ADS Beyond 300 User manual

http://www.adsequip.com/
User's Manual of
Beyond 300/301 Delivery System
P/N:8023488 Version:A Date:2017-08-01

User's Manual of Beyond 300/301 Delivery System
Copyrights
ADS makes no warranty of any kind with regard to this material ,
i n c l u ding but not limite d t o , t h e i m p l i e d w a r r a n t i e s o f
merchantability and fitness for a particular purpose. ADS shall not
be held liable for any errors contained herein or any consequential or
other damages concerning furnishing, performance or use of this
material. The information in this document is subject to change
without notice . If you find any problems in the documentation,
please report them to us in writing. ADS does not warrant that this
document is error-free.
No part of this document may be copied, reproduced, altered, or
transmitted in any form or by any means, electronic or mechanical,
including photocopying, recording, or by any information storage
and retrieval system, without prior written permission from ADS.
The document is intended to be used in conjunction with the
equipment.
Comments and feedback
Please contact us if you have any feedback and opinions on this
document:
Add: 12240 Eastend Ave,Chino,CA 91710
Tel: (626) 620 0456
E-mail : sales@adsequip.com
http://www.adsequip.com/
Product service
If you want to get service information, please contact the local
dealer authorized by ADS company.
Technical support tel: 800 488 9708

1 Introduction………………………………………………………… 1
1.1 Products information……………………………………………1
1.2 Symbols explanation……………………………………………1
2 Specifications………………………………………………………… 3
3 Application notice…………………………………………………… 3
3.1 Attent i o n s … …………………… … … … ………… … … 3
3.2 Warning information…………………………………………… 5
4 Sizes…………………………………………………………………6
5 Operations……………………………………………………………8
5.1 Water supply system……………………………………………8
5.2 Assistant's control………………………………………………9
5.3 Cuspidor (optional)……………………………………………10
5.4 Dentist unit……………………………………………………11
5.4.1 Level arm adjustments…………………………………………12
5.5 Control panel…………………………………………………13
5.5.1 System state……………………………………………………13
5.5.2 Chair position buttons…………………………………………14
5.5.3 Initial settings of the chair……………………………………14
5.5.4 Initial settings description……………………………………15
5.5.5 Chair positions description……………………………………16
5.6 Foot control…………………………………………………17
5.7 Junction box……………………………………………………18
6 Functions introduction……………………………………………19
6.1 Valves layout of delivery head…………………………………19
6.2 Functions introduction…………………………………………19
6.3 3-Way syringe adjustment……………………………………22
7 Maintenance and sterilization instructions………………………23
7.1 Handpiece tubing cleaning & maintenance……………………23
7.2 Clean HVE & SE valves………………………………………23
7.3 Clean dirty collector……………………………………………24
7.4 Clean the filter…………………………………………………24
8 Troubleshooting……………………………………………………24
User's Manual of Beyond 300/301 Delivery System
CONTENTS

1 P/N:8023488
Unit conversion
MPa
psi
0.2MPa
29psi
0.4MPa
58psi
0.6MPa
87psi
0.8MPa
116psi
User's Manual of Beyond 300/301 Delivery System
1 Introduction
1.1 Products information
•Name:Delivery system
•Model:Beyond 300/301
•Working conditions:
①Ambient temperature:5 ℃~40 ℃;
② Relative humidity:≤ 80%;
③ Nominal Voltage :AC110 V±22 V;
④ Nominal Frequency:50 Hz±1 Hz;
⑤ Water pressure:29psi~58psi;
⑥ Input air pressure :72.5psi~101.5psi, Air flow ≥50 L/min.
1.2 Symbols explanation
Symbols
Explanations
Manufacturer
EU Representative
Manufacturing date
Serial Number
Handle with Care
Keep dry
Table 1

2 P/N:8023488
User's Manual of Beyond 300/301 Delivery System
Symbols
Explanations
Refer to instructions
Notice
Refer to attatched document
Backrest Down
Avoid Electric Shock
Backrest Up
Ground protection
Chair Down
Setup Button
Chair Up
Spitting position
Program Chair Position 1
Program Chair Position 2
Program Chair Position 3
P

