Alpha Moisture Systems DS1200-PDT User manual

Model DS1200-PDT Dewpoint Hygrometer
-40 / +30°Cdp (50bar maximum pressure)
-60 / +20°Cdp (50bar maximum pressure)
User Manual
Englis

Model DS1200-PDT - Dewpoint Hygrometer
CERTIFICATE No. FM35600
BS EN ISO 9001:2015

Document Reference:
2249 - Issue 3
21/06/2018
Model DS1200-PDT Dewpoint Hygrometer
User Manual
©Alpha Moisture Systems Ltd.
This is a step by step instrucon manual to help you successfully set up
correctly the Model DS1200-PDT Dewpoint Hygrometer before use.
This manual should be kept with the instrument for future reference.
You must observe the safety informaon on pages 6 and 10 before
installaon.
Please read t is manual carefully from t e start.

4
Model DS1200-PDT - Dewpoint Hygrometer
Index
General Overview 5
Safety Informaon 6
Model DS1200 Control-Display Front Arrangement 7
PDT Arrangement 7
Features of the Model DS1200-PDT 7
DS1200 Mounng 8
Installing the Instrument into a Panel 8
Wiring DS1200 Power Supply 8
PDT Transmier Cable 8
Alarm Cable 9
Analogue Output Cable 9
DS1200 Connecons 9
PDT Connecons 10
Installing the Model PDT in a Air/Gas Sampling System 10-11
Piping Installaon Schemac 11-12
Installing and Commissioning the Model PDT Transmier 13
Seng the Alarm Trip Points 13
Dimensions 14
Speci caons 15-16
Contact Informaon 17

5
Model DS1200-PDT - Dewpoint Hygrometer
General Overview
Thank you for purchasing the Alp a Moisture Systems Model DS1200-PDT Dewpoint
Hygrometer.
We hope that you will agree that this instrument is easy to set up and very user friendly to
operate giving you reliable and accurate measurements quickly and efficiently.
The Model DS1200-PDT Dewpoint Hygrometer is a global popular choice of Dewpoint
Hygrometers (trace moisture analysers) that work on a basis of polymer sensors and are
used for the determinaon of moisture in most gases in the dewpoint or RH range.
They consist of a separately installed measuring cell of high grade stainless steel (available
through your local distributor or Alpha Moisture Systems where there is no distributor for
your locaon) a Transmier and Control-Display, that can be integrated in a Gas Handling
System*, and an in-line process source.
The Model DS1200-PDT Dewpoint Hygrometer is installed as a single channel analyser.
For addional accessories and parts, or technical help for the Model DS1200-PDT, or any
other Alpha Moisture Systems products, please contact your local distributor or contact
Alpha Moisture Systems - see last page for details.
*Alp a Moisture Systems can manufacture to your specificaons a Gas Sample System to
suit your exact requirements, please contact your local distributor for more informaon.
Visit our website www.amsystems.co.uk for more informaon.

6
Model DS1200-PDT - Dewpoint Hygrometer
Safety Informaon
The DS1200-PDT has been designed to be connected to hazardous electric voltages and
must be protected by a 5amp fused plug to the mains/AC supply.
Check to establish that all wiring and connecons are not damaged. If damage is
observed to any electrical wiring, connecons or damage to the apparatus they must not
be connected to the mains/AC supply but returned to your local distributor for rec caon.
Warnings
Risk of electric shock - Do not open any part of the enclosures of the Model DS1200-PDT
whilst connected to the mains/AC supply.
Do not connect the Model DS1200-PDT to the mains/AC supply unl all other connecons
and pipework is secure and have been tested.
Ignoring this safety informaon could result in severe personal injury and/or mechanical
damage to the Model DS1200-PDT.
The product speci caons must not be exceeded at anyme, as this may cause personal
injury, damage to the instrument and sensor or cause risk of re.
Do not connect the Model DS1200-PDT to any other device that is not recommended in
this manual.
Ensure that the Model DS1200-PDT does not come into direct contact with water or any
other liquids. See IP/NEMA protecon in the product speci caons.
To avoid the risk of electric shock, risk of damage or re, these safety instrucons and
guidelines must be followed. Only quali ed personnel/technicians should install these
units to the mains/AC supply and connecng pipework to ensure it is completely safe by
tesng and recording before use.
It is the responsibility of the customer to ensure safe working condions especially working
with hazardous gases and liquids. Leak tests should be carried out periodically by quali ed
personnel.
Safe Isolaon
Switch off at the mains/AC socket and remove the plugs before any maintenance is carried
out by a quali ed person. Always test components with an approved voltage meter before
handling to ensure it is completely dead.
YOU MUST ALSO READ THE WARNING INFORMATION ON PAGE 10.

