of a
Model 775 forced-air temperature management unit (with optional rolling stand, and other
accessories) and disposable components, including Bair Hugger forced-air blankets, 3MTM
Bair PawsTM patient warming gowns and the 3MTM 241TM blood/fluid warming set. You
can use the Model 775 temperature management unit in all clinical settings where the
patient may become too warm or too cold; including the operating room to provide patient
temperature management.
This manual includes operating instructions and unit specifications for the Model 775
temperature management unit. For information about using Bair Hugger blankets, Bair Paws
gowns, or the 241 blood/fluid warming set with the Model 775 temperature management
unit, refer to the “Instructions for Use” included with each of these disposable components.
Indications
The Bair Hugger Total Temperature Management system is intended to prevent and treat
hypothermia. In addition, the temperature management system can be used to provide patient
thermal comfort when conditions exist that may cause patients to become too warm or too
cold. The temperature management system can be used with adult and pediatric patients.
Contraindications
Do not apply heat to lower extremities during aortic cross-clamping. Thermal injury may
occur if heat is applied to ischemic limbs.
Warnings
1. Do not leave patients with poor perfusion unmonitored during prolonged warming
therapy sessions. Thermal injury may result.
2. The Bair Hugger temperature management unit has been designed to operate safely
ONLY with Arizant HealthcareTM disposable components. Use with other products
may cause thermal injury. (To the full extent permitted by law, the manufacturer
and/or importer declines all responsibility for thermal injury resulting from the unit
being used in conjunction with non-Arizant Healthcare products.)
3. Do not warm patients with the temperature management unit hose alone.
Thermal injury may result. Always connect the hose to a Bair Hugger
blanket or Bair Paws gown before providing therapy.
4. Do not place the non-perforated side of the blanket on the patient. Thermal injury may
result. Always place the perforated side (the side with small holes) towards the patient.

Bair Hugger Model 775 Temperature Management Unit Operator’s Manual
2
Arizant Healthcare Inc.
5. Do not continue temperature management therapy if the Over-temp indicator light
illuminates and the alarm sounds. Thermal injury may result. Unplug the unit, and
contact a qualified service technician.
6. Do not continue 241 blood/fluid warming therapy if the Over-temp indicator light
illuminates and the alarm sounds. Immediately stop fluid flow, and discard the
blood/fluid warming set. Unplug the temperature management unit, and contact a
qualified service technician.
7. Do not use a forced-air warming device over transdermal medication. Increased drug
delivery and patient injury or death may occur.
8. Do not allow the patient to lie on the warming unit hose or allow the hose to directly
contact the patient’s skin during patient warming; thermal injury may result.
9. Equipment not suitable for use in the presence of a flammable anesthetic mixture
with air or with oxygen or nitrous oxide.
10. Reusable blankets made from woven fabric, or blankets without discrete, visible
holes, can cause the safety system of this unit to fail, which may result in serious
thermal injury. This warming unit has been designed to operate safety ONLY with
Bair Hugger blankets and Bair Paws gowns.
Cautions
1. Except for specific blanket models, Bair Hugger blankets are not sterile and are all
intended for single patient use ONLY. Placing a sheet between the Bair Hugger
blanket and the patient does not prevent contamination of this product.
2. Monitor the temperature and cutaneous response of patients who are incapable of
reacting, communicating and/or who are without a sense of feeling every 10-20
minutes or according to institutional protocol. Monitor the patient’s vital signs
regularly. Adjust air temperature or discontinue therapy when the therapeutic goal
is reached or if vital sign instability occurs. Notify physician of vital sign instability
immediately.
3. Do not leave pediatric patients unattended during therapy.
4. Do not initiate temperature management therapy unless the temperature management
unit is safely placed on a hard surface or securely mounted. Otherwise, injury may
result.
5. To prevent tipping, clamp the Model 775 temperature management unit to an IV pole
at a height that provides stability. We recommend clamping the unit no higher than
44" (112 cm) on an IV pole with a minimum 28" (71 cm) diameter wheelbase. Failure
to do so may result in IV pole tipping, catheter site trauma, and patient injury.