3 P/N:8023488
User's Manual of Beyond 300/301 Delivery System
Light illuminance
8000-40000 Lux
Max load of delivery head
2.5kg
Ambient temperature
5 ℃~40 ℃
Air supply pressure
72.5psi~101.5psi
Water supply pressure
29~58psi
2 Specifications
Table 2
3 Application notice
3.1 Attentions
Before operating the unit, please read the manual carefully
•Input voltage : AC110V,60Hz.
•Make sure the unit is in good grounding .
•The unit needs professional maintenance..
• All the electrical components and connections should be made by
an authorized technician.
•All power and water,air supply must be cut off after work finished
every day.
•The equipment should not be operated in the environment of high
temperature, high humidity, vibration, flammability, unbalance, or
dusty.
•No object can be touched in the unit’s working range.
•Do not touch the bulb and the refector of the dental light when the
equipment is working.
•This equipment can not be used in the environment of strong
magnetic, electromagnetic.

4 P/N:8023488
• The equipment should not be used for purposes other than intended use .
• The water source of the system should be purified tap water, the water used
shall conform to the local water standards, water hardness ≤ 450 mg/L .
• The compressed air used for the system requires no water, no oil, and no
sanitary defects. The air used in dentistry shall comply with local standards.
• Sterilize the water tubings with once a day with standard disinfectant. Water
bottles are only available for use with distilled water.
•Exhaust water in the air regulator in time.
•Clean or replace water filter core in time.
•The filter should be cleaned every week .
•The power must be cut off when replacing electrical components.
• The power must be cut off before maintenance and cleaning.
• If you want to install functional accessories in the machine, such as scaler,
curing light or other accessories ,they must be high quality products that passed
the safety certification .
• Transportation, storage environment requirements :
a. Ambient temperature range :-40 ~ +70°C,
b. Relative humidity range :20%~90%, Including condensation ,
c. Air pressure range :12.5~15psi.
•Protect against rain and vibration in transportation, handle with care.
•Packaged equipment should be stored indoor with less than 80% relative
humidity , non-corrosive gases, and in good ventilation.
• The waste water, waste disposal must meet the local environmental protection
requirements.
• The system should be operated by professional dentist and people with
training, or under the guidance of professional dentist. The unauthorized, the
untrained,and non-professionals are not free to carry out any operations.
• Equipment maintenance must be done by the company's professional
technicians, unauthorized dismantling and maintenance may cause fatal harm
to the machine, and no longer enjoy the company's warranty service.
User's Manual of Beyond 300/301 Delivery System

5 P/N:8023488
User's Manual of Beyond 300/301 Delivery System
3.2 Warning Information
•You should read and understand the contents of the instructions carefully
before operating the equipment.
•Operation should follow all warnings and cautions on the instrument .
•The power socket cable should be in accordance with the standard
configuration and the ground wire must be secure .
•Delivery head shall not be stacked with too heavy items (max load 2.5kg) .
•Air pressure of the system has been set before leaving factory, non-
professionals are not free to adjust.
•Turn off the main switch of water, gas, power when get off work every day.
•The handpiece shall supply water only when the foot switch is in the state
of water supply. (only for 4-pin handpiece tube)
•Hospitals or individual clinics who use the equipment must be equipped
with or connected to an amalgam separation device in the waste discharge
system .
•The water and air must be exhausted and then cut off the water source and
electricity when maintaning or repairing; switch on the power, water and
air before the use.
•The 3-way syringe handle and syringe tip should be put into the disinfection
bag to disinfect, and then 134 ℃ (pressure 29.8psi) high pressure steam
disinfection, disinfection time is no less than 3 minutes.
•Please read the user’s guide before cleaning and disinfecting the high speed
turbine and pneumatic micromotor.
•The cover clean with wet cloth and neutral detergent to ensure smooth and
elastic leather surface free from corrosion; ABS parts clean with a soft cloth
dampened with soap water; the metal baking parts wash with soft cloth
dipped in soapy water or wipe with a wax spray.
•Take an extra care with a fiber-optic handpiece to avoid damaging the
shining tip and make sure the tip will not come into contact with the treated
mixture, keep a certain distance or use a transparent sheet before the first 5
seconds of treatment.
•Any traces left by the mixture on the tool must be removed immediately,
remove the fiber-optic handpiece and clean it with a cloth dipped in alcohol.
•Don’t point the light of the fiber-optic handpiece to the patient's eyes for the
light may harm to patient's eyes,even cause temporary blindness!

