Philips EverFlo User manual

1053621 R06
LZ 07/20/2017
EN-INTL, DE
EverFlo
EverFlo Q
Respironics Inc.
1001 Murry Ridge Lane
Murrysville, PA 15668 USA
Respironics Deutschland GmbH & Co. KG
Gewerbestrasse 17
82211 Herrsching, Deutschland
USER MANUAL | BENUTZERHANDBUCH
1041621

1
EverFlo / EverFlo Q User Manual
Table of ConTenTs
Symbols Glossary ....................................................................................................................2
Abbreviations ...........................................................................................................................3
Classications ...........................................................................................................................3
How to Contact Philips Respironics..................................................................................4
Chapter 1: Introduction.........................................................................................................4
Chapter 2: Operating Instructions.....................................................................................8
Chapter 3: Cleaning, Disinfecting, and Maintenance.............................................. 10
Chapter 4: Alarms and Troubleshooting...................................................................... 11
Chapter 5: Specications................................................................................................... 14
Appendix A: EMC Information......................................................................................... 17
Limited Warranty ................................................................................................................. 19
EverFlo / EverFlo Q User Manual
©Koninklijke Philips N.V., 2017. All rights reserved.

2
EverFlo / EverFlo QUser Manual
Symbols Glossary
Symbol Title and Meaning
Refer to the instruction manual
To signify that the instruction manual must be read.
Type BF applied part
To identify a type BF applied part complying with IEC 60601-1.
Class II equipment (Double Insulated)
To identify equipment meeting the safety requirements specied for Class II equipment.
No smoking
To prohibit smoking.
No open ame
To prohibit smoking and all forms of open ame.
No oil or grease
Do not disassemble
Alarm
Indicates an alarm condition.
AC power (Alternating current)
Indicates on the rating plate that the equipment is suitable for alternating current only; to identify relevant
terminals.
Catalogue number
Indicates the manufacturer’s catalogue number so the medical device can be identied.
Serial number
Identies the manufacturer’s serial number for the medical device.
On (Power)
Indicates connection to the mains.
O
Indicates disconnection from the mains.
IP21 Drip proof equipment

3
EverFlo / EverFlo Q User Manual
The EverFlo / EverFlo Q Oxygen Concentrator is classied as:
- IEC Class II Internally Powered Equipment
- Type BF Applied Part
- IP21 Drip Proof Equipment
- Not suitable for use in the presence of a flammable
anesthetic mixture with air or with oxygen or nitrous
oxide.
- Continuous Operation
Symbol Title and Meaning
Malfunction, general, failure
Indicates that a failure or other malfunction has occurred.
Compliant with the Waste Electrical and Electronic Equipment/Restriction of the Use of Certain
Hazardous Substances in Electrical and Electronic Equipment (WEEE/RoHS) recycling directives
Temperature Limit
Indicates the temperature limits to which the medical device can be safely exposed.
Humidity Limitation
Indicates the range of humidity to which the medical device can be safely exposed.
Atmospheric Pressure Limitation
Indicates the acceptable upper and lower limits of atmospheric pressure.
Manufacturer
Authorized Representative in the European Community
Indicates the Authorized Representative in the European Community.
Abbreviations Classifications
LED Light Emitting Diode
LPM Liters per Minute
OPI Oxygen Percentage
Indicator

4
EverFlo / EverFlo QUser Manual
How to Contact Philips Respironics
To have your device serviced, contact your equipment provider. If you need to contact Philips Respironics directly,
call the Customer Service department at +1-724-387-4000 or +49 8152 93060. You can also use one of the
following addresses:
Respironics Inc.
10
01 Murry Ridge Lane
Murr
ysville, PA 15668 USA
Respironics Deutschland GmbH & Co. KG
Gewerbestrasse 17
82211 Herrsching, Germany
Chapter 1: Introduction
Your health care professional has determined that supplemental oxygen is of benet to you and has prescribed an
oxygen concentrator set at a specic ow setting to meet your needs. DO NOT change the ow settings unless your
health care professional tells you to do so. Please read and understand this entire manual before using the device.
Intended Use
The EverFlo / EverFlo Q Oxygen Concentrator is intended to provide supplemental oxygen to persons requiring
oxygen therapy. The device is not intended to be life supporting or life sustaining.
The device is for use in the home or hospital/institutional environment.
About Your EverFlo / EverFlo Q
The device produces concentrated oxygen from room air for delivery to a patient requiring low ow oxygen
therapy. The oxygen from the air is concentrated using a molecular sieve and a pressure swing adsorption process.
Your home care provider will show you how to operate the concentrator and will be available to answer any
questions. If you have additional questions or problems, contact your home care provider.
Parts of Your Concentrator
Control Panel
On/O Switch
Flow Meter
Oxygen Outlet
Port
Humidier Bottle
Mounting Strap
Humidier
Connector
Tube
Filter Door
Air Inlet Filter

