9)With the instrument in the VERTICAL position, biefly
press the button, the message Analyzing reference will
appear on the display;
10)Wait until the display displays the Insert sample
message;
Reading –Apparent Color
While Insert sample is on the display:
11)Open the measuring cover and remove the cuvette
from the compartment;
12)With the aid of the syringe, collect al least 15mL of the
sample to be analyzed;
13)With the sample in the syringe, fill in the glass cuvette
(12ml);
14)Close the cuvette with its cover;
15)Clean and dry puter wall of the cuvette using a cleaning
wipe or soft towel paper;
16)Insert the cuvette into the measuring compartment;
17)Measuring alignment: for better accuracy insert the
cuvette always in the same position;
18)Close the measuring cover on the glass cuvette to
avoid interference in the readings;
19)Briefly press the Power-Measuring button and wait.
The display will display the Message Analyzing
Sample, followed by the measurement result in PCU;
20)At the end of the measurement, discard the sample
analyzed and immediately wash the glass cuvette with
the distilled or deionized water.