3
Bair Hugger Model 775 Temperature Management Unit Operator’s Manual
English 771022D
6. Electrical shock hazard. Do not disassemble the temperature management unit unless
you are a qualified service technician. There are electrically live parts within the unit
when it is connected to a power source, even when the unit is in Standby mode.
Notices
1. The Bair Hugger temperature management unit meets medical electronic interference
requirements. If radio frequency interference with other equipment should occur,
connect the unit to a different power source.
2. Federal law (USA) restricts this device to sale by or on the order of a licensed
healthcare professional.
3. To reliably ground this Bair Hugger warming unit, only connect to receptacles
marked “Hospital Only” or “Hospital Grade.”
Proper Use and Maintenance
Arizant Healthcare assumes no responsibility for the reliability, performance, or safety of
the temperature management unit if the following events occur:
• Modications or repairs are performed by unqualied personnel.
• e unit is used in a manner other than that described in the Operator’s or Service
Manual.
• e unit is installed in an environment that does not meet the appropriate electrical
and grounding requirements.
Read Before Servicing Equipment
All repair, calibration and servicing of the temperature management unit require the skill
of a qualified, medical equipment service technician who is familiar with good practice for
medical device repair. If service does not require the manufacturer’s attention, the Model
775 Service Manual provides the technical information needed to service the unit. Perform
all repairs and maintenance in accordance with the instructions in the Service Manual. For
additional service information please contact Arizant Healthcare Inc.

Bair Hugger Model 775 Temperature Management Unit Operator’s Manual
4
Arizant Healthcare Inc.
Overview and Operation
Unit Power-On-Reset
The Model 775 temperature management unit automatically performs the following
power-on-reset sequence after you connect the unit to a properly grounded power source:
• Performs all self-test functions.
• Illuminates indicator lights and all pixels in the alphanumeric display momentarily.
• Displays the text “MD 700’s” and the software revision level in the alphanumeric
display.
• Sounds the alarm (three low-level clicks).
• Enters the Standby mode with the high fan setting selected by default.
Note: e low fan setting indicator light will not illuminate during power up.
If the temperature management unit looses power for less than 1 second, the unit’s software
will return the unit to the operating mode you selected prior to the power loss. However, if
the temperature management unit loses power for longer than 1 second, the unit’s software
will reset when you restore power. The unit will then enter the Standby mode.
Figure A. Model 775 temperature
management unit – front view
Figure B. Model 775 temperature
management unit – rear view
Carrying handle
Alphanumeric
display
Control Panel Hose connection
Power cord
connection
Hose
IV pole clamp

5
Bair Hugger Model 775 Temperature Management Unit Operator’s Manual
English 771022D
Overview of Control Panel
Figure C. Model 775 temperature management unit control panel
Model 775
43˚C
STANDBY
TEMP IN
RANGE
FAULT
OVER
TEMP
AMBIENT
H
I
G
H
L
O
W
M
E
D
High Fan
Setting
Alphanumeric Display
Low Fan
Setting
Temperature Modes
alphanumeric display
The alphanumeric display shows the temperature at the Bair Hugger blanket or Bair Paws
gown end of the temperature management unit hose in degrees Celsius.
temperature modes
• Press the 32˚C (Low), 38˚C (Med), or 43˚C (High) button to select the desired
temperature.
• Press the Ambient button to supply room temperature air.
When you select a temperature mode, the following events occur:
• Corresponding temperature indicator and fan setting lights illuminate.
• Blower operates at selected fan speed.
• Fan setting indicator light illuminates.
• Temperature at the blanket or gown end of the hose appears in the alphanumeric display.
• Heater activates except in Ambient mode.
• Temperature mode timer activates (or resets itself if changing from one temperature
mode to another).
airflow modes
The Model 775 has two preset fan speed settings: a default or high fan setting ( )and a
reduced or low fan setting ( ). The Model 775 will reset/default to the high fan setting
each time the unit is powered up or Standby is selected. The low fan setting ( ) may be
pre-selected while in Standby prior to selecting the desired temperature.