5
EverFlo / EverFlo Q User Manual
EverFlo models that are compatible with the UltraFill Oxygen Filling Station will include an additional connection
port for connection to the UltraFill Oxygen Filling Station. The performance of the EverFlo is not aected when the
port is connected to the UltraFill Oxygen Filling Station. The port is for connection to the UltraFill Oxygen Filling
Station only, not for connecting the patient oxygen cannula. The patient oxygen cannula is connected to the
EverFlo as shown later in this manual.
When the EverFlo is connected to the UltraFill Oxygen Filling Station, the maximum ow of the device that can be
delivered to the patient will be reduced from 5 LPM to 3 LPM.
Note: When connected to the UltraFill Oxygen Filling Station, the EverFlo device disables its no ow and high ow
alarms when a cylinder is in the process of being lled. If you are breathing from the EverFlo at this time and an
occlusion occurs in the cannula, the EverFlo will not sense the disruption of the ow of oxygen to you. If you are
breathing from the EverFlo device while lling a cylinder, ensure that you place your oxygen cannula in a position
to avoid it being kinked or crushed. Refer to the Troubleshooting Guide for more information.
Accessory Equipment and Replacement Parts
Contact your home care provider if you have questions about this equipment. Use only the following
PhilipsRespironics accessories and replacement parts with this device:
- Air Inlet Filter
- Humidier Connector Tube
- The EverFlo device is compatible with all bottle style humidiers supplied through Philips Respironics.
Accessories provided with the concentrator are to be specied for use at oxygen ows of 1 to 5 liters per minute
and a maximum pressure of 6.5 PSIG (44.8 kPa). In addition, cannulas must be a minimum of 2.13 meters (7 feet) in
length and a maximum of 45.7 meters (150 feet).
The organization that provides this equipment to the patient for use is accountable for ensuring the compatibility
of the accessories used to connect the patient to the oxygen concentrator to the requirements of ISO 80601-2-69. In
order to meet the requirements of ISO 80601-2-69, the accessory cannula must have a re stop device that stops re
and ow of oxygen to the patient.
Contact your home care provider if you have any questions concerning the use of accessories.
WARNING: The use of incompatible parts or accessories can result in degraded performance.

