Agilent Technologies 5DX Specifications

Agilent Technologies
Medalist 5DX Automated
X-ray Inspection System
Operator Instructions


Operator Instructions 3
Contents
1Emergency Stop 5
2Hardware Overview 6
Medalist 5DX Diagram 6
System Light Tower 7
Operator Control Panel 8
3Operator Instructions 9
Startup the System 9
Verify the Safety Interlocks are Closed 9
Verify the X-Ray Has Been Enabled 9
Log on to Microsoft Windows XP 9
Log on to the Medalist 5DX User Interface 9
Run Testing 11
End Testing 13
Log Off Microsoft Windows 13
4What to do in Case of a Problem 14
Panel Retrieval 14
Other Errors 15

4 Operator Instructions

Emergency Stop
Operator Instructions 5
1 Emergency Stop
CAUTION
The emergency stop buttons are intended for use in emergency situations only
and should not be used for normal system shut down. Repeated misuse of the
emergency stop buttons will eventually cause damage to certain components or
the loss of data or both.
An emergency stop button is used to quickly shut down power to the
Medalist 5DX. The Medalist 5DX is equipped with two emergency
stop buttons located on either side of the Medalist 5DX main cabinet,
centered above the outer barrier panels. Whenever one of the
emergency stop buttons is pressed, all power is immediately removed
from the Medalist 5DX subsystems, with the exception of the system
controller and the monitor. Call a service technician to restart the
system after an emergency button has been pressed.
Push down on one of the emergency stop
buttons in an emergency situation.
The emergency stop button locks
OFF when pushed.

6 Operator Instructions
2 Hardware Overview
Medalist 5DX Diagram
The Medalist 5DX Automated X-ray Inspection System provides a
comprehensive test and inspection process to discover solder defects
on a circuit board.
Figure 1 Medalist 5DX system
1System controller
2Emergency stop (EMO) button
3Panel handler access door
4Connector for anti-static wrist strap
5System light tower
6Monitor
7Operator control panel
8Keyboard and mouse tray
9Power distribution unit
2
6
7
1
3
5
4
9
2
3
8

Hardware Overview
Operator Instructions 7
System Light Tower
The system light tower is permanently attached to the electronics
cabinet. The system light tower is used to visually indicate the
operating condition of the Medalist 5DX.
Figure 2 System light tower
1Buzzer —If enabled, sounds when there is a hardware failure.
2Red Lamp —System has a hardware failure and requires
attention.
3Yellow Lamp —System is waiting for user input.
4Green Lamp —System is operating.
1
2
3
4

8 Operator Instructions
Operator Control Panel
The operator control panel provides an X-ray Control key and X-ray
control buttons as well as status LEDs (see Figure 3). It is attached
to the keyboard tray.
Figure 3 Operator control panel
1X-ray Control key (Enable (1)/ Disable (0))
2X-ray Control enable button
3X-ray Control disable button
4X-rays LED
5Safety Interlocks LED
1
2
3
4
5

Operator Instructions
Operator Instructions 9
3 Operator Instructions
Startup the System
The Medalist 5DX is operated through the use of a user interface
running on the system controller. Before starting the system
software, verify that the X-ray has been enabled.
Verify the Safety Interlocks are Closed
On the operator control panel (see Figure 3) verify that the Safety
Interlocks green LED is lit. If it is not, call a service technician to
resolve the problem.
Verify the X-Ray Has Been Enabled
On the operator control panel (see Figure 3) verify that the X-ray
Control key is in the enabled position (1) and press the X-ray Control
enable button (1). Verify that the X-rays red LED is lit.
Log on to Microsoft Windows XP
Log on to Microsoft Windows using the Windows user name and
password provided by your administrator. When you log on, the
Medalist 5DX software starts automatically.
Log on to the Medalist 5DX User Interface
Log on to the Medalist 5DX user interface using the user name
provided by your administrator. Press the ‘Enter’ key to Login.
Figure 4 Login

10 Operator Instructions
The system will perform a system startup. This may take a few
minutes (see Figure 5).
Figure 5 Startup progress

Operator Instructions
Operator Instructions 11
Run Testing
1Depending on how the software has been set up, you may be
asked to choose the project to load. If prompted, select the
appropriate project and click OK (see Figure 6). If not, continue
with Step 2.
Figure 6 Enter project pop-up
2Depending on how the software has been set up, you may be
asked to enter the panel serial number. If prompted, enter the
panel serial number and click OK (see Figure 7). If not, continue
with step 3.
Figure 7 Enter serial number pop-up
3Load the panel into the machine as shown (see Figure 8).

12 Operator Instructions
Figure 8 Load the panel into the machine
Once a panel is loaded, the inspection will begin.
Figure 9 Inspection in progress

Operator Instructions
Operator Instructions 13
When the inspection is finished, the panel will automatically be
unloaded and the system will wait for the panel to be removed
from the machine.
4After the panel is removed, continue repeating steps 2 and 3 until
testing is complete.
End Testing
To exit the Medalist 5DX user interface:
1Press Cancel at the serial number prompt.
2Press Yes when asked if you want to abort (see Figure 10).
Figure 10 Stop testing pop-up
Log Off Microsoft Windows
1Click Start > Log Off.
2Click Log Off when you are asked if you really want to log off.

14 Operator Instructions
4 What to do in Case of a Problem
Panel Retrieval
Occasionally it is necessary to enter into the 5DX Cabinet to retrieve
a dropped or jammed panel. The Medalist 5DX will display an error
message if panel retrieval is necessary. Figure 11 shows the error
message that can occur when trying to load a panel:
Figure 11 Error message when trying to load a panel
Figure 12 shows the error message that can occur when trying to
unload a panel:
Figure 12 Error message when trying to unload a panel
To recover from these errors:
1Click OK to dismiss the error message.
2Turn the X-ray Control key to the disable (0) position on the
operator control panel.
3Unlock and open the top sliding access panels.

What to do in Case of a Problem
Operator Instructions 15
4Reach in and remove the panel.
5Close and lock all access panels.
NOTE
If you are unable to easily remove the panel, the panel clamps may be locked in
the down position, or the panel in place sensor may be extended. Ask a service
technician for assistance.
6Verify that the emergency stop buttons are in the up position. If
one is in the down position, call a service technician to restart
the system.
7Verify that the Safety Interlocks green LED is lit on the operator
control panel.
8Turn the X-ray Control key to the enable (1) position on the
operator control panel.
9Press the X-ray Control on (1) button on the operator control
panel.
10 Verify that the X-ray on light is now red on the operator control
panel.
11 You are now ready to continue testing as before.
Other Errors
If an error occurs during testing, an appropriate message will be
displayed describing the nature of the problem, and possible steps to
take to fix it. (Figure 13 shows one of many possible error messages).
Specific directions on the next steps to take are provided in the
dialog box and must be followed by the appropriate level of user and
not ignored.
NOTE
Do not dismiss error messages. Call a programmer or a service trained
technician to assist you.
Figure 13 Sample error message

www.agilent.com
© Agilent Technologies, Inc. 2008
Printed in Malaysia June 2008
N7200-90090 Rev. A
Other manuals for 5DX
2
Table of contents
Other Agilent Technologies Diagnostic Equipment manuals