AEMC MN124 User manual

99-MAN 100316.v1 09/06
OPERATION
Making Measurements with the AC Current Probe Model MN124
• Connect the black (S2) and red (S1) terminals to the 200mV range of your DMM or instrument. The
MN124 has a ratio of 1:1. This means that for 100AAC in the conductor around which the probe is
clamped, 100mVAC will come out of the probe leads to your DMM or instrument. The output is 1mVAC
per Amp. Clamp the probe around the conductor. (e.g., 59mV = 59A)
• For best accuracy, avoid if possible, the proximity of other conductors which may create noise.
Tips for Making Precise Measurements
• When using a current probe with a meter, it is important to select the range that provides the best
resolution. Failure to do this may result in measurement errors.
• Make sure that probe jaw mating surfaces are free of dust and contamination. Contaminants cause air
gaps between the jaws, increasing the phase shift between primary and secondary. It is very critical for
power measurement.
MAINTENANCE:
Warning
• For maintenance use only original factory replacement parts.
• To avoid electrical shock, do not attempt to perform any servicing unless you are qualified to do so.
• To avoid electrical shock and/or damage to the instrument, do not get water or other foreign agents into
the probe.
Cleaning
To ensure optimum performance, it is important to keep the probe jaw mating surfaces clean at all times.
Failure to do so may result in error in readings. To clean the probe jaws, use very fine sand paper (fine
600) to avoid scratching the jaw, then gently clean with a soft oiled cloth.
REPAIR AND CALIBRATION
You must contact our Service Center for a Customer Service Authorization number (CSA#). This will
ensure that when your instrument arrives, it will be tracked and processed promptly. Please write the CSA#
on the outside of the shipping container. If the instrument is returned for calibration, we need to know if you
want a standard calibration, or a calibration traceable to N.I.S.T. (includes calibration certificate plus
recorded calibration data).
Chauvin Arnoux®, Inc. d.b.a. AEMC®Instruments
15 Faraday Drive •Dover, NH 03820 USA
Tel: (800) 945-2362 (Ext. 360)
(603) 749-6434 (Ext. 360)
Fax: (603) 742-2346 or (603) 749-6309
[email protected]
(Or contact your authorized distributor)
Costs for repair, standard calibration, and calibration traceable to N.I.S.T. are available.
NOTE: All customers must obtain a CSA# before returning any instrument.
TECHNICAL AND SALES ASSISTANCE
If you are experiencing any technical problems, or require any assistance with the proper use or
application of this instrument, please call our technical hotline:
(800) 343-1391 • (508) 698-2115 • Fax (508) 698-2118
Chauvin Arnoux®, Inc. d.b.a. AEMC®Instruments
[email protected]
www.aemc.com
AC Current Probe
Model MN124
User Manual
DESCRIPTION
The MN124 (Catalog #2129.13) is a high accuracy current probe for tight areas such as crowded
wiring. Extends DMM AC measurements to 100AAC. The Model MN124 offers a 5 ft lead with safety
4mm banana plug.
WARNING
These safety warnings are provided to ensure the safety of personnel and proper operation of the
instrument.
• Read the instruction manual completely and follow all the safety information before attempting
to use or service this instrument.
• Use caution on any circuit: Potentially high voltages and currents may be present and may
pose a shock hazard.
• Read the Safety Specifications section prior to using the current probe. Never exceed the
maximum voltage ratings given.
• Safety is the responsibility of the operator.
• ALWAYS connect the current probe to the display device before clamping the probe onto the
sample being tested.
• ALWAYS inspect the instrument, probe, probe cable, and output terminals prior to use.
Replace any defective parts immediately.
• NEVER use the current probe on electrical conductors rated above 300V in overvoltage
category III (CAT III). Use only on insulated conductors.
INTERNATIONAL ELECTRICAL SYMBOLS
This symbol signifies that the current probe is protected by double or reinforced insulation. Use
only factory specified replacement parts when servicing the instrument.
This symbol signifies CAUTION! and requests that the user refer to the user manual before
using the instrument.
This is a type B current sensor. Do not apply around or remove from HAZARDOUS LIVE
conductors without additional protective means (de-energizing the circuit or wearing protective
