Alliance Bio Expertise DOSYWEL User manual

DOSYWEL
Alliance Bio Expertise –Le Tremplin ZA de Courbouton 35480 –Guipry
+33 (0)2 40 51 79 53 –contact@abioexpertise.com
Revision 3

UM A.B.E. - DOSYWEL
2
Basic edition:
Pages
Version
Date
Written by:
Verified by:
sections
modified
32
1
10/09/2015
A. PINEDA
A. BLANCHARD
-
Updates:
32
2
16/11/2015
A. PINEDA
A. BLANCHARD
-
New
A.B.E
version
3
17/07/2018
A.BLANCHARD
N.CHEROTZKY
-

UM A.B.E. –DOSYWEL
3
Table of contents
General Information .......................................................................................... 5
Specifications Dosywel ..................................................................................................5
Overview ...............................................................................................................................5
Standard accessories shipped with your Dosywel............................................6
Dosywel optional accessories & consumable......................................................6
Setting up your Dosywel ................................................................................... 8
Installation and operating environment..................................................................8
Unpacking ............................................................................................................................8
Setting pump's jaw...........................................................................................................8
Fixing the tubing set ........................................................................................................9
Supply systems ................................................................................................ 10
Power supply ................................................................................................................... 10
Getting Started your Dosywel ......................................................................... 11
Functions of the ambidextrous controller........................................................... 11
Screen of your Dosywel .............................................................................................. 12
Colours code of your screen..................................................................................... 12
Programming your Dosywel ............................................................................ 13
Create a program........................................................................................................... 13
Edit the Program’s name............................................................................................ 13
Dosage settings.............................................................................................................. 14
Pump’s number (ONLY available if you connect a 2nd pump for turbo
dispensing) ....................................................................................................................... 15
Choice of pump speed ................................................................................................ 15
Verification test (calibration)..................................................................................... 16
Diluent choice .................................................................................................................. 17
Verification mode setting (calibration).................................................................. 18
Use of turbo mode ........................................................................................................ 19
Settings up Users........................................................................................................... 20
Programs management ................................................................................... 21
Editing a program .......................................................................................................... 21
Elimination of a program............................................................................................ 21
Routine use:reminder of a program..................................................................... 21
End of the cycle .............................................................................................................. 22
End of the dilution:......................................................................................................... 22
Settings menu.................................................................................................. 22
Date/Time settings ....................................................................................................... 22
Languages settings....................................................................................................... 22
USB data export.............................................................................................................. 22
Advanced options.......................................................................................................... 23

UM A.B.E. - DOSYWEL
4
Passwords use ............................................................................................................... 23
Screen colour settings.................................................................................................23
Upkeep and maintenance ............................................................................... 24
Upkeep and daily cleaning ......................................................................................... 24
External body ................................................................................................................... 24
Annual upkeep................................................................................................................. 24
Symbols Legend............................................................................................................. 25
Quick-Start Programming.......................................................................................... 26
Quick-Start set the parameters................................................................................ 27

UM A.B.E. - DOSYWEL
5
General Information
Specifications Dosywel
Model
DOSYWEL
Manufacturer
A.B.E.
Country of origin
100% France
Power supply
100-240 V 50-60 Hz
Volume Range
1 ml to 9999 ml
Distribution Speed
Adjustable: 0 to 100%
Internal tube dimensions
1.6/3.2/4.8/6.4 mm
External tube dimensions
1.6 mm
Pause time
1 to 60 seconds
Calibration mode
Yes
Connectivity
Polywel 3D (distribution
automate XYZ)
Communication
USB, Ethernet, RS232
Memory of 19 dilution programs
Traceability (USB, printer, Excel)
Turbo mode
Overview
Dimensions: W x D x H: 21 x 20 x 24 mm
Lightness:3.8 kg

UM A.B.E. - DOSYWEL
6
Standard accessories shipped with your Dosywel
Dosywel optional accessories & consumable
Additional Tubing
Product Code
Product Description
Qty
DOSW2001
Set of tubing 6.4mm with dispensing
nozzle for DOSYWEL
1
DOSW2002
Set of tubing 4.8mm with dispensing
nozzle for DOSYWEL
1
DOSW2003
Set of tubing 3.2mm with dispensing
nozzle for DOSYWEL
1
DILW3001
Rotor (pump head) tubing 6.4mm
5
DILW3002
Rotor (pump head) tubing 4.8mm
5
DILW3003
Rotor (pump head) tubing 3.2mm
5
DILW2006
Priming nozzle for bottle (1L or 2L with
filter)
1
DOSYWEL Accessories
Product Code
Product Description
Qty
DILW2031
Bar code / QR code reader (1D/2D) for
DILUWEL+
1
DILW2011
Ticket printer (USB)
1
DILW3005
Ticket printer rolls - 2100 labels
3
Dosywel
Set of tubing
3.2mm
Ref. DILW2002
Mains supply unit
120W
Power cable
User Manuel
Certification
METRO report