6 P/N:8023488
Beyond 300
User's Manual of Beyond 300/301 Delivery System
4 Sizes(unit:inches)

7 P/N:8023488
Beyond 301
User's Manual of Beyond 300/301 Delivery System

8 P/N:8023488
5 Operations
5.1 Water supply system
A 1.5L quick-disconnect water bottle mounted on the front rotary arm ,provide
simple and convenient solution to control water quality.
Figure 1
User's Manual of Beyond 300/301 Delivery System
1. Air Q/D: This air output provides 80 psi
output to run accessories for the user.
Accepts a 3/8in. male quick-disconnect
fitting.
2. Water Q/D: This water output provides 40
psi output to run accessories for the user.
Accepts a 1/4in. male quick-disconnect
fitting.
3. Flow Control: This needle vale adjusts the
water output from the Water Q/D. Rotate
counterclockwise to increase output and
rotate clockwise to decrease output.
4. City/Bottle Selector: This routing valve
controls the water systerm input. With the
toggle up in the “City”position, water enters
the dental unit from the office plumbing.
With the toggle down in the”Bottle”position,
water enters the dental unit from self-
contained water system.
CAUTION: Only water bottles supplied by the manufacturer can be
used .Do not use soft drink bottles which are thin walled and may rupture
when under pressure.
WARNING: Do not attempt to adjust the water bottle pressure. Bottle
pressure is factory pre-set 40psi.Pressurizing the water bottle over 40 psi
may cause the bottle to rupture.

9 P/N:8023488
User's Manual of Beyond 300/301 Delivery System
Figure 2
5.2 Assistant's control
The assistant's control can rotate as the figure shown.

10 P/N:8023488
A
B
User's Manual of Beyond 300/301 Delivery System
5.3 Cuspidor (optional)
The cuspidorl can rotate as
the figure shown.
Figure 3
Figure 4
•To realize cup filling function and
bowl rinsing function, just press the
two buttons beside the cup filling
nozzle.
•The bowl rinsing button is designed
to have off-delay function.
•A. Cup filling button
• B. Bowl rinsing button
Note: Beyond 300 system
is non-cuspidor type.

11 P/N:8023488
User's Manual of Beyond 300/301 Delivery System
Figure 5 Adjust the needle valves
•If you want to change the duration
of cup filling or bowl rinsing, you
need to remove the two nozzles
and take off the upper shell of the
cuspidor body,rotate the needle
valves clockwwise to extend the
time,counterclockwise to shorten
the time.
5.4 Dentist unit
The unit can rotate as the
figure shown .
Figure 6

12 P/N:8023488
2. Remove the end cover of level
arm, adjust the tightness of the
screw to get suitable balance force.
Not e: L oosen the s crews
counterclockwise, tighten the
screws clockwise.
Figure 8
Tools:
5mm Allen wrench
Tools:
3mm Allen wrench
Figure 7
1. Remove the plugs, change the
tightness of the three screws to
adjust the damping force of level
arm .
Note: 1、3 of damping adjustment,
2 of stop adjustment.Loosen the
screws in counterclockwise, tighten
in clockwise. 12
3
User's Manual of Beyond 300/301 Delivery System
5.4.1 Level arm adjustment

13 P/N:8023488
5.5 Control panel
Main control panel
Figure 9
Assistant's control panel
5.5.1 System state
The light-on of control panel indicates a normal working state, and you
can operate normally
User's Manual of Beyond 300/301 Delivery System