clothing suitable for high voltage work).
In conformity with WEEE 2002/96/EC
RECEIVING YOUR SHIPMENT
Upon receiving your shipment, make sure that the contents are consistent with the packing list. Notify
your distributor of any missing items. If the equipment appears to be damaged, file a claim immediately
with the carrier and notify your distributor at once, giving a detailed description of any damage.
PACKAGING
The AC Current Probe MN124 is shipped with this instruction manual and a product warranty and
registration card.
INSTRUMENT COMPATIBILITY
The Model MN124 is compatible with any voltmeter, multimeter, or other current measuring instruments
with an input impedance of less than >1MΩ. To achieve the stated accuracy, use the MN124 with a
voltmeter having an accuracy of 0.75% or better. The accuracy of the meter must be added to the probe
accuracy.
ELECTRICAL SPECIFICATIONS
Current Range: 1A to 100AAC, continuous
Output Signal: 1mVAC/AAC (100mV @ 100A)
Accuracy *:
1 to 100A: 2% ± 0.5mV
(with non inductive load)
Phase Shift: N/A
(*Reference conditions: 23°C ± 3°K, 20 to
70% RH, external magnetic field < 40 A/m,
50 to 60Hz sine wave, no DC component, no
external current carrying conductor, test
sample centered.) Load impedance 1MΩ.
Overload:
150A continuously
Frequency Range:
45 to 65Hz
Load Impedance:
1MΩ
Working Voltage:
300V on insulated conductor
Common Mode Voltage:
100VAC Cat. III
MECHANICAL SPECIFICATIONS
Operating Temperature:
-13° to 122°F (-25° to 50°C)
Storage Temperature:
-40° to 176°F (-40° to 80°C)
Maximum Cable Diameter:
0.43” Ø max. (11mm)
Dimensions:
1.4 x 4.53 x 0.87"
(36 x 115 x 22mm)
Weight:
160 g (6 oz)
Colors:
Dark gray handles with red cover
Polycarbonate Material:
Handle: 10% Fiberglass charged polycarbonate
UL 94 V0
Output:
Double/reinforced insulated 5 ft (1.5 m) lead with
safety 4mm banana plug
SAFETY SPECIFICATIONS
Electrical:
300V working voltage on insulated conductor
100V max common mode between output and
ground, Cat. III
3kV 50/60Hz dielectric for 1mn
ORDERING INFORMATION
AC Current Probe MN124...............Cat #2129.13
Accessories:
Banana plug adapter
(to non-recessed plug) ..................... Cat #1017.45
Black
Red
0.87" (22mm)
1.36" (35mm)
4.53" (115mm)
Ø=0.43" max
(11mm)
1.4" (36mm)
ELECTRICAL SPECIFICATIONS
Current Range: 1A to 100AAC, continuous
Output Signal: 1mVAC/AAC (100mV @ 100A)
Accuracy *:
1 to 100A: 2% ± 0.5mV
(with non inductive load)
Phase Shift: N/A
(*Reference conditions: 23°C ± 3°K, 20 to
70% RH, external magnetic field < 40 A/m,
50 to 60Hz sine wave, no DC component, no
external current carrying conductor, test
sample centered.) Load impedance 1MΩ.
Overload:
150A continuously
Frequency Range:
45 to 65Hz
Load Impedance:
1MΩ
Working Voltage:
300V on insulated conductor
Common Mode Voltage:
100VAC Cat. III
MECHANICAL SPECIFICATIONS
Operating Temperature:
-13° to 122°F (-25° to 50°C)
Storage Temperature:
-40° to 176°F (-40° to 80°C)
Maximum Cable Diameter:
0.43” Ø max. (11mm)
Dimensions:
1.4 x 4.53 x 0.87"
(36 x 115 x 22mm)
Weight:
160 g (6 oz)
Colors:
Dark gray handles with red cover
Polycarbonate Material:
Handle: 10% Fiberglass charged polycarbonate
UL 94 V0
Output:
Double/reinforced insulated 5 ft (1.5 m) lead with
safety 4mm banana plug
SAFETY SPECIFICATIONS
Electrical:
300V working voltage on insulated conductor
100V max common mode between output and
ground, Cat. III
3kV 50/60Hz dielectric for 1mn
ORDERING INFORMATION
AC Current Probe MN124...............Cat #2129.13
Accessories:
Banana plug adapter
(to non-recessed plug) ..................... Cat #1017.45
Black
Red
0.87" (22mm)
1.36" (35mm)
4.53" (115mm)
Ø=0.43" max
(11mm)
1.4" (36mm)
99-MAN 100316.v1 09/06
OPERATION
Making Measurements with the AC Current Probe Model MN124
• Connect the black (S2) and red (S1) terminals to the 200mV range of your DMM or instrument. The
MN124 has a ratio of 1:1. This means that for 100AAC in the conductor around which the probe is
clamped, 100mVAC will come out of the probe leads to your DMM or instrument. The output is 1mVAC
per Amp. Clamp the probe around the conductor. (e.g., 59mV = 59A)
• For best accuracy, avoid if possible, the proximity of other conductors which may create noise.
Tips for Making Precise Measurements
• When using a current probe with a meter, it is important to select the range that provides the best
resolution. Failure to do this may result in measurement errors.
• Make sure that probe jaw mating surfaces are free of dust and contamination. Contaminants cause air
