
MAINTENANCE MANUAL
ATC-600A-2
2-2-2
Page 2
Nov 1/05
(2) ESD
CAUTION: THE CALIBRATION PROCEDURES SHOULD ONLY BE PERFORMED IN
AN ESD ENVIRONMENT. ALL PERSONNEL PERFORMING THE
CALIBRATION PROCEDURES SHOULD HAVE KNOWLEDGE OF
ACCEPTED ESD PRACTICES AND/OR BE ESD CERTIFIED.
CAUTION
THIS EQUIPMENT CONTAINS PARTS
SENSITIVE TO DAMAGE
BY ELECTROSTATIC DISCHARGE (ESD)
(3) EMC and Safety Compliance
All assemblies, cables, connectors, plastic fasteners, gaskets, fingerstock and
miscellaneous hardware within the Test Set are configured to satisfy the safety and
EMC compliance standards.
CAUTION: UPON COMPLETION OF ANY MAINTENANCE ACTION; ALL
ASSEMBLIES, CABLES, CONNECTORS, PLASTIC FASTENERS,
GASKETS, FINGERSTOCK AND MISCELLANEOUS HARDWARE MUST
BE CONFIGURED AS INSTALLED AT THE FACTORY.
C. Requirements
(1) Test Set Configuration
The ATC-600A-2 must be installed according to the Installation Calibration procedure
in the ATC-600A-2 Operation Manual.
(2) Test Equipment
Appendix B contains a comprehensive list of test equipment suitable for performing
any procedure contained in this manual. Other equipment meeting specifications listed
in Appendix B may be substituted in place of recommended models.
NOTE: For certain procedures in this manual, the test equipment listed in Appendix B
may exceed the minimum required specifications.
(3) Disassembly
Remove lid from ATC-600A-2 to perform the Verification Procedures.
Remove lid and case from ATC-600A-2 to perform the Calibration Procedures.
(4) Environment
For best results, the calibration environmental conditions should be identical to the
environmental conditions at the normal operating location.
D. Procedure Instruction
It is strongly recommended that personnel thoroughly read and understand all steps of the
procedures to be performed and be familiar with the circuit under test. Knowledge of
power, frequency and waveform to be expected at each test point is recommended.
NOTE: When one circuit provides the same pulse characteristic for different pulses, it is
necessary to test the specifications for that characteristic only once. Pulse
spacings are measured from leading edge to leading edge at the 50% amplitude
points. Pulse widths are measured from leading edge to trailing edge
at the 50% amplitude points.