Bair Hugger Model 775 Temperature Management Unit Operator’s Manual
6
Arizant Healthcare Inc.
standby
Upon power up, the warming unit defaults to Standby and the high fan setting. When the
unit is in 32˚C (Low), 38˚C (Med), or 43˚C (High) or Ambient mode, press the Standby
button to place the unit in Standby mode.
When you select the Standby mode, the following events occur:
• Standby indicator light illuminates.
• Blower and heater are turned OFF.
• Alphanumeric display deactivates.
• Alarm and fault detection functions remain active.
• Operating timer pauses.
• Airow mode resets/defaults to the high fan setting.
temp in range indicator light
The Temp in Range indicator light illuminates when the temperature at the blanket
or gown end of the hose is ±1.5°C of the selected setting; this indicator light does not
illuminate in the Ambient mode.
fault indicator light
When a system fault occurs, the amber Fault indicator light flashes and an alarm
sounds.
Please refer to What to Do in Case of a Fault Condition on page 9 for additional
information.
over-temp indicator light
If the unit senses an over-temperature condition, the red Over-temp indicator light
flashes and an alarm sounds.
Please refer to What to Do in Case of an Over-Temperature Condition on page 9 for
additional information.

7
Bair Hugger Model 775 Temperature Management Unit Operator’s Manual
English 771022D
Mounting the Temperature Management Unit to an IV Pole
caution
To prevent tipping, clamp the Model 775 temperature management unit to an IV pole
at a height that provides stability. We recommend clamping the unit no higher than 44"
(112 cm) from the floor on an IV pole with a minimum 28" (71 cm) diameter wheelbase.
Failure to do so may result in IV pole tipping, catheter site trauma and patient injury.
method
To mount the Model 775 temperature management unit to an IV pole, simply turn the
clamp handle clockwise to tighten the clamp on the IV pole and counterclockwise to
release it. Ensure the IV pole bumper rests against the pole. See Figure D.
28 in. (71 cm) minimum
44 in.
(112 cm)
maximum
Figure D. Model 775 on an IV Pole
IV Pole
Bumper
Clamp
Handle

Bair Hugger Model 775 Temperature Management Unit Operator’s Manual
8
Arizant Healthcare Inc.
Instructions for Use
The instructions below describe how to operate the Model 775 temperature management
unit. For information about using Bair Hugger blankets, Bair Paws gowns, or the 241
blood/fluid warming set with the Model 775 unit, refer to the Instructions for Use
included with each of these disposable components.
1. If the Model 775 temperature management unit is not mounted on an IV pole or
Bair Hugger rolling stand, place the unit on a flat, hard, dry surface, such as a table,
before beginning temperature management therapy. Do not place the unit on a soft,
uneven or visibly wet surface such as a bed, or the air intake may become blocked and
cause the unit to overheat.
2. Insert the end of the temperature
management unit hose into the hose port
of a Bair Hugger blanket or Bair Paws
gown. Use a twisting motion to ensure
a snug fit. A visual marker (Figure E) is
located around the mid section of the hose
end to guide the depth of hose insertion.
3. Connect the temperature management unit to a properly grounded power source.
The unit will be in Standby mode and the Standby indicator light will illuminate.
The high fan setting ( ) will be pre-selected by default with its indicator light
illuminated. The low fan setting ( ) may be pre-selected prior to selecting the
desired temperature mode.
4. Press the appropriate button (i.e., 32˚C, 38˚C, 43˚C, or Ambient) to select the desired
temperature. This will turn the blower and heater ON. When the unit reaches the
selected temperature, the Temp in Range indicator light will illuminate; this indicator
light does not illuminate in the Ambient mode.
5. Monitor the temperature and cutaneous response of patients who are incapable of
reacting, communicating and/or who are without a sense of feeling every 10-20
minutes or according to institutional protocol. Monitor the patient's vital signs
regularly. Adjust air temperature or discontinue therapy when the therapeutic goal
is reached or if vital sign instability occurs. Notify physician of vital sign instability
immediately.
6. Press the Standby button when temperature management therapy is complete and
discard disposable components.
7. Unplug the temperature management unit to disconnect it from the power source.
Note: At any time, the fan speed setting can be switched between high fan setting ( )
and low fan setting ( ) by pressing the corresponding button. However, these buttons
will not place the unit into Standby.
Hose port
Hose
Insert marker
Figure E.