6
EverFlo / EverFlo QUser Manual
Warnings and Cautions
Warnings
A warning represents the possibility of harm to the operator or patient.
• For proper operation, your concentrator requires unobstructed ventilation. The ventilation ports are located
at the rear base of the device and at the side air inlet filter. Keep the device at least 15 to 30 cm away from
walls, furniture, and especially curtains that could impede adequate airflow to the device. Do not place the
concentrator in a small closed space (such as a closet). The device should not be used adjacent to or stacked
with other equipment. For more information, contact your home care provider.
• Do not remove the covers of this device. Servicing must be referred to an authorized and trained Philips
Respironics home care provider.
• In the event of an equipment alarm or if you are experiencing any signs of discomfort consult your home care
provider and/or your health care professional immediately.
• Oxygen generated by this concentrator is supplemental and should not be considered life supporting or life
sustaining. In certain circumstances oxygen therapy can be hazardous; any user should seek medical advice
prior to using this device.
• Where the prescribing health care professional has determined that an interruption in the supply of oxygen,
for any reason, may have serious consequences to the user, an alternate source of oxygen should be available
for immediate use.
• Oxygen vigorously accelerates combustion and should be kept away from heat or open flame. Not suitable for
use in the presence of a flammable anesthetic mixture with air or with oxygen or nitrous oxide.
• Oxygen makes it easier for a fire to start and spread. Do not leave the nasal cannula or mask on bed coverings
or chair cushions, if the oxygen concentrator is turned on, but not in use; the oxygen will make the materials
flammable. Turn the oxygen concentrator off when not in use to prevent oxygen enrichment.
• Accessories used to connect the patient to the oxygen concentrator must be compatible to the requirements
of ISO 80601-2-69. Application accessories shall include a means to reduce the propagation of fire, including
having a fire stop device that stops fire and flow of oxygen to the patient.
• Do not smoke, allow others to smoke, or have open flames near the concentrator when it is in use. Smoking
during oxygen therapy is dangerous and is likely to result in facial burns or death.
• Do not use oil or grease on the concentrator or its components as these substances, when combined with
oxygen, can greatly increase the potential for a fire hazard and personal injury.
• Do not use the oxygen concentrator if either the plug or power cord is damaged. Do not use extension cords
or electrical adapters.
• Do not attempt to clean the concentrator while it is plugged into an electrical outlet.
• Device operation above or outside of the voltage, LPM, temperature, humidity and/or altitude values specified
may decrease oxygen concentration levels.
• Your home care provider is responsible for performing appropriate preventive maintenance at the intervals
recommended by the device manufacturer.
• Portable and mobile RF communications equipment can affect Medical Electrical Equipment. See the EMC
section of this manual for distances to observe between RF communications equipment and the device to
avoid interference.
• Medical Electrical Equipment needs special precautions regarding EMC and needs to be installed and put into
service according to the EMC information provided in this manual.

7
EverFlo / EverFlo Q User Manual
• The EverFlo is designed to meet current electromagnetic compatibility requirements. However, if you
suspect operation of this device is interfered by or interferes with the normal operation of other electronic
devices (such as TV, radio or other household appliances), try relocating the appliance or the device until the
interference stops, or plugging the device into a different power outlet controlled by a separate circuit breaker
or fuse.
• Be aware that the electrical cord and/or tubing could present a tripping or strangulation hazard.
• The No-Flow alarm is disabled when the low-flow meter is installed.
• Do not modify this system or equipment in any way. Modifications could result in hazards to the user.
• Place the device in a location so as to avoid pollutants and fumes.
• The oxygen delivery settings of the oxygen concentrator should be periodically reassessed for the
effectiveness of the therapy.
• If you feel discomfort or are experiencing a medical emergency while undergoing oxygen therapy, seek
medical assistance immediately to avoid harm.
• To ensure that you receive the correct therapeutic amount of oxygen delivery according to your medical
condition, the Philips Respironics EverFlo device must be used:
- Only after one or more settings have been individually determined or prescribed for you at your specific
activity levels
- With the specific combination of parts and accessories that are in line with the specification of the oxygen
concentrator manufacturer and that were used while your settings were determined
• Use only water-based lotions or salves that are oxygen compatible during setup or use during oxygen therapy.
To avoid the risk of fire and burns, never use petroleum or oil-based lotions or salves.
• Do not lubricate fittings, connections, tubing, or other accessories of the oxygen concentrator to avoid the risk
of fire and burns.
• Geriatric or any other patient unable to communicate discomfort, or hear or see the alarms while using this
device, may require additional monitoring.
• Use only power cords supplied by Philips Respironics for this device. Use of power cords not supplied by
Philips Respironics may cause overheating or damage to the device and may result in increased emissions or
decreased immunity of the equipment or system.
• The use of incompatible parts or accessories can result in degraded performance.
Cautions
A caution represents the possibility of damage to the equipment.
• Do not place liquids on or near the device.
• If liquid is spilled on the device, turn the power off and unplug from electrical outlet before attempting to
clean up spill. Call your home care provider if device does not continue to work properly.