gaps between the jaws, increasing the phase shift between primary and secondary. It is very critical for
power measurement.
MAINTENANCE:
Warning
• For maintenance use only original factory replacement parts.
• To avoid electrical shock, do not attempt to perform any servicing unless you are qualified to do so.
• To avoid electrical shock and/or damage to the instrument, do not get water or other foreign agents into
the probe.
Cleaning
To ensure optimum performance, it is important to keep the probe jaw mating surfaces clean at all times.
Failure to do so may result in error in readings. To clean the probe jaws, use very fine sand paper (fine
600) to avoid scratching the jaw, then gently clean with a soft oiled cloth.
REPAIR AND CALIBRATION
You must contact our Service Center for a Customer Service Authorization number (CSA#). This will
ensure that when your instrument arrives, it will be tracked and processed promptly. Please write the CSA#
on the outside of the shipping container. If the instrument is returned for calibration, we need to know if you
want a standard calibration, or a calibration traceable to N.I.S.T. (includes calibration certificate plus
recorded calibration data).
Chauvin Arnoux®, Inc. d.b.a. AEMC®Instruments
15 Faraday Drive •Dover, NH 03820 USA
Tel: (800) 945-2362 (Ext. 360)
(603) 749-6434 (Ext. 360)
Fax: (603) 742-2346 or (603) 749-6309
[email protected]
(Or contact your authorized distributor)
Costs for repair, standard calibration, and calibration traceable to N.I.S.T. are available.
NOTE: All customers must obtain a CSA# before returning any instrument.
TECHNICAL AND SALES ASSISTANCE
If you are experiencing any technical problems, or require any assistance with the proper use or
application of this instrument, please call our technical hotline:
(800) 343-1391 • (508) 698-2115 • Fax (508) 698-2118
Chauvin Arnoux®, Inc. d.b.a. AEMC®Instruments
[email protected]
www.aemc.com
AC Current Probe
Model MN124
User Manual
DESCRIPTION
The MN124 (Catalog #2129.13) is a high accuracy current probe for tight areas such as crowded
wiring. Extends DMM AC measurements to 100AAC. The Model MN124 offers a 5 ft lead with safety
4mm banana plug.
WARNING
These safety warnings are provided to ensure the safety of personnel and proper operation of the
instrument.
• Read the instruction manual completely and follow all the safety information before attempting
to use or service this instrument.
• Use caution on any circuit: Potentially high voltages and currents may be present and may
pose a shock hazard.
• Read the Safety Specifications section prior to using the current probe. Never exceed the
maximum voltage ratings given.
• Safety is the responsibility of the operator.
• ALWAYS connect the current probe to the display device before clamping the probe onto the
sample being tested.
• ALWAYS inspect the instrument, probe, probe cable, and output terminals prior to use.
Replace any defective parts immediately.
• NEVER use the current probe on electrical conductors rated above 300V in overvoltage
category III (CAT III). Use only on insulated conductors.
INTERNATIONAL ELECTRICAL SYMBOLS
This symbol signifies that the current probe is protected by double or reinforced insulation. Use
only factory specified replacement parts when servicing the instrument.
This symbol signifies CAUTION! and requests that the user refer to the user manual before
using the instrument.
This is a type B current sensor. Do not apply around or remove from HAZARDOUS LIVE
conductors without additional protective means (de-energizing the circuit or wearing protective
clothing suitable for high voltage work).
In conformity with WEEE 2002/96/EC
RECEIVING YOUR SHIPMENT
Upon receiving your shipment, make sure that the contents are consistent with the packing list. Notify
your distributor of any missing items. If the equipment appears to be damaged, file a claim immediately
with the carrier and notify your distributor at once, giving a detailed description of any damage.
PACKAGING
The AC Current Probe MN124 is shipped with this instruction manual and a product warranty and
registration card.
INSTRUMENT COMPATIBILITY
The Model MN124 is compatible with any voltmeter, multimeter, or other current measuring instruments
with an input impedance of less than >1MΩ. To achieve the stated accuracy, use the MN124 with a
voltmeter having an accuracy of 0.75% or better. The accuracy of the meter must be added to the probe
accuracy.