9
Bair Hugger Model 775 Temperature Management Unit Operator’s Manual
English 771022D
Viewing the Temperature Mode Timer
The temperature mode timer records the amount of time the temperature management
unit has been operating in the selected temperature mode. The timer resets itself every
time you select a different temperature mode.
To view the temperature mode timer, press the button and hold for 2 seconds for the
current temperature mode. The alphanumeric display will show the temperature mode
time, then go back to showing the temperature mode setting.
What to Do in Case of an Over-Temperature Condition
If an over-temperature condition occurs, the red Over-temp indicator light flashes and
an alarm sounds. The unit automatically turns the heater, blower, and operating mode
indicator lights OFF. The control panel will not respond to commands.
Note: Pressing any button will silence the alarm.
if an over-temperature condition occurs:
1. Discontinue all temperature management therapy. If you are using the 241
blood/fluid warming system, immediately stop fluid flow and discard the
blood/fluid warming set.
2. Unplug the temperature management unit.
3. Contact a qualified service technician.
What to Do in Case of a Fault Condition
The Model 775 temperature management unit’s software recognizes several non-hazardous
conditions and reports those conditions as faults. When a system fault occurs, the software
stores the fault code in memory, the amber Fault indicator light flashes, and an alarm
sounds. The control panel will not respond to commands.
Note: Pressing any button will silence the alarm.
if a fault condition occurs:
1. Unplug the temperature management unit and wait for 5 minutes.
2. Reconnect the temperature management unit to a grounded power source.
The unit will perform the normal power-on-reset sequence and then enter the
Standby mode.
3. Reselect the temperature setting.
4. If the unit does not return to normal operation, contact a qualified service
technician.

Bair Hugger Model 775 Temperature Management Unit Operator’s Manual
10
Arizant Healthcare Inc.
General Maintenance
Cleaning the Cabinet and Hose
warning
Do not immerse the cabinet or hose while cleaning. Moisture will damage the components,
and thermal injury may result.
precautions
• Do not use a dripping wet cloth to clean the cabinet. Moisture may seep into the
electrical contacts and damage the components.
• Do not use alcohol or other solvents to clean the cabinet. Solvents may damage the
labels and other plastic parts.
method
1. Disconnect the temperature management unit from the power source before cleaning.
2. Wipe the cabinet and the outside of the hose with a damp, soft cloth and a mild
detergent or antimicrobial spray.
3. Dry with a separate soft cloth.

11
Bair Hugger Model 775 Temperature Management Unit Operator’s Manual
English 771022D
Technical Service and Order Placement
Technical Service
usa germany
TEL: +1-952-947-1200 TEL: +49-4154-9934-0
+1-800-733-7775 FAX: +49-4154-9934-20
FAX: +1-952-947-1400
united kingdom france
TEL: +44-1924-200-550 TEL: +33-4-42-98-14-00
FAX: +44-1924-200-518 Indigo (in France only): 0-820-33-90-40
FAX: +33-4-42-98-14-01
order placement
usa
TEL: +1-952-947-1200
+1-800-733-7775
FAX: +1-952-947-1400
+1-800-775-0002
When You Call for Technical Support
We will need to know the serial number of your Bair Hugger temperature management
unit when you call us. The serial number label is located on the back of the
Bair Hugger temperature management unit.
In-Warranty Repair and Exchange
usa
Call Arizant Healthcare customer service if your Model 775 temperature management
unit requires factory service. A customer service representative will give you a Return
Authorization (RA) number. Please use this RA number on all correspondence concerning
your temperature management unit. Your customer service representative will also
send a shipping carton to you at no charge, if needed. Call your local supplier or sales
representative to inquire about borrowing a temperature management unit while we service
your unit. For more detailed instructions on returning units for service, please refer to the
Model 775 service manual.
outside of the usa
Contact your local distributor concerning in-warranty repair and exchange.