8
EverFlo / EverFlo QUser Manual
Chapter 2: Operating Instructions
Warning: Do not use extension cords or electrical adapters.
1. Select a location that allows the concentrator to draw in room air without
being restricted. Make sure that the device is at least 15 to 30 cm away
from walls, furniture, and especially curtains that could impede adequate
airflow to the device. Do not place the device near any heat source.
2. After reading this entire manual, plug the power cord into an electrical
outlet. Before plugging the unit into a wall outlet (AC power):
• Verify that the AC power cord is labeled with 120 VAC or 230 VAC.
• Verify that the AC power in the wall outlet matches the voltage that
is labeled on the AC power cord.
• If the AC power in the wall outlet matches the voltage labeled on the
AC power cord, plug the device into the AC wall outlet.
• If the AC power in the wall outlet does not match the voltage labeled
on the AC power cord, do not plug the device into the AC wall outlet.
Contact your health care professional for assistance.
3. Do either Step A or Step B below.
A. If you are not using a humidifier, connect your nasal cannula to the
Oxygen Outlet Port, as shown in the top illustration on the right.
B. If you are using a humidifier, follow the steps below:
1. Open the filter door on the back of the device as shown.
2. Remove the humidifier connector tube from the back of the
filter door and replace the filter door, as shown.
3. Fill your humidifier bottle according to the manufacturer’s
instructions.
4. Mount the filled humidifier on the top of the EverFlo / EverFlo Q
device inside the Velcro strap, as shown in the illustration on the
right.
5. Tighten the Velcro strap around the bottle and secure it
so it is held firmly in place.
6. Connect the humidifier connector tube (that you
retrieved from the filter door) to the Oxygen Outlet Port
(as shown in Step 3-A above).
7. Connect the other end of the humidifier connector tube
to the top of the humidifier with the elbow in the tubing
facing the front, as shown here.
8. Connect your cannula to the humidifier bottle according to the
humidifier bottle manufacturer’s specifications.
Oxygen Outlet Port
Step 3-A
Step 3-B1
Step 3-B4
Step 3-B7

9
EverFlo / EverFlo Q User Manual
4. Press the power switch to the On [I] position. Initially, all the LEDs will illuminate and the audible
alert will beep for a few seconds. Only the green LED should remain lit.
Note: If the device is stored at the minimum storage temperature between uses, please allow
2.5hours for the unit to adequately warm up.
Note: If the device is stored at the maximum storage temperature between uses, please allow
2.5hours for the unit to adequately cool down.
5. After turning on the Oxygen Concentrator, you can begin breathing from the device immediately;
however, allow at least 10 minutes for oxygen delivery to reach defined specifications.
6. Adjust the flow to the prescribed setting by turning the knob on the top of the flow meter until the
ball is centered on the line marking the specific flow rate.
Flow Meter Knob
7. Be sure oxygen is flowing through the cannula. If it is not, refer to the Troubleshooting Guide in this
manual.

10
EverFlo / EverFlo QUser Manual
8. Put on the cannula as directed by your home care provider.
9. When you are not using the oxygen concentrator, press the power switch to the Off [O] position.
Chapter 3: Cleaning, Disinfecting, and Maintenance
Cleaning and Disinfecting the Device
Warning: It is important to unplug the device before you perform any cleaning and disinfecting.
Warning: Do not remove the covers of this device. Servicing must be referred to an authorized and trained
Philips Respironics home care provider.
Caution: Excess moisture may impair the proper operation of the device.
Cleaning
The exterior covers of the device should be cleaned weekly and between patient use and as needed by performing
the following steps:
1. Turn the device off and disconnect from the power source before cleaning.
2. Clean the device exterior, including the filter door, using a damp cloth with a mild household cleaner and wipe
it dry.
Disinfection
The exterior covers of the device should be disinfected between patient use as follows:
1. Clean the device as indicated above.
2. To disinfect, use a household chlorine bleach containing 8.25% sodium hypochlorite. Combine 10 parts water
to 1 part bleach.
3. Using a damp cloth with the bleach solution, wipe the exterior surfaces.
4. Allow the surface to remain damp for 2 minutes. Wipe dry as necessary.
Humidier, Cannula and Tubing
Clean and replace the humidier, cannula and tubing as instructed by the manufacturer and your equipment
provider.
Filters
The EverFlo air inlet lter should be routinely replaced by the provider as deemed necessary.
Service
The EverFlo / EverFlo Q Oxygen Concentrator contains no user-serviceable parts.