UM A.B.E. - DOSYWEL
7
DILW2013
Foot switch (USB)
1
DILW2012
Keypad for Diluwel/Dosywel
1
DILW2033
Keyboard for Diluwel/Dosywel -
Waterproof & rollable
1
DILW2001
Additional pump (for turbo)
1

UM A.B.E. - DOSYWEL
8
Setting up your Dosywel
Installation and operating environment
Environmental
Conditions:
- Temperature: 5 °C to 45 °C (41 °F to 113 °F);
- Maximum relative humidity 80%;
- Indoor utilization.
Electrical
environment:
According to EN61010-1 ED. 2011, international
standard about the safety requirements for
electrical equipment for measurement, control, and
laboratory use standard.
Installation
Category II
Pollution category 2 environments.
Electromagnetic
compatibility (EMC)
Performance:
According to EN61010-1 ED. 2011, international
standard about the safety requirements for
electrical equipment for measurement, control, and
laboratory use standard. Requirements about the
CEM (emission and immunity).
Unpacking
•Unpack in a clean and accessible place;
•Take the accessories out of the box:
-Power chord
-Tubing set 3.2mm
-User Manual
-Any additional ordered accessory or consumable.
•Take the Dosywel out of the box by lifting it from the bottom.
•Place the Dosywel on a stable and horizontal support preferably;
•Make sure, no elements are remaining in the box before disposal.
Setting pump's jaw
The pump's jaw is set in standard mode on the pipe 3.2 mm from both left
and right sides, to prevent any moving from the pipe or is damaged during
its use. The pump's jaw can be set in other dimensions (8.0, 0.5, 4.8 & 6.4
mm) by scrolling the knob at the bottom of the jaw.

UM A.B.E. - DOSYWEL
9
Fixing the tubing set
•Sterilize the tubing set and the dispensing nozzle before use;
•Open pump jaw and insert the rotor section which is indicated by two
metal closures into the pump head rotor;
•close the jaw and make sure the pump is properly adjusted;

UM A.B.E. - DOSYWEL
10
Supply systems
Power supply
The equipment must be connected to an electrical installation
that complies with the applicable standards in the country. The
installation must be equipped with protections against over-
currents, over-voltages, ground faults.
Do not install the equipment next to a source of heat or source
of humidity.
For your own safety, the equipment must be switched off and
disconnected from the mains before any technical intervention.
Connecting any other type of power supply may disrupt the
machine's operation or the surrounding machines.
•Make sure the electric supply matches the specifications of the
manufacturer's plate;
•The power supply of the unit must be assured by a grounded
outlet, and a safety device must guarantee automatic power-off
in case of faulty insulation. A power supply equipped with a
differential circuit breaker, powerful enough, would meet this
requirement. To ensure the power, switch off of the unit;
•the outlet must be immediately identifiable and easily reachable
by the operator.
•connect the power supply to the Dosywel’s back in the location
indicated by the symbol 24V:

UM A.B.E. - DOSYWEL
11
Getting Started your Dosywel
Functions of the ambidextrous controller
Button
General Features
Modality
Powering
Briefly pressing of the button
(1 Sec.) to turn on and off.
Access Settings menu
Long pressing of the button
(3 Secs.)
Changing the program
Briefly pressing of the button
(1 Sec.)
Validation/selecting a program
Briefly pressing of the button
(1 Sec.)
Go to General menu / Escape/
Delete.
Briefly pressing of the button
(1 sec.)

UM A.B.E. - DOSYWEL
12
Screen of your Dosywel
Figure 1. Principal screen’s work.
1Validate/Select –2Menu/Cancel –3On/Off –4Program Switch/Parameters
- 1Program name –2Pumps selected/Diluent name –3 Status,
Stop/Play/Pause way –4Dose delivered out of –5Interval (1 to 60 seconds).
Colours code of your screen
The orange screen appears when:
- A program was deleted.
- The declared pump is not found.
Green screen appears when:
- The dosage is ended.
To pause/cancel dilution:
- Click one/two times on .

UM A.B.E. - DOSYWEL
13
Programming your Dosywel
Create a program
You can create up to 19 programs. To create a program from the program 1,
follow the following procedure:
Edit the Program’s name
oPress briefly on to access to the programming menu;
oselect option, by a long pressing (2-3 seconds) on one of the
sides knobs.
oselect program number, scrolling and then clicking on the knob
oplace the cursor on the option program's name in the Program Editor
screen;
oclick once on the knob's side button to give a name to your program;
oenter the name using the knob or the keypad. To clear or go back,
press on .
Note: you must always write at least one character.
you don’t, the program will never exist.
oclick on the knob's side button to confirm the information entered.
Figure 2. Steps to appoint a program.
Always give a
name or a
number to create
a program.

UM A.B.E. - DOSYWEL
14
Dosage settings
oPlace the cursor over the “Dosage “pictogram in the Program
Editor screen, by scrolling and clicking on the knob;
oEnter the desired volume range from 1 ml to 9999 ml, by scrolling the
knob and click to confirm.
oselect the number of doses from 1 to 999 and click on the knob to
validate
ochoose the interval time between each dose from 1.0 to 60.0
seconds;
oclick on the knob to confirm the interval time entered.
Figure 3. Steps to choose the dosage’s volume.