7
Model DS1200-PDT - Dewpoint Hygrometer
Model DS1200 Control-Display Front Arrangement
LED Alarm Indicators. Sounders
can also be connected at the rear
of the unit. See page 9 diagram.
Menu
navigaon
keys
Con rmaon
key
PDT Transmier Arrangement
See Pages 14 - 16 for dimensions, diagrams and tec nical specificaons
Permeable
Sensor Guard
SUPPLIED WITH A CERTIFICATE OF CALIBRATION
Iden caon Label
with Serial Number
Cable
Connecon
Features of t e Model DS1200-PDT
DS1200 Control-Display
• The Model DS1200-PDT is a single channel analyser.
• DS1200 has a 4-digit 14 segment LED display
• DS1200 has 2 adjustable Alarm set points and an analogue output. See pages 9 and 13.
• DS1200 has mA Output (terminals 11 and 12)
• Universal Voltage Supply (terminals 31 & 32)
PDT Transmier
• Ranges available: -40 to +30°C or -60 to +20°C (-40 to +86°F or -76 to +68°F) Dewpoint
• Two wire 4-20mA transmier for connuous measurement of moisture in a process
gas or compressed air
• Reliable, accurate and easy to install

8
Model DS1200-PDT - Dewpoint Hygrometer
DS1200 Mounng
• Only technicians who are familiar with the technical terms, warnings, and instrucons
in the manual and who are able to follow these should connect the DS1200.
• Should there be any doubt as to the correct handling of the DS1200, please contact
your local distributor.
• Mounng and connecon of the DS1200 should comply with the naonal legislaon
for the mounng of electric materials, i.e. wire cross-secon, protecve fuse, and
locaon.
Descripons of Input / Output and supply connecons are shown in the block diagram
on page 9 and the instrument top label.
• The maximum size of the protecve fuse is 5A and, together with the power switch, it
should be easily accessible and close to the DS1200. The power switch should be
marked with a label indicang it will turn OFF the voltage to the DS1200.
• To be mounted in front panels. The included rubber packing must be mounted between
the panel cutout and the display front to obtain IP65 (NEMA 4) ingress protecon.
Installing t e Instrument into a Panel
• Make a cut-out in the donor panel 92.0/92.8 x 45.0/45.6mm (DIN 43700).
• The maximum panel thickness is 10mm and, if an effecve IP65 weatherproof seal is
required, the minimum recommended panel thickness is 1.6mm.
• Pass the instrument cabinet through the cut-out in the donor panel and slide the panel
clamp over the instrument, from the back.
• Turn the Red panel clamp screws unl the instrument is clamped in posion. The
screws must be ghtened sufficiently to affect a seal between the front of the donor
panel and the back of the instrument bezel, but never over ghtened to the point of
fracturing the panel clamp or instrument case.
NOTE Wires are retained by screws. Ensure t at t e exposed secon of t e wire is fully
inserted and t at no loose strands are exposed.
Wiring DS1200 Power Supply
• Connect the power supply cable to the 2 terminal block marked 31 and 32 no polarity.
• The power supply should be 22 to 253 VAC @ 50/60Hz or 20 to 300 VDC.
PDT Transmier Cable
• Connect the transmier cable to connector pins 45 and 46, ensuring that the red wire
connects to pin 46. See Diagrams on pages 9 and 10.
Observe that the cage is securely clamped onto the bootlace ferrules on the cable
provided.
• Route the sensor cable to the intended site of the sensor.
Note : - Do not install t e transmier at t is me.
Wait unl t e commissioning stage as described later

9
Model DS1200-PDT - Dewpoint Hygrometer
Alarm Cable
• Make the appropriate connecons, nong the normally open and normally closed relay
contact posions.
Analogue Output Cable
• Make the appropriate connecons, ensuring that the correct polarity and the
maximum load speci caon is strictly observed.
DS1200 Connecons:
BLUE wire from the
PDT Cable connected
to pin 45 on the rear
of DS1200
RED wire from the
PDT Cable connected
to pin 46 on the rear
of DS1200
Polarity is universal
and can be wired +/-
or -/+

10
Model DS1200-PDT - Dewpoint Hygrometer
M12 connector plug
If no connecon cable is ordered, the sensor will be supplied with a M12 connector plug.
The user can connect the supply and signal cables as indicated in the connecon diagram
below.
PDT Connecons:
Pin 2 to Pin 45 on
DS1200
Pin 3 to Pin 46 on
DS1200
Installing t e Model PDT in a Air/Gas Sampling System
Warning: D n t exceed a pressure f 50 bar.
Observe the o erating ranges of the sensor. The robes are damaged if they are
overheated.
Observe maximum storage and trans ort tem erature as well as maximum
o erating tem erature (i.e. rotect the instrument from direct sunlight).
Imp rtant: Before installation, bleed com ressed air systems in order to remove
condensate and articles to avoid contamination.