11
EverFlo / EverFlo Q User Manual
Chapter 4: Alarms and Troubleshooting
Alarm and Indicators
The device has an audible alarm and three LED indicators, as shown below.
Note: All EverFlo alarms are low priority alarms. The alarm system should be veried before use and
between users by service personnel in accordance with the EverFlo service manual.
Audible Alarm / Colored LED Possible Cause Your Action
All 3 LEDs illuminate continuously
and the Audible Alarm is
sounding continuously.
The device has
detected a system
malfunction.
Immediately turn o the device,
connect to a back up oxygen
source, and call your home care
provider.
The Audible Alarm is sounding
continuously. None of the LEDs
are illuminated.
The device is turned on
but is not operating.
Often this indicates
that the device is not
plugged in or there is a
power failure.
Check the power outlet and verify
that the device is plugged in. If the
problem continues, connect to a
back up oxygen source and call
your home care provider.
Red LED illuminates continuously
and the Audible Alarm is
sounding continuously.
The device has
detected a system
malfunction.
Immediately turn o the device and
wait 5 minutes. Restart the device. If
the condition persists turn the unit
o, connect to a back up oxygen
source, and call your home care
provider.
NORMAL
Red Alarm Indicator
Yellow Check System Indicator
Green Power On Indicator

12
EverFlo / EverFlo QUser Manual
Yellow LED illuminates
continuously. The Red LED is
blinking and the Audible Alarm is
beeping periodically.
The device has
detected an impeded
oxygen ow condition.
Follow the troubleshooting guide
on the next page. Connect to a
back up oxygen source and call
your home care provider if your
troubleshooting actions fail to end
this alert condition.
Note: The No Oxygen Flow alarm
may take up to 1.5 minutes to
activate from the time the alarm
condition is present.
Yellow LED is blinking. The
Red LED is not illuminated and
the Audible Alarm is beeping
periodically.
The device has
detected a high
oxygen ow condition.
Follow the troubleshooting guide
on the next page. Connect to a
backup oxygen source, and call
your home care provider if your
troubleshooting actions fail to end
this alert condition.
Note: The High Oxygen Flow alarm
may take up to 1 minute to activate
from the time the alarm
condition is present.
Yellow LED illuminates
continuously. The Red LED is o
and the Audible Alarm is silent.
The device has
detected a low oxygen
condition (OPI units
only).
Continue using the unit but call
your home care provider about this
condition.
Note: The Low Oxygen alarm may
take up to 15 minutes to activate
from the time the alarm condition
is present.

13
EverFlo / EverFlo Q User Manual
Troubleshooting Guide
The table below lists common problems and actions you can take. If you are unable to resolve a problem, please
contact your equipment provider.
Problem Why it Happened What to Do
High oxygen ow indication is
activated.
(Yellow LED is blinking. The Red LED is
o and the Audible Alarm is beeping
periodically.)
The device has detected a high
oxygen ow condition.
Turn the ow rate down to your
prescribed level. Wait at least
2minutes. If the condition persists
turn the unit o, connect to a back
up oxygen source, and call your
home care provider.
Green LED illuminates continuously.
The other LEDs are o and the Audible
Alarm is silent.
The device is turned on and
working properly. Take no action.
The device is not working when it is
turned on.
(The Audible Alarm is sounding
continuously. All LEDs are o.)
The power cord plug is not
properly inserted into the
electrical outlet.
Make sure the device is properly
plugged in to the electrical outlet.
The unit is not receiving power
from the electrical outlet.
Check your household outlet fuse
or circuit.
Internal part failure.
Connect to a back up oxygen
source and contact your home
care provider.
The device is not working when it is
turned on.
(The Audible Alarm is sounding
continuously and all 3 LEDs are
illuminated.)
Internal part failure.
Connect to a back up oxygen
source and contact your home
care provider.
Impeded oxygen ow indication is
activated.
(The Yellow LED illuminates
continuously, the Red LED is blinking,
and the Audible Alarm is beeping
periodically.)
The ow meter knob is
completely closed.
Turn the ow meter knob
counterclockwise to center the
ball on the prescribed LPM ow.
The oxygen tubing is kinked
and blocking the delivery of
oxygen.
Check to see that the tubing is
not kinked or blocked. Replace if
necessary.
Limited oxygen ow to the user
without any
fault indication.
(All LEDs are o and the Audible Alarm
is silent.)
The oxygen tubing or cannula
is faulty.
Inspect and replace the items if
necessary.
There is a poor connection to a
device accessory.
Ensure that all connections are
free from leaks.