Bair Hugger Model 775 Temperature Management Unit Operator’s Manual
12
Arizant Healthcare Inc.
Specifications
Physical Characteristics
13" high x 13" deep x 14" wide
33 cm high x 33 cm deep x 36 cm wide
16 lb (7.3kg)
53 dBA (High fan setting)
48 dBA (Low fan setting)
Flexible, compatible with the Bair Hugger brand 241
blood/fluid warming system.
High efficiency 0.2 µm air filter
Every 12 months or 500 hours of use.
Can be clamped to an IV pole, placed on a hard surface,
or mounted to the rolling stand accessory.
Temperature Characteristics
15°C-25°C
Electronically controlled.
High fan setting: 1600 BTU/hr (average), 470 W (average)
Low fan setting: 1330 BTU/hr (average), 390 W (average)
Average temperatures at the end of the hose:
HIGH: 43° ± 1.5°C 109.4° ± 2.7°F
MED: 38° ± 1.5°C 100.4° ± 2.7°F
LOW: 32° ± 1.5°C 89.6° ± 2.7°F
Safety System
Independent electronic circuit; thermal cutoff shuts the
heater OFF to ensure hose end air remains below 56°C
(53°C ± 3°C typical); back-up over-temperature detection
at hose inlet.
Over-temperature (≤56°C, 53°C ± 3°C typical): red Over-
temp indicator light flashes, alarm sounds, heater and
blower shut down, operating indicator lights turn OFF,
control panel becomes unresponsive.
Fault: amber Fault indicator light flashes, alarm sounds.
Dual input fused lines.

13
Bair Hugger Model 775 Temperature Management Unit Operator’s Manual
English 771022D
Electrical Characteristics
1400 W Resistive
Meets UL 60601-1 and IEC 60601-1 requirements.
Operating speed: 4,700 rpm (high fan setting),
4,100 rpm (low fan setting)
Airflow: up to 48 cfm or 23 L/s
Peak: 1550 W
Average: 800 W
15 ft., SJT, 3 cond., 13 A
15 ft., SJT, 3 cond., 15 A
4.6 m, HAR, 3 cond., 10 A
110-120 VAC, 50/60 Hz, 11.7 A, or
220-240 VAC, 50/60 Hz, 7.2 A, or
100 VAC, 50/60 Hz, 15 A
12 A (110 - 120 VAC)
8 A (220 - 240 VAC)
15 A (100 VAC)
IEC 60601-1; EN 60601-1-2; UL 60601-1;CAN/CSA-
C22.2, No. 601.1, EN 55011; ASTM F2196-02
Classified under IEC 60601-1 Guidelines (and other
national versions of the Guidelines) as Class I, Type
BF, Ordinary equipment, Continuous operation.
Not suitable for use in the presence of flammable
anesthetic mixtures with air or with oxygen or nitrous
oxide. Classified by Underwriters Laboratories Inc.
with respect to electric shock, fire, and mechanical
hazards only, in accordance with UL 60601-1, ASTM
F2196-02, and Canadian/CSA C22.2, No. 601.1.
Classified under the Medical Device Directive as a
Class IIb device.
A qualified service technician can perform over-
temperature detection system testing, temperature
output testing, operating temperature calibration, and
fault code troubleshooting.

Bair Hugger Model 775 Temperature Management Unit Operator’s Manual
14
Arizant Healthcare Inc.
Definition of Symbols
The following symbols may appear on the product’s labeling or exterior packaging.
Equipotentiality plug (ground)
Fuse
!
Attention, consult appropriate documents
Nonexplosion proof
Dangerous voltage
Type BF equipment (patient applied)
VAC
V
Voltage, alternating current (AC)
Special refuse, discard separately
Protective earth ground
No free hosing
High fan setting
Low fan setting
Other manuals for Bair Hugger 775
3
Table of contents
Other 3M Temperature Controllers manuals
Popular Temperature Controllers manuals by other brands

Neomitis
Neomitis SR Instructions for installing and using

Linear Technology
Linear Technology DC1792A manual

Viessmann
Viessmann VITOTROL 100 operating instructions

ELKATHERM
ELKATHERM FTRFB-280 Series manual

Arroyo Instruments
Arroyo Instruments TECPak 585 Series user manual

Excelair
Excelair Dual Room Damper Package Installation

Elko
Elko RFTC-50/G/ice manual

Wavelength Electronics
Wavelength Electronics WTC3293 manual

Lumel
Lumel RE55 user manual

Greystone Energy Systems
Greystone Energy Systems TSAP Series installation instructions

Atmel
Atmel AT30TK175 Hardware user's guide

REGULA
REGULA grandis 650 HK Installation and operating instructions