14 P/N:8023488
5.5.3 Initial settings of the chair
5.5.2 Chair position buttons
Symbols
Functions
Symbols
Functions
Backrest Up
Set
Backrest Down
Spitting position
Chair Down
Program Chair Positions 1
Chair up
Program Chair Positions 1
Original position
Program Chair Positions 1
0
P
User's Manual of Beyond 300/301 Delivery System
Notice: The hydraulic system shall have the initial setting of the chair
in following situations:
First installation or replacement of mainboard
Insert/replace/repair any position feedback potentiometer or wiring
Reset or replace the limit switch.
Reset the motion range of the chair.
Caution: When setting up the system, do not extent the maximum
range of the cylinder to prevent the hydraulic pump from working
improperly!

15 P/N:8023488
5.5.4 Initial settings description
Process or or
• Move the chair and backrest to the working position, and adjust dental
light to the desired condition ( On or Off), press the “P” button 3s till
the indicator lights on (beep 1 time, then loose button so as not to
enter the original setting state), and enters the setup procedure.
• Press the “P1” button one time, and then press the “P “button
again(beep1-2times), setup finished and escape the setup procedure.
• P2, P3 position setup are same as the P1.
PP
User's Manual of Beyond 300/301 Delivery System
Hydraulic system initial settings :
1. Press “P” button about 12 second until the buzzer sound 2 short beeps (the
indicator flashes twice and then illuminates),enter initial setup state; press
P1,the buzzer sound 1short beep( P light off),escape setup state.
This step eliminates the former stroke parameters.
2. Press “P” button about 12 seconds until the buzzer sound 2 short beeps,
(the indicator flashes twice and then illuminates),enter initial setup state,
move the chair by pressing the four direction buttons to make chair down
and backrest up. Then press “O”button , and move the chair again to make
chair up to the highest and backrest down to the lowest. A t this time, press
“LP” and then finish the setting.

16 P/N:8023488
5.5.5 Chair positions description
. Backrest Up
Press this button continuously and the backrest will move upwards. When
the backrest reach the desired position, stop pressing the button and the
backrest will stop immediately.
. Backrest Down
Press this button continuously and the backrest will move downwards.
When the backrest reach the desired position, stop pressing the button and
the backrest will stop immediately.
. Chair up
Press this button continuously and the chair will move upwards. When the
backrest reach the desired position, stop pressing the button and the backrest
will stop immediately.
. Chair down
Press this button continuously and the chair will move downwards. When
the backrest reach the desired position, stop pressing the button and the
backrest will stop immediately.
User's Manual of Beyond 300/301 Delivery System
NOTE: If you press the button continuously, when the chair backrest
reach the limit position, it will automatically stop the movement.
NOTE: If you press the button continuously, when the chair backrest
reach the limit position, it will automatically stop the movement.
NOTE: If you press the button continuously, when the chair backrest
reach the limit position, it will automatically stop the movement.
NOTE: If you press the button continuously, when the chair backrest
reach the limit position, it will automatically stop the movement.

17 P/N:8023488
. Original position
Press the button, the chair and backrest will move to the original position
automatically(Original position : backrest highest , the chair lowest) and
light off . Press this button again , the chair returns to former position and
light on. The chair shall stop at a new position by halfway stop.
. Spitting button
Press the button, the chair and backrest will move to spitting position
automatically(Spitting position : backrest highest , the chair is still) and
light off . Press this button again , the chair returns to former position and
light on. The chair shall stop at a new position by halfway stop.
0
5.6 Foot control
Delivery systems are equipped with wet-dry,variable speed, disc type foot
controls. Foot pressure on any part of the control disc controls the flow of
air to the active handpiece.
Figure 10
Water Coolant On/Off:
This toggle interrupts the flow of water coolant to the handpieces when
performing a procedure that requires dry cutting.
User's Manual of Beyond 300/301 Delivery System
This manual suits for next models
1