14
EverFlo / EverFlo QUser Manual
Chapter 5: Specifications
Environmental
Operating Transport & Storage
Temperature 13 to 32° C -34 to 71° C
Relative Humidity 15 to 95%, noncondensing 15 to 95%, noncondensing
Altitude 0 to 2286 m N/A
Physical
Dimensions 58 cm x 38 cm x 24 cm
Weight 14 to 15 kg
Expected Service Life of
Device and Accessories
5 years
Maximum Outlet Pressure** 6.5 PSIG (44.8 kPa)
Sound Level Device: 50 dBA or less
Alarm: 60 dBA or greater
Operating Pressure 69.7 kPa to 101 kPa
Standards Compliance
This device is designed to conform to the following standards:
- IEC 60601-1 Medical Electrical Equipment - Part 1: General requirements for basic safety and essential
performance
- IEC 60601-1-2 Medical Electrical Equipment - Part 1-2: General Requirement for Safety - Collateral Standard:
Electromagnetic Compatibility - Requirements and Tests
- IEC 60601-1-6 Medical electrical equipment - Part 1-6: General Requirements for Basic Safety and Essential
Performance - Collateral Standard: Usability
- IEC 60601-1-8 Medical electrical equipment - Part 1-8: General Requirements for Basic Safety and Essential
Performance - Collateral Standard: General requirements, tests, and guidance for alarm systems in medical
electrical equipment and medical electrical systems
- IEC 60601-1-11 Medical electrical equipment - Part 1-11: General Requirements for Basic Safety and Essential
Performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical
systems used in the home healthcare environment
- ISO 80601-2-69 - Medical Electrical Equipment - Part 2-69: Particular Requirements for Basic Safety and
Essential Performance of Oxygen Concentrator Equipment
- ISO 8359 Oxygen Concentrators for Medical Use - Safety Requirements
- IEC 62366-1 Medical devices - Part 1: Application of usability engineering to medical devices
- ISO 10993-1 - Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing (Biocompatibility)

15
EverFlo / EverFlo Q User Manual
NOTE: The EverFlo has Essential Performance as dened in ISO 80601-2-69. The EverFlo will deliver oxygen in
both normal and single fault conditions per the specications in this manual, or in the case of a power
supply failure, low oxygen concentration, or device malfunction, an alarm condition will occur.
Electrical, AC Power Consumption
Models 1020000, 1020001
1020002, 1020003
1020002BR, 1020003BR
1020014, 1020015
1039362, 1039363
1039364, 1039365
120 VAC ±10%, 350 W, 60 Hz
1020004, 1020005 220 VAC ±10%, 320 W, 60 Hz
1020006,1020007, 1020008
1020009, 1020010, 1020011
1020012, 1020016, 1020017
1020020, 1039366, 1039367
1039368, 1039370, 1104000
1125558
230 VAC ±10%, <300 W, 50 Hz
1102443 230 VAC ±10%, <300 W, 60 Hz
1020013 220 VAC ±10%, <300W, 60 Hz
Oxygen
Oxygen Concentration*
(All Models except
as noted below)
90–96% from 1 to 5 LPM**
Models 1020007, 1020008
1039367, 1039368, 1104000 87–96% from 1 to 5 LPM**
* Device operation above or outside of the voltage, LPM, temperature, humidity and/or altitude values specified may
decrease oxygen concentration levels.
** Oxygen flow rate <1 LPM requires low range flow meter accessory.
Maximum output pressure limited to 6.5 PSIG (44.8 kPa).