UM A.B.E. - DOSYWEL
15
Pump’s number (ONLY available if you connect a 2nd pump for turbo
dispensing)
oCheck that the pump used is connected to your Dosywel;
oplace the cursor over the pump number option in the Program Editor
screen, by scrolling the knob. Click on the knob to validate
oclick on the knob’s side button to validate the pump number chosen.
Choice of pump speed
oPlace the cursor over Vi option (pump’s speed), scrolling one of the
knobs; click on the knob to enter the cell and modify the information.
oscroll one of the knobs to change the speed of the pump:0% to 130%,
knowing that 100% is the basic setting of the pump; click on the knob
to validate
Figure 4 Steps to choose the pump.

UM A.B.E. - DOSYWEL
16
Verification test (calibration)
oPlace the cursor over the Verification option, scrolling one of the
knobs;
oclick once on the knob’s side button to select the option;
oturn one of the knobs to select the requested volume of the dose: 1ml
to 1 000 ml;
oput the nozzle above a measuring cylinder;
oclick on the knob’s side button to validate the volume chosen. Note:
the pump delivers one dose.
oenter the measured volume, scrolling one of the knobs;
oclick on the knob’s side button to validate the measured volume;
oplace the cursor over by scrolling one of the knobs;
oclick on the knob's side button to confirm the information entered;
oplace the cursor over , and click on the knob's side button to
access to Diluent menu.
Figure 5 Steps to set the verification.

UM A.B.E. - DOSYWEL
17
Diluent choice
When selecting a pump, the machine automatically proposes the diluent
settings.
omodify the required parameters using the knob’s side button,
knowing that DI: Diluent name, No: batch number, Vo: Volume to
dispense;
oedit the diluent’s name, clicking on the one knob's button side;
otype the name scrolling the knob's button side. To clear or go back,
press on ;
oclick on the knob’s side button to validate the diluent name chosen;
oplace the cursor over No: Batch Diluent’s number by scrolling one of
the knobs;
oclick on the knob’s side button to edit the batch number;
otype the batch number scrolling the knob’s button side. To clear or go
back, press on ;
oclick on the knob’s side button to validate the batch diluent’s number
chosen;
oplace the cursor over Vo: Volume by scrolling one of the knobs;
oedit the volume, clicking on the one knob's button side;
otype the name scrolling the knob's button side. To clear or go back,
press on ;
oclick on the knob’s side button to validate the Volume (ml) chosen;
oplace the cursor over , and click on the knob's side button to confirm
the information entered.

UM A.B.E. - DOSYWEL
18
Figure 6| Steps to set the Diluent option.
Verification mode setting (calibration)
oPlace the cursor over Verification option, scrolling one of
the knobs;
oclick once on the knob’s side button to select the option;
oselect YES or NO in the automatic calibration, on the start of cycle,
scrolling one of the knobs. Note: if you select the option YES, to start
a dosage you have to carry out the calibration control (see page 17:
Verification test-calibration);
oclick once on the knob’s side button to select the option;
oplace the cursor over “Periode” option, scrolling one of the knobs;
oclick once on the knob’s side button to select each period you need to
carry out an automatic calibration. Note: if your settings are 10 doses,
each ten doses will appear a window to perform the calibration
control (see page 17: Manual calibration)
oplace the cursor over by scrolling one of the knobs;
oclick on the knob's side button to confirm the information entered.
oOn the Program Editor screen, place the cursor over by
scrolling one of the knobs;
oconfirm your new program by pressing the knob’s side button or

UM A.B.E. - DOSYWEL
19
Figure 7 Steps to set the verification at the start of the cycle.
Use of turbo mode
oPlace an additional pump (Dilupump, ref: DILW2001) provided by
A.B.E on the additional port of the Dosywel (see the back panel).
oUse a special adapter "y" for the turbo mode, in order to gather 2
tubings in one source.
oGo to page 15 to select the pumps (you should select 1, main one +
2, additional one : shown 12 on the screen).

UM A.B.E. - DOSYWEL
20
Settings up Users
To create a username:
Place the cursor over “Username ”option, scrolling the knobs, (INVITE
is the basic setting of the username);
click once on the knob’s side button;
press briefly on the knob’s side button;
click again on the knob’s side button;
write the name of the desired user using the knob’s side button or keyboad;
confirm your username by pressing the knob’s side button.
To delete a username created, keep pressing 2 secs on
Table of contents
Other Alliance Bio Expertise Laboratory Equipment manuals
Popular Laboratory Equipment manuals by other brands

Belden
Belden HIRSCHMANN RPI-P1-4PoE installation manual

Koehler
Koehler K1223 Series Operation and instruction manual

Globe Scientific
Globe Scientific GCM-12 quick start guide

Getinge
Getinge 86 SERIES Technical manual

CORNING
CORNING Everon 6000 user manual

Biocomp
Biocomp GRADIENT MASTER 108 operating manual