11
Model DS1200-PDT - Dewpoint Hygrometer
Installing t e Model PDT in a Air/Gas Sampling System - Connued
Piping Installaon Sc emac
The piping installaon schemac diagram below shows all components, which could be
used in a dry gas measurement applicaon although not all the items shown will be
required for every installaon.
The flow rate, although not crical to the sensor measurement, should be low enough
to avoid abrasion to the sensor surface without being so low as to extend the system
response me to an unacceptable level. In general, a flow rate of between 2 and 3
litres/minimum at NTP will give the right balance.
The sensor is a variable capacitor, which is directly affected by changes in paral pressure
of water vapour. These changes are proporonal to the dew/frost point temperature.
The measuring transmier can be installed directly into the process line or ducng, but
this can create problems with access for maintenance and calibraon. It is for these
reasons that we recommend that the transmier be installed in a bypass, fast loop or
total loss sample system where the transmier is accessible without interrupng the
main process flow line.

12
Model DS1200-PDT - Dewpoint Hygrometer
Piping Installaon Sc emac - Connued
1) Sample Isolaon Valve - This is a recommended item as it allows access to the
sample system without interrupng the main process line.
2) Sample Tube – This should be stainless steel for dry air or gas applicaons but copper
or carbon steel can be used where weer gases are to be measured. If any secon of
the sample tube must be flexible then PTFE should be used. In most cases, 3mm OD
(1/8”) is sufficient as it provides good system response me within minimum flow.
6mm OD (1/4”) tube can be used where pressure drops across the 3mm tube are too
high.
3) Filter Unit – A lter unit is recommended when the samples are likely to contain
parculate maer. If the air/gas sample contains heavy hydrocarbon condensate, the
lter must be of the coalescing type with a drain. The lter unit should be posioned
as close to the sample point as praccal.
4) Pressure Reducon Valve or Pressure Regulator – If the sample is to be measured at
atmospheric pressure then the valve 4A should be ed and 4B omied from the
system. If the sample is to be measured, at full line pressure and the exhaust vented
to atmosphere, then valve 4B should be ed and 4A omied from the system. If
measurements are to be taken at full line pressure and the sample is to be returned
to a part of the main line or a vent, which is at a pressure higher than atmospheric,
and the input to that line needs a controlled pressure then both 4A and 4B will
be required.
5) Sample Pressure Gauge – This is not a crical part of the moisture measurement
but may be required if Dew/Frost point measurements are to be made at higher
than atmospheric pressure.
6) Measuring Transmier – Model PDT with Connector, General Arrangement.
7) Transmier Holder – Transmier Holder General Arrangement.
8) Desiccant C amber – This item is required when the sampling is to be intermient.
When installed it prevents the ingress of wet air to the sample system, while the
sample is not flowing, improving the response me.
9) Flow Control Valve – This can be a separate item or combined with the flow indicator.
10) Flow Indicator – The recommended minimum sample flow is 2 to 3 litres per minute.
11) Sample Ex aust – The exhaust can be vented to atmosphere or returned to the
process line as discussed above.
All addional components are available t roug your local distributor.

13
Model DS1200-PDT - Dewpoint Hygrometer
Installing and Commissioning t e Model PDT Transmier
It is advisable to carry out an inial purge of the sample loop, before installing the
transmier, in order to reduce the possibility of sensor damage on start-up.
Refer to the sample system schemac on page 11. Open the inlet isolaon valve slowly,
unl a small flow of air/gas at atmospheric pressure flows through the inlet pipe work to
the transmier holder and exhausts through the sensor entry port of the transmier
holder.
Allow this purge to connue for about 15 to 20 minutes to remove any residual moisture
from the sample pipe work and components.
Close the inlet isolaon valve and install the transmier into the transmier holder. Locate
and secure the cable connector in posion on the transmier. Use the locking screw in
order to affect a weatherproof seal.
Open the inlet valve slowly again and, by opening all valves aer the transmier holder,
allow a low-pressure purge through the whole sample system. (Note. If a closed by-pass
loop is installed, this secon of the procedure is not possible).
Set the required pressures and flows within the sample loop.
This completes the installaon and commissioning but, on inial start-up, it could take
several hours for the system to reach equilibrium.
Seng t e Alarm Trip
Points
To acvate the quick alarm
sengs screen press either the ”
” or ” ” buons, while the
DS1200 is displaying the moisture
level.
The DS1200 will flash the “F.SET”
message followed by a selecon
menu for Alarm 1 or 2 relay. Scroll
to the required alarm and press
the “OK” buon.
The DS1200 then displays the
message “SETP” momentarily
before displaying the current set
point in % of scale.