16
EverFlo / EverFlo QUser Manual
Sound Level
Models Sound Level (when measured at 1m from front of device)
1020000, 1020001
1020002, 1020003
1020004, 1020005
1039362, 1039363
45 dBA typical
102002BR, 102003BR
1020006, 1020008
1020009, 1020010
1020011, 1020012
1020013, 1020016
1020017, 1020020
1039366, 1039364
1039365, 1039368
1039370, 1102443
1104000
43 dBA typical
1020007, 1020014
1020015, 1039367, 1125558 <40 dBA typical
Note: Maximum Sound Pressure Level is 51 dB(A) and Maximum Sound Power is 59 dB(A) at 3LPM and 5LPM with an uncertainty of
2dB(A). Measured according to noise test method given in ISO 80601-2-69:2014 using the basic standards ISO 3744 and ISO 4871.
Disposal
Separate collection for electrical and electronic equipment per EC Directive 2012/19/EU. Dispose of this device in
accordance with local regulations.

17
EverFlo / EverFlo Q User Manual
Appendix A: EMC Information
GuidanCe and ManufaCTurer’sdeClaraTion - eleCTroMaGneTiC eMissions: This device is intended for
use in the electromagnetic environment specied below. The user of this device should make sure it is used in such an
environment.
eMissions TesT CoMplianCe eleCTroMaGneTiC environMenT – GuidanCe
RF emissions
CISPR 11
Group 1 The device uses RF energy only for its internal function. Therefore, its
RF emissions are very low and are not likely to cause any interference
in nearby electronic equipment.
RF emissions
CISPR 11
Class B The Device is suitable for use in all establishments, including
domestic establishments and those directly connected to the public
low-voltage power supply network that supplies buildings used for
domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage uctuations/Flicker emissions
IEC 61000-3-3
Complies
GuidanCe and ManufaCTurer’sdeClaraTion - eleCTroMaGneTiC iMMuniTy:This device is intended for
use in the electromagnetic environment specied below. The user of this device should assure that it is used in such an
environment.
iMMuniTy TesT ieC 60601 TesT
level
CoMplianCe level eleCTroMaGneTiC environMenT –
GuidanCe
Electrostatic Discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete or ceramic
tile. If oors are covered with synthetic
material, the relative humidity should be at
least 30%.
Electrical Fast Transient/Burst
IEC 61000-4-4
±2 kV for power supply lines
±1 kV for input-output lines
±2 kV for Power Supply Lines
NA – Device does not have
user I/O lines that are longer
than 3m in length.
Mains power quality should be that of a
typical home or hospital environment.
Surge
IEC 61000-4-5
±1 kV Line to Line
±2 kV Line to Ground
±1 kV Line to Line
NA - The device is a Class II
device and does not connect
to earth ground.
Mains power quality should be that of a
typical home or hospital environment.
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
IEC 61000-4-11
<5% UT
(>95% dip in UT) for 0.5 cycle
40% UT
(60% dip in UT) for 5 cycles
70% UT
(30% dip in UT) for 25 cycles
<5% UT
(>95% dip in UT) for 5 sec
<5% UT
(>95% dip in UT) for 0.5 cycle
40% UT
(60% dip in UT) for 5 cycles
70% UT
(30% dip in UT) for 25 cycles
<5% UT
(>95% dip in UT) for 5 sec
Mains power quality should be that of a
typical home or hospital environment. If
the user of the device required continued
operation during power mains interruptions,
it is recommended that the device be
powered from an uninterruptible power
supply or a battery.
Power frequency (50/60 Hz)
magnetic eld
IEC 61000-4-8
3 A/m 3 A/m Power frequency magnetic elds should be
at levels characteristic of a typical location in
a typical home or hospital environment.
Note: UTis the a.c. mains voltage prior to application of the test level.