Other AEMC Measuring Instrument manuals

AEMC JM810A User manual

AEMC

AEMC JM810A User manual

AEMC 1510 User manual

AEMC

AEMC 1510 User manual

AEMC powerpad 3945 User manual

AEMC

AEMC powerpad 3945 User manual

AEMC 8220 User manual

AEMC

AEMC 8220 User manual

AEMC 6611 User manual

AEMC

AEMC 6611 User manual

AEMC MN352 User manual

AEMC

AEMC MN352 User manual

AEMC CL601 User manual

AEMC

AEMC CL601 User manual

AEMC 5070 User manual

AEMC

AEMC 5070 User manual

AEMC FlexProbe 24-3001 User manual

AEMC

AEMC FlexProbe 24-3001 User manual

AEMC MN375 User manual

AEMC

AEMC MN375 User manual

AEMC A196-24-BK User manual

AEMC

AEMC A196-24-BK User manual

AEMC SL261 User manual

AEMC

AEMC SL261 User manual

AEMC 603 User manual

AEMC

AEMC 603 User manual

AEMC 5050 User manual

AEMC

AEMC 5050 User manual

AEMC MN185 User manual

AEMC

AEMC MN185 User manual

AEMC 5100 User manual

AEMC

AEMC 5100 User manual

AEMC MN291 User manual

AEMC

AEMC MN291 User manual

AEMC SR631 User manual

AEMC

AEMC SR631 User manual

AEMC 6609 User manual

AEMC

AEMC 6609 User manual

AEMC PowerPad III 8336 User manual

AEMC

AEMC PowerPad III 8336 User manual

AEMC PowerPad 8335 Supplement

AEMC

AEMC PowerPad 8335 Supplement

AEMC SR701 User manual

AEMC

AEMC SR701 User manual

AEMC 6422 User manual

AEMC

AEMC 6422 User manual

AEMC 6416 User manual

AEMC

AEMC 6416 User manual

Popular Measuring Instrument manuals by other brands

Blue-White industries Sonic-Pro S6A Series operating manual

Blue-White industries

Blue-White industries Sonic-Pro S6A Series operating manual

ATEN EC1000 user manual

ATEN

ATEN EC1000 user manual

Kobold DAH operating instructions

Kobold

Kobold DAH operating instructions

Helwett Packard 4339B user guide

Helwett Packard

Helwett Packard 4339B user guide

Hanna Instruments HI702 instruction manual

Hanna Instruments

Hanna Instruments HI702 instruction manual

Kobold MAK operating instructions

Kobold

Kobold MAK operating instructions

SMA SUNNY HOME MANAGER 2.0 Quick reference guide

SMA

SMA SUNNY HOME MANAGER 2.0 Quick reference guide

WATANABE ELECTRIC INDUSTRY WPMZ-1 Series Quick instruction manual

WATANABE ELECTRIC INDUSTRY

WATANABE ELECTRIC INDUSTRY WPMZ-1 Series Quick instruction manual

Sper scientific 850013 instruction manual

Sper scientific

Sper scientific 850013 instruction manual

SBS 3500 instruction manual

SBS

SBS 3500 instruction manual

Kedida CT-6023 instruction manual

Kedida

Kedida CT-6023 instruction manual

Kvarts DRSB-01 quick start guide

Kvarts

Kvarts DRSB-01 quick start guide

Baumer CleverLevel LBFH/I operating instructions

Baumer

Baumer CleverLevel LBFH/I operating instructions

Hama TH-130 operating instructions

Hama

Hama TH-130 operating instructions

PCB Piezotronics 3991B112KG Installation and operating manual

PCB Piezotronics

PCB Piezotronics 3991B112KG Installation and operating manual

Gas Clip Technologies Infrared user manual

Gas Clip Technologies

Gas Clip Technologies Infrared user manual

BERTHOLD TECHNOLOGIES Junior LB 9509 operating manual

BERTHOLD TECHNOLOGIES

BERTHOLD TECHNOLOGIES Junior LB 9509 operating manual

Hydrotechnik MultiControl 4070 operating instructions

Hydrotechnik

Hydrotechnik MultiControl 4070 operating instructions

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.