14
Model DS1200-PDT - Dewpoint Hygrometer
Dimensions
DS1200
PDT Transmier
PDT Transmier Holder Model SH-PDT Flow T roug Sample Cell
3 (0.125)
34ø (1.34 dia)
54 (2.125)
15 (0.6)
58 (2.28)
68 (2.68)
10 (0.4)
43 (1.7)
43 (1.7)
44 (1.73)
54 (2.125)
2 Mounting Holes 5.5ø (0.22 dia.)
43 (1.7)
Note: The Assembly is shown with 1/4@ OD tube ngs. The dimensions across the tube ngs will
vary for all other size ngs.

15
Model DS1200-PDT - Dewpoint Hygrometer
Specificaons
DS1200
• Temperature operang range = -20°C to +60°C (-4°F to 140°F)
• Power supply = 20 – 300 VDC, or 22 to 253 VAC 50 – 60 Hz
• 3.2W internal power consumpon
• 3.5W maximum power consumpon
• Isolaon voltage test / operaon = 2.3KVAC / 250 VAC
• Signal to noise rao = Min 60dB (0-100kHz)
• Display Response me (0-90%, 100 to 10%) = 0.4s.
• Wire size, pin 45 & 46 (max) = 1 x 1.5mm2 stranded wire
• Wire size, others (max) = 1 x 2.5mm2 stranded wire
• Relave Humidity = < 95% RH
• Dimensions (HxWxD) = 48 x 96 x 120mm
• Cut out dimensions = 44.5 x 91.5mm
• Ingress Protecon = (Front panel) IP65/NEMA4
• Weight = 230g
• Input
oFrom Model PDT
oError detecon = Cable loop break
• Display
oDisplay Readout = PDT range
oScrolled error display
• Current Output
oProgrammable signal ranges = 0…20 or 4…20mADC (Factory set)
oLoad (max) 20mA / 800Ω / 16VDC
oLoad stability ≤ 0.01 % of span / 100 Ω
• Relay
oRelay funcon = setpoint
oHysterisis 1% of range
oMax voltage = 250 VRMS
oMax Current = 2A /AC
oMax AC power = 500VA
oMax current at 24 VDC = 1A

16
Model DS1200-PDT - Dewpoint Hygrometer
Output Signal : 4 to 20mA Linear to its range.
Operang Voltage : 10V - 30V DC.
Factory calibraon : Supplied with Cer cate of Calibraon traceable to NPL/NIST
Accuracy : ± 2°C dewpoint (NPL / NIST traceable)
Temperature compensaon : Temperature compensated for operang range.
Operang temperature : -30…70°C (ideal 0…50°C)
Storage Temperature : -40…80°C
Operang Pressure : From 1kPa (0.01 barA) to Maximum 5,000kPa (50 barA)
Sample Flow Rate : Independent but ideally 2 to 3 litres per minute.
Max: 25 litres/min.
Connecon : M12, 5-pole
EMC : DIN EN 61326
Transmier Enclosure : Zinc alloy, PC, ABS
Sensor Protecon : 316 Sintered stainless steel lter - 50 micron
Weatherproof Classi caon : IP65/NEMA4 when Connector mated to Transmier.
Mechanical Connecon : G 1/2”
Mechanical Warranty : 12 months in case of faulty workmanship and defecve parts.
Specificaons PDT Transmier
Specificaons PDT Transmier Holder
High Grade 316 Stainless Steel - Part Number SH-PDT

17
Model DS1200-PDT - Dewpoint Hygrometer
Notes:
Contact Informaon
Head Office:
Alp a Moisture Systems
Alpha House
96 City Road
Bradford
West Yorkshire
BD8 8ES
United Kingdom
Tel: +44 (0) 1274 733100
Fax: +44 (0) 1274 733200
Email: info@amsystems.co.uk
Web: www.amsystems.co.uk
Office Opening Hours:
Monday - T ursday
8.30am - 5.30pm GMT
Friday
8.30am - 5.00pm
Saturday and Sunday - Closed
Distributor Informaon:
Table of contents
Other Alpha Moisture Systems Measuring Instrument manuals

Alpha Moisture Systems
Alpha Moisture Systems SADPmini2 User manual

Alpha Moisture Systems
Alpha Moisture Systems dewTEC DS2000 User manual

Alpha Moisture Systems
Alpha Moisture Systems dewSMART DS1000 User manual

Alpha Moisture Systems
Alpha Moisture Systems SADPminiEx User manual

Alpha Moisture Systems
Alpha Moisture Systems SADPmini2 User manual

Alpha Moisture Systems
Alpha Moisture Systems DS1200-AMT Guide