18
EverFlo / EverFlo QUser Manual
GuidanCe and ManufaCTurer’sdeClaraTion - eleCTroMaGneTiC iMMuniTy:This device is intended for use in the
electromagnetic environment specied below. The user of this device should assure that it is used in such an environment.
iMMuniTy TesT ieC 60601 TesT level CoMplianCe level eleCTroMaGneTiC environMenT – GuidanCe
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to 80 MHz
3 V/m
80 MHz to 2.5 GHz
3 Vrms
3 V/m
Portable and mobile RF communications equipment should be
used no closer to any part of the device, including cables, than
the recommended separation distance calculated from the
equation applicable to the frequency of the transmitter.
Recommended separation distance:
d = 1.2
d = 1.2 80 MHz to 800 MHz
d = 2.3 800 MHz to 2.5 GHz
Where P is the maximum output power rating of the transmitter
in watts (W) according to the transmitter manufacturer and dis
the recommended separation distance in meters (m).
Field strengths from xed RF transmitters, as determined
by an electromagnetic site survey a, should be less than the
compliance level in each frequency range b.
Interference may occur in the vicinity of equipment marked with
the following symbol:
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is aected by absorption and reection from structures,
objects, and people.
a: Field strength from xed transmitters such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM
and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to
xed RF transmitters, an electromagnetic site survey should be considered. If the measured eld strength in the location in which the Device is
used exceeds the applicable RF compliance level above, the Device should be observed to verify normal operation. If abnormal performance is
observed, additional measures may be necessary, such as re-orienting or relocating the Device.
b: Over the frequency range 150 kHz to 80 MHz, the eld strengths should be less than 3 V/m.
reCoMMended separaTion disTanCes beTween porTable and Mobile rf CoMMuniCaTions equipMenT and The
deviCe:The Device is intended for use in the electromagnetic environment in which radiated RF disturbances are controlled.
The user of the Device can help prevent electromagnetic interference by maintaining a minimum distance between portable and
mobile RF communication equipment (transmitters) and the Device as recommended below, according to the maximum output
power of the communications equipment.
raTed MaxiMuM ouTpuT
power of TransMiTTer
(waTTs)
separaTion disTanCe aCCordinG To frequenCy of TransMiTTer
(MeTers)
150 khzTo 80 Mhz
d= 1.2
80 MhzTo 800 Mhz
d= 1.2
800 MhzTo 2.5 Ghz
d= 2.3
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73
11.2 1.2 2.3
10 3.8 3.8 7.27
100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended separation distance din meters (m) can be
estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is aected by absorption and reection from
structures, objects, and people.

19
EverFlo / EverFlo Q User Manual
Limited Warranty
Respironics, Inc. warrants that the system shall be free from defects of workmanship and materials and
will perform in accordance with the product specications for a period of three (3) years from the date
of sale by Respironics, Inc. to the dealer. Respironics warrants that the EverFlo / EverFlo Q units serviced
by Respironics, or an authorized service center, will be free from defects in serviced materials for a
period of 90 days and free from defects in workmanship for a period of 90 days from the time of service.
Respironics accessories are warranted to be free of defects in materials and workmanship for a period
of 90 days from the time of purchase. If the product fails to perform in accordance with the product
specications, Respironics, Inc. will repair or replace – at its option – the defective material or part.
Respironics, Inc. will pay customary freight charges from Respironics, Inc. to the dealer location only.
This warranty does not cover damage caused by accident, misuse, abuse, alteration, and other defects
not related to material or workmanship.
Respironics, Inc. disclaims all liability for economic loss, loss of prots, overhead, or consequential
damages which may be claimed to arise from any sale or use of this product. Some states do not allow
the exclusion or limitation of incidental or consequential damages, so the above limitation or exclusion
may not apply to you.
This warranty is given in lieu of all other express or implied warranties, including the implied warranties
of merchantability and tness for a particular purpose. In addition, in no event shall Respironics be liable
for lost prots, loss of good will, or incidental or consequential damages, even if Respironics has been
advised of the possibility of the same. Some states or provinces do not allow the exclusion of limitation
of implied warranties or the disclaimer of incidental and consequential damages. Accordingly, the laws
of your state or province may give you additional protections.
To exercise your rights under this warranty, contact your local authorized Respironics, Inc. dealer or
contact Respironics, Inc. at:
Respironics Inc.
10
01 Murry Ridge Lane
Murr
ysville, PA 15668 USA
Respironics Deutschland GmbH & Co. KG
Gewerbestrasse 17
82211 Herrsching, Germany
1-724-387-4000 +49 8152 93060
This manual suits for next models
1
Table of contents
Languages:
Other Philips Oxygen Equipment manuals