• Reuse of the infusion set may cause site irritation or infection and may lead
to inaccurate medication delivery.
• Be sure that the needle cover is removed before insertion is made.
W PRECAUTIONS
• Never try to fill or free a clogged tubing while the infusion set is inserted.
This may result in unpredictable medication flow.
• Do not leave air in the infusion set. Make sure to fill the tubing completely.
Consult pump manufacturer's instructions for use.
• Accu-Chek TenderLink infusion set is a single use item, which must be
disposed of after usage. Do not clean or re-sterilise.
• Check the infusion set frequently to ensure that the soft cannula remains
firmly in place. Replace with a new infusion set at a new site, if the infusion
set is not in place. Since the cannula is soft, it will not cause any pain if
it slips out, and this may happen without notice. The soft cannula must
always be completely inserted to receive the full amount of medication.
• Do not in any way put disinfectants, perfumes, deodorants or other products
containing disinfectants in contact with the luer lock connector or the
tubing. This may impact the integrity of the infusion set.